03 Additional Information.pdf
PDF 1 MB Posted
- Attached to
- Z2DA--MRI Trailer Relocate Federal contract opportunity
- Solicitation number
- 36C25923R0021
About this file
This solicitation is for a construction project to relocate a mobile MRI trailer at the Grand Junction, Colorado VA Medical Center. The scope of work includes selective demolition, construction of a new concrete pad and platform addition, installation of HVAC, electrical, and enclosure connecting the MRI trailer to an existing building. The period of performance is 180 calendar days from notice to proceed. The contractor must comply with all applicable federal, state, and local regulations including OSHA and VA safety requirements. The contractor will be responsible for construction management, scheduling, and safety plans.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| S05 - RFIs and Responses_2nd.pdf | ||
| S06 - 077100 ROOF SPECIALTIES.pdf | ||
| 36C25923R0021 0002.docx | DOCX document | |
| GMRVE-3000-D20-48-06 - GE Voyager-Explorer (004).pdf | ||
| GJ Geotech Report 2019.pdf | ||
| Site Utilities.pdf | ||
| S05 - RFIs and Responses.pdf | ||
| S04 - Pre-Bid Sign in.pdf | ||
| S-101.1 CONSTRUCTION.pdf | ||
| 051200 STRUCTURAL STEEL FRAMING_rev1.pdf | ||
| S04 - Site Visit Meeting Doc_MRI Trailer.pdf | ||
| 36C25923R0021 0001.docx | DOCX document | |
| 852.219-75 LOS Certification.docx | DOCX document | |
| 03 Plans.pdf | ||
| 03 Specifications.pdf | ||
| 36C25923R0021.docx | DOCX document |
Show all 16
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
GMRVE-3000-D20-48-06
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Site Planning Guide
GE SIGNA VOYAGER & EXPLORER 1.5 T
Mobile MRI Systems
48’ L x 8’-6” W x 13’-6” H USA Unit
© 2017 Advanced Mobility Specialty Vehicles, All Rights Reserved
This guide contains confidential information of Advanced Mobility Specialty Vehicles. You may not copy it or any part of it without the written permission of Advanced Mobility Specialty Vehicles. This manual may be used only by you, and only for the purposes for which it was intended. You may not disclose this manual or the confidential information it contains outside of your company. If you wish to copy any part of this manual, or to use it other than as described above, you must contact Advanced Mobility Specialty Vehicles seeking permission to do so.
North America
Corporate Headquarters Advanced Mobility Specialty Vehicles 500 Crossing Drive University Park IL. 60484
Europe
Advanced Mobility Specialty Vehicles, Ltd.
Unit 17, Nelson Way Tuscum Trading Estate, Camberley, Surrey
GU15 3DH
United Kingdom
(44) 01276.64490
Buys Ballotstraat 6 3261 LA Oud-Beijerland, Holland +31 (0) 186-614322 Fax. +31 (0) 186-619367 E-mail: smit@smit-mobile.nl
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
List of Revisions
Date Revision
Number
Revised by Description
11/01/2016 01 Bill Landgraf Initial Release
12/01/2016 02 Bill Landgraf Document Update
01/31/2017 03 Bob Slevnik Pg.9 Mag. Shielding
03/12/2017 04 Bill Landgraf Floor Height to 54”
04/12/2017 05 Bob Slevnik Floor Height to 56” REF (+/-1”) Pg. 17
3/1/2018 06 Chirag Patel Updated Document Number to New Numbering System
Notice In accordance with our policy of continued product improvement, Advanced Mobility Specialty
Vehicles reserves the right to make changes in the equipment, design, specifications and materials of the product described herein. Any problems or questions related to the components or systems covered in this booklet may be directed to:
Advanced Mobility Specialty Vehicles Attention: Service Department 16745 South Lathrop Avenue Harvey, Illinois 60426
USA
Telephone: (001) 800.839.0630 (24/7 Service) Fax: (001) 708.596.2480 http://www.Advanced Mobilitysv.com
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Table of Contents Introduction
Warnings & Safety Alert Conventions Support Pad Requirements
Trailer Weight Recommended Support Pad Support Pad Depth Support Pad Levelness Recommended Service Pad Steel Reinforced Concrete Pad Vehicle Access Recommended Attachment to Facility Exclusion Zone Vehicle Movement Vibration / Foundation Design Swing Clearance Note
Customer Power Requirements Lockout/Tagout Electrical Service Configuration Frequency Phase Balance Maximum voltage variation Connector Type Customer Facility Voltage Surges Power Source Monitoring (Facility Only) Ground Conductor Magnetic Shielding R.F. Shielding
Mobile Grounding Requirements Special Grounding Note:
Mobile Grounding Chart
Telephone and Data Service Requirements Telephone Service Data Service
Water and Chiller Requirements Humidifier Water Fill Water Supply Requirements (Sink Option) Waste Water Connections (Sink Option)
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Table of Figures Figure 1: Ground Connection Figure 2: Plan View and Pad Layout Figure 3: Right Side Elevation Figure 4: Left Side Elevation Figure 5: Stair/Lift/Wall Elevation Figure 6: Russellstoll Service Outlet Figure 7: Russellstoll Chart Figure 8: Crouse Hinds Service Outlet (Optional) Figure 9: Crouse Hinds Chart (Optional) Figure 10: Turning Requirements
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Introduction
The purpose of this document is to provide the basic information needed for site planning. For specific information not contained in this document, please contact Advanced Mobility Specialty
Vehicles.
The mobile unit requires sufficient room to be maneuvered and positioned for setup and takedown.
The mobile unit has many storage compartments and service doors that require access during these procedures as well as during operation. The platform lift, entry stair and optional platform require additional space on the passenger side of the mobile unit. Refer to the drawings provided for actual locations of doors, platform lift, and stair sizes and locations.
Warnings & Safety Alert Conventions Three types of statements are used throughout this document to warn the operator of potential situations. Always read these statements carefully and take the appropriate safety precautions to ensure a safe environment for all personnel and all property. The statements are as follows:
This type of notice indicates a potentially hazardous situation, which if not avoided, could result in injury or death to the operator of the mobile unit.
This type of notice indicates a potentially hazardous situation, which if not avoided, could result in irreparable damage to the mobile unit.
This type of notice is meant to inform the operator of useful information.
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Support Pad Requirements
The following is a list of recommendations and requirements for a concrete support pad. However, due to varying site conditions, the actual pad design should be prepared by an appropriately licensed structural or architectural engineer.
Trailer Weight The weight of the trailer should be considered in the design of the support and service pads. The overall weight of the trailer is approximately 60,000 lbs. The weight on the rear axles is approximately 38,400 lbs. The weight on the King Pin is approximately 21,880 lbs.
Recommended Support Pad A full pad measuring 52’-5” x 11’, located as shown on
Figure 2: Plan View and Pad Layout and Figure 3: Right Side Elevation as the cross-hatching, is the recommended support pad.
Support Pad Depth Recommendations for the width and length of the pad are given above. Based upon the weight distribution information given in Figure 3: Right Side Elevation and existing site conditions, the depth should be determined by a local contractor. It is recommended that non-ferrous reinforcement materials be used for pad reinforcement. GE must approve plans for pad construction.
Support Pad Levelness The support pad must be level to ensure proper operation of the MRI system. The pad must not exceed .125” deviation in 10’-0”. If the minimum support pads are selected, rather than the recommended single pad, they must also meet this specification.
Recommended Service Pad A service pad is recommended to provide adequate service access. The recommended size of the pad is 60’ x 18’. See Figure 2: Plan View and Pad Layout , for details.
Steel Reinforced Concrete Pad Nonferrous reinforcement materials are recommended. If ferrous materials are used contact GE for the maximum weight allowed per square foot.
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Vehicle Access A firm, level surface is required around the mobile unit to provide access to the tractor/trailer, aid with patient handling, servicing the unit, and delivering of cryogens.
Recommended Attachment to Facility An inflatable air bag or soft seal is recommended at the point of connection from the unit to the facility. Fixed or solid connections may hinder imaging quality. Contact OSV or the local GE representative prior to construction if the proposed connection varies from the recommended.
Exclusion Zone An area of 5’-0” x 5’-0”, located directly below the magnet vent should be fenced off to prevent injury in the event of magnet quenches. The helium gas must be allowed to vent, unrestricted, to a non-accessible area, allowing the helium gas to dissipate.
Vehicle Movement The MRI is very sensitive to vibration and moving metal. Consequently, all vehicle traffic shall be kept as far away as possible from the pad. Moving ferrous materials having the listed masses should be limited to areas as described in the GE site planning publication. Contact GE to obtain the latest version.
Vibration / Foundation Design Please contact GE Medical Systems for the latest system specific vibration requirements.
Swing Clearance Note Please verify the actual dimensions of the rearmost projections on the cab of your tractor to the centerline of tandem suspension or centerline of the fifth wheel plate on your tractor. Refer to
Figure 10: Turning Requirements for proper tractor sizing information.
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents
Customer Power Requirements
It is the operators’ responsibility to verify that the shore power receptacle is of the same type and voltage as the connection that is supplied by Advanced Mobility Specialty Vehicles. Failure to do this can result in injury or death to the operator of the mobile unit as well as irreparable damage to the mobile unit.
The standard connector for the unit is a Russellstoll DS2504MP000/DF2032 480V 200A Plug. An acceptable optional connector is the Crouse Hinds # AP 20457. If an existing site currently implements a different connector or connector configuration, please contact Advanced Mobility Specialty Vehicles in order to arrange for a compatible power connector before the unit leaves the facility.
Lockout/Tagout A Lockout/Tagout provision in accordance with OSHA Standard 1910.147 is required. The facility shore power disconnect device must be located within 40’- 0” of the unit and must provide for an effective lockout/tagout to facilitate safe service and maintenance of the unit.
Electrical Service A single electrical power source is required for operation of the MRI system.
480 Volt A.C., 3 Phase, 125 KVA, fused at 150 Amps
Configuration Three phase, five wire, wye connection, with ground and neutral (not less than AWG #1/0).
Frequency 60 Hz ±3Hz
Phase Balance The phase balance is + 2% maximum phase-to-phase line voltage difference lowest phase.
Maximum voltage variation The maximum voltage variation is ±10% from nominal steady state (under the worst case conditions of line voltage)
Connector Type The unit is supplied with a 50-foot cable and male connector. The connector is a Russellstoll 200 Amp plug #DS 2504 MP000/DF2032.
The optional Crouse Hinds connector male plug # AP20457 is also available.
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Customer Facility The facility must have the matching receptacle as specified in Figure 6: Russellstoll Service Outlet and Figure 7: Russellstoll Chart. The receptacle is a Russellstoll #DF 2504 FRAB0.
The optional Crouse Hinds facility receptacle # AR2041 may also be used as specified in Figure 8:
Crouse Hinds Service Outlet and Figure 9: Crouse Hinds Chart.
Voltage Surges Transient voltage variations caused by external loads must not:
• Exceed + 5%
• Exceed five cycles duration
• Occur more than ten times an hour.
Power Source Monitoring (Facility Only) NOTE: Perform a power audit first.
A power analyzer should be used to check the proposed Mobile MRI Series facility site power for average line voltage, surges, sags, reclosures, impulses, frequency and microcuts. A period that includes two weekends should be used to simulate several days of normal use. Analysis of the data and site history of any previous power problems with other X-ray systems or computer installations should be reviewed with your power and ground representative. Verify “brown-out” (low voltage) conditions, which may occur during summer months, will not exceed the allowable range.
Some analyzer models that are suitable for power monitoring are:
• Dranetz Model 658
• Dranetz Model 656A
• BMI 3630
• RPM
Ground Conductor An insulated ground conductor sized in accordance with National, State, and local codes shall be installed between the facility vault and the MRI System ground bus location in the power distribution unit.
Magnetic Shielding The MRI unit is equipped with magnetic shielding. The exclusion zone for cardiac pacemakers, neurostimulators, and other biostimulation devices is recommended at 5 gauss (0.5mT). Signs provided by GE are posted to the mobile unit at the time of manufacture; the purpose of these signs are to alert all who approach the unit of the 5 gauss (0.5mT) requirement. These appropriate warning signs are exclusively and permanently attached to the scan room doors.
The 1.0T and 1.5T magnet systems exclusion zone (5) gauss is restricted to within 6” of the exterior of the mobile unit.
R.F. Shielding The R.F. shield is included with the MRI system and will provide minimum level of attenuation:
100Mhz, Min.85db (plane wave).
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Mobile Grounding Requirements
Special Grounding Note:
The unit must have an earth driven ground rod within five (5) feet of the hospitable power receptacle. A grounding cable of a minimum #1/0 AWG must be connected between the grounding rod and the grounding pin of the hospital power receptacle. Another cable to be kept as short as possible, may also be connected between the ground stud on the Incoming Power Distribution Panel and an earth driven ground rod See Figure 1: Ground Connection below. A separate grounding conductor must still be run with the phase conductors to the source of power from the grounding pin of the hospital power receptacle in accordance with NEC 2002 Article 250-24.
Figure 1: Ground Connection
NOTE: For the associated drawing please refer to the following chart for grounding requirements.
Ground Stud
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Mobile Grounding Chart
RUSSEL STOLL
DS2504FP
FEMALE
RECEPTACLE
OR EQUIVALENT
DISCONNECT LOCATED
AT VAN CONNECTION
150 AMP
FEEDER
GROUND BUS BAR
FROM
FACILITY
SERVICE
EQUIPMENT
L1
L2
L3
N
GN
G
G
L1
L2
L3 N
CHASSIS GND
SERVICE EARTH
MAIN GROUNDING
AT FACILITY
SERVICE
EQUIPMENT
NOTE:
- ALL WORK TO BE DONE IN ACCORDANCE WITH NATIONAL AND LOCAL ELECTRICAL CODES INFORMATION SHOWN HERE IS ONLY A RECOMMENDATION. MUST BE VERIFIED FOR SITE NATIONAL AND LOCAL CODES.
- GROUND WIRES INSIDE ENCLOSURES TO BE TAPED GREEN FOR ENTIRE VISUAL LENGTH FOR IDENTIFICATION.
MAIN BONDING JUMPER BETWEEN GROUNDED (NEUTRAL) CONDUCTOR AND EQUIPMENT GROUNDING CONDUCTOR TO BE PROVIDED IN FACILITY SERVICE EQUIPMENT AND DOWNSTREAM AT SEPARATELY DERIVED SYSTEM TRANSFORMER
SECONDARY AS SHOWN.
GROUNDING
GROUNDING WIRE
Harry E. Rauworth Debra C. Balis April 22, 1999
MOBILE GROUNDING REQUIREMENTS
CHASSIS GND
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Telephone and Data Service Requirements
Telephone Service The mobile unit is supplied with three (3) telephone connections.
The connector type that is used and supplied by Advanced Mobility Specialty Vehicles is an all-weather Hubbell PH-6595 (inlet) with a model PH-6624 connector body.
The customer is required to purchase and install three (3) all weather Hubbell PH-6597 phone outlets for use at the site.
Three all-weather Hubbell PH-6599, 50’-0” telephone connecting cables are included with the unit.
If additional cables are required, the customer must purchase them.
Data Service The mobile unit is supplied with three data line connections that utilize RJ-45 outlets.
The customer is required to purchase the data connection cables for use with the data line connections. The data line connections each require a 50’-0” CAT-6 cable with RJ-45 connections.
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents
Water and Chiller Requirements
During winter conditions, provisions must be made to ensure that water lines do not freeze because of weather conditions.
Humidifier Water Fill The unit contains a water storage tank for the humidifier. This tank is located in the equipment room and must always contain water to insure the specified humidity level remains constant. There are two options for filling the tank:
• A ¾” garden hose coupling male threaded hose connection is located outside under the equipment room on the left side front near the fifth wheel.
• A fill port is located in the humidifier for manual fill capability.
Water Supply Requirements (Sink Option) The mobile unit will need a ¾” diameter, 20’-0” long hose terminated with a ¾” I.P.S male threaded hose connector located on the left side of the mobile unit. The facility must provide a ¾” female connecter and a water supply that meets the following specifications:
• A flow rate of 5 gallons per minute.
• 45-60 PSI.
• A maximum temperature of 70°F.
Waste Water Connections (Sink Option) A 1½” IPS male connection is required to sink wastewater.
Any pipes and drains within 20’-0” of the magnet iso-center should be of non-ferrous materials such as PVC, copper, or brass. All plumbing must comply with all applicable codes.
The mobile unit will need a 20’-0” long 1½” diameter hose that is terminated with a 1½” male threaded connector for sanitary wastewater drainage located on the left side of the mobile unit. The facility must provide means of sanitary wastewater drainage from the system, which complies with all local applicable codes.
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Figure 2: Plan View and Pad Layout
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Figure 3: Right Side Elevation
PASSENGER SIDE ELEVATION
ROLL
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Figure 4: Left Side Elevation
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Figure 5: Stair/Lift/Wall Elevation
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Figure 6: Russellstoll Service Outlet
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Figure 7: Russellstoll Chart
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Figure 8: Crouse Hinds Service Outlet (Optional)
O P T I O N A L
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
Figure 9: Crouse Hinds Chart (Optional)
O P T I O N A L
The information contained in this booklet is the property of AMST. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.
A minimum dimension of 79” is required from rearmost projection to centerline of kingpin. This provides swing clearance for generator set which is mounted on the front of the trailer. Hospital is responsible to ensure the access route is clear of obstructions when the trailer is scheduled to arrive or depart. The 50’ minimum outside turning radius shown here has been calculated using an international harvester (Navistar) tractor Model COF- 9670 with a 161” wheelbase. Turning radius will vary with towing tractor.
Customer must confirm the turning radius on their tractor and prepare each site with adequate space to accommodate it.
Figure 10: Turning Requirements