The FDA's Center for Devices and Radiological Health (CDRH), Office of Science and Engineering Laboratories (OSEL), is conducting market research to identify qualified sources for one portable hyperspectral camera with associated accessories, software, documentation, warranty, and technical support by August 7, 2026. This is a sources-sought notice preceding formal solicitation; no set-aside is anticipated.

The system must operate as a standalone unit with integrated display and user interface, requiring no external computer for normal operation. Minimum technical requirements include spectral wavelength range of 400 nm to 800 nm, 200 or more evenly distributed spectral bands, spectral bandwidth (FWHM) of 7 nm or less, image output of at least 512×512 pixels, minimum spectral resolution of 12 bits per pixel, and focus capability from approximately 15 centimeters to infinity. The camera must operate using internal rechargeable battery, support standard 120V, 60 Hz U.S. power charging, store collected data on portable media, and support data transfer to external computers.

The procurement supports FDA research activities for developing bench testing methods to evaluate the safety and effectiveness of digital pathology devices. Place of performance is FDA facilities. The requirement is classified under NAICS 333310 (Commercial and Service Industry Machinery Manufacturing) and PSC 6650 (Optical Instruments, Test Equipment, Components and Accessories). Vendors must submit responses in native Microsoft Excel format by the August 7 deadline. Small business respondents must address domestic manufacture, availability, country of origin, domestic content percentage, COTS status, production capacity, and supply chain constraints.

Notice text

The U.S. Food and Drug Administration (FDA), Center for Devices and Radiological Health (CDRH), Office of Science and Engineering Laboratories (OSEL), Division of Imaging, Diagnostics and Software Reliability (DIDSR), is conducting market research to identify qualified sources capable of providing a portable hyperspectral camera for scientific research and laboratory testing. The required system will support FDA research activities by enabling the development of bench testing methods used to evaluate the safety and effectiveness of digital pathology devices. The Government seeks a commercially available portable hyperspectral imaging system capable of capturing high-resolution spectral imaging data for laboratory and research applications.

The anticipated requirement consists of one (1) portable hyperspectral camera and associated commercial accessories, software, documentation, warranty, and technical support necessary for full operation. At a minimum, the system shall be capable of meeting the following salient characteristics:

  • The anticipated requirement consists of one (1) portable hyperspectral camera and associated commercial accessories, documentation, warranty, and technical support necessary for full operation. At a minimum, the system shall be capable of meeting the following salient characteristics: 
  • Operate over a spectral wavelength range of 400 nm to 800 nm. 
  • Provide 200 or more evenly distributed spectral bands. 
  • Provide a spectral bandwidth (FWHM) of 7 nm or less. 
  • Produce image output of at least 512 × 512 pixels. 
  • Provide a minimum spectral resolution of 12 bits per pixel. 
  • Focus on objects from approximately 15 centimeters to infinity. 
  • Operate as a standalone system with an integrated display and user interface without requiring an external computer for normal operation. 
  • Store collected data on portable media for transfer to external computers. 
  • Operate using an internal rechargeable battery. 
  • Utilize standard 120V, 60 Hz U.S. power for battery charging. 
  • Weigh no more than 2 kilograms. 
  • Include a minimum one-year manufacturer's warranty and technical support. 
  • Be delivered within approximately six (6) weeks after award. 
  • On-site installation is not required. Training is optional. Electronic or hard-copy user documentation shall be provided. 

This Sources Sought Notice is issued solely for market research purposes in accordance with FAR Part 10 to determine the availability of capable small and large business concerns. It does not constitute a solicitation or a commitment by the Government to issue a solicitation or award a contract. Responses will be used to assess market capability, available sources, and potential small business participation.

Files

File Type Posted
NEW_Sources_Sought_SAM_Attachment_V1_1_260621.docx DOCX document
Sources_Sought_Vendor_Response.xlsx XLSX spreadsheet

Similar open opportunities

Opportunity Agency Set-aside Responses due
Sources Sought for (ICP-OES) Spectrometer System with an Autosampler Annapolis, Maryland 21402, United States · NAICS 334516 Naval Supply Systems Command No set-aside
Goldeye Camera Washington, District of Columbia 20375, United States · NAICS 334516 Office of Naval Research Total Small Business
Camera system, surveill NAICS 333310 Construction and Equipment No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity