J065--Bio/Fume Hood Testing & Certification Source Sought Notice

Closed Pre-Solicitation Posted

Solicitation number
36C25025Q0318
Agency
Responses due
Set-aside
No set-aside

Opportunity facts

Place of performance
Detroit VA Medical Center 4646 John R. Street This Sources Sought Notice, United States

Notice details come from SAM.gov. Updated .

About this opportunity

The Department of Veterans Affairs Veterans Health Administration is seeking a contractor to provide comprehensive bio/fume hood testing and certification services at the Detroit VA Medical Center. The sources sought notice, identified by solicitation number 36C25025Q0318, requires a qualified vendor to perform semi-annual and annual testing of biological, chemical, and pharmacy hoods in various departments including PALMS, Pharmacy, Nuclear Medicine, and Oncology. Contractors must submit a capabilities statement, SAM UEI, company information, socioeconomic status, and GSA contract number (if applicable) by March 3, 2025, at 10:00 AM Eastern Time via email to Jeanne Ortiz. The testing services must meet VHA standards, Joint Commission, CETA guidelines, and USP 797/800 standards, with comprehensive evaluations including airflow tests, HEPA filter integrity, smoke pattern analysis, particle count surveys, viable air and surface sampling, and environmental monitoring. Responses will be voluntary, and the government will not compensate vendors for information submission.

The procurement is not set aside for any specific socioeconomic category, with the anticipated NAICS code 541380 suggesting professional, scientific, and technical services. The contract is structured as a firm-fixed-price agreement with a base year and four potential option years, to be executed Monday through Friday between 6:00 AM and 4:30 PM. Work must be completed within five business days of starting, with detailed reports required within 14 days of each inspection. Contractors will be responsible for providing all necessary personal protective equipment and must demonstrate the capability to perform precise technical testing across multiple hood types and locations. While no specific budget range is mentioned, the complexity of testing requirements suggests a substantial technical and logistical undertaking, with strict performance standards and comprehensive reporting expectations.

Notice text

This is a Sources Sought Notice for biological, chemical and pharmacy hood testing and certification services at the Detroit VA Medical Center in Detroit, Michigan. Contractor shall furnish all necessary labor, supervision, management, management support, transportation, supplies, equipment and materials necessary to furnish the following services.

The services required are provided below and subject to changes should a solicitation be issued as a result of this notice.

Please respond if you have the ability and intent to complete this requirement. Include the following information in your response:

SAM UEI (include within email response and in capability statement)

Capabilities Statement

Company name and address

Point of contact information (name, phone, email)

Socioeconomic size status (example: large, small, WOSB, SDVOSB, etc.)

FSS or GSA contract number, if applicable.

This Sources Sought Notice is intended to locate any business capable of performance.

The anticipated NAICS Code for this requirement is 541380.

This is not a Request for Quotation and quotes shall not be reviewed at this time.

The Government shall not be responsible for any costs incurred by the contractor because of this Sources Sought Notice.

This notice shall not be construed as a commitment by the Government to ultimately award a contract, nor does it restrict the Government to any particular acquisition approach. All information submitted in response to this announcement is voluntary; the Government will not pay for information requested nor will it compensate any respondent for any cost incurred in developing information provided to the Government.

All responses shall be submitted via email by Monday, March 3, 2025 at 10:00 AM Eastern Time (ET). Responses shall be submitted to Jeanne Ortiz at jeanne.ortiz@va.gov and shall include 36C25025Q0318 in the email subject line. PHONE CALLS SHALL NOT BE ACCEPTED FOR THIS NOTICE.

DRAFT STATEMENT OF WORK

GENERAL INFORMATION

The Detroit VA (DVA) Medical Center at 4646 John R. Street, Detroit, Michigan 48201, requires Hood Testing and Certification services. The resulting contract will be firm, fixed-price with a base year and four (4) anticipated option years, at Government discretion.

SCOPE OF WORK

Semi-Annual: The contractor shall provide all resources necessary to perform and complete the tasks described in this statement of work (SOW), except as may otherwise be specified. This Semi-Annual testing will be done in the calendar year months of May and November. The contractor shall provide all labor, materials, instrumentation, supervision, and all other resources required to perform semi-annual testing and certifications on the following hoods:

These rooms are in PALMS, Pharmacy, Nuclear Medicine, and Oncology

Annual: The contractor shall provide all resources necessary to perform and complete the tasks described in this statement of work (SOW), except as may otherwise be specified. The Annual testing will be done in the calendar year month of November. The contractor shall provide all labor, materials, instrumentation, supervision, and all other resources required to perform annual testing and certifications on:

Background Semi-Annual: Semi-Annual hoods and the IV Clean room are to be tested and certified for proper airflows on an annual and semi-annual basis. The contractor shall test and certify all hoods and the IV Clean room within the John D. Dingell VAMC. The contractor shall perform Semi-Annual testing in the calendar year months of November and May for the testing and certifications. The contractor will schedule the work to ensure there is not more than a 6-month overlap of inspections under the contract period of performance.

Background Annual: Annual hoods are to be tested and certified for proper airflows on an annual basis. The contractor shall test and certify all hoods within the John D. Dingell VAMC. The contractor shall perform the Annual testing in the calendar year month of November for testing and certifications.

Performance Period: The contractor shall complete the work required under this SOW within the month of November and again in the month of May under the contract period of performance. The work shall take place between the hours of 6:00 am to 4:30 pm, Monday through Friday. Some critical areas may require testing to be scheduled a day ahead of time in order to gain access. This will be scheduled through the COR. The contractor shall complete all testing within five business days from the start date of the initial work.

SPECIFIC MANDATORY TASK AND ASSOCIATED DELIVERABLES

The contractor shall provide the specific deliverables below within the performance period stated in Section B-5 of this SOW. The contractor shall begin the first semi-annual testing/certification of all the hoods and IV clean room in the month of November of the contract period of performance and complete the work within five business days from when the work was started. The contractor shall begin the second semi-annual testing/certification of all the hoods and IV clean room in the month of May of the contract period of performance and complete the work within five business days from when the work was started. The contractor must provide the Contracting Officer Representative (COR) at least one week notice before starting any testing. The contractor shall communicate with the COR daily during the inspections to inform of any issues with hoods tested that day. The contractor shall notify the COR of any hoods that do not meet certification criteria and supply the COR with a quote for repairs needed to make the hood compliant, prior to leaving the facility.

Testing of the Pharmacy IV clean room shall be done on a semi-annual schedule to coincide with the hood testing. The following testing and sampling are required:

Airflow Test

Citations: USP , CAG-001-2005, CAG-002-2006, Manufacturer Specifications

Methods: CAG-002-2006

Locations: LAFW

Procedure: Appropriate airflow parameters will be measured using a validated method which will be kept on file. Air velocity will be measured by thermal anemometer using equally spaced readings across a horizontal plane producing the most repeatable readings. The locations within the plane, and the level of the plane in relationship to the filter diffuser will be documented. Reported in feet per minute with % uniformity.

Pass/Fail Criteria:

PASS Average velocity and uniformity within manufacturer specifications

FAIL Average velocity and uniformity outside manufacturer specifications

HEPA Filter Integrity Test

Citations: USP , CAG-002-2006

Methods: CAG-002-2006, IEST-RP-CC034.2 Section 6.2.1 for filters that can be scan tested IEST-RP-CC034.2 Section 6.2.3 for filters that cannot be scan tested

Locations: LAFW

Procedure: HEPA filters will be leak tested using a validated method which will be kept on file. An aerosol challenge of 10-90 micrograms per liter of polydispersed dioctylphthalate, polyalphaolefin, or equivalent will be introduced upstream and filter efficacy verified by aerosol photometer scan. Certifier shall have capability to patch and repair HEPA filter up to 3% of the face area per side and not to exceed 1.5 inches in width or provide replacement HEPA filter with filter of like size and construction (Type C or K) on same day.

Pass/Fail Criteria:

PASS leakage of less than or equal to 0.01% of upstream concentration for filters that can be scanned, or 0.005% for filters that cannot be scanned

FAIL leakage of greater than 0.01% or patch size exceeds IEST-RP-CC034 (current version) limit

Airflow Smoke Pattern Test

Citations: USP , CAG-001-2005, CAG-002-2006

Methods: CAG-002-2006

Locations: LAFW

Procedure: Visual observation of airflow patterns, under dynamic operating conditions, demonstrating first-air flow through critical sites within the DCA. Minimal dead spots or reflux away from critical sites is allowable. Methods will be kept on file. Delivery velocity of smoke generator shall not overcome isolator airflow patterns. Results to be video recorded and provided to Facility Program Manager for IV/Sterile Preparations.

Pass/Fail Criteria:

PASS Airflow demonstrates smooth downflow with no dead spots or refluxing across critical sites.

FAIL Airflow does not meet direction or turbulence standards

Particle Count Survey

Citations: USP , CAG-001-2005, CAG-002-2006

Methods: CAG-002-2006, ISO 14644-1

Locations: LAFW

Procedure: Air particle counts will be measured using a validated method which will be kept on file. The pass-through will be tested under static (at-rest) condition and the main chamber under both static and dynamic operating conditions. A discrete particle counter capable of detecting particles of 0.5 micron and larger shall be used. Sampling probe is to be oriented upward and located no more than 12 above the work surface (static test) and 12 away from the critical site (dynamic test). A minimum sample of two liters of air over one minute will be taken for each testing location. Three particle counts will be taken from the geometric center of the pass-through and five particle counts each will be taken in the main chamber under static and dynamic conditions one near each corner,

and one in the geometric center. Particle counts will be reported individually for each sampling location and designated by location and testing condition.

Pass/Fail Criteria (each PEC):

PASS ISO Class 5 or better for all samples: less than or equal to 3,520 particles (0.5 micron and larger) per cubic meter

FAIL greater than 3,520 particles (0.5 micron and larger) per cubic meter for any sample

Viable Air Sampling

Citations: USP , USP , CAG-009-2011

Methods: CAG-009-2011

Locations: One of each bacterial and fungal near the geometric center of the work surface in the LAFW and outside the LAFW.

Procedure: All samples shall be collected under dynamic operating conditions by volumetric impaction. Bacterial samples will be collected using 55mm plates containing TSA with neutralizing agents (lecithin and polysorbate 80). Samples will be incubated inverted at 30-35° C for 48-72 hours. Fungal samples will be collected using 55mm plates containing MEA or SDA with neutralizing agents (lecithin and polysorbate 80). Samples will be incubated at 26-30° C for 5-7 days. Colony forming units will be assessed daily and counts will be adjusted per manufacturer MPN calculation or statistical correction factor and reported as CFU per cubic meter.

Pass/Fail Criteria:

PASS 1CFU per plate and no highly pathogenic organisms

FAIL >1 CFU per plate or 1 CFU of highly pathogenic organisms

Highly pathogenic organisms defined as gram-negative rods, coagulase-positive staphylococci, molds and yeasts

Pass/Fail Criteria (Outside CAI):

PASS 100 CFU per plate

FAIL >100 CFU per plate

Viable Surface Sampling

Citations: USP , USP , CAG-009-2011

Locations: One of each bacterial and fungal near the geometric center of the work surface in the LAFW, and outside the LAFW.

Methods: CAG-009-2011

Procedure: All samples shall be collected under dynamic operating conditions using convex agar plates by gently rolling across the surface once in a single direction. Surface to be disinfected by wiping with non-shedding wipe saturated with sterile 70% isopropyl alcohol (provided by facility) immediately after each sample is collected. Bacterial samples will be collected using 55mm plates containing TSA with neutralizing agents (lecithin and polysorbate 80). Samples will be incubated inverted at 30-35° C for 48-72 hours. Fungal samples will be collected using 55mm plates containing MEA or SDA with neutralizing agents (lecithin and polysorbate 80). Samples will be incubated at 26-30° C for 5-7 days. Colony forming units will be assessed daily and reported as CFU per plate.

Pass/Fail Criteria (CAI):

PASS 3 CFU per plate and no highly pathogenic organisms

FAIL >3 CFU per plate or 1 CFU of highly pathogenic organisms

Highly pathogenic organisms defined as gram-negative rods, coagulase-positive staphylococci, molds and yeasts

Pass/Fail Criteria (Outside CAI):

PASS 100 CFU per plate

FAIL >100 CFU per plate

Temperature and Humidity Tests: Ante Room, I.V. Room, Air Lock, and Chemo Room

Purpose: The purpose of these tests is to demonstrate the capability of the system to control temperature uniformity and humidity within buyer-specified limits.

Sound Level Tests: Ante Room, I.V. Room, Air Lock, and Chemo Room

Purpose: The purpose of this test is to establish the airborne sound pressure levels produced by the basic Cleanroom mechanical and electrical systems as experienced within the room and in adjacent external occupied areas, and to verify that performance meets the Buyer's specifications.

Light Level Tests: Ante Room, I.V. Room, Air Lock, and Chemo Room

Purpose: The purpose of this test is to verify that the specified lighting levels and uniformity of lighting within the Cleanroom has been met.

Room Pressurized Tests: Ante Room, I.V. Room, Air Lock, and Chemo Room

Purpose: To verify the capability of the Cleanroom system to maintain the specified pressure differential in the Cleanroom.

Locations of all areas and test locations will be required for each individual test.

CONFORMANCE STANDARDS

All testing of equipment, hoods, and areas shall meet any VHA standards, Joint Commission for Accreditation of Hospital Organization (JCAHO) or National Codes, Controlled Environment Testing Association (CETA) Sterile Compounding Facilities (or equivalent) guidelines, and USP 797/800 standards.

HOURS OF WORK

The Hood testing and Pharmacy IV clean-room testing will be performed Monday through Friday between the hours of 6:00 am to 4:30pm. Any time changes must be cleared thru the acting COR of the contract. All Federal Holidays are observed Detroit VA Medical Center, and any proposed work during these days must be cleared through the COR or Facilities Maintenance Service Chief.

GOVERNMENT RESPONSIBILITIES

The Government will provide access to rooms where hoods are located so contractor can perform testing/ certifications. The Government will supply room maps for all areas of work. The COR will provide an

escort for Contractor, if needed, to find rooms and locations.

G. SAFETY

The contractor shall provide all personal protective equipment (PPE) needed for their employees such as safety glasses, hearing protection, safety shoes, gloves and any other PPE.

H. REPORTS

The contractor shall be responsible for submitting detailed reports within 14 days of completion of each semi-annual inspection, for each type of hood. The report must include the type of hood, ID number, EE number, room number, air flow data and hood pass/fail.

Files

File Type Posted
36C25025Q0318.docx DOCX document

Similar open opportunities

Opportunity Agency Set-aside Responses due
J073--FY 27 Services | Kitchen Exhaust Hood and Ductwork Cleaning | Base + 4 | 561 East Orange VA Medical Center 385 Tremont Avenue East Orange, New Jersey 07018, United States · NAICS 811310 Veterans Integrated Service Network 2 No set-aside
H379-- Facility Hood Cleaning and Inspection for Nothern Arizona VA Healthcare System Department of Veterans Affairs Network Contracting Office 22 Prescott AZ 86313 · NAICS 561790 Veterans Integrated Service Network 22 Service-Disabled Veteran-Owned Small Business
Sterile Compounding Services Texas 78229, United States · NAICS 541380 Veterans Integrated Service Network 17 Service-Disabled Veteran-Owned Small Business
72-inch laboratory chemical fume hood system Washington, District of Columbia 20375, United States · NAICS 33399 Office of Naval Research No set-aside
6550--Cepheid GeneXpert Testing Supplies Beckley Vamc Clarksburg Vamc Vamhcs Washington Vamc · NAICS 325413 Veterans Integrated Service Network 5 No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity