Paragraph 1:

The Department of Veterans Affairs, Network Contracting Office (NCO) 6 is conducting market research to identify qualified businesses capable of providing an In-Vivo Radiation Measurement System. The system must perform real-time monitoring of radiation dose using small detectors that can be placed on a patient's skin or in a cavity, and determine the radiation dose received by cardiac devices like pacemakers. Key requirements include supporting a variety of radiation detectors covering a range of photon and electron energies, with a minimum of 6 measurement channels and wireless connectivity between the detectors and monitoring station. Respondents must provide information on their small business status, interest in providing a Federal Supply Schedule quote or open market offering, ownership of related businesses, active SAM registration, and a capabilities statement. Responses are due by May 29, 2024 at 11:00 AM EST.

Paragraph 2:

This opportunity is not set aside for any specific business size or type. The Pre-solicitation notice does not identify any potential incumbents or current contractors. The notice does not provide any information on the anticipated award value or budget range for this procurement. The key numerical details provided are the requirement for a system that can measure doses up to 30 Gy per detector, using detectors no larger than 2mm x 2mm.

Notice text

The Department of Veterans Affairs, Network Contracting Office (NCO) 6 on behalf of Richmond VAMC is conducting Market Research through this Sources Sought notice to help determine the availability of qualified Businesses, capable of providing In-Vivo Radiation Measurement System that meet the following:

Functions to be performed;

System for real-time monitoring of radiation dose using small radiation detectors that can be placed on the patient s skin or in a cavity. Also used to determine radiation dose received by cardiac devices (pacemakers, etc.) for evaluation of possible radiation damage to the device.

Performance required;

System must support a variety of radiation detectors covering a range of photon energies (6MV 18MV) and electron energies with both directional and angularly independence.

Must support a minimum of 6 channels to allow multiple radiation measurements to be acquired in a single treatment.

Must support a wireless connection between a remote connection for the measurement devices and a monitoring station.

Essential physical characteristics;

Remote station must be rechargeable.

System must be able to measure doses up to 30 Gy per detector.

Must include detectors no larger than 2mm x 2mm for point measurements.

All respondents shall indicate/provide the following:

Whether or not they are small business manufacturers of the products they wish to offer, or if they are a reseller of a small business manufacturer.

Is your company interested in providing a Federal Supply Schedule (FSS) quote or Open Market (indicate FSS contract #)?

Identify any other companies owned (wholly or in-part) by the owners of this business who provide goods or services under the same or a related NAICS codes.

Identify any other companies owned (wholly or in-part) by the owners of this business who provide goods or services that are registered under the same or a related NAICS code with the Center for Veterans Enterprise (CVE); Provide the certification type (SDVOSB/VOSB), Federal Identification Number, and state of incorporation for each.

Whether or not your company has an active SAM registration and the companies Unique Entity Identifier.

A copy of its capabilities statement.

Failure to provide the above information may result in the potential offeror being considered incapable of meeting the VA s needs.

If the contractor can provide the above items, the contractor shall provide a letter from the manufacturer stating they are an authorized distributor/licenser of their products. Failure to submit an authorized distributor letter shall result in the product(s) potentially offered to be considered grey market and shall not be considered technically acceptable.

The capability statements received from this market research are for planning purposes only and will assist the Government in planning its acquisition strategy. This is strictly market research, and the Government will not entertain questions concerning this market research. The Government will not assume any costs incurred by the contractor in the preparation of responses. All responses shall be submitted in writing no later than May 29, 2024, at 11:00 AM EST to the attention of: Natasha.Hawkins@va.gov

Please title the subject line for all responses as: In-Vivo Radiation Measurement System 36C246-24-AP-4038

Files

File Type Posted
36C24624Q0904.docx DOCX document

Notice history

Notice Type Posted
6515--In-Vivo Radiation Measurement System Latest solicitation Solicitation
This notice Original Pre-Solicitation

Similar open opportunities

Opportunity Agency Set-aside Responses due
H365--460-27-1-132-0001 | SPE 49273 | VISN 4 Radiation Safety Officer (RSO) | NCO 4 Services 3 (VA-27-00003148) 9 VA Medical Facilities In the NCO 4 Service Area · NAICS 541380 Veterans Integrated Service Network 4 Service-Disabled Veteran-Owned Small Business
Radiological Detection System Supplies NAICS 334519 Naval Sea Systems Command No set-aside
J065--FY27 (BASE 4) RaySearch America RayStation Treatment Planning System Service Contract NAICS 541519 Veterans Integrated Service Network 2 No set-aside
6515--NX EQ Image Guided Surgical Navigation System (VA-26-00083972) Nationwide, United States · NAICS 334510 VA Strategic Acquisition Center Frederick Service-Disabled Veteran-Owned Small Business
6505--FY27 Low Energy General Radiopharmaceuticals 36C25027Q0052 See Section 11 of the Cleveland and Kent, Ohio, United States · NAICS 325412 Veterans Integrated Service Network 10 No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity