ATTACHMENT 1
Chemical Parameters, Matrix and Detection Limits
Alpha Spectrometry Isotope Matrix Detection Limit Minimum Sample Size Americium (Am-241) Urine 0.05 pCi/L 1 Liter
Fecal 0.0005 pCi/g 50 grams Plutonium (Pu-238, Pu- 239, Pu-240, Pu-242)
Urine 0.05 pCi/L 1 Liter
Fecal 0.0005 pCi/g 50 grams Water 0.05 pCi/L 1 Liter
Uranium (U-238, U-235, U-234)
Urine 0.05 pCi/L 1 Liter
Fecal 0.0005 pCi/g 50 grams Water 0.05 pCi/L 1 Liter Soil 0.05 pCi/g 50 grams Sediment 0.05 pCi/g 50 grams Air Filter 0.0005 pCi/L air 1 filter Animal Tissue 0.005 pCi/g 20 grams Vegetation 0.01pCi/g 20 grams
Thorium (Th-230, Th-232, Th-228)
Urine 0.05 pCi/L 1 Liter
Fecal 0.0005 pCi/g 50 grams
Gas Flow Proportional Counters Isotope Matrix Detection Limit Minimum Sample Size Gross Alpha Water 3 pCi/L 1 Liter Gross Beta Water 4 pCi/L 1 Liter Strontium (Sr-90) Urine 5 pCi/L 1 Liter
Water 1 pCi/L 1 Liter Radium (Ra-228) Water 1 pCi/L 1 Liter
Liquid Scintillation Counter Isotope Matrix Detection Limit Minimum Sample Size Tritium Water 1000 pCi/L 0.1 Liter
Urine 0.03 uCi/L 50 milliliters
Scintillation Counter Isotope Matrix Detection Limit Minimum Sample Size Radium (Ra-226) Water 1 pCi/L 1 Liter
Urine 0.003 uCi/L 1 Liter Fecal 0.05 pCi/g 50 grams
Inductively Coupled Plasma-Mass Spectrometry Analysis (Mass determination) Isotope Matrix Detection Limit Minimum Sample Size Uranium (U-238, U-235) Water 0.5 ug/L 0.5 Liter
Soil 0.1 ug/g 50 grams Sediment 0.1 ug/g 50 grams Air Filter 0.1 ug/filter 1 filter Wipe 0.1 ug/wipe 1 wipe Urine 5 ng/L 1 Liter
Gamma Spectroscopy Isotope Matrix Detection Limit Minimum Sample
Size Gamma (various isotopes) Water 1 pCi/L 1 Liter
Urine 1 pCi/L 1 Liter Fecal 0.05 pCi/g 50 grams Soil 0.05 pCi/g 50 grams
TCLP for Analysis of TCLP Extract for Metals for samples containing radioactive materials PARAMETER METHOD Limit of
Quantitation (LOQ), mg/L
Minimum Sample Volume
Arsenic (As) 6010x or 6020xB 0.5 100 grams Barium (Ba) 6010x or 6020x6010B 10 100 grams Cadmium (Cd) 6010x or 6020x6010B 0.1 100 grams Chromium (Cr) 6010x or 6020x6010B 0.5 100 grams Lead (Pb) 6010x or 6020x6010B 0.5 100 grams Mercury (Hg) 7471xA 0.02 100 grams Selenium (Se) 6010x or 6020x6010B 0.1 100 grams Silver (Ag) 6010x or 6020x6010B 0.5 100 grams
Notes:
All Fecal and urine samples shall be reported in compliance with ANSI N13.30, Section
4.2 reporting requirements.
For all water analyses: methods used for analysis must be from approved methods listed in Title 40, CFR, Part 141.25 and 141.26 for Drinking Water Compliance. Methods must be specified and provided for review. Method substitutions must be approved prior to analysis.
All soil analyses for uranium and uranium ratios shall be prepared using a complete dissolution method to include hydrofluoric acid digestion. Preparation methods using a leach method will not be acceptable for these parameters and will be reanalyzed at contractor expense. Specify if a separate additional cost is required.
Detection Limits are highly matrix dependent; however, samples will be counted for the maximum count time to reasonably achieve specified detection level requirements.