ATTACHMENT 1

Chemical Parameters, Matrix and Detection Limits

Alpha Spectrometry Isotope Matrix Detection Limit Minimum Sample Size Americium (Am-241) Urine 0.05 pCi/L 1 Liter

Fecal 0.0005 pCi/g 50 grams Plutonium (Pu-238, Pu- 239, Pu-240, Pu-242)

Urine 0.05 pCi/L 1 Liter

Fecal 0.0005 pCi/g 50 grams Water 0.05 pCi/L 1 Liter

Uranium (U-238, U-235, U-234)

Urine 0.05 pCi/L 1 Liter

Fecal 0.0005 pCi/g 50 grams Water 0.05 pCi/L 1 Liter Soil 0.05 pCi/g 50 grams Sediment 0.05 pCi/g 50 grams Air Filter 0.0005 pCi/L air 1 filter Animal Tissue 0.005 pCi/g 20 grams Vegetation 0.01pCi/g 20 grams

Thorium (Th-230, Th-232, Th-228)

Urine 0.05 pCi/L 1 Liter

Fecal 0.0005 pCi/g 50 grams

Gas Flow Proportional Counters Isotope Matrix Detection Limit Minimum Sample Size Gross Alpha Water 3 pCi/L 1 Liter Gross Beta Water 4 pCi/L 1 Liter Strontium (Sr-90) Urine 5 pCi/L 1 Liter

Water 1 pCi/L 1 Liter Radium (Ra-228) Water 1 pCi/L 1 Liter

Liquid Scintillation Counter Isotope Matrix Detection Limit Minimum Sample Size Tritium Water 1000 pCi/L 0.1 Liter

Urine 0.03 uCi/L 50 milliliters

Scintillation Counter Isotope Matrix Detection Limit Minimum Sample Size Radium (Ra-226) Water 1 pCi/L 1 Liter

Urine 0.003 uCi/L 1 Liter Fecal 0.05 pCi/g 50 grams

Inductively Coupled Plasma-Mass Spectrometry Analysis (Mass determination) Isotope Matrix Detection Limit Minimum Sample Size Uranium (U-238, U-235) Water 0.5 ug/L 0.5 Liter

Soil 0.1 ug/g 50 grams Sediment 0.1 ug/g 50 grams Air Filter 0.1 ug/filter 1 filter Wipe 0.1 ug/wipe 1 wipe Urine 5 ng/L 1 Liter

Gamma Spectroscopy Isotope Matrix Detection Limit Minimum Sample

Size Gamma (various isotopes) Water 1 pCi/L 1 Liter

Urine 1 pCi/L 1 Liter Fecal 0.05 pCi/g 50 grams Soil 0.05 pCi/g 50 grams

TCLP for Analysis of TCLP Extract for Metals for samples containing radioactive materials PARAMETER METHOD Limit of

Quantitation (LOQ), mg/L

Minimum Sample Volume

Arsenic (As) 6010x or 6020xB 0.5 100 grams Barium (Ba) 6010x or 6020x6010B 10 100 grams Cadmium (Cd) 6010x or 6020x6010B 0.1 100 grams Chromium (Cr) 6010x or 6020x6010B 0.5 100 grams Lead (Pb) 6010x or 6020x6010B 0.5 100 grams Mercury (Hg) 7471xA 0.02 100 grams Selenium (Se) 6010x or 6020x6010B 0.1 100 grams Silver (Ag) 6010x or 6020x6010B 0.5 100 grams

Notes:

All Fecal and urine samples shall be reported in compliance with ANSI N13.30, Section

4.2 reporting requirements.

For all water analyses: methods used for analysis must be from approved methods listed in Title 40, CFR, Part 141.25 and 141.26 for Drinking Water Compliance. Methods must be specified and provided for review. Method substitutions must be approved prior to analysis.

All soil analyses for uranium and uranium ratios shall be prepared using a complete dissolution method to include hydrofluoric acid digestion. Preparation methods using a leach method will not be acceptable for these parameters and will be reanalyzed at contractor expense. Specify if a separate additional cost is required.

Detection Limits are highly matrix dependent; however, samples will be counted for the maximum count time to reasonably achieve specified detection level requirements.

Other isotopes of the same technologies may be added on a case by case basis with prior coordination between the contract lab and APHC (Prov).