The file contains Salient Characteristics for two specialized equipment pieces for the Department of Veterans Affairs Veterans Health Administration: an Ultra Low Temp RFID Freezer and an Ambient Cabinet, both designed for specimen storage and tracking. The Ultra Low Temp RFID Freezer must maintain a precise -80°C temperature, integrate with TrackCore system via RFID tracking, feature digital temperature monitoring, and have specific external dimensions of 78" x 23" x 38" and internal dimensions of 51.2" x 17.8" x 28.3". The Ambient Cabinet must maintain a temperature range of 12°C to 30°C, include RFID tracking capabilities, support device identifier barcode integration, and have external dimensions of 80" x 41" x 50" with internal side dimensions of 65" x 14" x 24". Both devices require electronic security features like door locks, temperature alarm systems, and connectivity for data transfer and monitoring. The solicitation (36C24725Q0716) is for commercial items under the NAICS code 493120 with a small business size standard of $36.5 Million, seeking quotes for these name-brand or equivalent medical storage and tracking systems.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Combined Synopsis-RFID Freezer Cabinet Combo.pdf | ||
| RFID Freezer and Cabinet QAs.docx | DOCX document | |
| RFID Freezer and Cabinet QAs.docx | DOCX document | |
| Combined Synopsis-RFID Freezer Cabinet Combo.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics – Ultra Low Temp RFID Freezer
• Must enable seamless integration with the TrackCore system/software via built-in RFID tracking functionality.
• Must include necessary connectivity features to facilitate data transfer to the TrackCore system for monitoring and logging purposes.
• Must reliably maintain a temperature of -80°C, with integrated temperature monitoring and alarm systems.
• Digital temperature readings should remain within a ±2°C range of the mercury (red-spirit filled) thermometer.
• External dimensions should approximate 78 inches in height, 23 inches in width, and 38 inches in depth to meet spatial requirements.
• Internal dimensions should be approximately 51.2 inches in height, 17.8 inches in width, and 28.3 inches in depth to accommodate specimen storage needs.
• Should feature adjustable shelving and drawer systems to support various sample sizes and types.
• Equipped with an electronic door lock to ensure security.
Salient Characteristics – Ambient Cabinet
• Must enable seamless integration with the TrackCore system/software through built-in RFID tracking capabilities.
• Must include the necessary connectivity features to support data transfer to the TrackCore system for monitoring and logging purposes.
• External dimensions should approximate 80 inches in height, 41 inches in width, and 50 inches in depth to meet spatial requirements.
• Internal dimensions on each side should measure approximately 65 inches in height, 14 inches in width, and 24 inches in depth to support specimen storage.
• Must maintain a consistent internal temperature range of 12°C to 30°C, with integrated temperature monitoring and alarm systems to ensure compliance.
• Should facilitate access tracking, including user identification and item movement records.
• Should support integration with unique device identifier barcodes for point-of-use scanning.
• Equipped with a locking mechanism to ensure security.