S02 DRAWINGS.pdf
PDF 14 MB Posted
- Solicitation number
- 36C24221B0032_1
About this file
This solicitation is for the renovation of approximately 900 square feet of existing linear accelerator (LINAC) room, control room, and corridor at the VA New York Harbor Healthcare System in Brooklyn, New York. The scope of work includes a full gut renovation of the spaces, fire suppression, electrical, plumbing, architectural, rigging, construction administration, demolition, asbestos abatement, fireproofing, metal framing, gypsum board, acoustical ceilings, flooring, painting, HVAC, medical gas, direct digital controls, wiring devices, communications, telephone/data, fire alarm, and other requirements. A site visit is scheduled for April 27, 2021 at 11:00 AM EST. This is a 100% set-aside for Service Disabled Veteran Owned Small Businesses. Interested contractors should review the attached solicitation for additional details on response dates and requirements. The soliciting agency is the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 2.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| AMENDMENT 0001.pdf | ||
| S02 Participation Goals for Minorities and Females.pdf | ||
| S02 SPECS.pdf | ||
| S02 VA ID PACKAGE 06-18-2019.pdf | ||
| S02 ALL POSTERS.pdf | ||
| S02 Safety or Environmental Violations and Experience Modification Rate.pdf | ||
| 36C24221B0032_1.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
TITLE
DESIGN ONCOLOGY LINAC SITE PREP
VA NY Harbor Healthcare System, Brooklyn Campus
MC TB
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
Fi le
P at h 6/
/2
:3
9:
P M
ISSUE FOR BID
Office of
Construction and Facilities Management
U.S. Department of Veterans Affairs
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT OF RECORD STAMPCONSULTANT
321 4 5 6 7 8 9 10
Project Director
23 MAR 2021Revisions: Date:
630A4-CSI-426
800 POLY PLACE, BROOKLYN, NY 11209-7104
DESIGN ONCOLOGY LINAC SITE PREP
THINKFORM DESIGN ARCHITECT LLC
38 EAST BROAD STREET
HOPEWELL, NJ 08525
T:855.821.0274 F:609.644.4397 Russell DiNardo, AIA NY 031521-1
VA NY HARBOR HEALTHCARE SYSTEM
BROOKLYN CAMPUSENGINEERING LLP
www.fellp.com
22 MULBERRY ST, SUITE 2A
MIDDLETOWN, NY 10940
T:845.343.1481 F:855.320.8735
21-03-BK
AutoCAD SHX Text X
AutoCAD SHX Text X
AutoCAD SHX Text
SECTION SHEET NUMBER
AutoCAD SHX Text
BUILDING ELEVATION SHEET NUMBER
AutoCAD SHX Text N
AutoCAD SHX Text X
AutoCAD SHX Text X
AutoCAD SHX Text E
AutoCAD SHX Text X
AutoCAD SHX Text X
AutoCAD SHX Text W
AutoCAD SHX Text S
AutoCAD SHX Text
DETAIL SHEET NUMBER
AutoCAD SHX Text X
AutoCAD SHX Text X
AutoCAD SHX Text
DETAIL NUMBER SHEET NUMBER
AutoCAD SHX Text
PARTITION TYPE
AutoCAD SHX Text
XX
AutoCAD SHX Text X
AutoCAD SHX Text X
AutoCAD SHX Text
WINDOW TYPE
AutoCAD SHX Text
REVISIONS
AutoCAD SHX Text
ROOM NAME ROOM NUMBER
AutoCAD SHX Text
AutoCAD SHX Text
KEYED NOTE
AutoCAD SHX Text
DATUM LINE
AutoCAD SHX Text
DOOR NUMBER
AutoCAD SHX Text
AutoCAD SHX Text
TITLE
AutoCAD SHX Text
AutoCAD SHX Text
INTERIOR ELEVATION SHEET NUMBER
AutoCAD SHX Text X
AutoCAD SHX Text
XX
AutoCAD SHX Text
SCALE
AutoCAD SHX Text 0-G001
AutoCAD SHX Text
TITLE SHEET
AutoCAD SHX Text GENERAL: 0-G001 - TITLE SHEET - TITLE SHEET TITLE SHEET 0-G002 - ABBREVIATIONS, CODE NOTES AND FIRE SAFETY/EGRESS - ABBREVIATIONS, CODE NOTES AND FIRE SAFETY/EGRESS ABBREVIATIONS, CODE NOTES AND FIRE SAFETY/EGRESS ABATEMENT: HA100 - GROUND FLOOR ASBESTOS ABATEMENT PLAN - GROUND FLOOR ASBESTOS ABATEMENT PLAN GROUND FLOOR ASBESTOS ABATEMENT PLAN HA101 - ASBESTOS DIAGRAMS AND NOTES - ASBESTOS DIAGRAMS AND NOTES ASBESTOS DIAGRAMS AND NOTES STRUCTURAL: S-100 - STRUCTURAL EXISTING CONDITIONS PLAN & NEW EQUIPMENT DIAGRAMS - STRUCTURAL EXISTING CONDITIONS PLAN & NEW EQUIPMENT DIAGRAMS STRUCTURAL EXISTING CONDITIONS PLAN & NEW EQUIPMENT DIAGRAMS S-101 - STRUCTURAL PARTIAL PLANS AT VAULT - STRUCTURAL PARTIAL PLANS AT VAULT STRUCTURAL PARTIAL PLANS AT VAULT S-201 - STRUCTURAL SECTIONS - STRUCTURAL SECTIONS STRUCTURAL SECTIONS S-400 - STRUCTURAL SPECIAL INSPECTIONS - STRUCTURAL SPECIAL INSPECTIONS STRUCTURAL SPECIAL INSPECTIONS ARCHITECTURAL: 1-AD101 - GROUND FLOOR DEMOLITION PLAN - GROUND FLOOR DEMOLITION PLAN GROUND FLOOR DEMOLITION PLAN 1-AS101 - GROUND FLOOR PLAN AND PARTITION TYPES - GROUND FLOOR PLAN AND PARTITION TYPES GROUND FLOOR PLAN AND PARTITION TYPES 1-AS102 - GROUND FLOOR REFLECTED CEILING PLAN AND DETAILS - GROUND FLOOR REFLECTED CEILING PLAN AND DETAILS GROUND FLOOR REFLECTED CEILING PLAN AND DETAILS 1-AS300 - DOOR SCHEDULE, FINISH SCHEDULE AND ROOM SECTION - DOOR SCHEDULE, FINISH SCHEDULE AND ROOM SECTION DOOR SCHEDULE, FINISH SCHEDULE AND ROOM SECTION 1-AS301 - MILLWORK DETAILS AND SIGNAGE - MILLWORK DETAILS AND SIGNAGE MILLWORK DETAILS AND SIGNAGE 1-AS600 - EQUIPMENT SCHEDULE AND RIGGING PLAN - EQUIPMENT SCHEDULE AND RIGGING PLAN EQUIPMENT SCHEDULE AND RIGGING PLAN MECHANICAL: MH-001 - MECHANICAL SYMBOLS, ABBREVIATIONS & NOTES - MECHANICAL SYMBOLS, ABBREVIATIONS & NOTES MECHANICAL SYMBOLS, ABBREVIATIONS & NOTES MD-101 - MECHANICAL REFLECTED CEILING & ROOF DEMOLITION PLANS - MECHANICAL REFLECTED CEILING & ROOF DEMOLITION PLANS MECHANICAL REFLECTED CEILING & ROOF DEMOLITION PLANS MH-101 - MECHANICAL REFLECTED CEILING & ROOF PLAN - MECHANICAL REFLECTED CEILING & ROOF PLAN MECHANICAL REFLECTED CEILING & ROOF PLAN MH-701 - MECHANICAL DETAILS - MECHANICAL DETAILS MECHANICAL DETAILS MH-702 - MECHANICAL DETAILS - MECHANICAL DETAILS MECHANICAL DETAILS MH-801 - MECHANICAL SCHEDULES - MECHANICAL SCHEDULES MECHANICAL SCHEDULES PLUMBING: PL-001 - PLUMBING SYMBOLS, ABBREVIATIONS & NOTES - PLUMBING SYMBOLS, ABBREVIATIONS & NOTES PLUMBING SYMBOLS, ABBREVIATIONS & NOTES PL-101 - PLUMBING SANITARY, MEDICAL GAS, & DOMESTIC WATER FLOOR PLANS - PLUMBING SANITARY, MEDICAL GAS, & DOMESTIC WATER FLOOR PLANS PLUMBING SANITARY, MEDICAL GAS, & DOMESTIC WATER FLOOR PLANS FIRE PROTECTION: FX-100 - FIRE PROTECTION SPRINKLER REFLECTED CEILING PLANS AND NOTES - FIRE PROTECTION SPRINKLER REFLECTED CEILING PLANS AND NOTES FIRE PROTECTION SPRINKLER REFLECTED CEILING PLANS AND NOTES ELECTRICAL: ES-001 - ELECTRICAL SYMBOLS, ABBREVIATIONS & NOTES - ELECTRICAL SYMBOLS, ABBREVIATIONS & NOTES ELECTRICAL SYMBOLS, ABBREVIATIONS & NOTES ED-101 - ELECTRICAL POWER DEMOLITION FLOOR & ROOF PLANS - ELECTRICAL POWER DEMOLITION FLOOR & ROOF PLANS ELECTRICAL POWER DEMOLITION FLOOR & ROOF PLANS ES-100 - ELECTRICAL BASEMENT, FIRST & PARTIAL FLOOR PLANS - ELECTRICAL BASEMENT, FIRST & PARTIAL FLOOR PLANS ELECTRICAL BASEMENT, FIRST & PARTIAL FLOOR PLANS ES-101 - ELECTRICAL POWER FLOOR & ROOF PLANS - ELECTRICAL POWER FLOOR & ROOF PLANS ELECTRICAL POWER FLOOR & ROOF PLANS ES-201 - ELECTRICAL LIGHTING PLANS - ELECTRICAL LIGHTING PLANS ELECTRICAL LIGHTING PLANS ES-600 - ELECTRICAL ONE-LINE RISER DIAGRAMS - ELECTRICAL ONE-LINE RISER DIAGRAMS ELECTRICAL ONE-LINE RISER DIAGRAMS ES-701 - ELECTRICAL DETAILS- ELECTRICAL DETAILSELECTRICAL DETAILS
AutoCAD SHX Text
DEPARTMENT OF VETERANS AFFAIRS
AutoCAD SHX Text
800 POLY PLACE, BROOKLYN, NY
AutoCAD SHX Text PROJECT No. 630A4-CSI-426
AutoCAD SHX Text
SITE MAP
AutoCAD SHX Text
NO SCALE
AutoCAD SHX Text
SYMBOLS
AutoCAD SHX Text
KEY MAP: BUILDING 1
AutoCAD SHX Text
SCALE: 1/32" = 1'-0"
AutoCAD SHX Text
PROPOSED AREA OF WORK
AutoCAD SHX Text
IMAGE: GOOGLE EARTH
AutoCAD SHX Text
POLY PLACE
AutoCAD SHX Text
7TH AVENUE
AutoCAD SHX Text
14TH AVENUE
AutoCAD SHX Text
VA NY HARBOR HEALTHCARE
AutoCAD SHX Text
BROOKLYN CAMPUS
AutoCAD SHX Text
800 POLY PLACE, BROOKLYN, NY
AutoCAD SHX Text
RADIATION ONCOLOGY SUITE
AutoCAD SHX Text
NORTH
AutoCAD SHX Text
BUILDING 1 : PARTIAL FIRST FLOOR
AutoCAD SHX Text
NORTH
AutoCAD SHX Text
DRAWING LIST
AutoCAD SHX Text
RADIATION ONCOLOGY SUITE
SAFETY/EGRESS - FIRST FL PART PLAN
MC TB
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
Fi le
P at h 6/
/2
:3
9:
P M
ISSUE FOR BID
Office of
Construction and Facilities Management
U.S. Department of Veterans Affairs
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT OF RECORD STAMPCONSULTANT
321 4 5 6 7 8 9 10
Project Director
23 MAR 2021Revisions: Date:
630A4-CSI-426
800 POLY PLACE, BROOKLYN, NY 11209-7104
DESIGN ONCOLOGY LINAC SITE PREP
THINKFORM DESIGN ARCHITECT LLC
38 EAST BROAD STREET
HOPEWELL, NJ 08525
T:855.821.0274 F:609.644.4397 Russell DiNardo, AIA NY 031521-1
VA NY HARBOR HEALTHCARE SYSTEM
BROOKLYN CAMPUSENGINEERING LLP
www.fellp.com
22 MULBERRY ST, SUITE 2A
MIDDLETOWN, NY 10940
T:845.343.1481 F:855.320.8735
21-03-BK
KEY MAP - BLDG 1, GROUND FLOOR
AutoCAD SHX Text
PATH 1 TRAVEL DISTANCE FROM LINAC ROOM TO EXIT: 158'-6"
AutoCAD SHX Text
PATH 2 TRAVEL DISTANCE FROM LINAC ROOM TO EXIT: 199"-0"
AutoCAD SHX Text NOTE: NO CHANGE TO THE EXISTING RADIOLOGY ONCOLOGY SUITE ESTABLISHED TRAVEL DISTANCES - MEASUREMENTS SHOWN FOR REFERENCE ONLY.
AutoCAD SHX Text A LABEL CLASS A DOOR CLASS A DOOR A/C AIR CONDITION AIR CONDITION A/E ARCHITECT/ENGINEER ARCHITECT/ENGINEER ABV ABOVE ABOVE ACT ACOUSTICAL CEILING TILE ACOUSTICAL CEILING TILE AD AREA DRAIN AREA DRAIN ADH ADHESIVE ADHESIVE ADJ ADJACENT ADJACENT ADMIN ADMINISTRATION ADMINISTRATION AFC ABOVE FINISHED COUNTER ABOVE FINISHED COUNTER AFF ABOVE FINISHED FLOOR ABOVE FINISHED FLOOR AHU AIR HANDLING UNIT AIR HANDLING UNIT ALT ALTERNATE ALTERNATE ALUM ALUMINUM ALUMINUM APPROX APPROXIMATE APPROXIMATE ARCH ARCHITECT ARCHITECT ASB ASBESTOS ASBESTOS ASPH ASPHALT ASPHALT ASSN ASSOCIATION ASSOCIATION AUTO AUTOMATIC AUTOMATIC AUX AUXILIARY AUXILIARY BAT BATTEN BATTEN BD BOARD BOARD BEV BEVEL BEVEL BKG BACKING BACKING BLDG BUILDING BUILDING BOT BOTTOM BOTTOM BTWN BETWEEN BETWEEN C TO C CENTER TO CENTER CENTER TO CENTER CAB CABINET CABINET CONC CONCRETE CONCRETE CIR CIRCLE CIRCLE CL CENTER LINE CENTER LINE CLDG CLADDING CLADDING CLG CEILING CEILING CLG HT CEILING HEIGHT CEILING HEIGHT CLR CLEAR CLEAR CM CENTIMETER CENTIMETER CMU CONCRETE MASONRY UNIT CONCRETE MASONRY UNIT CNTR COUNTER COUNTER COL COLUMN COLUMN CONC FLR CONCRETE FLOOR CONCRETE FLOOR CONSTR CONSTRUCTION CONSTRUCTION CONT CONTINUE CONTINUE COORD COORDINATE COORDINATE CPT CARPET CARPET CSWK CASEWORK CASEWORK CT CERAMIC TILE CERAMIC TILE CTRL CONTROL CONTROL DAT DATUM DATUM DC DIRECT CURRENT DIRECT CURRENT DEG DEGREE DEGREE DEL DELETE DELETE DEMO DEMOLITION DEMOLITION DEPT DEPARTMENT DEPARTMENT DTL DETAIL DETAIL DIFF DIFFERENCE DIFFERENCE DIM DIMENSION DIMENSION DIR DIRECTION DIRECTION DR DOOR, OR DRAIN DOOR, OR DRAIN DR OPNG DOOR OPENING DOOR OPENING E.J. EXPANSION JOINT EXPANSION JOINT EL ELEVATION ELEVATION ELEC ELECTRIC ELECTRIC ELEV ELEVATOR ELEVATOR EMER EMERGENCY EMERGENCY ENGR ENGINEER ENGINEER EQ EQUAL EQUAL EQUIP EQUIPMENT EQUIPMENT EQUIV EQUIVALENT EQUIVALENT ESP ESPECIALLY ESPECIALLY EXIST EXISTING EXISTING FA FIRE ALARM FIRE ALARM FAS FASCIA FASCIA FD FLOOR DRAIN FLOOR DRAIN FE FIRE EXTINGUISHER FIRE EXTINGUISHER FEC FIRE EXTINGUISHER CABINET FIRE EXTINGUISHER CABINET FIN FLR FINISH FLOOR FINISH FLOOR FIXT FIXTURE FIXTURE FLR FLOOR FLOOR FP FIRE PROTECTION OR FIREPROOF FIRE PROTECTION OR FIREPROOF FRT FIRE RETARDANT TREATED FIRE RETARDANT TREATED FRZ FREEZER FREEZER FT FEET OR FOOT FEET OR FOOT FURN FURNISH OR FURNITURE FURNISH OR FURNITURE GA GAGE OR GYPSUM ASSOCIATION GAGE OR GYPSUM ASSOCIATION GAL GALLON GALLON GALV GALVANIC OR GALVANIZED GALVANIC OR GALVANIZED GB GRAB BAR GRAB BAR GC GENERAL CONTRACTOR GENERAL CONTRACTOR GEN GENERAL OR GENERATOR GENERAL OR GENERATOR GL GLASS GLASS GLU LAM GLUED LAMINATED WOOD GLUED LAMINATED WOOD GT GROUT GROUT GYP GYPSUM GYPSUM GYP BD GYPSUM BOARD GYPSUM BOARD HB HOSE BIBB HOSE BIBB HC HOLLOW CORE HOLLOW CORE HDWD HARDWOOD HARDWOOD HM HOLLOW METAL HOLLOW METAL HT HEIGHT HEIGHT HVAC HEATING, VENTILATING, HEATING, VENTILATING, AND AIR CONDITIONING ID IDENTIFICATION IDENTIFICATION ILLUM ILLUMINATION ILLUMINATION INCL INCLUDED INCLUDED INFO INFORMATION INFORMATION INSUL INSULATION INSULATION INT INTERIOR INTERIOR JAN JANITOR JANITOR J-BOX JUNCTION BOX JUNCTION BOX LAB LABORATORY LABORATORY LAM LAMINATE LAMINATE LAV LAVATORY LAVATORY LDR LEADER LEADER LED LIGHT EMITTING DIODE LIGHT EMITTING DIODE LT LIGHT LIGHT LVD LOUVERED LOUVERED LVR LOUVER LOUVER LYR LAYER LAYER
AutoCAD SHX Text M METER METER MAINT MAINTENANCE MAINTENANCE MATL MATERIAL MATERIAL MAX MAXIMUM MAXIMUM MECH MECHANICAL MECHANICAL MEMO MEMORANDUM MEMORANDUM MEZZ MEZZANINE MEZZANINE MID MIDDLE MIDDLE MIRR MIRROR MIRROR MISC MISCELLANEOUS MISCELLANEOUS MLDG MOLDING (MOULDING) MOLDING (MOULDING) MLWK MILLWORK MILLWORK MO MASONRY OPENING MASONRY OPENING MTL METAL METAL MULL MULLION MULLION NA NOT APPLICABLE NOT APPLICABLE NAT NATURAL NATURAL NIC NOT IN CONTRACT NOT IN CONTRACT NO NUMBER NUMBER NOM NOMINAL NOMINAL NUM NUMERAL NUMERAL OC ON CENTER ON CENTER OCC OCCUPY OCCUPY OFF OFFICE OFFICE OPNG OPENING OPENING OPP OPPOSITE OPPOSITE OSHA OCCUPATIONAL SAFETY AND OCCUPATIONAL SAFETY AND HEALTH ADMINSTRATION OZ OUNCE OUNCE PART PARTIAL PARTIAL PAT PATTERN PATTERN PEN PENETRATE PENETRATE PERF PERFORATED PERFORATED PERP PERPENDICULAR PERPENDICULAR PLAS PLASTER OR PLASTIC PLASTER OR PLASTIC PLBG PLUMBING PLUMBING PLYWD PLYWOOD PLYWOOD PNL PANEL PANEL PRCST PRECAST PRECAST PREFAB PREFABRICATE PREFABRICATE PRELIM PRELIMINARY PRELIMINARY PREP PREPARATION PREPARATION PREV PREVIOUS PREVIOUS PROP PROPERTY PROPERTY PTD PAINTED PAINTED PVC POLYVINYL CHLORIDE (PLASTIC) POLYVINYL CHLORIDE (PLASTIC) QA QUALITY ASSURANCE QUALITY ASSURANCE QC QUALITY CONTROL QUALITY CONTROL QT QUARRY TILE QUARRY TILE QTB QUARRY TILE BASE QUARRY TILE BASE QTR QUARTER QUARTER QTY QUANTITY QUANTITY R RADIUS OR RISER RADIUS OR RISER RAD RADIATOR RADIATOR RB RESILIENT BASE OR RUBBER RESILIENT BASE OR RUBBER BASE RBR RUBBER RUBBER RCP REFLECTED CEILING PLAN REFLECTED CEILING PLAN RCVR RECEIVER RECEIVER RD ROAD OR ROOF DRAIN ROAD OR ROOF DRAIN REC RECESSED RECESSED RECT RECTANGLE RECTANGLE REF REFERENCE OR REFRIGERATOR REFERENCE OR REFRIGERATOR REINF REINFORCE REINFORCE REQ REQUIRE REQUIRE REQD REQUIRED REQUIRED RESIL RESILIENT RESILIENT RET RETURN RETURN REV REVISION REVISION RM ROOM ROOM RND ROUND ROUND RO ROUGH OPENING ROUGH OPENING SALV SALVAGE SALVAGE SBSTR SUBSTRATE SUBSTRATE SCHED SCHEDULE SCHEDULE SCHEM SCHEMATIC SCHEMATIC SECT SECTION SECTION SEP SEPARATE SEPARATE SF SQUARE FOOT (FEET) SQUARE FOOT (FEET) SLNT SEALANT SEALANT SMK SMOKE SMOKE SPEC SPECIFICATION SPECIFICATION SPKLR SPRINKLER SPRINKLER SPLY SUPPLY SUPPLY SQ SQUARE SQUARE SS STAINLESS STEEL STAINLESS STEEL STD STANDARD STANDARD STOR STORAGE STORAGE STR STRINGERS STRINGERS STRUCT STRUCTURAL STRUCTURAL SURF SURFACE SURFACE SUSP SUSPEND SUSPEND SYS SYSTEM SYSTEM T&G TONGUE AND GROOVE TONGUE AND GROOVE TECH TECHNICAL TECHNICAL TEL TELEPHONE TELEPHONE TEMP TEMPERATURE OR TEMPORARY TEMPERATURE OR TEMPORARY THK THICKNESS THICKNESS THRU THROUGH THROUGH TMPD TEMPERED TEMPERED TO TOP OF ___ TOP OF ___ TRANS TRANSOM TRANSOM TRTD TREATED TREATED TYP TYPICAL TYPICAL U HEAT TRANSFER COEFFICIENT HEAT TRANSFER COEFFICIENT UPS UNINTERRUPTIBLE POWER UNINTERRUPTIBLE POWER SUPPLY UTIL UTILITY UTILITY UV ULTRAVIOLET ULTRAVIOLET VAC VACUUM VACUUM VAR VARIES VARIES VCT VINYL COMPOSITION TILE VINYL COMPOSITION TILE VENT VENTILATION VENTILATION VERT VERTICAL VERTICAL VEST VESTIBULE VESTIBULE VIF VERIFY IN FIELD VERIFY IN FIELD VOL VOLUME VOLUME W/ WITH WITH W/O WITHOUT WITHOUT WD WOOD WOOD WSCT WAINSCOT WAINSCOT WT WEIGHT WEIGHT
AutoCAD SHX Text
ABBREVIATIONS
AutoCAD SHX Text
INFECTION CONTROL RISK ASSESSMENT
AutoCAD SHX Text A. 1. PRIMARY APPLICABLE CODES AND STANDARDS - EXISTING BUILDING / RECONSTRUCTION 1. PRIMARY APPLICABLE CODES AND STANDARDS - EXISTING BUILDING / RECONSTRUCTION THE DEPARTMENT OF VETERANS AFFAIRS HAS ADOPTED THE LATEST EDITION OF THE FOLLOWING CODES AND STANDARDS AS A MINIMUM FOR ALL PROJECTS PERFORMED IN THE MODERNIZATION, ALTERATION, ADDITION OR IMPROVEMENT OF ITS REAL PROPERTY AND THE CONSTRUCTION OF NEW STRUCTURE. a. NFPA 101 LIFE SAFETY CODE, 2018 EDITION b. NFPA NATIONAL FIRE CODES (WITH THE EXCEPTION OF NFPA 5000 AND NFPA 900) c. NATIONAL ELECTRIC CODE (NEC) d. INTERNATIONAL PLUMBING CODE (IPC) e. ASME BOILER AND PRESSURE VESSEL CODE f. ASME CODE FOR PRESSURE PIPING g. ARCHITECTURAL BARRIERS ACT ACCESSIBILITY STANDARDS (ABAAS) INCLUDING VA SUPPLEMENT, BARRIER FREE DESIGN GUIDE (PG-18-13) h. ENERGY POLICY ACT (EPAct) i. VA DIRECTIVES, DESIGN MANUALS, MASTER SPECIFICATIONS, VA NATIONAL CAD STANDARD APPLICATION GUIDE, AND OTHER GUIDANCE ON THE TECHNICAL INFORMATION LIBRARY (TIL) j. VA MENTAL HEALTH DESIGN GUIDE k. INTERNATIONAL BUILDING CODE (IBC) - ONLY WHEN SPECIFICALLY REFERENCE IN VA DESIGN DOCUMENTS l. OCCUPATIONAL, SAFETY AND HEALTH ADMINISTRATION (OSHA) STANDARDS 2. THE CONSTRUCTION CLASSIFICATION IS TYPE II (222). THE MINIMUM FIRE RATING FOR ALL LOAD-BEARING STRUCTURAL ELEMENTS IS 2HRS. IF DURING CONSTRUCTION, ANY DAMAGE TO EXISTING FIREPROOFING IS OBSERVED OR IF DISCOVERED THAT ANY FIRE PROTECTION IS MISSING, THE CONTRACTOR SHALL REPAIR TO MAINTAIN FULL PROTECTION. 3. RENOVATION TO THE EXISTING LINAC AND CONTROL ROOM AREA OCCUPIES APPROX 1,200 SQ FT OF THE EXISTING RADIATION ONCOLOGY SUITE ON THE FIRST FLOOR OF BUILDING 1.
AutoCAD SHX Text
CODE NOTES
AutoCAD SHX Text
TRAVEL DISTANCE
AutoCAD SHX Text 0-G002
AutoCAD SHX Text
ABBREVIATIONS, CODE NOTES AND FIRE SAFETY/ EGRESS
AutoCAD SHX Text
AutoCAD SHX Text G002
AutoCAD SHX Text 1/16" = 1'-0"
AutoCAD SHX Text
2 HOUR FIRE RESISTANCE RATED FIRE BARRIER
AutoCAD SHX Text NFPA 101 REVIEW B. CHAPTER 21 EXISTING AMBULATORY HEALTH CARE OCCUPANCIES CHAPTER 21 EXISTING AMBULATORY HEALTH CARE OCCUPANCIES 1. (21.1.1.1.1) THE REQUIREMENTS OF THIS CHAPTER SHALL APPLY TO EXISTING BUILDINGS OR PORTIONS THEREOF CURRENTLY OCCUPIED AS AN AMBULATORY HEALTH CARE OCCUPANCIES. 2. (21.1.1.4.4) CONSTRUCTION - SEE 4.6.10: BUILDINGS, OR PORTIONS OF BUILDINGS, SHALL BE PERMITTED TO BE OCCUPIED DURING CONSTRUCTION ONLY WHERE REQUIRED MEANS OF EGRESS AND REQUIRED FIRE PROTECTION FEATURES ARE IN PLACE AND CONTINUOUSLY MAINTAINED FOR THE PORTION OCCUPIED OR WHERE ALTERNATIVE LIFE SAFETY MEASURES ACCEPTABLE TO THE AUTHORITY HAVING JURISDICTION ARE IN PLACE. 3. (21.1.7) THE OCCUPANT LOAD SHALL BE DETERMINED ON THE BASIS OF OCCUPANT LOAD FACTORS OF TABLE 7.3.1.2. - NOTE NO CHANGE TO EXISTING OCCUPANCY IN THE RENOVATION OF THE LINAC ROOM.
AutoCAD SHX Text C. MEANS OF EGRESS 1. (21.2.1) ALL MEANS OF EGRESS SHALL COMPLY WITH CHAPTER 7, UNLESS OTHERWISE MODIFIED BY 21.2.2 THROUGH 21.2.11. 2. (21.2.3.2) CLEAR WIDTH FOR ANY CORRIDOR OR PASSAGEWAY REQUIRED FOR EXIT ACCESS SHALL NOT BE LESS THAN 44 INCHES. 3. (21.2.3.4) DOOR IN THE MEANS OF EGRESS FROM DIAGNOSTIC OR TREATMENT AREAS SHALL PROVIDE A CLEAR WIDTH OF NOT LESS THAN 32 INCHES. 4. (21.2.5.2) DEAD-END CORRIDORS SHALL NOT EXCEED 50 FT. 5. (21.2.6.2.2) THE MAXIMUM TRAVEL DISTANCE BETWEEN ANY POINT IN A ROOM AND AN EXIT SHALL (150 FT PER 21.2.6.2.1) BE PERMITTED TO INCREASE BY 50 FEET IN BLDGS PROTECTED THROUGHOUT BY AN APPROVED AUTOMATIC SPRINKLER SYSTEM. D. PROTECTION FROM HAZARDS 1. (21.3.2.1) HAZARDOUS AREAS INCLUDING, BUT NOT LIMITED TO , AREAS USED FOR GENERAL STORAGE, BOILER OR FURNACE ROOMS SHALL BE PROTECTED IN ACCORDANCE WITH SECTION 8.7. 1. (21.3.2.5) AREAS WHERE MEDICAL GAS IS STORED OR ADMINISTERED, AND THE OPERATION, MANAGEMENT, AND MAINTENANCE OF MEDICAL GASES SHALL BE IN ACCORDANCE WITH NFPA 99. E. INTERIOR FINISHES - CHAPTER 20 AND SECTION 10.2 1. (21.3.3.2.1) INTERIOR WALL AND CEILING FINISH MATERIAL COMPLYING WITH SECTION 10.2 SHALL BE CLASS A OR CLASS B IN EXITS AND IN EXIT ACCESS CORRIDORS. 2. (21.3.3.2.2) INTERIOR WALL AND CEILING FINISHES SHALL BE CLASS A, CLASS B OR CLASS C IN ALL AREAS OTHER THAN THOSE SPECIFIED IN 20.3.3.2.1.
AutoCAD SHX Text
CALL NORTH
AutoCAD SHX Text
1 HOUR FIRE RESISTANCE RATED FIRE BARRIER
AutoCAD SHX Text THE PURPOSE OF INFECTION CONTROL PROCEDURES IS TO MINIMIZE THE POTENTIAL FOR THE SPREAD OF INFECTIONS DUE TO DEGRADED AIR QUALITY, ENVIRONMENTAL CONTAMINATION, OR CONTAMINATION OF WATER. IT IS THE POLICY OF THIS MEDICAL CENTER THAT A SYSTEMATIC APPROACH BASED ON ASSESSMENT AND PLANNING WILL EFFECTIVELY MANAGE THE ISSUE OF TRANSMISSION OF INFECTIOUS DISEASES AND THE AGGRAVATION OF ALLERGIES. THE OVERALL APPROACH RELIES ON PRE-CONSTRUCTION ASSESSMENT, MANAGEMENT OF HVAC SYSTEMS, AND AGGRESSIVE USE OF A VARIETY OF BARRIERS. CONTRACTOR IS TO WORK CLOSELY WITH THE VA FACILITY MANAGEMENT SERVICE CENTER, THE INFECTION CONTROL PROGRAM STAFF AND THE SAFETY AND OCCUPATION HEALTH MANAGER TO HAVE IN PLACE A MONITORING A PROGRAM TO ASSURE RIGOROUS IMPLEMENTATION OF THE REQUIRED MEASURES. REFER TO PROJECT SPECIFICATIONS.
AutoCAD SHX Text
INFECTION CONTROL NOTES
AutoCAD SHX Text A. THE CONTRACTOR SHALL VERIFY ALL EXISTING CONDITIONS IN THE FIELD AND COORDINATE ALL THE CONTRACTOR SHALL VERIFY ALL EXISTING CONDITIONS IN THE FIELD AND COORDINATE ALL INFECTION CONTROL REQUIREMENTS WITH THE DEMOLITION AND NEW CONSTRUCTION. THE CONTRACTOR SHALL REPORT ALL DISCOVERIES TO THE OWNER'S REPRESENTATIVE FOR RESOLUTION PRIOR TO BEGINNING. B. CONTRACTOR TO INSTALL BARRIERS AND OTHER MEASURES. CONSTRUCTION BARRIER DOOR(S) MUST CONTRACTOR TO INSTALL BARRIERS AND OTHER MEASURES. CONSTRUCTION BARRIER DOOR(S) MUST BE SELF-CLOSING. PROVIDE DUST CONTROL AND CONTAINMENT STICKY MATS ON ALL ACCESS DOORS INTO CONSTRUCTION AREA. ANTE-ROOM REQUIRED AT CONSTRUCTION ENTRY. PROVIDE MANOMETER MONITORS FOR ALL CONSTRUCTION BARRIERS.
AutoCAD SHX Text C. CONTRACTOR TO DISPOSE OF MATERIALS BEFORE 8:00AM OR AFTER 4:00PM UNLESS ARRANGED WITH CONTRACTOR TO DISPOSE OF MATERIALS BEFORE 8:00AM OR AFTER 4:00PM UNLESS ARRANGED WITH THE VA TO DISPOSE OF DURING NORMAL HOURS. USE ROUTE AGREED UPON BY ICRA TEAM AND COTR FOR TRANSPORT OF TRANSPORT CONTAINERS THROUGH THE FACILITY. CONTRACTORS CAN ONLY DEVIATE TO NORMAL HOURS WITH THE APPROVAL OF THE INFECTION CONTROL NURSE AND COR DUE TO PARTICULAR CIRCUMSTANCES THAT MAY OBSTRUCT PROGRESS. A CLEAR PLAN WILL HAVE TO BE SUBMITTED AND PRECAUTIONARY INFECTION CONTROL MEASURES IMPLEMENTED. D. CONTRACTOR TO COVER CARTS AND WEAR BOOTIES WHEN DISPOSING MATERIALS. CONTRACTOR TO COVER CARTS AND WEAR BOOTIES WHEN DISPOSING MATERIALS. E. CONTRACTOR TO CHANGE MATS MULTIPLE TIMES A DAY AS NEEDED AND MOP FLOORS BEHIND THEM CONTRACTOR TO CHANGE MATS MULTIPLE TIMES A DAY AS NEEDED AND MOP FLOORS BEHIND THEM MULTIPLE TIMES A DAY. F. CONTRACTOR TO MAINTAIN CONTINUOUS NEGATIVE AIR PRESSURE WITHIN THE WORK SITE UTILIZING CONTRACTOR TO MAINTAIN CONTINUOUS NEGATIVE AIR PRESSURE WITHIN THE WORK SITE UTILIZING WINDOW EXHAUST FANS STRATEGICALLY LOCATED OR PROVIDE NEGATIVE PRESSURE AIR HEPA UNITS AND EXHAUST TO EXTERIOR OF THE BUILDING. CONTRACTOR TO INCLUDE MONITORING MANOMETER AT ANTEROOM SO THE NEGATIVE PRESSURE CAN BE INSPECTED.
AutoCAD SHX Text
NO SCALE
N
ASBESTOS ABATEMENT PLAN
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
Fi le
P at h 6/
/2
:3
9:
P M
ISSUE FOR BID
Office of
Construction and Facilities Management
U.S. Department of Veterans Affairs
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT OF RECORD STAMPCONSULTANT
321 4 5 6 7 8 9 10
Project Director
Revisions: Date:
630A4-CSI-426
800 POLY PLACE, BROOKLYN, NY 11209-7104
DESIGN ONCOLOGY LINAC SITE PREP
THINKFORM DESIGN ARCHITECT LLC
38 EAST BROAD STREET
HOPEWELL, NJ 08525
T:855.821.0274 F:609.644.4397 Russell DiNardo, AIA NY 031521-1
VA NY HARBOR HEALTHCARE SYSTEM
BROOKLYN CAMPUS
KEY MAP - BLDG 1, GROUND FLOOR
Drawing 01 of 02
TGE EDS
Egan Environmental Consulting, Inc.
23 MARCH 2021
AutoCAD SHX Text
LINEAR ACCEL
AutoCAD SHX Text G110
AutoCAD SHX Text
LOCKER
AutoCAD SHX Text G113
AutoCAD SHX Text
CONTROL
AutoCAD SHX Text G109
AutoCAD SHX Text
TOILET
AutoCAD SHX Text G112
AutoCAD SHX Text
TOILET
AutoCAD SHX Text G112A
AutoCAD SHX Text
RAD ONCOLOGIST
AutoCAD SHX Text G114
AutoCAD SHX Text
CHIEF ONCOLOGIST
AutoCAD SHX Text G115
AutoCAD SHX Text
RAD ONCOLOGIST
AutoCAD SHX Text G116
AutoCAD SHX Text
CONF
AutoCAD SHX Text G117A
AutoCAD SHX Text
OFFICE
AutoCAD SHX Text G117
AutoCAD SHX Text
CORR
AutoCAD SHX Text G110A
AutoCAD SHX Text
LINAC
AutoCAD SHX Text G111
AutoCAD SHX Text
CONTROL
AutoCAD SHX Text G104
AutoCAD SHX Text
TOILET
AutoCAD SHX Text G137
AutoCAD SHX Text
WAITING
AutoCAD SHX Text G136
AutoCAD SHX Text
CLOSET
AutoCAD SHX Text G136A
AutoCAD SHX Text
TEL CLOSET
AutoCAD SHX Text G201
AutoCAD SHX Text
STORAGE
AutoCAD SHX Text G202
AutoCAD SHX Text
RADIATION ONCOLOGY
AutoCAD SHX Text
WAITING
AutoCAD SHX Text G204
AutoCAD SHX Text
MECH
AutoCAD SHX Text G214
AutoCAD SHX Text
TOILET
AutoCAD SHX Text G231
AutoCAD SHX Text
ELEC
AutoCAD SHX Text G232
AutoCAD SHX Text
FEMALE
AutoCAD SHX Text
WAITING
AutoCAD SHX Text G230
AutoCAD SHX Text
CORRIDOR SE3
AutoCAD SHX Text G200C
AutoCAD SHX Text
CAT
AutoCAD SHX Text G103
AutoCAD SHX Text
CHAPEL
AutoCAD SHX Text
JAN
AutoCAD SHX Text LEAD (Pb) LINED WALL TO BE DEMOLISHED IN ACCORDANCE WITH 29 CFR 1926.62.
AutoCAD SHX Text
AutoCAD SHX Text
WORKER DECON. UNIT
AutoCAD SHX Text
WASTE DECON. UNIT
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text G101
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text H
AutoCAD SHX Text 1 HOUR FIRE RESISTANT RATED SMOKE BARRIER 1 HOUR FIRE RESISTANT RATED FIRE BARRIER 2 HOUR FIRE RESISTANT RATED FIRE BARRIER
AutoCAD SHX Text Asbestos-Containing Materials Quantity Table
AutoCAD SHX Text Location
AutoCAD SHX Text Material
AutoCAD SHX Text Quantity
AutoCAD SHX Text Ground Floor - Room G104
AutoCAD SHX Text
PIPE INSULATION
AutoCAD SHX Text
30 L.F. 4-8" 10 FITTINGS
AutoCAD SHX Text Ground Floor - Room G104
AutoCAD SHX Text
CONTAMINATED CEILING TILE
AutoCAD SHX Text
150 SQ. FT.
AutoCAD SHX Text Ground Floor - Room G101
AutoCAD SHX Text
FLOOR TILE & MASTIC
AutoCAD SHX Text
320 SQ. FT.
AutoCAD SHX Text Ground Floor - Room G101
AutoCAD SHX Text
PIPE INSULATION
AutoCAD SHX Text
30 L.F. 4-8" 10 FITTINGS
AutoCAD SHX Text Ground Floor - Room G101
AutoCAD SHX Text
CONTAMINATED CEILING TILE
AutoCAD SHX Text
320 SQ. FT.
AutoCAD SHX Text ASBESTOS ABATEMENT LEGEND: PIPE INSULATION TO BE REMOVED. FLOOR TILE & MASTIC TO BE REMOVED. CONTAMINATED CEILING TILE TO BE REMOVED. PATH OF WASTE OUTSIDE OF BUILDING. HEPA - FILTERED NEGATIVE AIR MACHINE
AutoCAD SHX Text
AutoCAD SHX Text H
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
HA100
AutoCAD SHX Text
GROUND FLOOR ASBESTOS ABATEMENT PLAN
AutoCAD SHX Text
AutoCAD SHX Text
HA-100
AutoCAD SHX Text 1/4" = 1'-0"
AutoCAD SHX Text
GROUND FLOOR
AutoCAD SHX Text
NORTH
AutoCAD SHX Text
NO SCALE
AutoCAD SHX Text N
ASBESTOS ABATEMENT CONSULTANT: 14 HIGH STREET MAHWAH, NEW JERSEY 07430 Tel: (914) 320-7924 Email: tom_egan2@yahoo.com
ASBESTOS ABATEMENT NOTES:
1. All asbestos removal areas indicated on the drawings and specification is for informational assistance to the contractor. All work areas having asbestos-containing materials (ACM's) must be field measured and verified for actual locations, size and quantities. Additional piping or equipment which may have asbestos-containing insulation may be present. The contractor must remove ACM's from all specified areas whether shown or accurately indicated. These specifications shall be used as a guide, and do not limit the contractor's responsibility to seek and remove all ACM's as specified.
2. Contractor shall use extreme care and caution during all abatement procedures. Contractor shall conduct removal of all materials from the site with a minimum disturbance and provide regular maintenance and protection of the building premises, staging area, transportatioon routes, etc. throughout the building.
3. All building finishes, construction equipment or other components of this building which are damaged or in anyway disrupted as a result of the execution of this contract shall be replaced or restored to original condition.
4. Contractor shall provide as many shifts of laborers as necessary to complete the project in the amount of time specified in the contract documents at no extra cost.
5. Asbestos removal shall be performed in accordance with all applicable federal, state and local laws, guidelines, regulations, orders and directives, including without limitations those of the responsible New York State agencies; U.S. Environmental Protection Agency (EPA), the U.S.
Department of Labor, the Occupational Safety and Health Administration (OSHA), National Institute of Occupational Safety and Health (NIOSH), and the U.S. Department of Veterans Affairs.
6. Contractor shall develop and implement written standard procedures for abatement work to ensure maximum protection and safeguards for asbestos exposure to the workers, employees, general public and the environment.
7. Contractor shall be responsible for removing any light fixtutres, suspended ceilings, walls, ductwork, or any obstructions necessary for the successful completion of the project. Contractor shall carefully remove, decontaminate and reinstall all removed fixtures, supporting grids or any other building structure removed by the contractor to access ACM's.
8. Air filtration device (AFD) shall be equipped with smoke/heat sensor systems that will shut down the device wihtin 30 seconds after fire and/or smoke detection. AFD sensor devices must comply with American National Standards NFPA Life Safety Code 101.
9. Asbestos removal drawings were prepared from existing drawings supplied by the owner and are provided for information only. The consultant makes no claims as to their accuracy or for quantities derived through their use.
10. Contractor shall provide all scaffolding, ladders, plywood, equipment, supplies, etc. as required and/or necessary to complete all work required by the contract documents.
11. Contractor shall provide all signs, labels, warnings and posted instructions that are necessary to protect, inform and warn people of the hazard from asbestos exposure.
12. Contractor shall post in a prominent and convenient place for workers, a copy of the latest applicable regulations from OSHA, EPA, NIOSH and
NYSDOL.
13. Contractor shall completely protect all controls, conduit, tubing, electrical panels, equipment, etc. within the work areas.
14. The VAMC and the asbestos abatement contractor shall mutually agree upon the location of the decontamination units for each phase during the preconstruction meeting unless shown on the drawings.
15. Submit shop drawings showing location and number of HEPA-filtered negative air machines, airlocks, and decontamination chambers. Submit calculations showing number of air changes for each containment area.
16. All asbestos removals shall be performed within a HEPA-filtered negative air pressure enclosure. Pipe insulation work shall be performed using the tent/glovebag method.
17. Specific areas on the 1st floor of Building 1 will require asbestos abatement to allow for the LINAC replacement project.
Asbestos-containing materials (ACM's) must be removed to allow for the renovation. ACM's include but are not limited to pipe insulation, floor tile and mastic.
18. Scheduling of work shall be coordinated with the V.A. to avoid interruptions to the normal activities of the V.A. Asbestos Abatement will be performed over a weekend. Work will begin at 5pm on a Friday and be completed by 5am Monday including clearance air testing and breakdown.
19. The contractor shall prepare the work areas for asbestos removal using a HEPA-filtered negative air pressure containment in accordance with this specification and applicable codes, rules and regulations.
20. Ceiling tiles must be removed from the work area following erection of the HEPA-filtered negative air pressure containment.
21. The contractor shall erect isolation barriers that separate the work areas from the remainder of the building. The barriers shall be constructed of a fire retardant sheathing material of at least 5/8" thickness on the work side of the studs. The isolation barrier shall be plasticized with two (2) layers of 6-mil fire retardant plastic sheeting. The isolation barriers shall be caulked at the sheathing seams and where the sheathing meets the floor, walls and ceiling. The isolation barrier shall be constructed using fire retardant 2"x4" framing, 16 inches center-to-center, and caulked at all seams.
22. All utilities and piping shall be locked out and tagged before the contractor is to begin abatement activities.
23. All asbestos-containing materials within the contract limits must be removed and disposed of as asbestos-contaminated waste unless specifically excluded in these specifications or on the drawings.
24. The black iron supporting the drop ceilings is to remain intact throughout asbestos abatement activities.
25. All work area decontamination units (personal and waste) shall be fully framed and sheathed with fire retardant wood because they will be located in areas of public access.
26. All utilities, mechanical systems and piping shall be locked out and tagged before the contractor is to begin abatement activities.
27. All asbestos-containing materials within the contract limits must be removed and disposed of as asbestos-contaminated waste unless specifically excluded in these specifications or on the drawings.
Drawing 02 of 02
HA101
Egan Environmental Consulting, Inc.
KEY MAP - BLDG 1, GROUND FLOOR
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
Fi le
P at h 6/
/2
:3
9:
P M
Office of Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT OF RECORD STAMPCONSULTANT
321 4 5 6 7 8 9 10
Project Director
23 MARCH 2021Revisions: Date:
630A4-CSI-426
800 POLY PLACE, BROOKLYN, NY 11209-7104
DESIGN ONCOLOGY LINAC SITE PREP
THINKFORM DESIGN ARCHITECT LLC
38 EAST BROAD STREET
HOPEWELL, NJ 08525
T:855.821.0274 F:609.644.4397 Russell DiNardo, AIA NY 031521-1
VA NY HARBOR HEALTHCARE SYSTEM
BROOKLYN CAMPUS
ISSUE FOR BID
TGE EDS
ASBESTOS DIAGRAMS AND NOTES
AutoCAD SHX Text
10' MIN
AutoCAD SHX Text
EQUIPMENT ROOM
AutoCAD SHX Text
12' MIN
AutoCAD SHX Text
CLEAN ROOM
AutoCAD SHX Text
4'MIN
AutoCAD SHX Text
SHOWER
AutoCAD SHX Text
ROOM
AutoCAD SHX Text
8" MAX
AutoCAD SHX Text
CURTAINED DOORWAY
AutoCAD SHX Text
(3 LAYERS OF 6 MIL
AutoCAD SHX Text
POLY SHEETING WITH
AutoCAD SHX Text
WEIGHTED FLAPS)
AutoCAD SHX Text
(TYP @ AT ENTRANCE
AutoCAD SHX Text
AND EXIT)
AutoCAD SHX Text
2"X4" STUDS
AutoCAD SHX Text
2"X4" STUDS
AutoCAD SHX Text @ 48 O.C
AutoCAD SHX Text
4' MIN
AutoCAD SHX Text
AIRLOCK OF 36" MIN
AutoCAD SHX Text
WITH CURTAINED DOORWAY
AutoCAD SHX Text
CLEAN ROOM
AutoCAD SHX Text
WORK AREA
AutoCAD SHX Text A
AutoCAD SHX Text A
AutoCAD SHX Text @ 48 O.C
AutoCAD SHX Text
2"X4" STUDS
AutoCAD SHX Text
CONSTRUCTION OF A TEMPORARY SOLID BARRIER
AutoCAD SHX Text
WORKER
AutoCAD SHX Text
10. Puts on any additional
AutoCAD SHX Text clothing -deck shoes, hard hat
AutoCAD SHX Text
11. Collects necessary tools
AutoCAD SHX Text proceeds to work area
AutoCAD SHX Text
WORKER
AutoCAD SHX Text
1. Enters clean room
AutoCAD SHX Text
2. removes street clothing, places in locker
AutoCAD SHX Text
3. Puts on nylon swim suit (optional)
AutoCAD SHX Text
4. Puts on clean overalls
AutoCAD SHX Text
5. If separate disposable foot covering are
AutoCAD SHX Text used, these are put on
AutoCAD SHX Text
6. Applies tape around ankles, wrists, etc.
AutoCAD SHX Text
7. Inspects respirator, puts it on, checks fit
AutoCAD SHX Text
8. Puts on hood over respirator headstraps
AutoCAD SHX Text
9. proceeds to equipment room
AutoCAD SHX Text
21. Dries off, dresses in clean
AutoCAD SHX Text coveralls or street clothing
AutoCAD SHX Text
22. Cleans and dries respirator, AutoCAD SHX Text replaces filter (if applicable)
AutoCAD SHX Text
WORKER
AutoCAD SHX Text filters (without removing)
AutoCAD SHX Text with soap and water
AutoCAD SHX Text
19. Washes swim suit
AutoCAD SHX Text
20. Thoroughly washes body
AutoCAD SHX Text and hair
AutoCAD SHX Text
WORKER
AutoCAD SHX Text
13. Removes all clothing
AutoCAD SHX Text except respirator
AutoCAD SHX Text
14. places disposable
AutoCAD SHX Text protective clothing
AutoCAD SHX Text in a bag or bin
AutoCAD SHX Text
15. Stores any other
AutoCAD SHX Text contaminated articles
AutoCAD SHX Text
16. Proceeds to shower
AutoCAD SHX Text
NOTE: INSIDE WORK AREA WORKERS SHALL EXIT TO OUTSIDE THOUGH
AutoCAD SHX Text
DECONTAMINATION FACILITIES ONLY. ALL WORKERS TO WEAR
AutoCAD SHX Text
WORKER
AutoCAD SHX Text
12. Brushes off
AutoCAD SHX Text contamination
AutoCAD SHX Text
WORK AREA
AutoCAD SHX Text
AIRLOCK
AutoCAD SHX Text
EQUIPMENT ROOM
AutoCAD SHX Text
AIRLOCK
AutoCAD SHX Text
AIRLOCK
AutoCAD SHX Text
CLEAN ROOM
AutoCAD SHX Text
SHOWER
AutoCAD SHX Text
ROOM
AutoCAD SHX Text
ALL ASBESTOS CONTAINING MATERIALS SHALL BE
AutoCAD SHX Text
PACKED IN APPROVED CONTAINERS PRIOR TO
AutoCAD SHX Text
MOVING TO BAG STORAGE
AutoCAD SHX Text
BAGGING CHAMBER
AutoCAD SHX Text
EQUIPMENT/ WASTE
AutoCAD SHX Text
WASH ROOM
AutoCAD SHX Text
SHOWER ROOM
AutoCAD SHX Text
CLEAN ROOM
AutoCAD SHX Text
WORK AREA
AutoCAD SHX Text
CLEAN AREA
AutoCAD SHX Text
AIRLOCK
AutoCAD SHX Text
AIRLOCK
AutoCAD SHX Text
AIRLOCK
AutoCAD SHX Text
AIRLOCK
AutoCAD SHX Text
WORK AREA
AutoCAD SHX Text
AIR LOCK
AutoCAD SHX Text
SHOWER/WASH ROOM
AutoCAD SHX Text
CLEAN ROOM
AutoCAD SHX Text
AIR LOCK
AutoCAD SHX Text
CLEAN AREA
AutoCAD SHX Text
CEILING/DECKING
AutoCAD SHX Text
CAULKING
AutoCAD SHX Text
2 LAYERS OF 6 MIL
AutoCAD SHX Text
WALLS AND FLOOR
AutoCAD SHX Text
WORK AREA
AutoCAD SHX Text
CLEAN SIDE
AutoCAD SHX Text
EXISTING WALL
AutoCAD SHX Text
ELEVATION
AutoCAD SHX Text
3/8" PLYWOOD SHEET
AutoCAD SHX Text
ON WORK AREA
AutoCAD SHX Text
CEILING DECKING
AutoCAD SHX Text 2"x4" @ 16" O.C(TYP.)
AutoCAD SHX Text
CONTINUOUS CAULKING
AutoCAD SHX Text
@ EDGES & SEAMS OF
AutoCAD SHX Text
PLYWOOD SHEETS
AutoCAD SHX Text
SECTION A-A
AutoCAD SHX Text
EXISTING WALL
AutoCAD SHX Text
FLOOR
AutoCAD SHX Text
FLOOR
AutoCAD SHX Text
USE OF GLOVE BAG ENCLOSURE FOR PIPE INSULATION ABATEMENT
AutoCAD SHX Text Staple and tape securely to insure that
AutoCAD SHX Text containment bag will not break as a result
AutoCAD SHX Text of vacuum and/or weight of debris within.
AutoCAD SHX Text Tape seal tightly around
AutoCAD SHX Text the pipe at both sides
AutoCAD SHX Text
WETTING
AutoCAD SHX Text
AGENT
AutoCAD SHX Text
SIDE
AutoCAD SHX Text
PORT
AutoCAD SHX Text
GLOVE
AutoCAD SHX Text
ARMS
AutoCAD SHX Text
TOOL
AutoCAD SHX Text
BAG
AutoCAD SHX Text Hepa vacuum (secure & seal nozzle
AutoCAD SHX Text w/tape after inserting into side
AutoCAD SHX Text port of containment bag)
AutoCAD SHX Text
CEILING
AutoCAD SHX Text
CENTER AFD EXHAUST TUBE.
AutoCAD SHX Text
FIRE RETARDENT PLYWOOD
AutoCAD SHX Text
BLASTGATE MANIFOLD AND ATTACH
AutoCAD SHX Text
FLEXIBLE AFD
AutoCAD SHX Text
EXHAUST DUCT
AutoCAD SHX Text
AFD
AutoCAD SHX Text
UNIT
AutoCAD SHX Text
FLOOR
AutoCAD SHX Text
TO ADDITIONAL
AutoCAD SHX Text
AFD UNIT
AutoCAD SHX Text
DUCT TAPE
AutoCAD SHX Text
DUCT TAPE
AutoCAD SHX Text
APPLY TWO LAYERS
AutoCAD SHX Text
OF SIX MIL PLASTIC
AutoCAD SHX Text
EXISTING WALL
AutoCAD SHX Text
EXISTING WALL
AutoCAD SHX Text
CLEAN AREA
AutoCAD SHX Text
CAUTION SIGN SHALL BE PLACED
AutoCAD SHX Text
ON THE CLEAN SIDE OF THE DOOR
AutoCAD SHX Text
CAUTION LABELING SHALL IN
AutoCAD SHX Text
ACCORDANCE WITH EPA FEDERAL
AutoCAD SHX Text
REGULATIONS
AutoCAD SHX Text
APPLY TWO SHEETS
AutoCAD SHX Text
OF SIX MILL POLY
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
EXISTING DOOR
AutoCAD SHX Text
BUCK
AutoCAD SHX Text
APPLY AND SECURE
AutoCAD SHX Text
SHEETS OF POLY
AutoCAD SHX Text
WITH DUCT TAPE
AutoCAD SHX Text
EXISTING DOOR TO
AutoCAD SHX Text
BE CLOSED AND SEALED
AutoCAD SHX Text
AN EMERGENCY EXIT SIGN IS TO
AutoCAD SHX Text
BE PLACED ON THE CONTAMINATED
AutoCAD SHX Text
SIDE AT ALL SPECIFIED LOCATIONS
AutoCAD SHX Text
EXISTING DOOR
AutoCAD SHX Text
THE PERIMETER OF THE DOOR IS TO
AutoCAD SHX Text
BE SEALED WITH DUCT TAPE ON THE
AutoCAD SHX Text
CONTAMINATED SIDE PRIOR TO APPLYING
AutoCAD SHX Text
THE TWO LAYERS OF POLY
AutoCAD SHX Text
APPLY AND SECURE SHEETS
AutoCAD SHX Text
OF POLY WITH DUCT TAPE
AutoCAD SHX Text
AS REQUIRED
AutoCAD SHX Text
EXISTING WALL
AutoCAD SHX Text
APPLY TWO SHEETS
AutoCAD SHX Text
OF SIX MIL POLY
AutoCAD SHX Text
EMERGENCY CRITICAL BARRIER AT DOOR
AutoCAD SHX Text
CONTAMINATED SIDE
AutoCAD SHX Text
CONSTRUCTION OF PERSONNEL DECONTAMINATION SYSTEM
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
POLY SHEETING ON
AutoCAD SHX Text
PARALLEL PERSONNEL AND WASTE DECON. ENCLOSURE SYSTEMS
AutoCAD SHX Text
PERSONNEL DECONTAMINATION SYSTEM
AutoCAD SHX Text
WINDOW AIR FILTRATION DEVICE EXHAUST
AutoCAD SHX Text
RESPIRATORS AND FULLY PROTECTIVE CLOTHING INSIDE THE
AutoCAD SHX Text
WORK AREA (INCLUDING HEAD AND FOOT WEAR)
AutoCAD SHX Text
17. Washes respirator and soaks
AutoCAD SHX Text
18. Removes respirator, washes
AutoCAD SHX Text
WASTE DECONTAMINATION ENCLOSURE SYSTEM
AutoCAD SHX Text
USING ADJUSTABLE CLAMP AND
AutoCAD SHX Text
DUCT TAPE. SEAL AIRTIGHT.
AutoCAD SHX Text
HOSE BEAD
AutoCAD SHX Text
BLASTGATE MANIFOLD AND AJUSTABLE CLAMP FOR WINDOW
AutoCAD SHX Text
ADJUSTABLE CLAMP WITH INTERNAL GASKET
AutoCAD SHX Text
12" I.D. GALVANIZED METAL
AutoCAD SHX Text
GALVANIZED METAL
AutoCAD SHX Text
GALVANIZED METAL
AutoCAD SHX Text
GALVANIZED METAL
AutoCAD SHX Text ASBESTOS ABATEMENT CONSULTANT: 14 HIGH STREET MAHWAH, NEW JERSEY 07430 Tel: (914) 320-7924 Email: tom_egan2@yahoo.com
AutoCAD SHX Text
NO SCALE
N
EXISTING ROOF SHEILDING SUPPORT PLAN - TO REMAIN4
FE PROJECT
NO: 20-175
KEY PLAN
Project Number
Building Number
Drawing Number
Project Title
Location
Issue Date Checked Drawn
Drawing Title
Approved:
VA FORM 08 - 6231
A
FOR BID
Office of
Construction and Facilities Management
U.S. Department of Veterans Affairs
B
C
D
E
F
A
B
C
D
E
F
21 4 5 6 7 8 9 10
PhaseARCHITECT OF RECORD STAMPCONSULTANT
321 4 5 6 7 8 9 10
Project Director
23 MARCH 2021Revisions: Date:
630A4-CSI-426
800 POLY PLACE, BROOKLYN, NY 11209-7104
DESIGN ONCOLOGY LINAC SITE PREP
THINKFORM DESIGN ARCHITECT LLC
38 EAST BROAD STREET
HOPEWELL, NJ 08525
T:855.821.0274 F:609.644.4397 Russell DiNardo, AIA NY 031521-1
VA NY HARBOR HEALTHCARE SYSTEM
BROOKLYN CAMPUSENGINEERING LLP
www.fellp.com
22 MULBERRY ST, SUITE 2A
MIDDLETOWN, NY 10940
T:845.343.1481 F:855.320.8735
EXISTING PARTIAL GROUND FLOOR PLAN1
SECTION VIEW5 SECTION VIEW6
SECTION VIEW2 SECTION VIEW3
EXISTING EXTERIOR EQUIPMENT SLAB7
SECTION E-E8
GENERAL NOTE:
TREATMENT ROOM FLOOR & PIT NOTES:
FLOOR LOADING NOTES:
STAND AND COUCH FLOOR LOADS
TREATMENT ROOM FLOOR PIT CABLE ACCESS10 SERVICE REQUIREMENTS11
NOTE: NOTE:
STAND AND COUCH FLOOR LOADS SECTION9
(E)RAMP DN
EXISTING 1'-0" THICK
SLAB-ON-GRADE (V.I.F.)
FIN. FLR. EL. = ±26'-0"
EXISTING 6" HIGH
DENSITY CONCRETE (V.I.F.)
INDICATES
ASSUMED EXISTING
LEAD SHIELDING TYP.REPLACE
EXISTING ±71
2"
LEAD SHIELDING
AS REQ'D AFTER
NEW SLAB IS IN PLACE
EXISTING 6" HIGH
DENSITY CONCRETE
EXISTING
4" LEAD
(V.I.F.)
S201S201
2' -1
1" 2'
-1 1"
1'-13 8" 1'-33
4" 2'-51
8" NEW PIT
5'-10"X2'-51
8"X1'-0" DEEP
±5'-3"
IN-FILL EXISTING
± 8" DEEP PIT
V.
I.F
.5' -1
0"
IN-FILL EXISTING
± 1'-1" DEEP PIT
S201
S201
S201
S201
W
X5
W
X5
±1 2'
-8
V.
I.F
C O
N C
R
ET
E W
AL
L
TO
B
E R
EM
O
VE
D
S201
S201
S201
PARTIAL FOUNDATION PLAN AT VAULT
SCALE: 14" = 1'-0"
NOTES:
- COORDINATE EQUIPMENT CENTERLINE WITH ARCH. & VARIAN.
- COORDINATE ALL DETAILS AND REQUIREMENTS WITH VARIAN DRAWINGS AND PRODUCT PLANNING GUIDE.
- REFER TO VARIAN DRAWINGS AND PRODUCT PLANNING GUIDE FOR ACCEPTABLE TOLERANCES AND EQUIPMENT ANCHORAGE REQUIREMENTS.
- DESIGN OF EQUPMENT ANCHORAGE IS BY VARIAN AND THE EQUIPMENT INSTALLER.
- COORDINATE OVERHEAD EQUIPMENT SUPPORT DETAILS WITH VARIAN.
S101
PARTIAL ROOF FRAMING PLAN AT VAULT
SCALE: 14" = 1'-0"
NOTES:
- DO NOT UNDERCUT OR UNDERMINE EXISTING BEAM BEARING LOCATIONS.
- FIELD-VERIFY ALL CONDITIONS.
- FIELD VERIFY WALL THICKNESS AT NEW BEAM BEARING LOCATIONS. NOTIFY THE E.O.R. IMMEDIATELY IF WALL THICKNESSES ARE
NOT AS SHOWN.
- THE CONTRACTOR SHALL PROVIDE ALL NECESSARY TEMPORARY SHORING. SUBMIT SIGNED AND SEALED TEMPORARY SHORING
PLANS AND PROPOSED SEQUENCE OF WORK TO THE ARCHITECT AND E.O.R., PRIOR TO PERFORMING ANY WORK. REFER TO
STRUCTURAL NOTES FOR ADDITIONAL REQUIREMENTS.
S101
2' -2
2'
-2
1' -1
0" 1'
-8
3' -2
1'
-8
1' -1
0"
±4'-0"
(E)W24X68
(E)W24X68
(E)W14X22
(E)W14X22
(E )W
X1
(E )W
X1
(E )W
X1
(E )W
X1
(E )W
X1
(E )W
X1
(E)W12X14
(E )W
X1
(E )W
X1
(E )W
X1
(E )W
X1
(E)1'-9" THK.
CONC.
ROOF SLAB
A
C
C B
B
DO NOT UNDERMINE
EXISTING BEAM BEARING
1'-11" V.I.F.
1' -9
V.
I.F
(E)RAMP DN
INDICATES
ASSUMED EXISTING
LEAD SHIELDING TYP.TEMPORARILY REMOVE
EXISTING LEAD
SHIELDING & REPLACE
WHEN FINISHED WITH WORK
EXISTING
4" LEAD
(V.I.F.)
±5'-3"
EXISTING
± 8" DEEP PIT
EXISTING
± 1'-1" DEEP PIT
PARTIAL FOUNDATION DEMO PLAN AT VAULT
SCALE: 14" = 1'-0"
NOTES:
- CONTRACTOR TO PROVIDE TEMPORARY SHORING AS REQ'D TYP. REFER TO STRUCTURAL NOTES FOR ADDITIONAL REQUIREMENTS.
S101
1'-11" V.I.F.
INDICATES AREA OF
SLAB TO BE CUT & REMOVED
INDICATES AREA OF
SLAB TO BE CUT
& REMOVED
INDICATES PORTION OF
EXISTING CONCRETE WALL
AND LEAD SHIELDING
TO BE REMOVED. TEMP.
SHORING OF ROOF IS TO
BE PROVIDED BY THE
CONTRACTOR. (SEE NOTES)
NEW EQUIPMENT
CENTERLINE
(COORD. W/
ARCH. & VARIAN)
EXISTING WALL
& FTG. TO REMAIN
5' -1
5'
-1
2'-33 4"
1'-0"
±8'-113 4" (V.I.F.)
±3 '-5
8"
(V .I.
F.
EXISTING 1'-0" THICK
SLAB-ON-GRADE (V.I.F.)
FIN. FLR. EL. = ±26'-0"
MODIFICATIONS
TO LEAD SUPPORT
/ BRACING, IF REQ'D,
IS BY OTHERS.
HSS6X6X1
FILLER
0"
CLR
(E)10" THK.
CONC.
ROOF SLAB
(E)1'-3" THK.
CONC.
ROOF SLAB
OCCUPANCY CATEGORY: IV
VAULT FLOOR LIVE LOAD: 60 PSF
GROUND SNOW LOAD: 25 PSF
THERMAL FACTOR: 1.0
EXPOSURE FACTOR: 1.0
IMPORTANCE FACTOR: 1.2
FLAT ROOF SNOW LOAD: 21 PSF (PLUS DRIFT)
BASIC WIND SPEED: 98 MPH (AT 33FT) (ASD)
IMPORTANCE FACTOR: 1.15
EXPOSURE: C
SITE CLASS: D (ASSUMED)
Ss: 0.2775 S1: 0.0714 Sds: 0.292 Sd1: 0.114
SEISMIC DESIGN CATEGORY: C
SEISMIC IMPORTANCE FACTOR: 1.5
VARIAN EQUIPMENT FLOOR LOADS
(SEE FOUNDATION PLAN):
LOCATION A: 945 LBS.
LOCATION B: 3,845 LBS.
LOCATION C: 1,754 LBS.
BUILDING DESIGN LOADS:
DESIGN STRESSES
CONCRETE (STRENGTH DESIGN) MINIMUM COMPRESSIVE STRENGTH IN 28 DAYS:
ALL INTERIOR CONCRETE-------------------------------------------------------------------------F'C=4,000PSI
REINFORCING BARS (ASTM A615 OR ASTM A706, GRADE 60)--------------------------------FY=60,000PSI
STRUCTURAL STEEL W AND S SHAPES
(ASTM A992 OR ASTM A572/50)-------------------------------------------------------------------FY=50,000PSI STRUCTURAL STEEL OTHER SHAPES (ASTM A36)------------------------------------------------FY=36,000PSI ANCHOR RODS (F1554, GRADE 36) UNLESS OTHERWISE NOTED-------------------------- FY=36,000PSI ASSUMED ALLOWABLE BEARING STRESS AT VAULT SLAB-----------------------------------2,000 PSF
CODES AND STANDARDS
1. BUILDING HAS BEEN DESIGNED TO, AND SHALL BE CONSTRUCTED IN ACCORDANCE WITH
THE FOLLOWING BUILDING CODES AND STANDARDS:
1.1. 2014 NEW YORK CITY BUILDING CODE
2. UNLESS EXPLICITLY MODIFIED IN THE CONTRACT DRAWINGS AND SPECIFICATIONS, THE
CONTRACTOR SHALL COMPLY WITH THE FOLLOWING STANDARDS:
2.1. ASCE/SEI 7-05, MINIMUM DESIGN LOADS FOR BUILDINGS AND OTHER STRUCTURES.
2.2. ACI 301-10, SPECIFICATIONS FOR STRUCTURAL CONCRETE.
2.3. ACI 318-14, BUILDING CODE REQUIREMENTS FOR STRUCTURAL CONCRETE.
2.4. ACI 530-13 / ASCE 5-13 / TMS 402-2013, BUILDING CODE REQUIREMENTS FOR MASONRY
STRUCTURES.
2.5. ANSI/AISC 341-10, SEISMIC PROVISIONS FOR STRUCTURAL STEEL BUILDINGS.
2.6. ANSI/AISC 360-10, SPECIFICATION FOR STRUCTURAL STEEL BUILDINGS.
2.7. AWS D1.4/D1.4M-2011, STRUCTURAL WELDING CODE - STEEL.
2.8. SJI 44TH EDITION CATALOG OF STANDARD SPECIFICATIONS AND LOAD TABLES FOR
STEEL JOISTS AND JOIST GIRDER.
2.9. SDI - NO. MOC2, SDI MANUAL OF CONSTRUCTION WITH STEEL DECK, 2006.
2.10. AISI S100-12, NORTH AMERICAN SPECIFICATION FOR DESIGN OF COLD-FORMED STEEL
STRUCTURE MEMBERS.
CONCRETE CONSTRUCTION
1. ALL CONCRETE CONSTRUCTION SHALL BE IN ACCORDANCE WITH THE LATEST BUILDING CODES REQUIREMENTS FOR REINFORCED CONCRETE ACI 318 AND ACI DETAIL MANUAL, EXCEPT THAT CONSTRUCTION AND REMOVAL OF
FORMS AND RESHORING SHALL BE INSPECTED BY THE CONTRACTOR'S ENGINEER.
2. REINFORCING STEEL SHALL HAVE THE FOLLOWING MINIMUM COVERAGE. PLACE BARS AS NEAR TO THE
CONCRETE SERVICE AS THE MINIMA PERMIT WHEREVER POSSIBLE, UNLESS NOTED OTHERWISE:
2.1. CONCRETE POURED AGAINST EARTH 3"
2.2. FORMED CONCRETE IN CONTACT WITH EARTH 2"
2.3. EXTERIOR FACE OF WALLS 2"
2.4. ALL OTHER WALL FACES AND SLABS 3
4"
2.5. BEAMS AND COLUMNS 1 12" (2" FOR EXTERIOR)
3. WELDED WIRE REINFORCEMENT FOR SLABS ON GROUND SHALL HAVE A MINIMUM COVERAGE OF 1" AND MAXIMUM TOP COVERAGE OF 1 12", UNLESS OTHERWISE NOTED. REINFORCEMENT SHALL BE POSITIVELY
SUPPORTED AND MAINTAINED IN THIS POSITION DURING PLACEMENT OF CONCRETE.
4. FURNISH BAR SUPPORTS WHERE NECESSARY DURING CONSTRUCTION.
5. PROVIDE PLASTIC-COATED (NOT PLASTIC-TIPPED) OR STAINLESS STEEL CHAIRS IN ALL CONCRETE EXPOSED TO
VIEW IN COMPLETED STRUCTURE.
6. PROVIDE PIPE SLEEVES AND INSERTS IN CONCRETE WORK WHERE REQUIRED. SEE ARCHITECTURAL AND
MECHANICAL DRAWINGS.
7. UNLESS NOTED OTHERWISE, PROVIDE THE FOLLOWING MINIMUM REINFORCING:
7.1. SLAB ON GROUND, SLAB ON DECK, AND TOPPINGS (2" MINIMUM): 6 X 6-W1.4 X W1.4 WELDED WIRE
REINFORCEMENT IN FLAT SHEETS.
7.2. WALLS: #4 @ 12" ON CENTER EACH WAY (FOR EACH 6" OF WALL THICKNESS).
8. CONSTRUCTION JOINTS SHALL BE POSITIONED SO AS NOT TO CHANGE THE STRUCTURAL DESIGN
REQUIREMENTS. FRAMED FLOORS AND ROOFS SHALL HAVE CONSTRUCTION JOINTS SO THAT THE RATIO OF LENGTH TO WIDTH OF A SINGLE POUR SHALL NOT EXCEED 2. THE LOCATION AND SIZE OF ALL CONSTRUCTION JOINTS SHALL BE APPROVED BY THE ENGINEER. SUBMIT PROPOSED POUR LAYOUT FOR ENGINEER'S REVIEW
AND APPROVAL TWO WEEKS PRIOR TO PLACING CONCRETE.
9. WELDING REINFORCING BARS (INCLUDING TACK WELDING) IS NOT PERMITTED WITHOUT PERMISSION OF ENGINEER IN WRITING. WHERE AND WHEN PERMITTED, WELDING REBARS SHALL COMPLY WITH ASTM A706 (FY=60 KSI) AND WELDING SHALL CONFORM TO AWS D1.4. WELDING SHALL BE PERFORMED BY CERTIFIED WELDERS.
10. PROVIDE HORIZONTAL KEYWAYS IN CONSTRUCTION JOINTS IN SUPPORTED SLABS AND WALL FOOTINGS;
MINIMUM 1 12" DEPTH WITH HEIGHT EQUAL TO ONE-THIRD OF MEMBER DEPTH, UNLESS OTHERWISE SHOWN OR
NOTED.
11. ALL EXPOSED CORNERS OF COLUMNS, WALLS, AND STEEL MEMBERS ENCASED IN CONCRETE ARE TO BE
CHAMFERED 45°. MINIMUM CHAMFER TO BE 12".
12. UNLESS NOTED OTHERWISE IN PROJECT SPECIFICATIONS OR DRAWINGS, ALL EXPOSED CONCRETE SUBJECTED TO FREEZING AND THAWING SHALL HAVE A MINIMUM CEMENT CONTENT OF 610 POUNDS PER YARD, A MAXIMUM
WATER/CEMENT RATIO OF 0.40, AND 6% ±1% OF ENTRAINED AIR.
13. SLAB THICKNESS INDICATED OVER STEEL FORM DECK INCLUDES FORM DEPTH. PROVIDE FORM DECK CONTINUOUS OVER AT LEAST THREE SUPPORTS AND WELD TO EACH SUPPORT AT MINIMUM 12" INTERVALS. USE DECKING THAT WILL SAFELY SUPPORT ALL CONSTRUCTION LOADS, INCLUDING WET WEIGHT OF CONCRETE. LIMIT DEFLECTION OF DECKING TO 1/240 OF SPAN WHEN LOADED WITH WET WEIGHT OF CONCRETE AND
CONSTRUCTION LOADS
14. BEND ALL HORIZONTAL WALL AND FOOTING BARS 1'-0" AROUND CORNERS OR PROVIDE CORNER BARS WITH 2'-0"
LAP.
15. KEY AND DOWEL ALL AREAWAYS AND OTHER PROJECTING ELEMENTS TO SUPPORTING WALLS WITH #4 @ 12" ON
CENTER EXTENDING 1'-0" INTO SUPPORTING WALLS UNLESS NOTED
16. PROVIDE FOUNDATION DOWELS FOR ALL WALLS, PIERS, AND COLUMNS SAME SIZE AND SPACING AS VERTICAL
STEEL.
17. HORIZONTAL FLOORS SUPPORTED BY DEFLECTING STRUCTURAL MEMBERS (STEEL JOISTS, UNSHORED STEEL BEAMS, ETC) SHALL BE FINISHED LEVEL. THE SLAB THICKNESS NOTED IS MINIMUM. ADD CONCRETE AS NECESSARY TO OVERCOME MEMBER DEFLECTIONS. SHORED CONSTRUCTION SHALL BE FINISHED TO A
CONSTANT DEPTH.
18. ALL BARS INTERRUPTED BY STRUCTURAL STEEL SHALL EXTEND TO WITHIN 1" OF STRUCTURAL STEEL FLANGE OR
WEB AND HAVE A 90° HOOK UNLESS OTHERWISE SHOWN.
19. DRAWINGS SHOW TYPICAL REINFORCING CONDITIONS. CONTRACTOR SHALL PREPARE DETAILED PLACEMENT DRAWINGS OF ALL CONDITIONS SHOWING QUANTITY, SPACING, SIZES, CLEARANCES, LAPS, INTERSECTIONS, AND COVERAGE REQUIRED BY THE STRUCTURAL DETAILS, APPLICABLE CODE, AND TRADE STANDARDS. CONTRACTOR SHALL NOTIFY REINFORCING INSPECTOR OF ANY ADJUSTMENTS FROM TYPICAL CONDITIONS WHICH ARE PROPOSED IN PLACEMENT DRAWINGS TO FACILITATE FIELD PLACEMENT OR REINFORCING STEEL AND
CONCRETE.
20. BAR BENDS ARE TO BE MADE COLD. BARS SHALL NOT BE BENT AFTER ANY PORTION OF THE BAR IS ENCASED IN
CONCRETE.
21. SPLICES (GRADE 60 DEFORMED BARS):
21.1. LAP ALL COMPRESSION SPLICES 30 BAR DIAMETERS OF THE LARGER BAR.
21.2. LAP ALL TENSION SPLICES IN ACCORDANCE WITH THE…
This is the start of the file's text. The full file is on GovTribe.