This document is a technical drawing set for the HFIR Instrument CG-2 GP-SANS V-Cavity Magnetic Casing Assembly, a component for Oak Ridge National Laboratory's High Flux Isotope Reactor.
The assembly consists of multiple detailed engineering drawings (CG2-M-20-U-0018 through CG2-M-20-U-0030) that specify fabrication requirements for the complete casing system. Key components include: V-cavity aluminum walls (6061 aluminum), steel walls (A36 steel), magnetic casing magnet assemblies incorporating neodymium block magnets, borated aluminum shielding panels (31% boron-10 composition), steel lifting blocks, temporary aluminum plates for assembly, and various fasteners including stainless steel hex head cap screws and American Drill and Bushing swivel hoist rings. The assembly weighs approximately 71 pounds and uses electroless nickel coating per ASTM B733 specifications with 0.001-inch thickness classification SC3 (moderate service), Type 1, post-heat treatment Class 1. Critical fabrication notes specify handling precautions for neodymium magnets, material cleanliness requirements using water-soluble non-sulfurous cutting oils, deionized water rinsing, and specific packaging in polyethylene bags to prevent vacuum volume contamination. All drawings reference drawing number CG2-M-20-U-0018 as the parent assembly with revision 0 dated October 22, 2025, and prohibit vendor substitution without prior engineering approval. Tolerances follow ASME Y14.100 drafting standards and Y14.5 geometric dimensioning specifications.
View the file
Other files
Show all 12
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
7 6 5 4 3 2 18
A
C
D
A
B
C
D
B
( )6.43
( )59.06
( )3.49
NO REPRESENTATION OR WARRANTY, EXPRESSED OR IMPLIED, IS MADE AS TO THE ACCURACY, COMPLETENESS OR USEFULNESS OF THE INFORMATION OR STATEMENTS CONTAINED IN THESE DRAWINGS, OR THAT THE DISCLOSURE OF ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS MAY NOT INFRINGE PRIVATE RIGHTS OF OTHERS. NO LIABILITY IS ASSUMED
WITH RESPECT TO THE USE OF, OR FOR DAMAGES RESULTING FROM THE USE OF, ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS. DRAWINGS MADE AVAILABLE FOR INFORMATION TO BIDDER ARE NOT TO BE USED FOR OTHER PURPOSES AND ARE TO BE RETURNED UPON REQUEST OF THE FORWARDING CONTRACTOR.
THIS DRAWING IS THE PROPERTY OF OAK RIDGE NATIONAL LABORATORY.
APPROVAL BLOCK
REV DESCRIPTION OF CHANGE DES CHK ENG AP1 AP2 AP3 DATE
0 ORIGINAL ISSUE PER DCN CG2-DCN-00-0006 10/22/2025
NEXT USED ASSEMBLY
UNITS: INCHES
DEFAULT TOLERANCES
X.XX �#�.01
X.XXX �#�.005
ANGULAR �#�.5
BREAK ALL SHARP EDGES
EXCEPT AS NOTED: .06 MAX
SURFACE
ROUGHNESS: 125 RMS U.N.O.
DRAFTING STD: ASME Y14.100
GDT: ASME Y14.5
WELD SYMBOLS: AWS A2.4
THIRD
ANGLE
PROJECTION
HFIR INSTRUMENT CG-2 GP-SANS
20 - OPTICAL
V-CAVITY MAGNETIC CASING ASSEMBLY
DRAWING NUMBER REV
CG2-M-20-U-0018 0
SIZE SCALE WEIGHT SHEET
D 1:4 71 LB 1 of 1VENDOR
MATERIAL
SPECIFICATION
DESCRIPTION PART NUMBER QTY
FIND
NO.
V-CAVITY MAGNETIC CASING MAGNET ASSEMBLY CG2-M-20-U-0023 2 1
V-CAVITY CASING STEEL WALL & LIFTING BLOCK ASSEMBLY CG2-M-20-U-0026 1 2
STEEL
A36 (K02600)
STEEL WALL FOR MAGNETIC CASING CG2-M-20-U-0024 1 3
McMASTER-CARR
WWW.MCMASTER.COM
18-8 STAINLESS STEEL
ASME B18.3, ASTM F879
FHSCS, 1/4- 20 UNC-3A X 0.75" LG 92210A540 8 4
MCMASTER-CARR
WWW.MCMASTER.COM
316 SS
FHSHC SS #8-32 UNC X 3/4 LG 92210A197 16 5
AMERICAN DRILL AND BUSHING
https://americandrillbushing.com/
ALLOY STEEL
AMERICAN DRILL AND BUSHING SWIVEL HOIST RING 23053 23053 2 6
AL EN6061 31% BORON-10
BORATED ALUMINUM FOR V-CAVITY SHIELDING STEEL WALL SIDE CG2-M-20-U-0029 2 7
AL EN6061 31% BORON-10
BORATED ALUMINUM FOR V-CAVITY SHIELDING ALUMINUM WALL SIDE CG2-M-20-U-0030 2 8
NOTES:
1. VENDORS AND VENDOR SPECIFIC ITEMS MAY NOT BE SUBSTITUTED WITHOUT PRIOR ENGINEERING APPROVAL.
2. CAUTION: HANDLING OF NEODYMIUM MAGNETS CAN CAUSE INJURY, REVIEW SAFETY DATA SHEET
AND ENSURE APPROPRIATE CONTROL OF HAZARDS ARE IN PLACE PRIOR TO HANDLING.
SCALE 1:2
6s9 Initials
A
C
D
A
B
C
D
B
98X 0.31
99X 0.28
99X 0.06
99X 0.15
0.75
59.06
0.65
99X 0.53
6X 0.38
2X 12.81
2X
396X R0.0156
6x 0.38
2X 0.25
11.64
23.37
35.1
46.83
12.22
23.95
35.69
47.42
NO REPRESENTATION OR WARRANTY, EXPRESSED OR IMPLIED, IS MADE AS TO THE ACCURACY, COMPLETENESS OR USEFULNESS OF THE INFORMATION OR STATEMENTS CONTAINED IN THESE DRAWINGS, OR THAT THE DISCLOSURE OF ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS MAY NOT INFRINGE PRIVATE RIGHTS OF OTHERS. NO LIABILITY IS ASSUMED
WITH RESPECT TO THE USE OF, OR FOR DAMAGES RESULTING FROM THE USE OF, ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS. DRAWINGS MADE AVAILABLE FOR INFORMATION TO BIDDER ARE NOT TO BE USED FOR OTHER PURPOSES AND ARE TO BE RETURNED UPON REQUEST OF THE FORWARDING CONTRACTOR.
REV DESCRIPTION OF CHANGE DES CHK ENG AP1 AP2 AP3 DATE
0 ORIGINAL ISSUE PER DCN CG2-DCN-00-0006 10/22/2025
NEXT USED ASSEMBLY
CG2-M-20-U-0023
ANGULAR �#�.5�$�
BREAK ALL SHARP EDGES
EXCEPT AS NOTED: .06 MAX
SURFACE
ROUGHNESS: 125 RMS U.N.O.
DRAFTING STD: ASME Y14.100
GDT: ASME Y14.5
WELD SYMBOLS: AWS A2.4
THIRD
ANGLE
PROJECTION
HFIR INSTRUMENT CG-2 GP-SANS
20 - OPTICAL
V-CAVITY ALUMINUM WALL FOR MAGNETIC CASING
DRAWING NUMBER REV
CG2-M-20-U-0019 0
SIZE SCALE WEIGHT SHEET
D 1:4 6 LB 1 of 1
SPECIFICATIONMATERIAL
6061ALUMINUM
FRONT VIEW
BACK VIEW
NOTES:
1 METAL STAMP OR ETCH " CG2-M-20-U-0019 " IN 1/4 INCH MINIMUM HIGH LETTERS ON BOTTOM SURFACE.
2. REMOVE ALL CONTAMINATION E.G. SCALE, DIRT, GREASE AND OIL PRIOR TO FABRICATION BY WIPING
MATERIAL DOWN WITH A CLEAN CLOTH OR PAPER TOWEL.
3. USE ONLY WATER SOLUBLE, NON-SULFUROUS BEARING CUTTING OILS FOR MACHINING.
4. WASH ALL PARTS WITH DETERGENT TO REMOVE CUTTING OILS AND RINSE WITH DE-IONIZED WATER.
5. PART SHALL BE PACKAGED IN A WAY TO PROTECT ALL SURFACES AND TO STOP CONTAMINATION FROM
ENTERING VACUUM VOLUME. SEALING ITEMS IN A POLYETHYLENE BAG IS PREFERRED.
SEE DETAIL A 2X
1/4-20 UNC - 2B TAP ↧ 0.48
#7 DRILL ( 0.201 ) ↧ 0.603
4X
8-32 UNC - 2B TAP ↧ 0.6
#29 DRILL ( 0.136 ) ↧ 0.7 -( 4 ) HOLES
2X
1/4-20 UNC - 2B TAP ↧ 0.482
#7 DRILL ( 0.201 ) ↧ 1
DETAIL A
SCALE 3:1
2X
1/4-20 UNC - 2B TAP ↧ 0.482
#7 DRILL ( 0.201 ) ↧ 0.603
4X
8-32 UNC - 2B TAP ↧ 0.6
#29 DRILL ( 0.136 ) ↧ 0.7 -( 4 ) HOLES
2X
1/4-20 UNC - 2B TAP ↧ 0.48
#7 DRILL ( 0.201 ) ↧ 0.603
4 3 2 1
A
C
D
4 3 2 1
A
B
C
D
B
NO REPRESENTATION OR WARRANTY, EXPRESSED OR IMPLIED, IS MADE AS TO THE ACCURACY, COMPLETENESS OR USEFULNESS OF THE INFORMATION OR STATEMENTS CONTAINED IN THESE DRAWINGS, OR THAT THE DISCLOSURE OF ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS MAY NOT INFRINGE PRIVATE RIGHTS OF OTHERS. NO LIABILITY IS ASSUMED
WITH RESPECT TO THE USE OF, OR FOR DAMAGES RESULTING FROM THE USE OF, ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS. DRAWINGS MADE AVAILABLE FOR INFORMATION TO BIDDER ARE NOT TO BE USED FOR OTHER PURPOSES AND ARE TO BE RETURNED UPON REQUEST OF THE FORWARDING CONTRACTOR.
REV DESCRIPTION OF CHANGE DES CHK ENG AP1 AP2 AP3 DATE
0 ORIGINAL ISSUE PER DCN CG2-DCN-00-0006 10/29/2025
NEXT USED ASSEMBLY
CG2-M-20-U-0026
ANGULAR �#�.5�$�
BREAK ALL SHARP EDGES
EXCEPT AS NOTED: .06 MAX
SURFACE
ROUGHNESS: 125 RMS U.N.O.
DRAFTING STD: ASME Y14.100
GDT: ASME Y14.5
WELD SYMBOLS: AWS A2.4
THIRD
ANGLE
PROJECTION
HFIR INSTRUMENT CG-2 GP-SANS
20 - OPTICAL
V-CAVITY & MAGNETIC HOUSING LIFTING BLOCK
DRAWING NUMBER REV
CG2-M-20-U-0022 0
SIZE SCALE WEIGHT SHEET
C 2/1 1 LB 1 of 1
SPECIFICATIONMATERIAL
A36STEEL
TAP DRILL THRU
3/8-16 UNC-2B THRU
A
C
D
A
B
C
D
B
( )0.8
( )2.62
( )59.3
NO REPRESENTATION OR WARRANTY, EXPRESSED OR IMPLIED, IS MADE AS TO THE ACCURACY, COMPLETENESS OR USEFULNESS OF THE INFORMATION OR STATEMENTS CONTAINED IN THESE DRAWINGS, OR THAT THE DISCLOSURE OF ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS MAY NOT INFRINGE PRIVATE RIGHTS OF OTHERS. NO LIABILITY IS ASSUMED
WITH RESPECT TO THE USE OF, OR FOR DAMAGES RESULTING FROM THE USE OF, ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS. DRAWINGS MADE AVAILABLE FOR INFORMATION TO BIDDER ARE NOT TO BE USED FOR OTHER PURPOSES AND ARE TO BE RETURNED UPON REQUEST OF THE FORWARDING CONTRACTOR.
REV DESCRIPTION OF CHANGE DES CHK ENG AP1 AP2 AP3 DATE
0 ORIGINAL ISSUE PER DCN CG2-DCN-00-0006 10/22/2025
NEXT USED ASSEMBLY
CG2-M-20-U-0018
ANGULAR �#�.5�$�
BREAK ALL SHARP EDGES
EXCEPT AS NOTED: .06 MAX
SURFACE
ROUGHNESS: 125 RMS U.N.O.
DRAFTING STD: ASME Y14.100
GDT: ASME Y14.5
WELD SYMBOLS: AWS A2.4
THIRD
ANGLE
PROJECTION
HFIR INSTRUMENT CG-2 GP-SANS
20 - OPTICAL
V-CAVITY MAGNETIC CASING MAGNET ASSEMBLY
DRAWING NUMBER REV
CG2-M-20-U-0023 0
SIZE SCALE WEIGHT SHEET
D 3:10 12 LB 1 of 1VENDOR
MATERIAL
SPECIFICATION
DESCRIPTION PART NUMBER QTY
FIND
NO.
ALUMINUM
V-CAVITY ALUMINUM WALL FOR MAGNETIC CASING CG2-M-20-U-0019 1 1
K&J MAGNETICS INC.
https://www.kjmagnetics.com/
NEODYMIUM
NEODYMIUM BLOCK MAGNET - 1/2" LEN X 1/4" WID X 1/2" THI B84Y0 99 2
AL 6061 (A96061)
Temporary Plates for Magnetic Casing Steel Wall CG2-M-20-U-0025 2 3
MCMASTER-CARR
https://www.mcmaster.com/
Thumb Screw, SS, 1/4"-20, 1/2" Long 91746A421 4 4
NOTES:
1. VENDORS AND VENDOR SPECIFIC ITEMS MAY NOT BE SUBSTITUTED WITHOUT PRIOR ENGINEERING APPROVAL.
2. CAUTION: HANDLING OF NEODYMIUM MAGNETS CAN CAUSE INJURY, REVIEW SAFETY DATA SHEET
AND ENSURE APPROPRIATE CONTROL OF HAZARDS ARE IN PLACE PRIOR TO HANDLING.
A
C
D
A
B
C
D
B
0.25
2.66
7.28
11.64
12.22
23.37
23.95
29.53
35.1
35.69
46.83
47.42
51.77
56.4
58.81
59.06
2X 0.75
0.38
3.49
0.03
NO REPRESENTATION OR WARRANTY, EXPRESSED OR IMPLIED, IS MADE AS TO THE ACCURACY, COMPLETENESS OR USEFULNESS OF THE INFORMATION OR STATEMENTS CONTAINED IN THESE DRAWINGS, OR THAT THE DISCLOSURE OF ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS MAY NOT INFRINGE PRIVATE RIGHTS OF OTHERS. NO LIABILITY IS ASSUMED
WITH RESPECT TO THE USE OF, OR FOR DAMAGES RESULTING FROM THE USE OF, ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS. DRAWINGS MADE AVAILABLE FOR INFORMATION TO BIDDER ARE NOT TO BE USED FOR OTHER PURPOSES AND ARE TO BE RETURNED UPON REQUEST OF THE FORWARDING CONTRACTOR.
REV DESCRIPTION OF CHANGE DES CHK ENG AP1 AP2 AP3 DATE
0 ORIGINAL ISSUE PER DCN CG2-DCN-00-0006 10/27/2025
NEXT USED ASSEMBLY
CG2-M-20-U-0026
ANGULAR �#�.5�$�
BREAK ALL SHARP EDGES
EXCEPT AS NOTED: .06 MAX
SURFACE
ROUGHNESS: 125 RMS U.N.O.
DRAFTING STD: ASME Y14.100
GDT: ASME Y14.5
WELD SYMBOLS: AWS A2.4
THIRD
ANGLE
PROJECTION
HFIR INSTRUMENT CG-2 GP-SANS
20 - OPTICAL
STEEL WALL FOR MAGNETIC CASING
DRAWING NUMBER REV
CG2-M-20-U-0024 0
SIZE SCALE WEIGHT SHEET
D 1/4 21 LB 1 of 1
SPECIFICATIONMATERIAL
A36 (K02600)STEEL
NOTES:
1. METAL STAMP OR ETCH " CG2-M-20-U-0024 " IN 1/4 INCH MINIMUM HIGH LETTERS IN AREA SHOWN.
2. REMOVE ALL CONTAMINATION E.G. SCALE, DIRT, GREASE AND OIL PRIOR TO FABRICATION BY WIPING
MATERIAL DOWN WITH A CLEAN CLOTH OR PAPER TOWEL.
3. USE ONLY WATER SOLUBLE, NON-SULFUROUS BEARING CUTTING OILS FOR MACHINING.
4. WASH ALL PARTS WITH DETERGENT TO REMOVE CUTTING OILS AND RINSE WITH DE-IONIZED WATER.
5. THIS ASSEMBLY SHALL BE ELECTROLESS NICKEL COATED IN ACCORDANCE WITH ASTM B733. THE COATING THICKNESS SHALL BE .001 INCHES CLASSIFICATION: SC3 (MODERATE SERVICE), TYPE1, POST HEAT TREATMENT CLASS 1. THE COATING SHALL COVER ALL EXPOSED AREAS. ALL THREADED AND MACHINED HOLES SHALL BE MASKED TO PREVENT COATING MATERIAL FROM IMPINGING UPON MACHINED AREAS.
SEE DETAIL B
DETAIL B
SCALE 1/2
SCALE 1/1
C
D
4 3 2 1
A
B
C
D
B
.13
.75
59.06
2X .352X
.25
NO REPRESENTATION OR WARRANTY, EXPRESSED OR IMPLIED, IS MADE AS TO THE ACCURACY, COMPLETENESS OR USEFULNESS OF THE INFORMATION OR STATEMENTS CONTAINED IN THESE DRAWINGS, OR THAT THE DISCLOSURE OF ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS MAY NOT INFRINGE PRIVATE RIGHTS OF OTHERS. NO LIABILITY IS ASSUMED
WITH RESPECT TO THE USE OF, OR FOR DAMAGES RESULTING FROM THE USE OF, ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS. DRAWINGS MADE AVAILABLE FOR INFORMATION TO BIDDER ARE NOT TO BE USED FOR OTHER PURPOSES AND ARE TO BE RETURNED UPON REQUEST OF THE FORWARDING CONTRACTOR.
REV DESCRIPTION OF CHANGE DES CHK ENG AP1 AP2 AP3 DATE
0 ORIGINAL ISSUE PER DCN CG2-DCN-00-0006 10/22/2025
NEXT USED ASSEMBLY
CG2-M-20-U-0023
ANGULAR �#�.5�$�
BREAK ALL SHARP EDGES
EXCEPT AS NOTED: .06 MAX
SURFACE
ROUGHNESS: 125 RMS U.N.O.
DRAFTING STD: ASME Y14.100
GDT: ASME Y14.5
WELD SYMBOLS: AWS A2.4
THIRD
ANGLE
PROJECTION
HFIR INSTRUMENT CG-2 GP-SANS
20 - OPTICAL
Temporary Plates for Magnetic Casing Steel Wall
DRAWING NUMBER REV
CG2-M-20-U-0025 0
SIZE SCALE WEIGHT SHEET
C 3/16 1 LB 1 of 1
SPECIFICATIONMATERIAL
---AL 6061 (A96061)
2X
1/4-20 UNC - 2B CLEAR
F DRILL ( .257 ) THRU
A
C
D
A
B
C
D
B
( )15.38
( )41.68
( )59.06
( ).79
( )3.49( )1.37
8.0.19
NO REPRESENTATION OR WARRANTY, EXPRESSED OR IMPLIED, IS MADE AS TO THE ACCURACY, COMPLETENESS OR USEFULNESS OF THE INFORMATION OR STATEMENTS CONTAINED IN THESE DRAWINGS, OR THAT THE DISCLOSURE OF ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS MAY NOT INFRINGE PRIVATE RIGHTS OF OTHERS. NO LIABILITY IS ASSUMED
WITH RESPECT TO THE USE OF, OR FOR DAMAGES RESULTING FROM THE USE OF, ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS. DRAWINGS MADE AVAILABLE FOR INFORMATION TO BIDDER ARE NOT TO BE USED FOR OTHER PURPOSES AND ARE TO BE RETURNED UPON REQUEST OF THE FORWARDING CONTRACTOR.
REV DESCRIPTION OF CHANGE DES CHK ENG AP1 AP2 AP3 DATE
0 ORIGINAL ISSUE PER DCN CG2-DCN-00-0006 10/22/2025
NEXT USED ASSEMBLY
CG2-M-20-U-0018
ANGULAR �#�.5
BREAK ALL SHARP EDGES
EXCEPT AS NOTED: .06 MAX
SURFACE
ROUGHNESS: 125 RMS U.N.O.
DRAFTING STD: ASME Y14.100
GDT: ASME Y14.5
WELD SYMBOLS: AWS A2.4
THIRD
ANGLE
PROJECTION
HFIR INSTRUMENT CG-2 GP-SANS
20 - OPTICAL
V-CAVITY CASING STEEL WALL & LIFTING BLOCK ASSEMBLY
DRAWING NUMBER REV
CG2-M-20-U-0026 0
SIZE SCALE WEIGHT SHEET
D 1:4 23 LB 1 of 1VENDOR
MATERIAL
SPECIFICATION
DESCRIPTION PART NUMBER QTY
FIND
NO.
STEEL
A36 (K02600)
STEEL WALL FOR MAGNETIC CASING CG2-M-20-U-0024 1 1
STEEL
A36
V-CAVITY & MAGNETIC HOUSING LIFTING BLOCK CG2-M-20-U-0022 2 2
NOTES:
1 VENDORS AND VENDOR SPECIFIC ITEMS MAY NOT BE SUBSTITUTED WITHOUT PRIOR ENGINEERING
APPROVAL.
2. THIS ASSEMBLY SHALL BE ELECTROLESS NICKEL COATED IN ACCORDANCE WITH ASTM B733. THE COATING THICKNESS SHALL BE .001 INCHES CLASSIFICATION: SC3 (MODERATE SERVICE), TYPE 1, POST HEAT TREATMENT CLASS 1. THE COATING SHALL COVER ALL EXPOSED AREAS. ALL THREADED AND MACHINED HOLES SHALL BE MASKED TO PREVENT COATING MATERIAL FROM IMPINGING UPON MACHINED
AREAS.
3 WELD ITEM 2 TO 1 PRIOR TO COATING IN NICKEL.
4. PRIOR TO WELDING AND ELECTROLESS NICKEL COATING CLEAN THE PARTS WITH A MILD DETERGENT AND RINSE WITH DEIONIZED WATER. WIPE SURFACES WITH ISOPROPYL ALCOHOL AND ALLOW TO AIR DRY.
5. ASSEMBLY SHALL BE PACKAGED IN A WAY TO PROTECT ALL SURFACES AND TO STOP CONTAMINATION
FROM ENTERING VACUUM VOLUME. SEALING ITEMS IN A POLYETHYLENE BAG IS PREFERRED.
C
D
4 3 2 1
A
B
C
D
B
10 2
7.28
29.53
51.77
59.06
0.06
NO REPRESENTATION OR WARRANTY, EXPRESSED OR IMPLIED, IS MADE AS TO THE ACCURACY, COMPLETENESS OR USEFULNESS OF THE INFORMATION OR STATEMENTS CONTAINED IN THESE DRAWINGS, OR THAT THE DISCLOSURE OF ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS MAY NOT INFRINGE PRIVATE RIGHTS OF OTHERS. NO LIABILITY IS ASSUMED
WITH RESPECT TO THE USE OF, OR FOR DAMAGES RESULTING FROM THE USE OF, ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS. DRAWINGS MADE AVAILABLE FOR INFORMATION TO BIDDER ARE NOT TO BE USED FOR OTHER PURPOSES AND ARE TO BE RETURNED UPON REQUEST OF THE FORWARDING CONTRACTOR.
REV DESCRIPTION OF CHANGE DES CHK ENG AP1 AP2 AP3 DATE
0 ORIGINAL ISSUE PER DCN CG2-DCN-00-0006 10/27/2025
NEXT USED ASSEMBLY
CG2-M-20-U-0018
ANGULAR �#�.5�$�
BREAK ALL SHARP EDGES
EXCEPT AS NOTED: 1.5 MAX
SURFACE
ROUGHNESS: 3.2 Ra U.N.O.
DRAFTING STD: ASME Y14.100
GDT: ASME Y14.5
WELD SYMBOLS: AWS A2.4
THIRD
ANGLE
PROJECTION
HFIR INSTRUMENT CG-2 GP-SANS
20 - OPTICAL
BORATED ALUMINUM FOR V-CAVITY SHIELDING STEEL WALL SIDE
DRAWING NUMBER REV
CG2-M-20-U-0029 0
SIZE SCALE WEIGHT SHEET
C 1:5 0 LB 1 of 1
SPECIFICATIONMATERIAL
---AL EN6061 31% BORON-10
NOTES:
1. MATERIAL MUST BE BORAL FROM ALUMECO WITH 31% BORON AND A MINIMUM OF 1 MM THICKNESS
3X
TAP DRILL THRU
5/8-11 UNC-2B
C
D
4 3 2 1
A
B
C
D
B
12.81
46.25
59.06
0 1
0.06
NO REPRESENTATION OR WARRANTY, EXPRESSED OR IMPLIED, IS MADE AS TO THE ACCURACY, COMPLETENESS OR USEFULNESS OF THE INFORMATION OR STATEMENTS CONTAINED IN THESE DRAWINGS, OR THAT THE DISCLOSURE OF ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS MAY NOT INFRINGE PRIVATE RIGHTS OF OTHERS. NO LIABILITY IS ASSUMED
WITH RESPECT TO THE USE OF, OR FOR DAMAGES RESULTING FROM THE USE OF, ANY INFORMATION, APPARATUS, METHOD OR PROCESS DISCLOSED IN THESE DRAWINGS. DRAWINGS MADE AVAILABLE FOR INFORMATION TO BIDDER ARE NOT TO BE USED FOR OTHER PURPOSES AND ARE TO BE RETURNED UPON REQUEST OF THE FORWARDING CONTRACTOR.
REV DESCRIPTION OF CHANGE DES CHK ENG AP1 AP2 AP3 DATE
0 ORIGINAL ISSUE PER DCN CG2-DCN-00-0006 10/27/2025
NEXT USED ASSEMBLY
CG2-M-20-U-0018
ANGULAR �#�.5�$�
BREAK ALL SHARP EDGES
EXCEPT AS NOTED: 1.5 MAX
SURFACE
ROUGHNESS: 3.2 Ra U.N.O.
DRAFTING STD: ASME Y14.100
GDT: ASME Y14.5
WELD SYMBOLS: AWS A2.4
THIRD
ANGLE
PROJECTION
HFIR INSTRUMENT CG-2 GP-SANS
20 - OPTICAL
BORATED ALUMINUM FOR V-CAVITY SHIELDING ALUMINUM WALL SIDE
DRAWING NUMBER REV
CG2-M-20-U-0030 0
SIZE SCALE WEIGHT SHEET
C 3/16 0 LB 1 of 1
SPECIFICATIONMATERIAL
---AL EN6061 31% BORON-10
NOTES:
1. MATERIAL MUST BE BORAL FROM ALUMECO WITH 31% BORON AND A MINIMUM OF 1 MM THICKNESS
| Sheet 1 |
| Views |
| FRONT |
| TOP |
| DWG_ISO |
| Zones |
| D8 |
| D7 |
| D6 |
| D5 |
| D4 |
| D3 |
| D2 |
| D1 |
| C8 |
| C7 |
| C6 |
| C5 |
| C4 |
| C3 |
| C2 |
| C1 |
| B8 |
| B7 |
| B6 |
| B5 |
| B4 |
| B3 |
| B2 |
| B1 |
| A8 |
| A7 |
| A6 |
| A5 |
| A4 |
| A3 |
| A2 |
| A1 |
| CG2-M-20-U-0030-R00.pdf |
| Sheet 1 |
| Views |
| FRONT |
| TOP |
| DWG_ISO |
| Zones |
| D4 |
| D3 |
| D2 |
| D1 |
| C4 |
| C3 |
| C2 |
| C1 |
| B4 |
| B3 |
| B2 |
| B1 |
| A4 |
| A3 |
| A2 |
| A1 |
| CG2-M-20-U-0029-R00.pdf |
| Sheet 1 |
| Views |
| FRONT |
| TOP |
| DWG_ISO |
| Zones |
| D4 |
| D3 |
| D2 |
| D1 |
| C4 |
| C3 |
| C2 |
| C1 |
| B4 |
| B3 |
| B2 |
| B1 |
| A4 |
| A3 |
| A2 |
| A1 |
| CG2-M-20-U-0026-R00.pdf |
| Sheet 1 |
| Views |
| FRONT |
| RIGHT |
| DWG_ISO |
| Zones |
| D8 |
| D7 |
| D6 |
| D5 |
| D4 |
| D3 |
| D2 |
| D1 |
| C8 |
| C7 |
| C6 |
| C5 |
| C4 |
| C3 |
| C2 |
| C1 |
| B8 |
| B7 |
| B6 |
| B5 |
| B4 |
| B3 |
| B2 |
| B1 |
| A8 |
| A7 |
| A6 |
| A5 |
| A4 |
| A3 |
| A2 |
| A1 |
| CG2-M-20-U-0025-R00.pdf |
| Sheet 1 |
| Views |
| FRONT |
| TOP |
| DWG_ISO |
| Zones |
| D4 |
| D3 |
| D2 |
| D1 |
| C4 |
| C3 |
| C2 |
| C1 |
| B4 |
| B3 |
| B2 |
| B1 |
| A4 |
| A3 |
| A2 |
| A1 |
| CG2-M-20-U-0024-R00.pdf |
| Sheet 1 |
| Views |
| new_view_4 |
| new_view_5 |
| B |
| new_view_13 |
| Zones |
| D8 |
| D7 |
| D6 |
| D5 |
| D4 |
| D3 |
| D2 |
| D1 |
| C8 |
| C7 |
| C6 |
| C5 |
| C4 |
| C3 |
| C2 |
| C1 |
| B8 |
| B7 |
| B6 |
| B5 |
| B4 |
| B3 |
| B2 |
| B1 |
| A8 |
| A7 |
| A6 |
| A5 |
| A4 |
| A3 |
| A2 |
| A1 |
| CG2-M-20-U-0023-R00.pdf |
| Sheet 1 |
| Views |
| FRONT |
| TOP |
| Explode |
| Zones |
| D8 |
| D7 |
| D6 |
| D5 |
| D4 |
| D3 |
| D2 |
| D1 |
| C8 |
| C7 |
| C6 |
| C5 |
| C4 |
| C3 |
| C2 |
| C1 |
| B8 |
| B7 |
| B6 |
| B5 |
| B4 |
| B3 |
| B2 |
| B1 |
| A8 |
| A7 |
| A6 |
| A5 |
| A4 |
| A3 |
| A2 |
| A1 |
| CG2-M-20-U-0022-R00.pdf |
| Sheet 1 |
| Views |
| FRONT |
| TOP |
| new_view_4 |
| Zones |
| D4 |
| D3 |
| D2 |
| D1 |
| C4 |
| C3 |
| C2 |
| C1 |
| B4 |
| B3 |
| B2 |
| B1 |
| A4 |
| A3 |
| A2 |
| A1 |
| CG2-M-20-U-0019-R00.pdf |
| Sheet 1 |
| Views |
| FRONT |
| BOTTOM |
| new_view_5 |
| A |
| BACK |
| Zones |
| D8 |
| D7 |
| D6 |
| D5 |
| D4 |
| D3 |
| D2 |
| D1 |
| C8 |
| C7 |
| C6 |
| C5 |
| C4 |
| C3 |
| C2 |
| C1 |
| B8 |
| B7 |
| B6 |
| B5 |
| B4 |
| B3 |
| B2 |
| B1 |
| A8 |
| A7 |
| A6 |
| A5 |
| A4 |
| A3 |
| A2 |
| A1 |
File details come from the government source that posted it. Updated .