This is a technical specification document for the Department of Energy's procurement of a low temperature scanning tunneling microscope (STM) to support fundamental research in condensed matter physics and quantum information science.

The document outlines 18 specific technical requirements including: base temperature capability of 1.7K or less, uniaxial magnetic field of at least 5T, compatibility with Scienta Omicron flag-style sample holders, coarse positioning capabilities of 4x4x10 mm, qPlus AFM functionality, atomic spatial resolution with z height noise under 1 pm, liquid helium/nitrogen hold time exceeding 100 hours, and compatibility with Nanonis controller. Additional requirements cover UHV interface for PLD chamber connection, NRTL certification, shipping terms (Incoterms DDP), one-year warranty with parts and service, and onsite operational training at delivery. The microscope must demonstrate turnkey microscopy and tunneling spectroscopy of Pb or NbSe2 films during installation at base temperature.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specification Document: Low Temperature Scanning Tunneling Microscope (STM)

The Company is seeking to purchase a low temperature scanning tunneling microscope (STM) to support fundamental research in condensed matter physics and quantum information science. The microscope must include each the following capabilities and must meet the listed technical specifications:

1. Turnkey microscopy and tunneling spectroscopy of Pb or NbSe2 films demonstrated during the installation by vendor’s service engineer(s) at base temperature.

2. Base temperature of 1.7K or less.

3. Uniaxial magnetic field of at least 5T

4. Compatible with Scienta Omicron flag-style sample holders

5. Coarse positioning of at least 4x4x10 mm and low temperature scan range of at least

1.2x1.2x0.2 microns.

6. qPlus AFM functionality for non-contact AFM.

7. At least 4 additional sample contacts for device transport measurement

8. Atomic spatial resolution with z height noise less than 1 pm at the base temperature

9. Tunneling current less than 1 pA for scanning at the base temperature

10. Liquid helium and liquid nitrogen hold time longer than 100 hours

11. Include or compatible with Nanonis controller

12. Optical access to microscope for in situ imaging with spatial resolution of at least 10 microns.

13. Passive or active vibration isolation system sufficient to allow.

14. UHV interface for connection with existing PLD chamber.

15. NRTL certification of all components prior to delivery of instrument or all necessary information and feedback to complete NRTL certification at the Company’s site.

16. All shipping, insurance, and transportation costs (including import duty) – Incoterms

DDP.

17. One year warranty including all parts and service needed to guarantee microscope performance for the first year after delivery.

18. Onsite training at time of delivery to include all necessary training for operation of microscope.