Medivator Specs.pdf
PDF 691 KB Posted
- Attached to
- Z1DA--608-23-116 SPS Compliance Corrections Construction Federal contract opportunity
- Solicitation number
- 36C24123B0006
About this file
This document contains reprocessor specifications for a Medivators ADVATION PLUS system and information on a related federal contract opportunity for SPS Compliance Corrections Construction with the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 1.
The reprocessor specifications include requirements for electrical rating, power rating, air and water pressure ranges, water flow range, operating temperature range, humidity range, temperature range, dimensions, weight, and clearances. The specifications require a laboratory equipment that operates on 100-240 VAC, 50/60 Hz power with a 1000 Watt rating. It must be capable of functioning within an air pressure range of 87-145 psi, water pressure range of 35-87 psi, and water flow range of 2-8 Liters/minute. The operating and storage temperature ranges specified are 30-40C and 50-80.6F (10-27C) respectively. The unit dimensions are listed as 45"W x 31"D x 62"H and it weighs approximately 400 lbs.
The related federal contract opportunity is solicitation number 36C24123B0006 for SPS Compliance Corrections Construction with the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 1. No further details are provided on the requirements or scope of the opportunity in this document.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C24123B0006 0002.docx | DOCX document | |
| 36C24123B0006 0001.docx | DOCX document | |
| Facility drawings with CSB - bare.pdf | ||
| Med Req 50098-1520-EN US R.pdf | ||
| VFD SW addition Aquavar AV2V2S050D.pdf | ||
| Technical-Information-SSH - existing pumps set 1022-9SH184JM2-2.pdf | ||
| PRE-AWARD CONTRACTOR EVALUATION FORM.pdf | ||
| 36C24123B0006.docx | DOCX document | |
| EndoDry Site Requirements.pdf | ||
| Instrument Air compressor spec sheet.pdf | ||
| New medivators installation 50098-15.pdf | ||
| Sterlizer utility requirements.jpg | JPG image | |
| 608-23-116 project sketches.pdf | ||
| SOW - 608-23-116 SPS steam and sterilization r1.docx | DOCX document | |
| Medivator installation specs.pdf | ||
| VFD old AC shop abb acs550-u1-046a-2.pdf | ||
| PH-WHV-Spec-sheet-LR expansion tank - replace with WHV 75 or eqiv.pdf | ||
| SAM WD NH NH20230021 revised 040723.pdf |
Show all 18
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
REPROCESSOR SPECIFICATIONS
General Specifications
The basic reprocessor specifications required for installation and operation of the Medivators ADVANTAGE PLUS system are summarized in Table 1.
Table 1 Reprocessor Specifications
Equipment Type Laboratory Equipment
Electrical rating 100-240 VAC, 50/60 Hz
Power rating 1000 Watts
Air pressure range 87 —145 psi (6 - 10 bar)
Water pressure range 35 - 87 psi (2 - 6 bar)
Water flow range 2-8 Liters/minute
Water temperature operating range
30°C - 40°C, set mixing valve to 35°C
Humidity range for operation
20-80%, non-condensing
Temperature range for operation
50°F - 80.6°F (10°C - 27°C)
Environment Indoor use only
Installation/Over voltage category
II
Pollution degree 2
Connection to power supply Cord connection
Dimensions 45"W X 31" DX 62" H (inches)
114 WX 79 DX 157 H (cm)
Weight f 400 lbs. (182 kg)
Dimensions and Clearances
Q a a
• 000
Figure 1. Front Dimensions
Figure 2. Side Dimensions & Clearance Air Compressor