Medivator installation specs.pdf
PDF 681 KB Posted
- Attached to
- Z1DA--608-23-116 SPS Compliance Corrections Construction Federal contract opportunity
- Solicitation number
- 36C24123B0006
About this file
This document provides installation specifications for the Medivators ADVANTAGE PLUS System. Key details include electrical requirements of 1000 watts at 100-240 volts AC, potable water delivered at a minimum of 5 L/min at 30 psi and ideally 35C delivered via a mixing valve, drainage of at least 12 L/min, and room ventilation of a minimum 10 air exchanges per hour. The system must connect to a compressed air source of at least 40 L/min at 6 bar. Storage conditions between 0-60C and below 80% maximum relative humidity are specified. Transportation requires removing the base cover if encountering floor level changes and placing on a pallet with kraft liner when moving.
The related federal contract opportunity is solicitation number 36C24123B0006 for Z1DA--608-23-116 SPS Compliance Corrections Construction from the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 1. No further details are provided in the opportunity description.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C24123B0006 0002.docx | DOCX document | |
| 36C24123B0006 0001.docx | DOCX document | |
| Facility drawings with CSB - bare.pdf | ||
| Med Req 50098-1520-EN US R.pdf | ||
| VFD SW addition Aquavar AV2V2S050D.pdf | ||
| Technical-Information-SSH - existing pumps set 1022-9SH184JM2-2.pdf | ||
| PRE-AWARD CONTRACTOR EVALUATION FORM.pdf | ||
| 36C24123B0006.docx | DOCX document | |
| EndoDry Site Requirements.pdf | ||
| Instrument Air compressor spec sheet.pdf | ||
| New medivators installation 50098-15.pdf | ||
| Medivator Specs.pdf | ||
| Sterlizer utility requirements.jpg | JPG image | |
| 608-23-116 project sketches.pdf | ||
| SOW - 608-23-116 SPS steam and sterilization r1.docx | DOCX document | |
| VFD old AC shop abb acs550-u1-046a-2.pdf | ||
| PH-WHV-Spec-sheet-LR expansion tank - replace with WHV 75 or eqiv.pdf | ||
| SAM WD NH NH20230021 revised 040723.pdf |
Show all 18
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
INSTALLATION SPECIFICATIONS
Electrical
The Medivators ADVANTAGE PLUS System must be connected to a single electrical outlet capable of supplying 1000 watts at 100-240 volts AC, 50/60 Hz. The outlet must be properly grounded.
Water
Potable water is the minimum standard. For optimum performance, water must be delivered at a minimum flow rate of 5 L/min (1.32 gpm) at a dynamic pressure of 30 psi (2 bar) during the basin fill. For optimum performance water must be delivered at 35°C by use of a mixing valve (required). The mixing valve must be purchased separately and installed by the facility. A back syphonage prevention device must be purchased separately and installed between the water line & the reprocessor to prevent contamination of the water supply in the event of a sudden drop in water pressure.
Drain
For optimum performance the Medivators ADVANTAGE PLUS System should be connected to a vented sanitary drain system capable of draining at a minimum rate of 12 L/min (3.17 gpm).
Room Ventilation
The Medivators ADVANTAGE PLUS System must be installed in a room with a ventilation system capable of delivering a minimum of 10 air exchanges per hour.
An optional vapor management system is available for ductless fume removal.
The Medivators ADVANTAGE PLUS System must be connected to an air supply for proper operation. The air supply must be capable of delivering house compressed air (dry / oil free) at a minimum of 40 L/min (10.57 gpm) at 6 bar (87 psi).
An optional air compressor system is available for the Medivators ADVANTAGE PLUS reprocessor. The air compressor delivers proper air flows and pressures for correct system operation. The optional air compressor may be plugged into the same outlet as the Medivators ADVANTAGE PLUS unit provided that the outlet is rated for 120 VAC, 20 Ampere service.
Otherwise, the optional air compressor must be plugged into a separate 120 VAC, 15 Ampere circuit independent of the circuit to which the Medivators ADVANTAGE PLUS is attached.
STORAGE
If the Medivators ADVANTAGE PLUS System will be out of service for an extended period of time, the following steps must be followed:
• Remove the side covers, purge all liquids from the system, and then replace the covers.
• Place a covering over the unit to protect it from dust and moisture.
• The storage temperature must not go below 0°C (32°F), or exceed 60°C (140°F).
• The maximum relative humidity must not exceed 80% during storage.
TRANSPORT
The Medivators ADVANTAGE PLUS System can be safely moved on a smooth, level floor.
However, if you encounter any thresholds or floor level changes, the base cover must be removed to avoid damage. For all other moves, place the Medivators ADVANTAGE PLUS System on a pallet with a kraft liner.
Before moving Medivators ADVANTAGE PLUS System, ensure the electrical cord, drain line and water supply are either disconnected or are appropriate lengths to accommodate the relocation of the machine.