This document presents detailed specifications for a tube furnace with specific technical capabilities and requirements. The furnace must operate with temperature uniformity of +/- 10 °C between 300˚C and 1300˚C, reaching a maximum temperature of 1600˚C. It requires a work zone supporting 100 mm with a weight capacity of 250 lbs and has a machine footprint of 27.8" depth, 37" width, and 31.1" height.
The specification mandates several additional features including NRTL certification, safety switch interlocks, programmable temperature profiles, integral over-temperature protection, and a single heating zone design. Other required features include work tube adaptors for 4" tubes, digital over-temperature protection, a touchscreen temperature programmer, and CSA Inspection L1/L2 certification. The document emphasizes safety and precise temperature control, indicating this is likely for a research, industrial, or laboratory environment that demands high-precision thermal processing equipment.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Questions and Answers.docx | DOCX document | |
| Attachment A-Terms and Conditions.pdf | ||
| Attachment B-Prime-Flowdown-Clauses-BSD-CS-2258-10.01.2025.pdf | ||
| Bill of Materials.pdf | ||
| RFP 418338.pdf | ||
| Attachment D-Reps-and-Certs-Abbreviated - 10.01.2025.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Tube Furnace Specifications Capabilities Temperature Uniformity
+/- 10 °C at 300˚C - 1300˚C
Max Temp 1600˚C
Work Zone Must support 100 mm Single zone furnace
Work Zone Weight Cap 250 lbs
Machine Footprint Depth 27.8”
Width 37”
Height 31.1”
Does not include any clearances for work or maintenance zones.
Additional Features Upgrades
• Must include NRTL certification
• Must have included safety switch interlocks to turn off heating elements when furnace opened
• Must include programable temperature profiles and ramp rates
• Integral over temperature and run-away temperature protection
• Tube furnace needs to be of a single heating zone design over the required length
• Work tube adaptors for other tube dimension (4” tubes)
• Digital over temperature protection
• Touchscreen temperature programmer
• CSA Inspection L1/L2