About this file
This technical data sheet from Vaisala details specifications for the BAROCAP Barometer PTB110, which is being procured by the Texas Water Development Board (TWDB) under solicitation #580-25-IFB-3461 for 30 units across Travis County and Austin City jurisdictions. The PTB110 is an industrial barometer designed for accurate pressure measurements at room temperature and environmental monitoring across wide temperature ranges. The device features multiple pressure range options from 500-1100 hPa, output voltage of 0-2.5V or 0-5V, and frequency output of 500-1100 Hz. Bids must be submitted by February 24, 2025, at 2:00pm Central Time, with a question and answer period running February 18-25.
The barometer utilizes Vaisala's BAROCAP silicon capacitive sensor technology and offers accuracy of ±0.3 hPa at +20°C with excellent long-term stability of ±0.1 hPa/year. Technical specifications include IP32 rating, operating temperature range of -40 to +60°C, power consumption of 0.10W at 12V, and dimensions of 97.3 × 68.4 × 28.1 mm. The device is mountable on 35mm DIN rail and includes features like external On/Off control and traceable calibration certification. The procurement follows FOB Destination freight terms and Net30 payment terms, with no specific budget range disclosed. Only exact conforming bids for the specified model will be considered under Texas Government Code Section 2155.067.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
BAROCAPâ Barometer PTB110
For Industrial Use
Features
• Vaisala BAROCAPâ sensor
• Several pressure ranges
• Accuracy ±0.3 hPa at +20 °C
• Long-term stability
• On/Off control with external trigger
• Output voltage 0 … 2.5 or
0 … 5 VDC
• Current consumption less than 4 mA
• Mountable on 35 mm wide DIN rail
• Traceable calibration (certificate included)
Vaisala BAROCAPâ Barometer PTB110 is designed both for accurate barometric pressure measurements at room temperature and for general environmental pressure monitoring over a wide temperature range.
Vaisala BAROCAP Technology PTB110 uses the Vaisala BAROCAP sensor, a silicon capacitive absolute pressure sensor developed by Vaisala for barometric pressure measurement applications. The sensor combines the outstanding elasticity characteristics and mechanical stability of single-crystal silicon with the proven capacitive detection principle.
Applications PTB110 is suitable for a variety of applications, such as environmental pressure monitoring, data buoys, laser interferometers, and agriculture and hydrology. The compact PTB110 is ideal for data logger applications as it has low power consumption. The external On/Off control is practical when electricity supply is limited.
Accuracy and Stability The excellent long-term stability of the barometer minimizes or even removes the need for field adjustment in many applications.
Technical Data
Measurement Performance Pressure range (1 hPa = 1 mbar) 500 … 1 100 hPa
600 … 1100 hPa 800 … 1100 hPa 800 … 1060 hPa 600 … 1060 hPa
Resolution 0.1 hPa
Load resistance 10 000 Ω minimum
Load capacitance 47 nF maximum
Settling time to full accuracy after startup
1 s
Response time to full accuracy after a pressure step
500 ms
Acceleration sensitivity Negligible
Accuracy
Linearity 1) ±0.25 hPa
Hysteresis 1) ±0.03 hPa
Repeatability 1) ±0.03 hPa
Pressure calibration uncertainty 2) ±0.15 hPa
Voltage calibration uncertainty ± 0.7 mV
Frequency calibration uncertainty ± 0.3 Hz
Accuracy at +20 °C (+68 °F) 3) ±0.3 hPa
Total Accuracy at
+15 … +25 °C (+59 … +77 °F) ±0.3 hPa
0 … +40 °C (+32 … +104 °F) ±0.6 hPa
−20 … +45 °C (−4 … +113 °F) ±1.0 hPa
−40 … +60 °C (−40 … +140 °F) ±1.5 hPa
Long-term stability ±0.1 hPa / year
1) Defined as ±2 standard deviation limits of end-point non-linearity, hysteresis error, or repeatability error.
2) Defined as ±2 standard deviation limits of inaccuracy of the working standard including traceability to international standards.
3) Defined as the root sum of the squares (RSS) of end-point non-linearity, hysteresis error, repeatability error, and calibration uncertainty at room temperature.
Operating Environment Operating temperature −40 … +60 °C (−40 … +140 °F)
Storage temperature −40 … +60 °C (−40 … +140 °F)
Operating humidity Non-condensing
EMC compliance EN/IEC 61326-1, Electrical equipment for measurement, control and laboratory use - industrial environment
Mechanical Specifications IP rating IP32
Dimensions (H × W × D) 97.3 × 68.4 × 28.1 mm (3.83 × 2.69 × 1.10 in)
Weight 90 g (3.2 oz)
Materials
Housing cover Plastic ABS/PC blend
Mounting plate Aluminum
Inputs and Outputs Supply voltage 10 … 30 VDC
Supply voltage control With TTL-level (Transistor- Transistor Logic) trigger
Supply voltage sensitivity Negligible
Average power consumption 0.10 W at 12 V
Output voltage 0 … 2.5 VDC
0 … 5 VDC
Output frequency 500 … 1100 Hz
Pressure connector M5 (10 … 32) internal thread
Pressure fitting Barbed fitting for ⅛ in
Minimum pressure limit 0 hPa abs
Maximum pressure limit 2000 hPa abs
Electrical connector A removable connector for 5 wires
(AWG 28 … 16)
Terminals Pin 1: External triggering Pin 2: Signal ground Pin 3: Supply ground Pin 4: Supply voltage Pin 5: Voltage/Frequency output
Published by Vaisala | B210681EN-E © Vaisala 2018
All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. Any reproduction, transfer, distribution or storage of information contained in this document is strictly prohibited. All specifications — technical included — are subject to change without notice.www.vaisala.com http://www.vaisala.com