This technical evaluation document outlines requirements for a handheld x-ray fluorescence (XRF) analyzer to be provided under a solicitation from the Consumer Product Safety Commission. The analyzer will be evaluated based on its ability to accurately measure certified reference materials including various polymers, PVC, glass, and paint films within specified concentration ranges. It must also correctly identify sample matrices and meet targets when used in an automated "test all" mode. Additional requirements include the analyzer weighing no more than 3 pounds with its battery installed. The related solicitation seeks the Thermo Niton XRF XL5 Plus brand or equal for use in evaluating consumer products for hazardous substances.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
ATTACHMENT 2 – Technical Evaluation Handheld XRF Analyzer
Attachment 2
Technical Evaluation: Handheld X-ray Fluorescence
(XRF) Analyzer
The XRF analyzer will be evaluated and performance validated based on the following three (3) factors (A, B, and C):
FACTOR A: The XRF will be evaluated based on measurement results obtained using the following reference standard materials:
Subfactor 1: ERM®-EC681k Polyethylene (high level)
Subfactor 2: ERM®-EC680k Polyethylene (low level)
Subfactor 3: NIST Standard Reference Material® 2861 PVC
Subfactor 4: Restriction of Hazardous Substances (RoHS) 2-3 Glass Standard
Subfactor 5: NIST Standard Reference Material® 2569 Level 3 Lead Paint Films
Subfactor 6: NIST Standard Reference Material® 1412a Glass Standard.
Measurement results shall be within the following low/high limit specifications outlined in the tables below for each of the reference standard materials:
Subfactor 1: ERM®-EC681k Polyethylene (high level) Standard
ERM®-EC681k
Element Certified Value (ppm or mg/kg)
Low Limit (ppm or mg/kg)
High Limit (ppm or mg/kg)
Lead (Pb) 98 74 122
Arsenic (As) 29 15 44
Cadmium (Cd) 137 96 178
Chromium (Cr) 100 50 150
Measurement time will be a minimum of 60 seconds on optimum filter setting configuration for each element
Subfactor 2: ERM®-EC680k Polyethylene (low level) Standard
ERM®-EC680k
Element Certified Value (ppm or mg/kg)
Low Limit (ppm or mg/kg)
High Limit (ppm or mg/kg)
Lead (Pb) 13.6 6 20 configuration for each element
Subfactor 3: NIST Standard Reference Material® 2861 PVC
NIST Standard Reference Material® 2861
Element Certified Value (ppm or mg/kg)
Low Limit (ppm or mg/kg)
High Limit (ppm or mg/kg)
Lead (Pb) 88.3 70 106 configuration
Subfactor 4: RoHS 2-3 Glass Standard
RoHS 2-3 Glass Standard
Element Certified Value (ppm or mg/kg)
Low Limit (ppm or mg/kg)
High Limit (ppm or mg/kg)
Lead (Pb) 1000 Must have a detectable measurement reading >100ppm.
configuration
Subfactor 5: NIST Standard Reference Material® 2569 Level 3 Lead Paint Films
NIST Standard Reference Material® 2569 Level 3 Lead Paint Films
Element Certified Value (ppm or mg/kg)
Low Limit (ppm or mg/kg)
High Limit (ppm or mg/kg)
Lead (Pb) 314 Must have a detectable measurement reading.
Measurement time will be 120 seconds on optimum filter setting configuration
Subfactor 6: NIST Standard Reference Material® 1412a Glass Standard
NIST Standard Reference Material® 1412a Glass Standard
Element Certified Value (ppm or mg/kg)
Low Limit (ppm or mg/kg)
High Limit (ppm or mg/kg)
Aluminum (Al) 46,300 Must have a detectable measurement reading >10,000ppm.
Magnesium (Mg) 23,300 Must have a detectable measurement reading >5,000ppm.
configuration. Up to 300 seconds maximum scan time allowed.
FACTOR B: Evaluate the use of the “test all” mode or equivalent for sample analysis.
The “test all” mode or equivalent will automatically select the mode for the user when the user is uncertain of the appropriate mode to choose. The “test all” mode or equivalent will determine the sample matrix (i.e. metal (or alloy), plastic (or polymer), PVC plastic
(or PVC polymer)) and apply the appropriate algorithms for elemental detection and quantification. The following reference standards will be used to evaluate:
Subfactor 1: ERM®-EC681k Polyethylene (high level)
Subfactor 2: NIST Standard Reference Material® 2861 PVC
Subfactor 3: MBH Analytical Tin Check Sample (180-606)
Subfactor 4: XOS Coated Reference Sample 900071-08 ZINC
The “test all” mode or equivalent will correctly identify the sample matrix and shall provide measurement results within the following low/high limit specifications outlined in the tables below for each of the reference standard materials:
Subfactor 1: ERM®-EC681k Polyethylene (high level) Standard
ERM®-EC681k
Element Sample Matrix
ID
Certified Value (ppm or mg/kg)
Low Limit (ppm or mg/kg)
High Limit (ppm or mg/kg)
Lead (Pb)
Plastic/Polymer
98 74 122
Arsenic (As) 29 15 44
Cadmium (Cd) 137 96 178
Chromium (Cr) 100 50 150
Measurement time will be a minimum of 60 seconds on optimum filter setting configuration for each element
Subfactor 2: NIST Standard Reference Material® 2861 PVC
NIST Standard Reference Material® 2861
Element Sample Matrix ID
Certified Value (ppm or mg/kg)
Low Limit (ppm or mg/kg)
High Limit (ppm or mg/kg)
Lead (Pb) PVC Plastic/ PVC Polymer
88.3 70 106 filter setting configuration
Subfactor 3: MBH Analytical Tin Check Sample (180-606)
MBH Analytical Tin Check Sample (180-606)
Element Sample Matrix ID
Composition
Low Limit (ppm or mg/kg)
High Limit (ppm or mg/kg)
Lead (Pb) Metal/Alloy 0.13 Must have a detectable measurement reading >100ppm.
Subfactor 4: XOS Coated Reference Sample 900071-08 ZINC
XOS Coated Reference Sample 900071-08 ZINC
Element Sample Matrix ID
Certified Value (ppm or mg/kg)
Low Limit (ppm or mg/kg)
High Limit (ppm or mg/kg)
Lead (Pb) Record 287 Must have a detectable measurement reading >90ppm
(0.0090%)
Measurement time will be a minimum of 120 seconds on optimum
FACTOR C: Weight of XRF analyzer (with battery):
No more than 3lbs.