BAM-1022 Continuous Particulate Monitor Bid Specifications
This is a bid specification document from the Alabama Department of Environmental Management (ADEM) for the procurement of a BAM-1022 Continuous Particulate Monitor or equivalent. The document establishes comprehensive technical and general specifications for an EPA Class III Federal Equivalency Method (EQPM-1013-209) monitor utilizing beta-attenuation technology to measure PM2.5 particulate matter concentrations. The monitor must operate within relative humidity levels from 0% to 90% non-condensing, collect samples on mechanically-advanced filter tape, automatically measure and collect hourly averages with user-selectable short-term averages of 5 to 60 minutes, and continuously store at least 22,500 hourly averages. The equipment must include a vacuum pump and sample inlet heater, be housed in a weatherproof enclosure with a separable pump box, and transmit data hourly via Ethernet connection using a wireless modem with data accessibility through USB port or remote access via Comet software.
The monitor specifications require a measurement range of -15 to 10,000 micrograms per cubic meter, operating temperature capability from -30 to +50.0 degrees Celsius, and a sample flow rate of 16.7 liters per minute with ±2.0% tolerance. The equipment must achieve a lower detectable limit of less than 1.0 microgram per cubic meter for 24-hour readings and less than 4.8 micrograms per cubic meter for 1-hour readings, with measurement and data resolution of 0.1 and 1.0 micrograms per cubic meter respectively. The monitor requires 115VAC power, includes a 14C beta source of 60±15 Ci, and utilizes a scintillation photomultiplier tube detector. The equipment must meet or exceed US-EPA Class III PM2.5 Federal Equivalency Method standards for both additive and multiplicative bias. Delivery is required to Gina Curvin at 1350 Coliseum Boulevard, Montgomery, AL 36110.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| ITB Standard Boilerplate.pdf | ||
| Additional Terms & Conditions.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
BAM-1022 CONTINUOUS PARTICULATE (PM2.5) MONITOR
General Specifications:
• The monitor must be an Environmental Protection Agency (EPA) Class III Federal Equivalency Method
(EQPM-1013-209) and must include any accessories necessary to maintain this equivalency.
• The monitor must use the Beta-Attenuation process for determining particulate matter (PM) concentrations.
• The monitor must utilize a very sharp cut cyclone to limit PM size to 2.5 microns or smaller.
• The monitor must have the ability to operate during relative humidity levels from 0% to 90% non-condensing.
• The sample must be collected on filter tape; the tape must be mechanically aligned under the sample inlet after completing a beta ray transmission reading to determine the initial filter tape concentration (zero) and a post sample reading must be collected to determine PM2.5 concentrations of the sample.
• Must have the ability to sample continuously with automatic tape advance.
• Must have the ability to automatically measure and collect hourly averages with user selectable short-term averages of 5 to 60 minutes.
• Must have at least one isolated analog output; switch selectable 0 to 1 VDC, 0 to 2.5 VDC or 0-5 VDC.
• Must have the ability to internally store at least 22,500 hourly averages.
• Must be able to transmit data hourly over Ethernet connection using a wireless modem.
• Must have all data files accessible and retrievable via USB port or access files remotely using Comet software.
• Monitor to be supplied with vacuum pump and sample inlet heater.
• Monitor is integrated into its own, weatherproof enclosure and sits on a separable, pump box.
Technical Specifications:
• RANGES: -15 microgram per cubic meter – 10,000 microgram per cubic meter.
• OPERATING TEMPERATURE: -30 to + 50.0 degrees Celsius.
• SAMPLE FLOW RATE: 16.7 liters/minute +/- 2.0%.
• LOWER DETECTABLE LIMIT: < 1.0 microgram per cubic meter for 24-hour readings and <
4.8 micrograms per cubic meter for 1-hour readings.
• SPAN CHECK: Manual audits performed with zero and span foils (included).
• POWER REQUIREMENTS: 115VAC
• MEASUREMENT RESOLUTION: 0.1 micrograms per cubic meter or better
• DATA RESOLUTION: 1.0 micrograms per cubic meter or better
• ACCURACY: Must meet or exceed US-EPA Class III PM2.5 FEM standards for additive and multiplicative bias
• BETA SOURCE: 14C, 60 µCi ± 15 µCi
• DETECTOR TYPE: Scintillation Photomultiplier Tube
• DATA STORAGE: 22,500 records (2.6 years of 1 record/hour).