81029AttH3 - Technical Specifications - QA QC Criteria.doc

DOC document 238 KB Posted

Attached to
Laboratory Analytical Services for Fish TissuesBid Documents State and local contract opportunity
Solicitation number
RFP 25-81029
Issued by
Marion County, Indiana

About this file

This is a Technical Specifications document (Attachment H3) detailing Quality Assurance/Quality Control criteria for laboratory analytical services for the Indiana Department of Environmental Management Office of Water Quality (IDEM/OWQ). The document outlines four Data Quality Assessment (DQA) levels, with OWQ samples being treated at DQA4 level unless otherwise specified. The specifications detail requirements for data quality control, sampling procedures, analysis methods, and documentation requirements, with contractors required to maintain QA/QC documentation for five years after contract expiration.

The document specifies detailed analytical control criteria, including preservation and holding times, calibration requirements, and procedures for handling out-of-control situations. It includes specific quality control measures such as Laboratory Reagent Blank testing, Internal Standards, Surrogate Standards, and Matrix Spike/Matrix Spike Duplicate analyses. The specifications also include a comprehensive table of method quality control criteria for various test methods, with requirements for initial calibration, method detection limits, and quality control standards on a quarterly basis.

View the file

Other files for this state and local contract opportunity

Other files attached to Laboratory Analytical Services for Fish TissuesBid Documents, newest first.
File Type Posted
81029AttD - Cost Proposal Template.xlsx XLSX spreadsheet
81029AttG_-_QandA_Response2.xlsx XLSX spreadsheet
81029AttH5 - Technical Specifications - Reporting.doc DOC document
81029AttL Current Laboratory Services Contract Charges.xls XLS spreadsheet
Pre-Proposal Network Opportunities Submission.xlsx XLSX spreadsheet
81029 Pre-Proposal Conference Slides.pptx PPTX presentation
81029Addendum 1.docx DOCX document
81029AttA1 - IVOSB_08222023.docx DOCX document
81029AttB Professional-Services-Contract-Template.docx DOCX document
81029AttE - Business Proposal updated.docx DOCX document
81029AttF - Technical Proposal Template.doc DOC document
81029AttH1 - Technical Specifications.docx DOCX document
81029AttH7 - Technical Spec - Test Method References.docx DOCX document
81029AttI - Pre-proposal Network Form_08242021.docx DOCX document
81029AttJ - Attestation Form_09132022 (2).docx DOCX document
81029AttM IntenttoRespondForm.docx DOCX document
81029AttA - MWBE_08222023.docx DOCX document
81029AttA1 - IVOSB_08222023.docx DOCX document
81029AttC - Indiana Economic Impact Form (1).xls XLS spreadsheet
81029AttD - Cost Proposal Template.xlsx XLSX spreadsheet
81029AttE - Business Proposal.docx DOCX document
81029AttF - Technical Proposal Template.doc DOC document
81029AttH2 - Technical Specifications - Definitions.doc DOC document
81029AttH6 - Technical Specifications - Tasks List1.docx DOCX document
81029AttJ - Attestation Form_09132022 (2).docx DOCX document
81029AttB Professional-Services-Contract-Template.docx DOCX document
81029AttH1 - Technical Specifications.docx DOCX document
81029AttH3 - Technical Specifications - QA QC Criteria.doc DOC document
81029AttH7 - Technical Spec - Test Method References.docx DOCX document
81029AttM IntenttoRespondForm.docx DOCX document
81029AttA1 - IVOSB_08222023.docx DOCX document
81029AttH5 - Technical Specifications - Reporting.doc DOC document
81029AttH7 - Technical Spec - Test Method References.docx DOCX document
81029AttI - Pre-proposal Network Form_08242021.docx DOCX document
81029AttF - Technical Proposal Template.doc DOC document
81029AttG - QandA_.xlsx XLSX spreadsheet
81029AttH1 - Technical Specifications.docx DOCX document
81029AttH2 - Technical Specifications - Definitions.doc DOC document
81029AttJ - Attestation Form_09132022 (2).docx DOCX document
81029AttK - Electronic Data Reporting Standards.doc DOC document
RFP Boilerplate.pdf PDF
81029AttA - MWBE_08222023.docx DOCX document
81029AttC - Indiana Economic Impact Form (1).xls XLS spreadsheet
81029AttD - Cost Proposal Template.xlsx XLSX spreadsheet
81029AttE - Business Proposal.docx DOCX document
81029AttH - Reference Check.docx DOCX document
81029AttH3 - Technical Specifications - QA QC Criteria.doc DOC document
81029AttH4 - Tech Spec - Methods & Analytical Parameters.doc DOC document
81029AttH6 - Technical Specifications - Tasks List1.docx DOCX document
81029AttM IntenttoRespondForm.docx DOCX document
Show all 50

Laboratory Analytical Services for Fish TissuesBid Documents has more files on GovTribe.

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

RFP-25-81029

Attachment H3

Technical Specifications

Quality Assurance/Quality Control Criteria

A. Data Quality Assessment Data Quality Assessment provides guidance as to the quality of data necessary to meet a desired purpose. Sampling events and specific projects must define the goals of data collection and interpretation. Table 1, below, specifies OWQ data quality assessments (DQA) for sampling, chain-of-custody, Vendor analyses, and reporting. OWQ’s samples for the purposes of this contract are treated at Data Quality Assessment 4 (DQA4), unless otherwise specified in a Task or authorized by the responsible gatekeeper (IDEM/OWQ contact person) prior to analysis by the Contractor.

Table 1.: Data Quality Assessments

DQA LEVEL
DATA TYPE
DESCRIPTION
1
Screening Data
The results are usually generated onsite and have no QC checks. Analytical results, which have no QC checks or no precision or accuracy information or no detection limit calculations, but just numbers, are included in this category. Primarily, onsite data are used for presurveys and for preliminary rapid assessment.
2
Field Analysis Data
Data is recorded in the field or laboratory on calibrated or standardized equipment. Field duplicates are measured on a regular periodic basis. Calculations may be done in the field or later at the office. Analytical results, which have limited QC checks, are included in this category. Detection limits and ranges have been set for each analysis. The QC checks information for field or laboratory results is useable for estimating precision, accuracy, and completeness for the project. Data from this category are used independently for rapid assessment and preliminary decisions.
3
Laboratory Analytical Data
Analytical results include QC check samples for each batch of samples from which precision, accuracy, and completeness can be determined. Method detection limits (MDLs) have been determined using 40 CFR Part 136 Appendix B. Additionally, all reporting information required in the laboratory contract, and in the IDEM Surface Water Quality Monitoring and TMDL QAPP, especially Table A9-1, are included in the analytical data reports. Raw data, chromatograms, spectrograms, and bench sheets are not included as part of the analytical report, but are maintained by the contract laboratory for easy retrieval and review. Data can be elevated from DQA Level 3 to DQA Level 4 by the inclusion of this information in the data report and the QC data are reported using CLP forms or CLP format. Data falling under this category are considered as complete, legally defensible, and used for regulatory decisions.
4
Enforcement

Data Analytical results mostly meet the USEPA required Contract Laboratory Program (CLP) data analysis, Contract Required Quantification Limits (CRQL), and validation procedures. QC data are reported on CLP forms or CLP format. Raw data, chromatograms, spectrograms, and bench sheets are included as part of the analytical report. Additionally, all reporting information required in the laboratory contract, and in the IDEM Surface Water Quality Monitoring and TMDL QAPP, especially Table A9-1, are included in the analytical data reports. Data falling under this category are considered as complete, legally quantitative in value, and used for regulatory decisions.

B. Quality Assurance/Quality Control and Documentation 1.

Overview Quality control, in sampling and analysis, is a systematic approach to identifying, measuring, and minimizing errors introduced during sample acquisition and laboratory analysis. Quality control is a component of the overall process of quality assurance. Laboratories must not only look at completion of a method as satisfying the general requirements for quality control, but must also evaluate their overall processes and the quality of their performance over time. Only by maintaining an evaluation of an analytical methodology over time, can a single sample result be validated.

2.

Quality Assurance/Quality Control Plan Contractors must have and maintain a documented Quality Assurance/Quality Control (QA/QC) Program, capable of demonstrating that data has a specified degree of precision and reliability. An acceptable QA/QC program would be one patterned after a publication such as the “Handbook for Analytical Quality Control in Water and Wastewater Laboratories”, USEPA 600/4-79/019. Contractors must be able to validate each method used and each analysis performed by that method using the QA/QC Program.

QA/QC measures must be documented. All documentation must be maintained and made available for the use of IDEM/OWQ for five (5) years after the expiration date of this Contract. QA/QC documentation must be submitted as required I the following sections.

Contractors must maintain and document continual evaluation of the accuracy and precision of an analytical procedure and the ability of individual analysts to meet laboratory performance for a procedure.

QA/QC Criteria a.

Required Preservation and Holding Times

Analysis of a sample must begin within holding time. The beginning of an analysis or the end of the sample holding time is considered to be the time at which the Laboratory Reagent Blank is analyzed.

b. Analytical QA/QC Criteria

Contractors must follow the QA/QC criteria specified in the analytical method being performed. In the event that a method does not specify QA/QC criteria the following minimum criteria must be applied:

1. Initial calibration or linear calibration at least once per year.

2. Method detection limit determined in accordance with 40CFR, Part 136, Appendix B, at least once per year.

3. Continuing calibration check per sample batch. Typically a continuing calibration check should be run even with analytical procedures requiring a linear calibration.

4. Calibration blank per sample batch.

5. Laboratory fortified blank per sample batch.

6. Quality control standard on a quarterly basis.

c. Procedural calibration standards

Several methods use procedural calibration standards that do not require a separate LFB and CCC. If methods requiring an LFB and CCC do not state that the LFB and CCC are equivalent, a separate LFB and CCC must be analyzed, unless approved by the responsible IDEM/OWQ contact.

c.

Discrepancies

When a method and Table 2 of this Attachment are in direct conflict over QA/QC procedures or Table 2 of this Attachment lists QA/QC procedures not found in the method, Table 2 of this Attachment takes precedence. Offerors may utilize this judgment without penalty for a first occurrence of a direct conflict: however, Offerors must notify the responsible IDEM/OWQ gatekeeper after the first occurrence of a direct conflict. Failure to notify IDEM/OWQ after the first occurrence of a direct conflict may result in denial of future payments. No notification is required for QA/QC retroactively applied to pre-1991 methods as listed in Table 2 of this Attachment.

4.

Analytical Control a.

General Criteria

This section does not constitute all conditions under which an analysis would be deemed out‑of‑control or for remedies that a Contractor must employ to bring an analysis into control. The respective method and Table 2 of this Attachment must be consulted for these criteria: however, the conditions listed below supersede any method criteria, when in conflict with the method. Contractors are responsible for maintaining analytical controls necessary to meet the method QC specifications listed in Table 2 of this Attachment, and/or good judgment.

Good judgment is keeping consistent with the quality control necessary to justify a single sample result. For example, this is reflected in the retroactive application of method independent QA/QC criteria by IDEM/OWQ. There is no excuse for failure to monitor and address analytical control conditions by a Contractor.

IDEM/OWQ will be the sole judge of compliance with analytical control. Contractors found to not be employing and meeting the specifications outlined in the respective method, Table 2 of this Attachment, or employing good judgment, will be notified of the specific conditions and given the opportunity to correct any deficiency(s). Failure to correct the specified deficiency(s) may result in denial of payment for affected analyses and may result in termination of the Contractor’s contract in part or whole.

b.

Out-of-Control

1) Control Criteria

Analyses must be conducted in-control, as specified in the method and this Section. If an analysis or instrument is found to be out-of-control, the analyst must perform the corrective action measures necessary to bring the system back into control. All analytical and QA/QC samples analyzed since the last acceptable QC criteria must be reanalyzed after the analysis is brought into control. Contractors must document and report out-of-control operations, QC parameters, remedies, and reanalyses in the analytical report Narrative.

2) Holding Times

When samples are analyzed out of holding time, the analysis is deemed out-of-control, unless the analysis is approved by the IDEM/OWQ contact person.

3) Laboratory Reagent Blank (LRB) and Calibration Blanks The LRB or method blank and initial or continuing calibrations must be in control, prior to conducting an analytical run; otherwise the analysis is deemed out‑of‑control.

4) Internal Standards (IS) and Surrogate Standard (SS)

If a single IS or SS does not meet recovery criteria for a method during a batch run, the analysis will be considered in control: however, the analyst must take remedial measures to identify and correct the problem prior to future batches.

If the IS or SS retention window is not met per the analytical method, the analysis is deemed out‑of‑control. The IS retention time must be brought back into control and all samples reanalyzed since the last IS which met retention time criteria.

If two IS or SS, in a sample, do not meet recovery criteria, the analysis is deemed out‑of‑control.

If one or more IS or SS in consecutive samples do not meet recovery criteria, the analysis is deemed out‑of‑control.

If an analysis is out‑of‑control due to the failure of an IS or SS, a QC check sample must be analyzed following the last out‑of‑control sample. If the QC check sample meet recovery criteria, then the analysis is in control and the recovery failure may be due to matrix interferences. The Contractor should contact the responsible IDEM/OWQ contact person to see if reanalysis of the out‑of‑control sample(s) is required.

5) Matrix Spike/Matrix Spike Duplicate (MS/MSD)

Whenever the analytical procedure is “out of control,” the problem must be found, corrected, and the analysis repeated. The flagging of data as being “out-of-control without repeat analysis, is not acceptable. Analytical results reported when the procedure is operating “out-of-control” will be refused and payment will be withheld unless approved in writing by IDEM. Written approval may be given only in those situations where the results are needed and the sample cannot be analyzed again due to insufficient amount of sample remaining or a proper justification can be made using precision and accuracy data obtained by the method. Written approval will be entirely at the discretion of IDEM.

6) Laboratory Fortified Matrix (LFM)

If the analyses of an LFM does not agree with the true value within: +/- 20% for inorganic, volatile, and metals analyses; and +/-30% for pesticide, semivolatile and PCB analyses, and the LRB and CCC are in control, then the percent recovery failure is due to matrix interferences. Document the LFM matrix interference in the Narrative.

7) High Biased Samples

If a QC sample indicates a biased high result and the sample results are below detection limits for the all target compounds, reanalysis is not required.

Table 2: METHOD QUALITY CONTROL CRITERIA

Test
VERSION : DATE
IC
MDLs
LCR or LDR
CALIBRATION BLANK
LRB, MB or PB
LCS or LFB
MS/MSD

(LFM)

CCC
IPC or LPC
QCS
Additional QC Requirements and Comments
methods not listed
NA
R
R, Y
R, Y
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
120.1
:1982
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
130.1
:1971
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
150.1
:1982
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
160.1
:1971
QTRLY
Laboratories should be analyzing replicates on a at a rate of 1 per batch or 10%, whichever is greater.. Replicates should maintain an agreement of +/‑ 10% between replicates
160.2
:1971
QTRLY
Laboratories should be analyzing replicates on a at a rate of 1 per batch or 10%, whichever is greater.. Replicates should maintain an agreement of +/‑ 10% between replicates
160.3
:1971
QTRLY
Laboratories should be analyzing replicates on a at a rate of 1 per batch or 10%, whichever is greater.. Replicates should maintain an agreement of +/‑ 10% between replicates
160.4
:1971
QTRLY
Laboratories should be analyzing replicates on a at a rate of 1 per batch or 10%, whichever is greater.. Replicates should maintain an agreement of +/‑ 10% between replicates
160.5
:1974

QTRLY

Laboratories should be analyzing replicates on a at a rate of 1 per batch or 10%, whichever is greater.. Replicates should maintain an agreement of +/‑ 10% between replicates

180.1
2.0:1993
R
R, Y
R, 6M
R, AC, 10S, End Batch
R, Batch

R, AC

Verify 0.02 NTU resolution at three points from 0<X<1NTU, using freshly prepared Formazin on a quarterly basis. Maintain control charts on performance.

200.2
2.8:1994

0.02 NTU verification (180.1) must be conducted every 12hr when analyzing turbidity for omission of sample digestion.

200.7
4.4:1994
R
R, Y
R, Y
R, AC, 10S, End Batch
R, Batch, 20S
R, Batch
R, Batch, 10%
R, AC, 10S, End Batch
QTRLY
0.02 NTU verification (180.1) must be conducted every 12hr when analyzing turbidity for omission of sample digestion. Serial Dilution required at 10X MDL.
200.8
5.3:1994
R
R, Y
R, Y
R, AC, 10S, End Batch
R, Batch, 20S
R, Batch
R, Batch, 10%
R, AC, 10S, End Batch
QTRLY
0.02 NTU verification (180.1) must be conducted every 12hr when analyzing turbidity for omission of sample digestion.
200.9
2.2:1994
R
R, Y
R, Y
R, AC, 10S, End Batch
R, Batch, 20S
R, Batch
R, Batch, 10%
R, AC, 10S, End Batch
QTRLY
0.02 NTU verification (180.1) must be conducted every 12hr when analyzing turbidity for omission of sample digestion. Serial Dilution required at 10X MDL.
218.5
:1982
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
218.6
3.3:1994
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R,Batch
R, Batch, QReps
R, 10%
R, Daily, 20
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
245.1
3.0:1994
R
R, Y
R, Y
R, AC, 10S, End Batch
R, Batch, 20S
R, Batch
R, Batch, 10%
R, AC, 10S, End Batch
QTRLY
245.2
1974
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
245.5
2.3:1991
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, Batch, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
300.0
2.1:1993
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R,Batch
R, Batch, QReps
R, 10%
R, Daily, 20
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
310.1
:1978
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
3111B
22nd ed.:2012
R
R, Y
R, Y
R, AC, 10S, End Batch
R, Batch, 20S
R, Batch
R, Batch, 10%
R, AC, 10S, End Batch
QTRLY
0.02 NTU verification (180.1) must be conducted every 12hr when analyzing turbidity for omission of sample digestion. Serial Dilution required at 10X MDL.
3112B
22nd ed.:2012
R
R, Y
R, Y
R, AC, 10S, End Batch
R, Batch, 20S
R, Batch
R, Batch, 10%
R, AC, 10S, End Batch
QTRLY
0.02 NTU verification (180.1) must be conducted every 12hr when analyzing turbidity for omission of sample digestion. Serial Dilution required at 10X MDL.
3113B
22nd ed.:2012
R
R, Y
R, Y
R, AC, 10S, End Batch
R, Batch, 20S
R, Batch
R, Batch, 10%
R, AC, 10S, End Batch
QTRLY
0.02 NTU verification (180.1) must be conducted every 12hr when analyzing turbidity for omission of sample digestion. Serial Dilution required at 10X MDL.
3114
22nd ed.:2012
R
R, Y
R, Y
R, AC, 10S, End Batch
R, Batch, 20S
R, Batch
R, Batch, 10%
R, AC, 10S, End Batch
QTRLY
0.02 NTU verification (180.1) must be conducted every 12hr when analyzing turbidity for omission of sample digestion. Serial Dilution required at 10X MDL.
325.2
:1978
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
330.1
:1978
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
330.5
:1978
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
335.4
1.0:1993
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
350.1
2.0:1993
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
350.3
:1974
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
351.2
2.0:1993
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
352.1
:1971
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
353.2
2.0:1993
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
354.1
:1971
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
365.1
2.0:1993
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
365.2
:1971
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
370.1
:1978
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
375.2
2.0:1993
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
376.1
:1978
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
405.1
:1974
R, Batch
R, Batch

QTRLY

Seed Blank & Glucose‑Glutamic Acid check required per batch. Labs must maintain control charts of blank and check solution performance. Control limits must be established in accordance with 5210B, p 6a.

410.1
:1978
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
See sections 8 & 9 of the method
410.4
2.0:1993
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
413.2
:1978
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
Labs must utilize reference oil (section 6.4) to develop initial percent recoveries based on a minimum of 10 replicates performed on separate days. Labs must update percent recoveries every 20 to 30 check samples.
415.1
:1974
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
418.1
:1978
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
420.4
1.0:1993
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
524.2
4.1:1995
R
R, Y
R, Before Analysis, Batch, 12hr
R, Batch, 20S, 12hr
See Sect 9.10
R, 12hr
R, DAILY
QTRLY
525.2
2.0:1995
R
R, Y
R, Before Analysis, Batch, 12hr
R, Batch, 20S, 12hr
R, Batch, 20S, 12hr
R, 12hr
QTRLY
Surrogate(s): DFTPP, endrin, 4,4'‑DDT
601.0
:2012
R
R, Y

R, Before Analysis, Batch

R, 10%, Monthly
R, DAILY
QTRLY
Laboratory Control Standard required 10% of samples
602.0
:2012
R
R, Y

R, Before Analysis, Batch

R, 10%, Monthly
R, DAILY
QTRLY
Laboratory Control Standard required 10% of samples
603.0
:2012
R
R, Y

R, Before Analysis, Batch

R, 10%, Monthly
R, DAILY
QTRLY
Laboratory Control Standard required 10% of samples
608.0
:2012
R
R, Y

R, Before Analysis, Batch

R, 10%, Monthly
R, DAILY
QTRLY
Laboratory Control Standard required 10% of samples
610.0
:2012
R
R, Y

R, Before Analysis, Batch

R, 10%, Monthly
R, DAILY
QTRLY
Laboratory Control Standard required 10% of samples
624.0
:2012
R
R, Y

R, Before Analysis, Batch

R, 5%, Monthly
R, DAILY
QTRLY
Laboratory Control Standard required 5% of samples
625.0
:2012
R
R, Y

R, Before Analysis, Batch

R, 5%, Monthly
R, DAILY
QTRLY
Laboratory Control Standard required 5% of samples

Internal Standard(s) and Surrogate(s): R, See Table 8

1613.0
REV B:1994
R
R, Y
R, After LFB, 12hr, 10S
R, Batch

R, 12hr

QTRLY

1631.0
REV E:2002
R
R, Y
R, in Triplicate Monthly
R, 12hr, 10S

R, 12hr, 10S

QTRLY

1632.0
REV A:1998
R
R, Y
R, After LFB, 12hr, 10S
R, 12hr, 10S

R, 12hr, 10S

QTRLY

1636.0
REV A:1996
R
R, Y
After CCC
R, After LFB, 12hr, 10S
R, 12hr, 10S

R, 12hr, 10S

QTRLY

1637.0
REV A:1996
R
R, Y
After CCC
R, After LFB, 12hr, 10S
R, 12hr, 10S

R, 12hr, 10S

QTRLY

1638.0
:1996
R
R, Y
After CCC
R, After LFB, 12hr, 10S
R, 12hr, 10S

R, 12hr, 10S

QTRLY

1639.0
:1996
R
R, Y
After CCC
R, After LFB, 12hr, 10S
R, 12hr, 10S

R, 12hr, 10S

QTRLY

1640.0
:1997
R
R, Y
After CCC
R, After LFB, 12hr, 10S
R, 12hr, 10S

R, 12hr, 10S

QTRLY

1668.0
:1999
R
R, Y

R, After LFB, 12hr, 20S

All samples w/labeled cmpds.
R, 12hr, 10S

QTRLY

2130
22nd ed.:2012
R
R, Y
R, 6M
R, AC, 10S, End Batch
R, Batch

R, AC

Verify 0.02 NTU resolution at three points from 0<X<1NTU, using freshly prepared Formazin on a quarterly basis. Maintain control charts on performance.

2320B
22nd ed.:2012
R
R, Y
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, AC
R, CB, 10S, End Batch
QTRLY
2540B, C, D, E, F
22nd ed.:2012
R

Contractors should run sample replicates on a 10% basis. Replicates should agree within +/-10%.

4500CL(‑)‑D, E
22nd ed.:2012
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, Batch, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
4500NH3-D, F
22nd ed.:2012
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, Batch, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
4500N(Org)-B, C
22nd ed.:2012
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, Batch, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
4500CN‑C, E, F, G, I
22nd ed.:2012
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, Batch, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
4500F(‑)‑B, C, D, E
22nd ed.:2012
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, Batch, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
4500NO2‑B
22nd ed.:2012
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, Batch, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
4500NO3‑D, E, F
22nd ed.:2012
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, Batch, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
4500P‑E, F
22nd ed.:2012
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, Batch, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
4500Si-D
22nd ed.:2012
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
5210B, C
22nd ed.:2012

R, Batch

Seed Blank & Glucose‑Glutamic Acid check required per batch. Labs must maintain control charts of blank and check solution performance. Control limits must be established in accordance with 5210B, p 6a.

5220B
22nd ed.:2012
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, Batch, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
5310
22nd ed.:2012
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
6010C
3.0:2000
R
R, Y
R, Y
R, AC, 10S, End Batch
R, Batch, 20S
R, Batch
R, Batch, 10%
R, AC, 10S, End Batch
QTRLY
0.02 NTU verification (180.1) must be conducted every 12hr when analyzing turbidity for omission of sample digestion.
6020A
6.0:2004
R
R, Y
R, Y
R, AC, 10S, End Batch
R, Batch, 20S
R, Batch
R, Batch, 10%
R, AC, 10S, End Batch
QTRLY
0.02 NTU verification (180.1) must be conducted every 12hr when analyzing turbidity for omission of sample digestion.
7110B
22nd ed.:2012
R

R, Batch

R, Batch, 10%

Run one (1) duplicate per 10 samples. RPD must be <= 30 %

8081B
2.0:2000
R
R, Y

R, Before Analysis, Batch

R, 5%, Monthly
R, DAILY
QTRLY
Laboratory Control Standard required 5% of samples

Internal Standard(s) and Surrogate(s): R, See Table 8

8082A
1:2000
R
R, Y

R, Before Analysis, Batch

R, 5%, Monthly
R, DAILY
QTRLY
Laboratory Control Standard required 5% of samples

Internal Standard(s) and Surrogate(s): R, See Table 8

8260B
2.0:1996
R
R, Y

R, Before Analysis, Batch

R, 5%, Monthly
R, DAILY
QTRLY
Laboratory Control Standard required 5% of samples

Internal Standard(s) and Surrogate(s): R, See Table 8

8270D
4.0:1998
R
R, Y

R, Before Analysis, Batch

R, 5%, Monthly
R, DAILY
QTRLY
Laboratory Control Standard required 5% of samples

Internal Standard(s) and Surrogate(s): R, See Table 8

9010B
2.0:1996
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
9036
0:1986
R
R, 6M
R, 6M
R, AC, 10S, End Batch
R, Batch
R, Batch
R, 10%
R, AC
R, CB, 10S, End Batch
QTRLY
See sections 9 & 10 of the method
1633A
1.0:Dec. 2024
R
R

6M R

6M

R, AC, 10S, End Batch
R, Batch
R, Batch
R, Batch
R, Batch, 10%
R, CB, 10S, End Batch
9221A, B, C, E
22nd ed.:2012

R, AC, 10S, End Batch

R, Batch

QTRLY
Duplicate analyses must be conducted on 10% of all samples and all positive routine samples. QC criteria found in 9020B must be

followed. The most recent QC results determined in 9020B must be reported with each IDEM/OWQ sample set.

9222A, B, C, D
22nd ed.:2012

R, AC, 10S, End Batch

R, Batch

QTRLY
Duplicate analyses must be conducted on 10% of all samples and all positive routine samples. QC criteria found in 9020B must be

followed. The most recent QC results determined in 9020B must be reported with each IDEM/OWQ sample set.

9310
0:1986
R
R, Batch
R, 3 per Batch

R, Batch, 10%

Run one (1) duplicate per 10 samples. RPD must be <= 30 %

9320
0:1986
R
R, Batch
R, 3 per Batch

R, Batch, 10%

Run one (1) duplicate per 10 samples. RPD must be <= 30 %

RFP-25-81029

Technical Specifications H3

File details come from the government source that posted it. Updated .