Installation Information for ProMCS1500

Clients:

1. You will need to have a gang-box installed location of Server; and from the gang-box (Per Site Assessment)

a. Run 1ea CAT5 or CAT6 low voltage cable from Network Closest to server location in each room.

b. Run a CAT5 or CAT6 low voltage cable from server location to location of the elert button.

c. Run a CAT5 or CAT6 low voltage cable from server location to location of the speaker in each room.

d. Emergency Power Outlet required at location of Server in each room

2. The gang-box outlined in #1 will need:

a. 1ea 3-Port face plate

b. 3ea CAT5 or CAT6 quick-connect ports

3. At the location of where the elert button will be mounted you will need:

a. 1ea Gang-box

b. make sure the CAT5e/CAT6 cable is terminated with a CAT5e/CAT6 RJ45 connector.

***All Low Voltage Termination should follow the TYPE B Wiring (Diagram Below):

ELERT BUTTON

1. At the location of where speaker will be mounted you will need:

a. make sure the CAT5e/CAT6 cable is terminated with a CAT5e/CAT6 RJ45 connector.

1-Gang Orange **(In Wall where Elert Button is going)

One CAT5e or CAT6 Cable from location of ProMCS1500 to location of Elert Button terminated with RJ45 connector.

2. Emergency Power Outlet required at location of Server in each room.

3. Client will mount 19” GCX rail in each room. (Rails will be provided by CACI)

Speaker

1. At the location of where speaker will be mounted (refer to site assessment) you will need:

a. make sure the CAT5e/CAT6 cable is terminated with a CAT5e/CAT6 RJ45 connector.

Additional Installation Information:

1. Client or CACI will mount GCX rails in each room. (Rails will be provided by CACI)

WALL TEMPLATE FOR ICU ROOM

FOOTWALL

GCX Channel

1 Mount the 7-inch or 13-inch or 19-inch GCX Channel to the wall so that its edge is approximately 4 inches from the wall receptacle.

2 Follow manufacturer’s installation instructions to secure the channel to wall.

Mounting the GCX Channel

Network Jack Identification: jacks do not have to be in any specific order.

• Data 1

• Spkr/Mic

• Elert

CACI:

1. CACI will install all equipment in each room.

a. ProMCS1500

b. Elert Button

c. Speaker

2. CACI will test all equipment with Core for connectivity.

3. CACI will train POC onsite on equipment and how to use Medialink Toolsuite Software to configure equipment and test equipment.

4. CACI will work with Network/IT team to connect cart to Client’s wireless endpoint and test cart with Core.

VARIOUS INSTALL CABLING DIAGRAMS