RFI questions Phase I.docx

DOCX document 14 KB Posted

Attached to
Z1DA--523A4-24-005 Design Build Replace EP Lab West Roxbury
Solicitation number
36C24123R0185
Issued by

About this file

This document contains questions from the Department of Veterans Affairs regarding a solicitation to design and build a replacement electrophysiology lab at the VA Medical Center in West Roxbury, Massachusetts. The questions address how the patient lift will be supported, whether it can be mounted to an existing unistrut grid for medical equipment, and if not, whether the roof structure has a concrete deck where lift supports can be anchored. Respondents are informed that the general contractor will be responsible for the design and installation of the patient lift, and no details on lift support will be provided by the VA as this is a design-build project. The related solicitation is identified as number 36C24123R0185 to replace the EP lab at the West Roxbury VAMC, issued by the Department of Veterans Affairs Veterans Health Administration VISN 1.

View the file

Other files

File Type Posted
36C24123R0185 0001.docx DOCX document
P03 SSJ 523A4-24-005-Signed CO signed.pdf PDF
MA WD MA20230001.txt TXT text file
S02 Reference Drawing Existing.pdf PDF
S02 - Past Performance Questionnaire - Sample.pdf PDF
S02 Typical DRAWINGS-Philips AZURION-New Equipment.pdf PDF
36C24123R0185.docx DOCX document
S02 523A4-24-005 SOW revised.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

QUESTIONS FOR 523A-24-005 DESIGN BUILD REPLACE EP LAB AT VAMC WEST ROXBURY

What will the patient lift be supported from?

· Will the patient lift be mounted onto the Unistrut grid provided for the Phillips Equipment?

· If not, does the roof structure have a concrete deck for the lift supports to be anchored into?

· Please provide details regarding how the patient lift is to be supported?

The design and installation of patient lift shall be provided by GC. Whether the unistrut grid of equipment can be used or not will be determined in the design phase. There is no detail to be provided by VA regarding patient lift support as it is a design-built project.