This document is a Water Bath Specification Sheet for laboratory equipment, specifically targeting the Illinois Department of Transportation (IDOT), District 6. The delivery instructions indicate the final destination is to Grant Gruen at the IDOT District 6 office located in Springfield, Illinois.
The specification sheet details a large lab water bath designed to support ASTM D5581 and ASTM C88 standards, with a 14-gallon minimum volume and dimensions of 20" x 20" x 10" deep. Key technical features include a digital controller with a temperature range from ambient to 180 degrees Fahrenheit, precise temperature control of +/- 0.1%, a water and corrosion-resistant front panel display, dual digital temperature readout, and 110-volt 60 Hz power input. The water bath is fully insulated, with stainless steel exposed areas, an interior stainless-steel shelf, and a magnetic stirring bar to facilitate water circulation during testing, specifically configured to accommodate 16 4" Marshall specimens simultaneously.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| StandardTermsandConditions~7.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Large Lab Water Bath Specification Sheet
· Supports the following standards ASTM D5581 and ASTM C88
· Minimum volume of 14 gallons with a minimum 20” by 20” by 10” deep to allow for 16 4” Marshall specimens to be tested simultaneously.
· Digital controller
· Temperature range of ambient to 180 degrees Fahrenheit with +/- 0.1 % of input span.
· Front panel display that is water and corrosion resistant.
· Dual digital display showing set point and process temperature
· 110 volt 60 Hz input
· Interior stainless-steel shelf that allows 2” of circulating water above and below the specimens.
· Magnetic stirring bar to induce water flow during operation.
· Fully insulated unit with exposed areas being stainless steel.
Delivery Instructions:
Illinois Department of Transportation District 6 Attn: Grant Gruen 3215 Executive Park Drive Springfield, Illinois 62703