W81XWH22Q0989
SECTION SF 30 BLOCK 14 CONTINUATION PAGE
SUMMARY OF CHANGES
SECTION SF 30 - BLOCK 14 CONTINUATION PAGE
The following have been added by full text:
ADDITIONAL INFORMATION Q & AS
Lithium-Ion Battery Thermal/Humidity Test Chamber
1. What is the max payload?
Answer: 75lb
2. How will the payload be distributed?
Answer: Evenly in the center of the chamber.
3. Capacity of the floor and (2) shelves?
Answer: 75lb
4. Please describe the Protective enclosure requested.
Answer: Capable of extinguishing or containing a medical device battery fire.
5. Does the MRDC Test Branch Building have a loading dock and forklift?
Answer: Yes
SECTION SF 1449 - CONTINUATION SHEET
The following have been modified:
PERFORMANCE WORK STATEMENT
Lithium-Ion Battery Thermal/Humidity Test Chamber For the U.S. Army Medical Research and Development Command (USAMRDC) Fort Detrick, MD
1. Scope: The U.S. Army Medical Research and Development Command (USAMRDC) is Procuring a Lithium-Ion Battery Thermal/Humidity Test Chamber. The Thermal/Humidity Test chamber for Lithium-Ion Batteries allows the U.S Army Medical Research and Development Command to Conduct Research, Development, Test, and Evaluation services to the USAMRDC labs and mission partners.
2. Requirements: The Contractor will provide and Install Thermal/Humidity Test chamber for Lithium-Ion Batteries that meets all the below listed Minimum Essential Characteristics. Contractor will also provide a 3 year Comprehensive Warranty.
Minimum Essential Characteristics (MEC)
• Protective Enclosure/Structure: External structure to contain any fire or explosion
· Ventilation System: To remove all gases from inside the chamber prior to opening the door
· Pressure Release and Alarm
• Fire Safety System: Autonomous fire suppression system that does not utilize halon
· Hydrogen and Carbon Monoxide Gas Monitoring and Alarms
· Test Article Surface Cooling: To prevent thermal runaway, to cool cells or packs if overheated
· Spark Resistant Interior
· Emergency Stop Switch
· Operating Temperature Range: -51°C to 75°C
• Humidity Range: 10% to 95% RH
• Control Tolerance: ±0.5°C
• Limit Controller: Independent high and low temperature limits. Triggers an audible alarm and shuts down the chamber.
• Input Power Requirements: 230V ±10%, 60 Hz, 3 Phase, Max Current Draw 39 A
• Internal Dimensions: Minimum 32” W x 32” D x 32” H
· Payload weight: 200 pounds (minimum)
• Exterior Dimensions (including all venting, attachments, etc.): Maximum 64"W x 70"D x 121"H
• Window Viewing Area: Minimum 10" x 10" located on the front of chamber. Window wiper desired on the inside for visibility during humidity testing.
• Access ports:
· Minimum of two access ports located in the front of the chamber below the window to allow for technician access and manipulation of test item during testing.
· Cable port for wired test instrumentation of test items
• Shelving: Two adjustable stainless steel wire shelves. Each shelf encompassing the entire work area.
• Interior light: Interior lighting with the control switch located on the outside.
• Non-CFC Cascade Refrigeration
• 3 Year Comprehensive Warranty
• Chamber condensation drain
• Casters
• Humidity Water Recirculation System
· Proposals should include all costs associated with system transportation, shipping, and installation
3. Deliverables. The Contractor shall provide all deliverables identified in this Section of the PWS. Additionally, the contractors shall provide a 3-year comprehensive Warranty. All quotes must include all costs associated with system transportation, shipping, and installation.
4. Government Property: The Contractor shall comply with all Government property requirements and regulations as necessary. The Government shall not be responsible in any way for any damage to the contractor’s supplies, materials, equipment or the contractor’s employee’s personal belongings due to fire, theft, accident or other disaster.
5. Other Direct Costs: There is no long distance, local travel, or other ODCs authorized for this contract.
6. Common Access Card (CAC): The vendor shall not have access to any military computer systems in performance of these contract duties. Therefore, CAC cards are not required by the vendor personnel. However, all vendor employees accessing the Fort Detrick installation for installation and or onsite customer support of this Ethernet data line will be required to conform to the Fort Detrick policies for access to the installation, which will include obtaining an NAC-level clearance and appropriate visitor access passes from the Fort Detrick Provost Marshall.
7. Performance Assessment: The review/inspection and acceptance of final product will be conducted by a USAMRDC appointed representative upon completion of delivery and installation of final product at USAMRDC Test Branch, Building 1056 Patchel Street, Fort Detrick, MD 21702.