This document contains questions and answers regarding a solicitation for a rack washer. The solicitation seeks a fully operational rack washer to be installed at the Walter Reed Army Institute of Research in Forest Glen Annex, Maryland within four months of award. Key requirements include a minimum interior chamber size of 60" wide by 90" long by 86" high, stainless steel construction with a minimum of 304 grade, six oscillating wash arms, integrated hydrogen peroxide generator, and a two-year warranty. Bids are due by December 11, 2020. The site visit was held on December 2, 2020 and questions were due by December 4, 2020. The rack washer will be furnished and installed for the Department of the Army Medical Command.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| QandA.pdf | ||
| QandA.pdf | ||
| Exhibit A.png | PNG image | |
| QA.pdf | ||
| QandA.pdf | ||
| B08- Solicitation 11-10-20.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
W81XWH21Q0006
QUESTION & ANSWERS
1. What are the dimensions and weights of the largest components they need to wash?
NHP Rack: 38"D x 74"W x 82"H Rodent: 19"D x 69"W x 82"H NHP packs would generally be the heaviest item. They weigh several hundred pounds.
2. What are the dimensions of the existing washer doors, overall opening?
60"W x 96"H
3. Can someone share the existing pit details?
There is a large pit that houses 2 rack washers. The section of the pit for this rack washer is approximately 16"D x 146"L x 89"W
4. The existing washer is around 140” long (chamber door to door). The specification for the new washer asks for 90”… this will leave a large portion of the pit open/unused. We have the ability to custom build chambers to fit the existing pit. Would this be preferable to the end user instead of reducing the chamber volume?
The specifications states the minimum length of the interior compartment is 60” W x 90” L x 86” H (inside door to door).
5. The solicitation asks to provide an integral air compressor if one is needed. Is there house supplied air that can be used? We are providing the large sterilizer in the area that was recently awarded and that utility is present at that unit.
House air is available.
6. Please confirm if warranty is one year or two year.
Two Year Per the PWS: 6.3.7. Warranty (Deliverable 3) The contractor shall propose a two-year warranty and service/maintenance agreement providing a “next business day” response time.
7. Please confirm if 15 minute rodent cage cycle is acceptable.
MECs state 10 minute rodent cage cycle and 30 minute nonhuman primate rack cycle.
8. Since the largest item washed are NHP racks with dimensions 38”D x 74”W x 82”H. The 38”D will easily fit into a chamber with 48” wide load clearance. Please confirm the 48” wide load clearance chamber is acceptable rather than the 60” wide chamber.
Yes, a minimum of 48" wide x 85" H is acceptable
9. Is a minimum of eight (8) configurable cycles acceptable?
MECs call for minimum of ten (10) configurable cycles
10. Shall the washer be fabricated with non-proprietary parts available from North American suppliers?
That is not a requirement. The quote must be incompliance with the Buy American Act.
11. We are preparing our proposal for this solicitation. I have tried to connect toCaution-www.ecmra.mil < Caution- http://www.ecmra.mil > as per the instructions on page 11 of the specifications under 6.4 Management Reports and Plans. I had our accounting department also try to connect. The message that immediately appears is as follows:
“This website may be impersonating Caution-www.ecmra.mil < Caution-http://www.ecmra.mil > to steal your personal or financial information.”
Our accounting department received additional information that this page was de-commissioned as of 6/19/2020. I respectfully request your input so we know how to proceed.
Please use website https://www.sam.gov
12. SPECIFICATION: Interior compartment/chamber size shall not be less than: 60” W x 90” L x 86” H (inside door to door). The floor shall be made of no less than 3mm thick stainless steel with a greater than 200 lb. point load QUESTION(S): From the best of our knowledge there is not a washer on the market with a 60” door equipped with an inflatable door gasket. Door sizes are in between 40” to 50”. Considering that the largest item to wash is the NHP rack that is 39” wide, please confirm that any door larger than 40” is acceptable.
Interior compartment/chamber size shall not be less than: 60” W x 90” L x 86” H (inside door to door).
The floor shall be made of no less than 3mm thick stainless steel with a greater than 200 lb. point load
13. SPECIFICATION: Interior compartment/chamber size shall not be less than: 60” W x 90” L x 86” H (inside door to door). The floor shall be made of no less than 3mm thick stainless steel with a greater than 200 lb. point load.
QUESTION(S): The existing washer/pit are much longer than 90”. We believe that the best design for your washroom is a washer that uses the entire pit length. With a short (90”) washer there will be three main issues: cover the pit gap (extend the floor or the grate), extend the existing modular wall and move one of the hoods at the entrance/exit of the washer. Please confirm if a chamber that uses the entire length of the pit is required.
Vendor choice on length as long as the minimum length is met.
14. SPECIFICATION: The interior of the rack washer shall be all stainless steel construction. All wetted parts shall be stainless steel or appropriate polymeric materials. All structural support including sump support frame shall be stainless steel.
QUESTION(S): Please identify what grade of stainless steel must be used for the chamber. 316L stainless steel provides a much better resistance to corrosion than other stainless-steel grades.
MEC states stainless steel as a minimum. Would prefer a minimum of 304 or better.
http://www.ecmra.mil/ https://www.sam.gov/
15. SPECIFICATION: The chamber shall be insulated on all sides to reduce hot surfaces and heat loss. The chamber doors shall be insulated with tempered glass viewing windows and side mounted hinges.
Rack Washer shall provide a visual indication of machine status during wash phases QUESTION(S): Please confirm the door has to be made of stainless steel with inspection windows vs full glass door.
Requires a tempered glass viewing window and the interior of the door shall be stainless steel. Size of window and exterior door construction is vendor choice.
16. SPECIFICATION: Wash and rinse nozzles shall be mounted on no less than 6 oscillating arms. The machine shall have separate wash and rinse spray nozzles shall be positioned to reach all load surfaces, shelving, and caging surfaces.
QUESTION(S): Are rotating arms permitted?
MECs call for 6 oscillating arms
17. SPECIFICATION: Washer shall be capable of flushing auto-watering lines on racks with fresh house hot water during the final rinse phase. The system shall be supplied with a quick-disconnect hose and a pressure reducing station for the house hot water supply line.
QUESTION(S): There are two design available for the auto-watering flushing:
1: Use the water coming from the facility (temperature, pressure and filtration grade depends on the facility water line) Per MECs, "...flushing auto-watering lines on racks with fresh house hot water during the final rinse phase." (Hot Water: 1 ½” NPT, 50 GPM flow rate @ 70 psi @ 130oF). House water is softened and treated with UV.
2: Use water coming from a side tank, with a dedicated pump, dedicated heat exchanger and filter.
Which design do you need?
Shall contain dosing pump for a minimum of three chemicals with ability to add additional pumps.
18. SPECIFICATION: Shall contain dosing pump for a minimum of three chemicals with ability to add additional pumps.
QUESTION(S): Please identify what chemicals you use and for each its function.
Currently only acid and alkaline are in use.
19. SPECIFICATION: The sump shall contain a stainless steel, steam heat exchanger to maintain recirculated solution temperatures complete with steam supply and condensate controls and be easily removable for maintenance and cleaning.
QUESTION(S): Please confirm that the heat exchanger has to be located outside the chamber (not in the sump), in the technical area.
Not specified. Vendor must demonstrate placement of sump meets MEC requirements.
20. Is a portable H2O2 generator with its own controls, that would roll up to the soiled side of the rack washer and connect via external ports and fitting on the rack washer be acceptable?
No. MEC states "Built-in H202 generator included and integrated in the equipment’s technical compartment."
21. There is conflict in the answers provided for question 8 and question 12. Conflicts listed in blue.
Our question number 12 was not answered directly in regards to any door larger than 40” being acceptable. Please confirm that a door width larger than 40" wide is acceptable.
The answer for question number 12 states - Interior compartment/chamber size shall not be less than:
60” W x 90” L x 86” H (inside door to door).The floor shall be made of no less than 3mm thick stainless steel with a greater than 200 lb. point load while the answer for question number 8 states that a chamber 48" wide is acceptable Yes, a minimum of 48" wide x 85" H is acceptable.
Yes, a door width of larger than 40" is acceptable. Would prefer door size of 48" wide x 85" high or greater.
Question #12 is regarding interior compartment of chamber while question #8 is regarding the door.
22. Can you clarify if a compartment/chamber that is smaller than 60” wide with a door that is larger than 40” wide is acceptable?
The minimum width of the compartment is 60", thus a door of 40" or greater is possible. Per the MECs, "Interior compartment/chamber size shall not be less than: 60” W x 90” L x 86” H (inside door to door).
The floor shall be made of no less than 3mm thick stainless steel with a greater than 200 lb. point load."
23. Based on the answers received so far I would say that the department wants the compartment/chamber (wall to wall) to be at least 60” wide even if the door is anywhere between 40" to 48" wide.
The minimum width of the door must be 40" to accommodate the NHP racks. Would prefer if the door was 48" wide or greater.