Technical Evaluation Spreadsheet in accordance with FAR ADDENDUM TO 52.212‐2
W81K0418Q0004, USMEPCOM Audiometric Booths
Required Capabilities IAW
Solicitation
Manufacturer
Manufacturer's Part Number
Meets Requirements
Yes/No Comments
The audiometric booth shall be of a soundproof single wall construction
(minimal thickness of four inches), The booth shall be mounted on vibration isolation rails.
The ventilation system shall be capable of a complete air exchange within the booth every two minutes. Background noise suppression and noise generated from the ventilation system shall comply with American National
Standard Institute (ANSI) Standard, “Criteria for Permissible Ambient
Noise during Audiometric Testing, S3.1‐1999.”
The booth shall have a door with a double magnetic gasket, with minimum dimensions of 30 inches wide and 71 inches high and maximum dimensions of 34 inches wide and 75 inches high, and mounted "centered" on the width end of the booth.
Double glazed acoustical safety glass observation window in the door, with minimum dimensions of
22‐ inches wide and 16‐inches high and maximum dimensions of 26‐ inches wide and 20‐inches high.
A minimum of two (one for four‐ position booths) recessed light fixtures in the ceiling of the booth, with an on/off control switch
(collocated with ventilation switch) mounted on the front of the booth, either inside or outside of the booth, near the door handle side of the booth door opening.
The fixtures shall employ either incandescent light bulbs consuming at no more than 55‐watts each or light emitting diode (LED) fixtures providing the equivalent level of light output.