Amendment 0001
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
W40M USA HLTH CONTRACTING ACT
SOUTHERN REG CONTR OFC
POC: ALEX J. HOUSTON
3551 ROGER BROOK DR BLDG 3600
FORT SAM HOUSTON TX 78234-6200
0011070833
AMENDMENT OF SOLICITATION/MODIFICATION OF CONTRACT
Except as provided herein, all terms and conditions of the document referenced in Item 9A or 10A, as heretofore changed, remains unchanged and in full force and effect.
15A. NAME AND TITLE OF SIGNER (Type or print)
30-105-04EXCEPTION TO SF 30
APPROVED BY OIRM 11-84
STANDARD FORM 30 (Rev. 10-83)
Prescribed by GSA
FAR (48 CFR) 53.243
1. The purpose of this Amendment 0001 is to incorporate a slightly modif ied Statement of Objectives and present revised 52.212-1 &
52.212-2 Addendums. The conformed copy of the solicitation has been provided to help identify w here changes w ere made. Contractors are reminded it is their responsibility to review FedbizOpps regularly for changes to the solicitation.
2. Revised the 52.212-1 & 52.212-2 Addendums.
3. Extended the solicitation close date to 25 September 2017, 12:00 Noon CST.
All other terms and conditions remain the same.
1. CONTRACT ID CODE PAGE OF PAGES
J 1 25
16A. NAME AND TITLE OF CONTRACTING OFFICER (Type or print)
16C. DATE SIGNED
BY 13-Sep-2017
16B. UNITED STATES OF AMERICA15C. DATE SIGNED15B. CONTRACTOR/OFFEROR
(Signature of Contracting Officer)(Signature of person authorized to sign)
8. NAME AND ADDRESS OF CONTRACTOR (No., Street, County, State and Zip Code) X W81K00-17-T-0403
X 9B. DATED (SEE ITEM 11)
08-Sep-2017
10B. DATED (SEE ITEM 13)
9A. AMENDMENT OF SOLICITATION NO.
11. THIS ITEM ONLY APPLIES TO AMENDMENTS OF SOLICITATIONS
X The above numbered solicitation is amended as set forth in Item 14. The hour and date specified for receipt of Offer X is extended, is not extended.
Offer must acknowledge receipt of this amendment prior to the hour and date specified in the solicitation or as amended by one of the following methods:
(a) By completing Items 8 and 15, and returning 1 copies of the amendment; (b) By acknowledging receipt of this amendment on each copy of the offer submitted;
or (c) By separate letter or telegram which includes a reference to the solicitation and amendment numbers. FAILURE OF YOUR ACKNOWLEDGMENT TO BE
RECEIVED AT THE PLACE DESIGNATED FOR THE RECEIPT OF OFFERS PRIOR TO THE HOUR AND DATE SPECIFIED MAY RESULT IN
REJECTION OF YOUR OFFER. If by virtue of this amendment you desire to change an offer already submitted, such change may be made by telegram or letter, provided each telegram or letter makes reference to the solicitation and this amendment, and is received prior to the opening hour and date specified.
12. ACCOUNTING AND APPROPRIATION DATA (If required)
13. THIS ITEM APPLIES ONLY TO MODIFICATIONS OF CONTRACTS/ORDERS.
IT MODIFIES THE CONTRACT/ORDER NO. AS DESCRIBED IN ITEM 14.
A. THIS CHANGE ORDER IS ISSUED PURSUANT TO: (Specify authority) THE CHANGES SET FORTH IN ITEM 14 ARE MADE IN THE
CONTRACT ORDER NO. IN ITEM 10A.
B. THE ABOVE NUMBERED CONTRACT/ORDER IS MODIFIED TO REFLECT THE ADMINISTRATIVE CHANGES (such as changes in paying office, appropriation date, etc.) SET FORTH IN ITEM 14, PURSUANT TO THE AUTHORITY OF FAR 43.103(B).
C. THIS SUPPLEMENTAL AGREEMENT IS ENTERED INTO PURSUANT TO AUTHORITY OF:
D. OTHER (Specify type of modification and authority)
E. IMPORTANT: Contractor is not, is required to sign this document and return copies to the issuing office.
14. DESCRIPTION OF AMENDMENT/MODIFICATION (Organized by UCF section headings, including solicitation/contract subject matter where feasible.)
10A. MOD. OF CONTRACT/ORDER NO.
2. AMENDMENT/MODIFICATION NO. 5. PROJECT NO.(If applicable)
6. ISSUED BY
3. EFFECTIVE DATE
13-Sep-2017
CODE
W40M USA HLTH CONTRACTING ACT
SOUTHERN REG CONTR OFC
3551 ROGER BROOK DR BLDG 3600
FORT SAM HOUSTON TX 78234-6200
W81K00 7. ADMINISTERED BY (If other than item 6)
4. REQUISITION/PURCHASE REQ. NO.
CODE W81K00
FACILITY CODECODE
EMAIL:TEL:
W81K00-17-T-0403
SECTION SF 30 BLOCK 14 CONTINUATION PAGE
SUMMARY OF CHANGES
SECTION SF 1449 - CONTINUATION SHEET
The following have been added by full text:
STATEMENT OF OBJECTIVES
The successful offeror shall furnish all equipment, reagents, consumables, controls, and any necessary services (e.g., maintenance of furnished equipment) to the Department of Pathology at General Leonard Wood Army Community
Hospital, Fort Leonard Wood, Missouri for the purpose of immunoassay testing under a Cost-Per-Reportable (CPR)
Agreement. All equipment and reagents/consumables shall be supplied and serviced through the Contractor to include all parts, equipment, supplies, and other services required to ensure GLWACH is able to meet its objectives as listed herein.
The contractor shall provide a lease/rental basis for one (1) fully automated discrete random access multiplex immunoassay analyzer and two (2) fully automated enzyme immunoassay (EIA) analyzers, training, supplies, reagents, unscheduled maintenance, preventive maintenance for the equipment and telephonic assistance.
Contractor shall include delivery, installation, all calibrators/standards, reagents/reagent packs, and quality control materials necessary to perform the initial method validation per CLIA/NCCLS QM guidelines and analyzer removal at the termination of the contract. The analyzers will be located at the following location:
General Leonard Wood Army Community Hospital Department of Pathology 4430 Missouri Ave.
Fort Leonard Wood, Missouri 65473
Period of Performance:
Base: 1 October 2017 – 30 September 2018 Option 1: 1 October 2018 – 30 September 2019 Option 2: 1 October 2019 – 30 September 2020 Option 3: 1 October 2020 – 30 September 2021 Option 4: 1 October 2021 – 30 September 2022
1.0 MULTIPLEX IMMUNOASSAY ANALYZER AND REAGENT REQUIRED FEATURES
1.1 ANALYZER REQUIRED FEATURES: The contractor must provide the following at a minimum for ASIP operations at the GLWACH laboratory.
1.1.1 The immunoassay analyzer system must be capable of interfacing with the laboratory’s Laboratory
Information System (LIS) Composite Health Care System (CHCS). The contractor must have an established driver compatible for interfacing with CHCS. The interface between CHCS and the analyzer must be bi-directional.
1.1.2 The immunoassay analyzer system must include a data management system for control of the operation of the analyzer, management of the quality control program, management of the on-instrument useful life of the reagents/reagent packs, and management of patient results; and a printer for the production of hard copy patient reports and other reports generated via the analyzer’s data management system.
1.1.3 The analyzer must be random access, discrete analyzer with STAT interrupt capability and a primary tube sampling system.
1.1.4 Immunoassay analyzer with multiplex methodology is required to permit simultaneous detection and identification of antibodies outlined in Table 1.
1.1.5 Initial startup from the shutdown state must not exceed 20 minutes.
1.1.6 The analyzer must come equipped with an un-interrupted power supply capable of providing all necessary electrical power to all components of the blood bank system, for a minimum of thirty (30) minutes.
1.1.7 The analyzer must allow for growth (i.e., increased numbers of patient testing or addition of additional assays to include vitamin D).
1.1.8 The analyzer and all reagents shall be FDA approved. Analyzers and associated parts and accessories shall be new. No used, refurbished or like-new equipment will be provided in support of the contract at any time.
1.1.9 The analyzer shall have a minimum capacity of 50 specimen samples and a minimum capacity of 5 onboard reagents.
1.1.10 The analyzer must process a minimum of 90 samples per hour.
1.1.11 The analyzer must perform a minimum of 1800 tests per hour.
1.1.12 The analyzer shall have the ability to test multiple analytes from a single tube.
1.1.13 Lab Layout: The offeror shall provide a diagram or other graphical representation demonstrating how its proposed equipment fits within the dimensions below with its quote.
1.2 REAGENT REQUIRED FEATURE: The contractor must provide the following at a minimum for ASIP
1.2.1 The analyzer must have the ability to maintain reagents on-board at the proper storage temperature.
1.2.2 The mandatory tests outlined in Table 1 must be approved by the FDA for serum and plasma samples.
1.2.3 Reagents/consumables must meet the requirements of the GLWACH mission, workload and test menu outlined in Table 2.
1.2.4 Reagents must have minimum shelf life upon receipt by the Government of at least four (4) weeks.
1.2.5 All reagents provided will be identical to those used in the FDA 510K approval application. The reagents/reagent packs will be provided as follows:
1.2.5.1 Be provided by the manufacturer of the equipment.
1.2.5.2 Be marked with the required storage temperature.
1.2.5.3 Be marked with the lot number and date of expiration.
Enzy me
Imm unoa ssay
Anal yzer
Enzyme Immunoassay Analyzer
Multiplex Immunoassay Analyzer
1.2.6 Reagents must be shipped in a manner that will preserve the proper storage temperature during transportation and be delivered on time at the appropriate storage temperature.
1.2.7 The contractor shall provide the reagents/reagent packs, calibrators/standards, user-replaceable maintenance items and analyzer specific tools/supplies.
1.2.8 Emergency orders will be delivered within 24 hours after the order is placed by 1400 Monday through
Thursday.
1.2.9 The Government retains the right to modify the quantity of each reagent/reagent pack based on changing patient population and test requests.
1.2.10 All reagents shall be FDA-approved and should exhibit at least 90% sensitivity and 90% specificity.
1.2.11 Reagents used for troubleshooting analytical systems shall be reimbursed (or replaced) by the Contractor at no additional cost.
1.2.12 Delivery of the reagents will be on an “As Needed” basis by the individual delivery orders issued by the contracting officer. The Contracting Office or their representative shall be the only entity authorized to issue orders against the resulting contract.
1.2.13 Delivery of reagents will not exceed 7 days from the date of request.
1.2.14 The contractor shall be responsible for the collection and transportation of equipment upon termination of the contract.
1.2.15 Contractor shall provide Safety Data Sheets (SDS) for each item delivered under this contract that is considered to be HAZARDOUS MATERIAL by the United States. The SDS shall contain information on the hazards associated with each specific chemical or material. The SDS shall contain, as a minimum, the following information: identification of the hazardous material and the specific hazard, the required practices for the safe use of the product, first aid precautions and what to do in the event of a spill or other mishap or accident.
Table 1
Mandatory Tests for Analyzer A FDA Approved for Serum Samples
Measles IgG Rubella IgG Varicella IgG
Anti-Nuclear Antibody Screen
(ANA) to include:
Ribosomal P SS-A SS-B
Centromere B dsDNA Chromatin DNP
Sm Sm/RNP Scl-70
Jo-1
Table 2
Annual Volume-Core Laboratory
TEST BASE OPTION 1 OPTION 2 OPTION 3 OPTION 4
Measles IgG 40,000 42,000 44,100 46,305 48,620
Rubella IgG 40,000 42,000 44,100 46,305 48,620
Rubeola IgG 40,000 42,000 44,100 46,305 48,620
ANA Screen 600 630 662 695 730
2.0 ENZYME IMMUNOASSAY (EIA) ANALYZERS AND REAGENT REQUIRED FEATURE
2.1 ANALYZER REQUIRED FEATURE: The contractor must provide the following at a minimum for ASIP
2.1.1 The immunoassay analyzes must be capable of interfacing with the laboratory’s Laboratory Information
System (LIS) Composite Health Care System (CHCS). The contractor must have an established driver compatible for interfacing with CHCS. The interface between CHCS and the analyzers must be bi-directional.
2.1.2 The analyzers must include a data management system for control of the operation of the analyzers, management of the quality control program, management of the on-instrument useful life of the reagents/reagent packs, and management of patient results; and a printer for the production of hard copy patient reports and other reports generated via the analyzer’s data management system.
2.1.3 The analyzers must be random access, discrete analyzers with STAT interrupt capability and a primary tube
2.1.4 Two analyzers with microplate enzyme immunoassay (EIA) methodology are required to perform tests outlined in Table 3.
2.1.5 Initial startup from the shutdown state must not exceed 20 minutes.
2.1.6 The analyzers must come equipped with an un-interrupted power supply capable of providing all necessary electrical power to all components of the analyzers, for a minimum of thirty (30) minutes.
2.1.7 The analyzers must allow for growth (i.e., increased numbers of patient testing or addition of additional assays to include QuantiFERON TB Gold).
2.1.8 The analyzers and all reagents shall be FDA approved. Analyzers and associated parts and accessories shall be new. No used, refurbished or like-new equipment will be provided in support of the contract at any time.
2.1.9 The analyzers shall have a minimum capacity of 120 specimen samples.
2.1.10 N/A
2.1.11 The analyzers shall have the ability to test multiple analytes from a single tube.
2.2 REAGENT REQUIRED FEATURES
2.2.1 N/A
2.2.2 The mandatory tests outlined in Table 3 must be approved by the FDA for serum samples.
2.2.3 Reagents/consumables must meet the requirements of the GLWACH mission, workload and test menu outlined in Table 4.
2.2.4 Reagents must have minimum shelf life upon receipt by the Government of at least four (4) weeks, or if this shelf life cannot be met, the Contractor must identify the guaranteed shelf life after receipt that can be accommodated.
2.2.5 All reagents provided will be identical to those used in the FDA 510K approval application. The
2.2.5.1 Be provided by the manufacturer of the equipment.
2.2.5.2 Be marked with the required storage temperature.
2.2.5.3 Be marked with lot number and expiration date.
2.2.6 Reagents must be shipped in a manner that will preserve the proper storage temperature during transportation
2.2.7 The contractor shall provide the reagents/reagent packs, calibrators/standards, user-replaceable maintenance
2.2.8 Emergency orders will be delivered within 24 hours after the order is placed by 1400 Monday through
2.2.9 The Government retains the right to modify the quantity of each reagent/reagent pack based on changing
2.2.10 All reagents shall be FDA-approved and should exhibit at least 90% sensitivity and 90% specificity.
2.2.11 Reagents used for troubleshooting analytical systems shall be reimbursed (or replaced) by the Contractor at
2.2.12 Delivery of the reagents will be on an “As Needed” basis by the individual delivery orders issued by the contracting officer. The Contracting Office or their representative shall be the only entity authorized to issue orders
2.2.13 Delivery of reagents will not exceed 7 days from the date of request.
2.2.14 The contractor shall be responsible for the collection and transportation of equipment upon termination of the
2.2.15 Contractor shall provide Safety Data Sheets (SDS) for each item delivered under this contract that is hazards associated with each specific chemical or material. The SDS shall contain, as a minimum, the following information: identification of the hazardous material and the specific hazard, the required practices for the safe use of the product, first aid precautions and what to do in the event of a spill or other mishap or accident.
Table 3
Mandatory Tests for Analyzers B Approved by the FDA for Serum Samples
QuantiFERON TB Gold HBc IgM Anti-HBc
HBsAg HBsAb HCV
HAV IgM HAVT
Table 4
TEST BASE OPTION 1 OPTION 2 OPTION 3 OPTION 4
Anti-HBc 700 735 772 811 852
HBc IgM 1,200 1,260 1,323 1,389 1,458
HBsAg 1,800 1,890 1,985 2,084 2,188
HBsAb 23,000 24,150 25,358 26,626 27,957
HCV 2,600 2,730 2,867 3,010 3,161
HAV IgM 1,300 1,365 1,433 1,505 1,580
HAVT 23,000 24,150 25,358 26,626 27,957
QuantiFERON TB
Gold
7,200 7,560 7,938 8,335 8,752
3.0 MAINTENANCE AND SERVICE:
3.1 MAINTENANCE:
3.1.1 At the start of the contract, the Contractor shall have a contract representative and alternate responsible for scheduling and coordination of service calls. The name and phone number of the designated individual and alternate shall be provided to the Chief, Medical Maintenance Branch at (573) 596-0482.
3.1.2 Contractor shall deliver all new equipment to the Property Management Section at the beginning of the contract. Clinical Engineering Section will perform acceptance inspection and establish maintenance records for equipment. These records will be maintained for the duration of the contract period. Clinical Engineering will maintain historical data that shows the acceptance inspection and services performed on the equipment by Medical
Maintenance and/or the MFG as long as the equipment is in the Government Facility.
3.1.3 Any and all equipment contracts to include, lease, cost per test, loan, etc., must be reviewed and approved by the Chief, Clinical Engineering and reviewed by the PBO for purposes of maintenance and accountability, whether accomplished by in-house maintenance or contract personnel.
3.1.4 The GLWACH Medical Maintenance Branch will first evaluate any equipment malfunction and facilitate repairs (screening process). If a service call is necessary, the Contractor shall provide unlimited service calls. Only the Medical Maintenance Branch will determine if the required repair service is emergency or routine.
3.1.5 The contractor shall be responsible for furnishing maintenance, services, repairs and parts as required in order to maintain the system is operating in the proper working condition in accordance with the Original Equipment
Manufacture’s (OEM) specifications.
3.1.6 The contractor is required to maintain all equipment installed under this contract except for repairs necessitated by willful damage or negligence on the part of the U.S. Government.
3.1.7 The contractor shall provide a full unscheduled and preventive maintenance service for the performance period at no additional cost. Service shall include repair and replacement of defective parts, complete maintenance program (as required by the manufacturer's manuals), and hotline telephone service to assist operators in troubleshooting problems i n accordance with paragraph 3.4.2 of this Statement of Work.
3.2 REPAIR:
3.2.1 Upon notification of equipment failure, the Contractor shall respond to telephonic requests for unscheduled repair within one (1) working day. Should the analyzer become inoperable, the Contractor’s technical specialist must be on-site within eight (8) hours of initial notification.
3.2.2 All work shall be performed in a professional manner by an authorized service representative or have two (2) years of maintenance experience on the contracted equipment. If any deficiencies are found due to negligence of the service representative, the contractor shall be required to correct the deficiency to a fully operational status in accordance with manufacturer specifications at no additional cost to the U.S. Government.
3.2.3 Services are to be performed during the month(s) of October through September. Contractor shall respond within 48-hours, during normal duty hours of 8:00 AM and 4:00 PM excluding Federal Public Holidays.
3.2.4 LEGAL PUBLIC HOLIDAYS:
New Year’s Day, January 1
Martin Luther King's Birthday, 3rd Monday in January
Washington's Birthday, 3rd Monday in February
Memorial Day, last Monday in May
Independence Day, July 4
Labor Day, 1st Monday in September
Columbus Day, 2nd Monday in October
Veteran's Day, November 11
Thanksgiving Day, 4th Thursday in November
Christmas Day, December 25
3.2.5 Contractor's representative shall report to Medical Maintenance, 4430 Missouri Avenue, Room No. 034, General Leonard Wood Army Community Hospital, Telephone No. 573-596-0482, prior to commencing services during normal operating hours. The contractor will get a security badge from Logistics Division, Room No. 034.
During other than normal operating hours, contractor's representative shall report to Staff Duty, Front Desk, 4430
Missouri Avenue, General Leonard Wood Army Community Hospital. The service representative will obtain a badge before reporting to service area.
3.2.6 Contractor's service representative shall report in person or telephonically notify the maintenance manager, Building No. 310, 4430 Missouri Avenue, Medical Maintenance Department, Telephone No. 573-596-0482 prior to commencing services during normal duty operating hours of 8:00 AM and 4:00 PM.
3.2.7 The government and the contractor’s service representative shall exchange any hazard communication information before the commencement of any repair.
3.2.8 When required, the contractor’s service representative will comply with the Office of Safety and Health
Administration lockout/tag out standards while performing maintenance on equipment.
3.2.9 Upon completion of services by the contractor’s service representative, a written service report shall be provided to the maintenance manager or the AOD. The service report shall provide detailed information regarding the cause of the equipment malfunction and corrective action taken. Include, at a minimum, 1) the time required to complete the work 2) price of labor (hourly rate) 3) list of parts replaced with part numbers and prices for each part
4) Nomenclature and Serial Number 5) Date 6) Name of the Technician.
3.2.10 In the event all information is not available to the contractor's service representative when services are performed, the initial service report shall include all available information. The contractor shall provide the balance of the required information to the maintenance manager no later than 10 days after services are completed.
3.2.11 In the event of calibration, after performing Calibration/Verification/Certification services, the contractor’s service representative will affix and/or update DD Form 2163 (Medical Equipment Verification/Certification). The contractor shall complete DD Form 2163 in accordance with the instructions provided in Technical Bulletin (TB)
38-750-2, or by the maintenance activity’s internal SOP.
3.2.12 If the contractor's calibration equipment produces a printed summary of the calibration procedure used, attach the printed summary to the DD Form 2164. Ensure that the heading of the DD Form 2164 is filled out and that the form is properly signed.
3.2.13 Contractor’s service representative will be factory trained and have a minimum of two years of experience working on the contracted equipment.
3.3 Not used.
3.4 SERVICE:
3.4.1 The GLWACH Medical Maintenance Branch will first evaluate any equipment malfunction and facilitate repairs (screening process). If a service call is necessary, the Contractor shall provide unlimited service calls. Only the Medical Maintenance Branch will determine if the required repair service is emergency or routine.
3.4.2 The Contractor shall provide direct customer service and support, 24 hours/7 days a week, to include hotline telephone service and on-call emergency repair service support, to assist operators i n correcting equipment operation problems.
3.4.3 Preventive maintenance services shall be performed in accordance with the manufacturer's standards/procedures. A preventive maintenance service shall include, but is not limited to, safety, calibration, complete operational testing, lubrication, adjustments and cleaning of equipment to which the operator does not have access. This also includes the installation of all non-operator parts required to ensure proper operation.
3.4.4 The contractor shall notify the Point of Contact of the exact date and time for performance of a preventive maintenance service at least two (2) days prior to the scheduled preventive maintenance services.
3.4.5 The contractor's qualified service technician shall be onsite to provide routine services, service requiring immediate repair and one (1), two (2) or four (4) scheduled services for preventive maintenance (as specified in the
Contractor's instrument maintenance manual), safety and calibration as required by this performance based work statement. The contractor will take every reasonable action to have the equipment repaired and operational within
48 hours after initial response by the contractor.
3.4.6 The Contractor shall perform routine repair service between 8:00 AM and 4:15 PM Monday through Friday, excluding Legal Public Holidays. Contractor response time shall be within 48 clock hours after telephonic notification excluding weekends and Legal Public Holidays.
3.4.7 Repair completion time for a routine repair service shall not be more than 96 clock hours after telephonic
3.4.8 Upon analyzer failure between 0730-1 630 Monday-Fridays, the Contractor's qualified service technician shall be on-site to provide unlimited emergency service (service requiring immediate repair; an unscheduled maintenance/service requirement) within 24 hours.
3.4.9 Should the analyzers at GLWACH become inoperable, the Contractor's technical repair specialist should be on-site within eight (8) hours of initial notification and the repair of the analzyer completed within one (1) working day of initial notification. If deemed necessary after consultation with the contractor, if analyzer cannot be repaired within one (1) working day of the initial notification, contractor will work with GLWACH for determining the identification of alternative sources of testing. The Contractor will pay all costs associated with the provision of this service, for as long as the analyzer remains inoperable.
3.4.10 If the equipment is determined to be out of service for longer than 48 hours (from the time of service technician's arrival on the site), the contractor shall notify the Lab Manager as to the reason(s) (i.e. non-availability of parts, etc.) for non-operability.
3.4.11 The Contractor shall provide all necessary user-replaceable maintenance items that will be required to support the analyzer, and any analyzer specific tools/supplies necessary to perform operator-level maintenance tasks, at no additional cost to the Government.
3.4.12 Contractor's representative shall report to Medical Maintenance, 4430 Missouri Avenue, Room No. 034, General Leonard Wood Army Community Hospital, Telephone No. 573-596-0482, prior to commencing services during normal operating hours. The contractor will get a security badge from Logistics Division, Room No. 034.
During other than normal operating hours, contractor's representative shall report to Staff Duty, Front Desk, 4430
Missouri Avenue, General Leonard Wood Army Community Hospital. The service representative will obtain a
3.4.13 Service reports. Upon completion of services, a written service report shall be provided to the medical equipment maintenance manager, or the government medical equipment repairer assigned to the service call.
3.4.14 The service report shall indicate the Contract Number, service performed (e.g., PM and/or repair), provide detailed information regarding the cause of equipment malfunction, corrective action taken, to include the time required to complete the work. The service report shall also include the estimated price of labor (hourly labor rate);
a list of parts replaced with price for each part, the unit Site ID and ECN number, and the name of the technician performing the service.
3.4.15 After normal duty hours, service tickets shall be given to the Government AOD/SDNCO, at the Information/ reception desk. Faxed reports are acceptable with the concurrence of the Government.
3.4.16 In the event that all information is not available to contractor's representative when services are performed, the initial service report shall include all information available. Contractor shall provide the balance of the required information in writing to the Chief, Medical Equipment Branch, not later than 10 days after services are completed.
The remaining information should be mailed to: Commander, USA Medical Department Activity, ATTN: MCXP-
LO-MM, Service Reports, 126 Missouri Avenue, Ft. Leonard Wood, MO 65473-8952 or Faxed to 573-596-1786, ATTN: Medical Maintenance Branch. This address is not to be used for any purpose other than as stated in the paragraph and C.5.8.2.
3.4.17 Contractor shall provide a copy of the invoice to the Chief, Medical Maintenance Branch, for verification of services performed. This is in addition to the requirements of block 13 of the DD Form 1155, Order for Supplies or
Services. The copy of the invoice can be mailed to Commander, USA Medical Department Activity, and ATTN:
MCXP-LO-MM, 126 Missouri Avenue, Ft. Leonard Wood, MO 65473-8952 or Faxed to 573-596-1786, ATTN:
Service Report.
3.4.18 The Contractor shall be responsible for all items of Government equipment throughout the period they are under his control. The Contractor shall be responsible for any contractor-caused damage to Government-owned property and shall replace or repair at the contractor’s expense. The Contracting Officer will approve replacements or repairs.
3.4.19 The Contractor shall ensure all areas where equipment is serviced on-site are left in a clean, neat, safe, and orderly condition. To prevent safety hazards, no equipment shall be left pulled out or apart at the end of the workday. All equipment under repair shall be secured at all times.
3.4.20 If the Contractor determines that the item being serviced is uneconomically repairable, the Chief, Medical
Maintenance Branch shall be notified. The Chief, Medical Maintenance Branch, GLWACH will determine economic reparability in accordance with US Army Directives.
4.0 TRAINING:
4.1.1 The Contractor will provide initial on-site operator training to all technicians assigned to the Pathology
Laboratory prior to initiation of use of the analyzer for patient testing. In addition, the Contractor will provide at least two (2) training slots per analyzer at the Contractor's training site at no additional cost to the U.S. Government.
The Contractor may charge the Government only for training services exceeding the basic requirement for a given year. Such training will be provided during designated implementation weeks and at mutually times, between the hours of 7 am - 7 pm local time.
4.1.2 Training shall include basic operation of the analyzers, troubleshooting procedures, performance of operator-level periodic preventive maintenance procedures and use of the data management/quality control software.
5.0 Delivery and Installation:
5.1.1 The Analyzer will be delivered to the location as indicated below within 30 calendar days after contract award. Equipment setup, on-site training, method validation studies, and analytical measurement range/linearity verification, shall (must) be completed by Contractor personnel within a period of time to ensure that equipment is fully operational on or before 1 October 2017. During the initial implementation, a dedicated technical representative must be available on site throughout the entire implementation period.
Substitutions must be coordinated with the COR.
Contractor shall provide a milestone chart with their quote that depicts significant events from the time of award through the "Go-Live" date.
5.1.2 PRIMARY LOCATIONS OF COMPLETE SYSTEMS:
4430 Missouri Avenue
Building 310
Fort Leonard Wood, Missouri 65473
5.1.2.3 Exact installation locations at the primary location shall be coordinated by the Contractor with the point of contacts listed below:
Laboratory Manager
Tel: 573-596-1509
The successful offeror shall furnish all equipment, reagents, consumables, controls, and any necessary services (e.g., maintenance of furnished equipment) to the Department of Pathology at General Leonard Wood Army Community
Hospital, Fort Leonard Wood, Missouri for the purpose of immunoassay testing under a Cost-Per-Reportable (CPR)
Agreement. All equipment and reagents/consumables shall be supplied and serviced through the Contractor to include all parts, equipment, supplies, and other services required to ensure GLWACH is able to meet its objectives as listed herein.
The contractor shall provide a lease/rental basis for one (1) fully automated discrete random access multiplex immunoassay analyzer and two (2) fully automated enzyme immunoassay (EIA) analyzers, training, supplies, reagents, unscheduled maintenance, preventive maintenance for the equipment and telephonic assistance.
Contractor shall include delivery, installation, all calibrators/standards, reagents/reagent packs, and quality control materials necessary to perform the initial method validation per CLIA/NCCLS QM guidelines and analyzer removal at the termination of the contract. The analyzers will be located at the following location:
4430 Missouri Ave.
Fort Leonard Wood, Missouri 65473
Period of Performance:
Base: 1 October 2017 – 30 September 2018 Option 1: 1 October 2018 – 30 September 2019 Option 2: 1 October 2019 – 30 September 2020 Option 3: 1 October 2020 – 30 September 2021 Option 4: 1 October 2021 – 30 September 2022
1.0 MULTIPLEX IMMUNOASSAY ANALYZER AND REAGENT REQUIRED FEATURES
1.1 ANALYZER REQUIRED FEATURES: The contractor must provide the following at a minimum for ASIP
1.1.1 The immunoassay analyzer system must be capable of interfacing with the laboratory’s Laboratory
Information System (LIS) Composite Health Care System (CHCS). The contractor must have an established driver compatible for interfacing with CHCS. The interface between CHCS and the analyzer must be bi-directional.
1.1.2 The immunoassay analyzer system must include a data management system for control of the operation of the analyzer, management of the quality control program, management of the on-instrument useful life of the reagents/reagent packs, and management of patient results; and a printer for the production of hard copy patient reports and other reports generated via the analyzer’s data management system.
1.1.3 The analyzer must be random access, discrete analyzer with STAT interrupt capability and a primary tube
1.1.4 Immunoassay analyzer with multiplex methodology is required to permit simultaneous detection and identification of antibodies outlined in Table 1.
1.1.5 Initial startup from the shutdown state must not exceed 20 minutes.
1.1.6 The analyzer must come equipped with an un-interrupted power supply capable of providing all necessary electrical power to all components of the blood bank system, for a minimum of thirty (30) minutes.
1.1.7 The analyzer must allow for growth (i.e., increased numbers of patient testing or addition of additional assays to include vitamin D).
1.1.8 The analyzer and all reagents shall be FDA approved. Analyzers and associated parts and accessories shall be new. No used, refurbished or like-new equipment will be provided in support of the contract at any time.
1.1.9 The analyzer shall have a minimum capacity of 50 specimen samples and a minimum capacity of 5 onboard reagents.
1.1.10 The analyzer must process a minimum of 90 samples per hour.
1.1.11 The analyzer must perform a minimum of 1800 tests per hour.
1.1.12 The analyzer shall have the ability to test multiple analytes from a single tube.
1.1.13 Lab Layout: The offeror shall provide a diagram or other graphical representation demonstrating how its proposed equipment fits within the dimensions below with its quote.
1.2 REAGENT REQUIRED FEATURE: The contractor must provide the following at a minimum for ASIP
1.2.1 The analyzer must have the ability to maintain reagents on-board at the proper storage temperature.
1.2.2 The mandatory tests outlined in Table 1 must be approved by the FDA for serum and plasma samples.
1.2.3 Reagents/consumables must meet the requirements of the GLWACH mission, workload and test menu outlined in Table 2.
1.2.4 Reagents must have minimum shelf life upon receipt by the Government of at least four (4) weeks.
1.2.5 All reagents provided will be identical to those used in the FDA 510K approval application. The
1.2.5.1 Be provided by the manufacturer of the equipment.
1.2.5.2 Be marked with the required storage temperature.
1.2.5.3 Be marked with the lot number and date of expiration.
Enzy me
Imm unoa ssay
Anal yzer
Enzyme Immunoassay Analyzer
Multiplex Immunoassay Analyzer
1.2.6 Reagents must be shipped in a manner that will preserve the proper storage temperature during transportation
1.2.7 The contractor shall provide the reagents/reagent packs, calibrators/standards, user-replaceable maintenance
1.2.8 Emergency orders will be delivered within 24 hours after the order is placed by 1400 Monday through
1.2.9 The Government retains the right to modify the quantity of each reagent/reagent pack based on changing
1.2.10 All reagents shall be FDA-approved and should exhibit at least 90% sensitivity and 90% specificity.
1.2.11 Reagents used for troubleshooting analytical systems shall be reimbursed (or replaced) by the Contractor at
1.2.12 Delivery of the reagents will be on an “As Needed” basis by the individual delivery orders issued by the contracting officer. The Contracting Office or their representative shall be the only entity authorized to issue orders
1.2.13 Delivery of reagents will not exceed 7 days from the date of request.
1.2.14 The contractor shall be responsible for the collection and transportation of equipment upon termination of the
1.2.15 Contractor shall provide Safety Data Sheets (SDS) for each item delivered under this contract that is hazards associated with each specific chemical or material. The SDS shall contain, as a minimum, the following information: identification of the hazardous material and the specific hazard, the required practices for the safe use of the product, first aid precautions and what to do in the event of a spill or other mishap or accident.
Table 1
Mandatory Tests for Analyzer A FDA Approved for Serum Samples
Measles IgG Rubella IgG Varicella IgG
Anti-Nuclear Antibody Screen
(ANA) to include:
Ribosomal P SS-A SS-B
Centromere B dsDNA Chromatin DNP
Sm Sm/RNP Scl-70
Jo-1
Table 2
TEST BASE OPTION 1 OPTION 2 OPTION 3 OPTION 4
Measles IgG 40,000 42,000 44,100 46,305 48,620
Rubella IgG 40,000 42,000 44,100 46,305 48,620
Rubeola IgG 40,000 42,000 44,100 46,305 48,620
ANA Screen 600 630 662 695 730
2.0 ENZYME IMMUNOASSAY (EIA) ANALYZERS AND REAGENT REQUIRED FEATURE
2.1 ANALYZER REQUIRED FEATURE: The contractor must provide the following at a minimum for ASIP
2.1.1 The immunoassay analyzes must be capable of interfacing with the laboratory’s Laboratory Information
System (LIS) Composite Health Care System (CHCS). The contractor must have an established driver compatible for interfacing with CHCS. The interface between CHCS and the analyzers must be bi-directional.
2.1.2 The analyzers must include a data management system for control of the operation of the analyzers, management of the quality control program, management of the on-instrument useful life of the reagents/reagent packs, and management of patient results; and a printer for the production of hard copy patient reports and other reports generated via the analyzer’s data management system.
2.1.3 The analyzers must be random access, discrete analyzers with STAT interrupt capability and a primary tube
2.1.4 Two analyzers with microplate enzyme immunoassay (EIA) methodology are required to perform tests outlined in Table 3.
2.1.5 Initial startup from the shutdown state must not exceed 20 minutes.
2.1.6 The analyzers must come equipped with an un-interrupted power supply capable of providing all necessary electrical power to all components of the analyzers, for a minimum of thirty (30) minutes.
2.1.7 The analyzers must allow for growth (i.e., increased numbers of patient testing or addition of additional assays to include QuantiFERON TB Gold).
2.1.8 The analyzers and all reagents shall be FDA approved. Analyzers and associated parts and accessories shall be new. No used, refurbished or like-new equipment will be provided in support of the contract at any time.
2.1.9 The analyzers shall have a minimum capacity of 120 specimen samples.
2.1.10 N/A
2.1.11 The analyzers shall have the ability to test multiple analytes from a single tube.
2.2 REAGENT REQUIRED FEATURES
2.2.1 N/A
2.2.2 The mandatory tests outlined in Table 3 must be approved by the FDA for serum samples.
2.2.3 Reagents/consumables must meet the requirements of the GLWACH mission, workload and test menu outlined in Table 4.
2.2.4 Reagents must have minimum shelf life upon receipt by the Government of at least four (4) weeks, or if this shelf life cannot be met, the Contractor must identify the guaranteed shelf life after receipt that can be accommodated.
2.2.5 All reagents provided will be identical to those used in the FDA 510K approval application. The
2.2.5.1 Be provided by the manufacturer of the equipment.
2.2.5.2 Be marked with the required storage temperature.
2.2.5.3 Be marked with lot number and expiration date.
2.2.6 Reagents must be shipped in a manner that will preserve the proper storage temperature during transportation
2.2.7 The contractor shall provide the reagents/reagent packs, calibrators/standards, user-replaceable maintenance
2.2.8 Emergency orders will be delivered within 24 hours after the order is placed by 1400 Monday through
2.2.9 The Government retains the right to modify the quantity of each reagent/reagent pack based on changing
2.2.10 All reagents shall be FDA-approved and should exhibit at least 90% sensitivity and 90% specificity.
2.2.11 Reagents used for troubleshooting analytical systems shall be reimbursed (or replaced) by the Contractor at
2.2.12 Delivery of the reagents will be on an “As Needed” basis by the individual delivery orders issued by the contracting officer. The Contracting Office or their representative shall be the only entity authorized to issue orders
2.2.13 Delivery of reagents will not exceed 7 days from the date of request.
2.2.14 The contractor shall be responsible for the collection and transportation of equipment upon termination of the
2.2.15 Contractor shall provide Safety Data Sheets (SDS) for each item delivered under this contract that is hazards associated with each specific chemical or material. The SDS shall contain, as a minimum, the following information: identification of the hazardous material and the specific hazard, the required practices for the safe use of the product, first aid precautions and what to do in the event of a spill or other mishap or accident.
Table 3
Mandatory Tests for Analyzers B Approved by the FDA for Serum Samples
QuantiFERON TB Gold HBc IgM Anti-HBc
HBsAg HBsAb HCV
HAV IgM HAVT
Table 4
TEST BASE OPTION 1 OPTION 2 OPTION 3 OPTION 4
Anti-HBc 700 735 772 811 852
HBc IgM 1,200 1,260 1,323 1,389 1,458
HBsAg 1,800 1,890 1,985 2,084 2,188
HBsAb 23,000 24,150 25,358 26,626 27,957
HCV 2,600 2,730 2,867 3,010 3,161
HAV IgM 1,300 1,365 1,433 1,505 1,580
HAVT 23,000 24,150 25,358 26,626 27,957
QuantiFERON TB
Gold
7,200 7,560 7,938 8,335 8,752
3.0 MAINTENANCE AND SERVICE:
3.1 MAINTENANCE:
3.1.1 At the start of the contract, the Contractor shall have a contract representative and alternate responsible for scheduling and coordination of service calls. The name and phone number of the designated individual and alternate shall be provided to the Chief, Medical Maintenance Branch at (573) 596-0482.
3.1.2 Contractor shall deliver all new equipment to the Property Management Section at the beginning of the contract. Clinical Engineering Section will perform acceptance inspection and establish maintenance records for equipment. These records will be maintained for the duration of the contract period. Clinical Engineering will maintain historical data that shows the acceptance inspection and services performed on the equipment by Medical
Maintenance and/or the MFG as long as the equipment is in the Government Facility.
3.1.3 Any and all equipment contracts to include, lease, cost per test, loan, etc., must be reviewed and approved by the Chief, Clinical Engineering and reviewed by the PBO for purposes of maintenance and accountability, whether accomplished by in-house maintenance or contract personnel.
3.1.4 The GLWACH Medical Maintenance Branch will first evaluate any equipment malfunction and facilitate repairs (screening process). If a service call is necessary, the Contractor shall provide unlimited service calls. Only the Medical Maintenance Branch will determine if the required repair service is emergency or routine.
3.1.5 The contractor shall be responsible for furnishing maintenance, services, repairs and parts as required in order to maintain the system is operating in the proper working condition in accordance with the Original Equipment
Manufacture’s (OEM) specifications.
3.1.6 The contractor is required to maintain all equipment installed under this contract except for repairs necessitated by willful damage or negligence on the part of the U.S. Government.
3.1.7 The contractor shall provide a full unscheduled and preventive maintenance service for the performance period at no additional cost. Service shall include repair and replacement of defective parts, complete maintenance program (as required by the manufacturer's manuals), and hotline telephone service to assist operators in troubleshooting problems i n accordance with paragraph 3.4.2 of this Statement of Work.
3.2 REPAIR:
3.2.1 Upon notification of equipment failure, the Contractor shall respond to telephonic requests for unscheduled repair within one (1) working day. Should the analyzer become inoperable, the Contractor’s technical specialist must be on-site within eight (8) hours of initial notification.
3.2.2 All work shall be performed in a professional manner by an authorized service representative or have two (2) years of maintenance experience on the contracted equipment. If any deficiencies are found due to negligence of the service representative, the contractor shall be required to correct the deficiency to a fully operational status in accordance with manufacturer specifications at no additional cost to the U.S. Government.
3.2.3 Services are to be performed during the month(s) of October through September. Contractor shall respond within 48-hours, during normal duty hours of 8:00 AM and 4:00 PM excluding Federal Public Holidays.
3.2.4 LEGAL PUBLIC HOLIDAYS:
New Year’s Day, January 1
Martin Luther King's Birthday, 3rd Monday in January
Washington's Birthday, 3rd Monday in February
Memorial Day, last Monday in May
Independence Day, July 4
Labor Day, 1st Monday in September
Columbus Day, 2nd Monday in October
Veteran's Day, November 11
Thanksgiving Day, 4th Thursday in November
Christmas Day, December 25
3.2.5 Contractor's representative shall report to Medical Maintenance, 4430 Missouri Avenue, Room No. 034, General Leonard Wood Army Community Hospital, Telephone No. 573-596-0482, prior to commencing services during normal operating hours. The contractor will get a security badge from Logistics Division, Room No. 034.
During other than normal operating hours, contractor's representative shall report to Staff Duty, Front Desk, 4430
Missouri Avenue, General Leonard Wood Army Community Hospital. The service representative will obtain a
3.2.6 Contractor's service representative shall report in person or telephonically notify the maintenance manager, Building No. 310, 4430 Missouri Avenue, Medical Maintenance Department, Telephone No. 573-596-0482 prior to commencing services during normal duty operating hours of 8:00 AM and 4:00 PM.
3.2.7 The government and the contractor’s service representative shall exchange any hazard communication information before the commencement of any repair.
3.2.8 When required, the contractor’s service representative will comply with the Office of Safety and Health
Administration lockout/tag out standards while performing maintenance on equipment.
3.2.9 Upon completion of services by the contractor’s service representative, a written service report shall be provided to the maintenance manager or the AOD. The service report shall provide detailed information regarding the cause of the equipment malfunction and corrective action taken. Include, at a minimum, 1) the time required to complete the work 2) price of labor (hourly rate) 3) list of parts replaced with part numbers and prices for each part
4) Nomenclature and Serial Number 5) Date 6) Name of the Technician.
3.2.10 In the event all information is not available to the contractor's service representative when services are performed, the initial service report shall include all available information. The contractor shall provide the balance of the required information to the maintenance manager no later than 10 days after services are completed.
3.2.11 In the event of calibration, after performing Calibration/Verification/Certification services, the contractor’s service representative will affix and/or update DD Form 2163 (Medical Equipment Verification/Certification). The contractor shall complete DD Form 2163 in accordance with the instructions provided in Technical Bulletin (TB)
38-750-2, or by the maintenance activity’s internal SOP.
3.2.12 If the contractor's calibration equipment produces a printed summary of the calibration procedure used, attach the printed summary to the DD Form 2164. Ensure that the heading of the DD Form 2164 is filled out and that the form is properly signed.
3.2.13 Contractor’s service representative will be factory trained and have a minimum of two years of experience working on the contracted equipment.
3.3 Not used.
3.4 SERVICE:
3.4.1 The GLWACH Medical Maintenance Branch will first evaluate any equipment malfunction and facilitate repairs (screening process). If a service call is necessary, the Contractor shall provide unlimited service calls. Only the Medical Maintenance Branch will determine if the required repair service is emergency or routine.
3.4.2 The Contractor shall provide direct customer service and support, 24 hours/7 days a week, to include hotline telephone service and on-call emergency repair service support, to assist operators i n correcting equipment operation problems.
3.4.3 Preventive maintenance services shall be performed in accordance with the manufacturer's standards/procedures.