SPEC SHEET NRMAF0002300275.pdf

PDF 105 KB Posted

Attached to
ONE (1) PH ANALYZER AND ONE (1) CARBON ANALYZER
Solicitation number
NRMAF0002300275S1C
Issued by

View the file

Other files

File Type Posted
SON NRMAF0002300275.pdf PDF
SAM COMBINED SYNOPSIS POSTING NRMAF0002300275.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SPECIFICATIONS AND DESCRIPTIONS

NR‐MAF‐00023 ‐ 00275

The following specifications must be met in order to fulfill this bona fide need, justified by this procurement request.

Specifications are as follows:

Spectrophotometric Seawater pH Analyzer Model No. AS‐pH2 With Agilent Cary ‐ High precision digital pump (48K) w/ 25 ml Syringe ‐ High quality multi‐port valve (12 Ports) ‐ Water jacketed spectrophotometric flow cell ‐ Data acquisition unit ‐ 0.2g pure mCP ‐ 1 sample bag ‐ Precision: better than ± 0.001 pH units.

‐ Accuracy: better than ± 0.003 pH units.

‐ Wavelengths for pH analysis: 434 nm, 578 nm, and 705 nm (user‐definable and expandable).

‐ Indicator for pH: m‐cresol purple.

‐ Digital Pump: Tecan Cavro XLP 6000 pump equipped with a 12‐port valve and a 10 mL syringe ‐ Spectrophotometric cell: 10 cm with thermostatic water jacket.

‐ Precision of temperate‐control system: < 0.1°C ‐ Automated: run 9 samples sequentially ‐ Typical sample volume: ~ 40 mL per analysis ‐ Time consumption: ~ 5 minutes per analysis ‐ Working environment: Both land and shipboard laboratories

Dissolved Inorganic Carbon Analyzer model AS‐C5 (with LI‐850) ‐ Typical sample volume: 1.2 mL per analysis ‐ Time consumption: ~3 minutes per analysis ‐ Precision: better than ± 0.15% or ± 3 μM for seawater ‐ Analytical range: 0.2 – 20 mM ‐ Only one standard is required 6. Working environment: both land and shipboard laboratories

File details come from the government source that posted it. Updated .