Spec Sheet IV Hood 03.pdf

PDF 2 MB Posted

Attached to
6515--Compounding Hoods (VA-23-00074588)
Solicitation number
36C24723Q0792
Issued by

View the file

Other files

File Type Posted
36C24723Q0792.docx DOCX document
Factory Specifications BSC.pdf PDF
Spec Sheet IV Hood.pdf PDF
Brochure IV Hood.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

NU-240 AireGard™ ES (Energy Saver) Bench Top Horizontal Laminar Airflow Workstation

NuAire® | 2100 Fernbrook Lane | Plymouth, MN 55447 | USA | 763-553-1270 | www.nuaire.com

© Copyright 2020 NuAire, Inc. All Rights Reserved

20-1478-F-G-EN-3-0919

NU-240 TABLE TOP HORIZONTAL LAMINAR AIRFLOW WORKSTATION

The NuAire NU-240 AireGard™ ES Horizontal Laminar Airflow Worksta-tion (LAFW) model NU-240 is designed as a table top clean bench that features the newest technologies in laminar airflow design, materials, and manufacturing processes. The clean workstation can be used where ISO Class 5 clean airflow is required for the preparation of in-jectable drugs, IV solutions, tissue culture, optics, microelectron-ics, etc. The clean workstation is used to protect work product from contamination. The clean workstation is available with an optional base stand, placing the work surface at 30” (762mm) or 36” (914mm).

A significant number of design innovations give the NuAire Laminar flow equipment superior performance qualities in airflow, lighting, noise levels, and vibration.

Electronic Control system monitors and displays airflow system performance via

PressureFlow - A specialized digital pressure transducer. Audible and visual alarms will trigger in the event of high or low airflow status.

On/Off Outlets

On/Off Blower

On/Off Light

On/Off UV Light

Silence Visual Alarm

Up Key

Down KeyAirflow Monitor

Features

• Control blower & lights via solid state switch.

• Disable audible alarm switch with ring back function.

• Control DC ECM motor with solid-state Motor Controller that provides automatic compensation (constant volume control) for both filter loading and line voltage variances.

• Security password protection of Work station use.

• Work station usage sync functions with blower, fluorescent light.

• Work station usage auto duration timers, fluorescent light.

NU-240 AireGard™ ES (Energy Saver) Bench Top Horizontal Laminar Airflow Workstation

NuAire® | 2100 Fernbrook Lane | Plymouth, MN 55447 | USA | 763-553-1270 | www.nuaire.com

© Copyright 2020 NuAire, Inc. All Rights Reserved

20-1478-F-G-EN-3-0919

NU-240 TABLE TOP HORIZONTAL LAMINAR AIRFLOW WORKSTATION

Your Local NuAire PartnerPopular Accessories/Built-in Options

Ergotron® Arm for Monitor and Keyboard/Mouse

IV Bar

- 3 Height Locations

- 6 Hooks

Telescoping Base Stand

- Work Surface Height Range

30” - 36” (762 - 914 mm)

- Feet Options: Leg Levelers or Casters

- Shelf, 8 or 12-inch width

- OSHPD OPM Seismic Approved

Quick Specifications

Model Width Size (nominal)

Electrical* (all sizes) Work Area Dimensions (W x D x H) [mm]

Exterior Dimensions (W x D x H) [mm]

Net Weight

NU-240-330 / E

3 ft

[0.9 m]

115 VAC / 60 Hz E: 230 VAC / 50 Hz-60 Hz

35 7/8 x 19 5/8 x 32 [911 x 498 x 813]

38 x 32 x 55 3/8

[965 x 813 x 1406] 330 lbs [150 kg]

NU-240-336

35 7/8 x 19 5/8 x 38 [911 x 498 x 965]

38 x 32 x 61 3/8

[965 x 813 x 1559] 345 lbs [156 kg]

NU-240-430 / E

4 ft

[1.2 m]

47 7/8 x 19 5/8 x 32 [1216 x 498 x 813]

50 x 32 x 55 3/8

[1270 x 813 x 1406] 390lbs [177kg]

NU-240-436

47 7/8 x 19 5/8 x 38 [1216 x 498 x 965]

50 x 32 x 61 3/8

[1270 x 813 x 1559] 405 lbs [184 kg]

NU-240-530 / E

5 ft

[1.5 m]

59 7/8 x 19 5/8 x 32 [1521 x 498 x 813]

62 x 32 x 55 3/8

[1575 x 813 x 1406] 460 lbs [209 kg]

NU-240-536

59 7/8 x 19 5/8 x 38 [1521 x 498 x 965]

62 x 32 x 61 3/8

[1575 x 813 x 1559] 480 lbs [218 kg]

NU-240-630 / E 6 ft [1.8 m]

71 7/8 x 19 5/8 x 32 [1826 x 498 x 813]

74 x 32 x 55 3/8

[1880 x 813 x 1406] 540 lbs [245 kg]

*Specify model with appropriate letter suffix for electrical specifications. “NU-340-330E” for 230 VAC / 50 - 60 Hz

Performance

True laminar horizontal air of 90 fpm (0.46 m/s)

250% Filter load capacity (±10 Year Average dependent on environment)

Meets performance requirements IEST- RP-CC002.4 and ISO 14644-1

ISO Class 5 Air Quality

Protection

HEPEX Zero Leak Airflow System

Motor

Ultra-High Efficiency DC ECM Motor with optimally determined forward-curved fan

Attenumont Vibration Control

Controls

AeroMax Electronic Control System w/ PressureFlow Monitor

Filter

3-inch (76 mm) Metal Frame Separatorless HEPA Filter 99.995% efficient @ 0.3 µ

Permanent positive pressure plenum with quick release supply filter removal

Pre-filter System

Ergonomics

ADA Compliant

Large User-friendly Work zone

Cleaning

Quick removal protective metal filter diffuser

Work zone back wall spill lip

Smooth interior

Features Design Your Own Nominal Width

3-foot (0.9 m)

4-foot (1.2 m)

5-foot (1.5 m)

6-foot (1.8 m)

Connections Kits for Extended Sizes

Filter Height

30-inches (762 mm)

36-inches (914 mm)

Work Surface

16/18 Gauge Type 304 Stainless Steel with PVC Core (standard)

Nevamar with PVC Core

Sidewalls

16/18 Gauge Type 304 Stainless Steel with Cord Pass Through on either Sidewall (standard)

Polycarbonate with Cord Pass Through on either Sidewall

Lighting

Fluorescent [90 fc (968 lux)] (standard)

LED [90 fc (968 lux)]

Options

IV Bar with 3 locations, 6 hooks

Outlets

Service Valves

Storage Drawers

Seismic Brackets

Accessories

Telescoping Base Stand (Casters, Leg Levelers)

Telescoping Base Stand (OSHPD OPM) (Leg Levelers)

Motorized Adjustable Height Base Stand (Casters, Leg Levelers)

Base Stand Shelf (8, 12-inch) [203, 305 mm]

Ergotron Computer Monitor/ Keyboard Arm

Adjustable Footrest

Stainless Steel Turntable

PROCESS #9-1236 Rev4 August/2019 Best Products. | Best Performance. | Best Protection. Page 1 of 4 NuAire, Inc. | 2100 Fernbrook Lane | Plymouth, MN 55447 | U.S.A | ph: 763.553.1270 | fx: 763.553.0459 | tf: 800.328.3352 | www.nuaire.com

PURCHASE SPECIFICATIONS: FOR NUAIRE AIREGARD™ ES ENERGY SAVER NU-240 HORIZONTAL LAMINAR FLOW WORK STATION

The intent herein is to provide a concise statement of requirements for a quality Horizontal Laminar Airflow Work Station which may be used to augment your purchase request/order.

The AIREGARD™ ES NU-240 meets the performance requirements of the IEST-RP-CC002.4 and ISO 14644-1.

NuAire sales representatives will be pleased to explain the importance of the performance and control affected by each of the following requirements. The NuAire AIREGARD™ ES NU-240 meets all of the requirements in the following SPECIFICATION.

1. Dimensions Inches (mm)

Overall Dimensions NU-240-330 NU-240-430 NU-240-530 NU-240-630

NU-240-336 NU-240-436 NU-240-536 -------

Width (W) 38 (965) 50 (1270) 62 (1575) 74 (1880)

Depth (D) 32 (813) 32 (813) 32 (813) 32 (813)

Height (H) w/30” HEPA filter 55 3/8 (1406) 55 3/8 (1406) 55 3/8 (1406) 55 3/8 (1406)

Height (H) w/36” HEPA filter 61 3/8 (1559) 61 3/8 (1559) 61 3/8 (1559) --------

Basestand, 30" to 36” W.S.

w/30” HEPA filter^

83 7/8 (2080) to 89 7/8 (2282)

83 7/8 (2080) to 89 7/8 (2282)

83 7/8 (2080) to 89 7/8 (2282)

83 7/8 (2080) to 89 7/8 (2282)

Basestand, 30 to 36” W.S.

w/36” HEPA filter^

89 7/8 (2232) to 95 7/8 (2435)

89 7/8 (2232) to 95 7/8 (2435)

89 7/8 (2232) to 95 7/8 (2435) --------

Interior Dimensions

Width (W) 35 7/8 (911) 47 7/8 (1216) 59 7/8 (1521) 71 7/8 (1826)

Depth (D) 19 5/8 (498) 19 5/8 (498) 19 5/8 (498) 19 5/8 (498)

Height (H) w/30” HEPA filter 32 (813) 32 (813) 32 (813) 32 (813)

Height (H) w/36” HEPA filter 38 (965) 38 (965) 38 (965) -------

^Measured using Telescoping Basestand with Leg Levelers

PROCESS #9-1236 Rev 4 August/2019 Best Products. | Best Performance. | Best Protection. Page 2 of 4 NuAire, Inc. | 2100 Fernbrook Lane | Plymouth, MN 55447 | U.S.A | ph: 763.553.1270 | fx: 763.553.0459 | tf: 800.328.3352 | www.nuaire.com

2. Work station shall provide airflow performance as specified.

a. Work station shall provide ISO 5 quality airflow over the entire work zone for product protection proven by particle test, (e.g. test conducted by NuAire) validated by NuAire.

* b. Work station work zone shall be constructed from 16/18 gauge, Type 304 stainless steel.

c. Supply HEPA filter 99.99% efficient @ 0.3 microns shall be of full work zone width and height.

* d. Supply HEPA filter shall be protected by a perforated metal diffuser.

e. Air Velocity from the supply filter shall average 80 to 100 FPM (.41 to .51 m/s) with no single point outside the

20% of average range measured in a horizontal plane defined by 6 inches (102mm) from the diffuser.

*Having all of these features is unique ONLY to NuAire work stations.

3.* The Work station shall be ergonomically designed for maximum user comfort and adjustability to meet the requirements of the American Disabilities Act (ADA).

• Work station shall have a large armrest/airfoil radius on the leading edge of the work surface designed for comfort while maximizing product protection performance.

• Work station shall have a centrally located control center that is easily serviced with quick disconnects.

• Work station shall have the capability of incorporating a user adjustable basestand or base storage Work station as an option.

• Work station shall have a large work surface (19-5/8” inch (498mm) depth) by work station width with raised rear spill lip for improved cleaning.

4.* The Work station shall have all positive pressure plenums surrounded by a vacuum relative to the room (the AIREGARD™ ES employs the HEPEX Zero Leak Airflow System).

5. Electrical power shall be supplied with a 12-foot (2.5m), 3-wire cord with molded plug. Electrical supply should be 115 VAC, 60 Hz (Current rating varies per Work station size. Reference electrical requirements) protected with thermal circuit breaker from distribution panel.

6. The Work station shall use a DC ECM Motor with an optimally determined forward-curved fan for each model size/width to maximize both energy efficiency and filter loading capacity.

7. The Work station shall have one internal electrical circuit for the blower and lights. The circuit shall be protected with fuse(s) located in the Control Center on the electronic module.

8.* The Work station shall be listed by Underwriters Laboratories to meet the requirements of both the U.S. and Canada for electrical/mechanical integrity.

9.* Work station shall contain the Aeromax™ control system consisting of electronic modules that will perform the following functions:

• Easy user interface via LED’s and function keys

• Control blower via solid state switch.

• Control lights via solid state switch.

• Disable audible alarm switch with ring back function.

• Control blower DC ECM motor with solid-state DC Motor Controller that provides automatic compensation (constant volume control) for both filter loading and line voltage variances.

• Monitor and display airflow system performance via PresurFlow™ monitor.

10.* Work station shall contain the Aeromax™ control system that provides the following optional functional features (included with Work station, but must be configured during certification):

• Security password protection of Work station use.

• Work station usage sync functions with blower, fluorescent light.

• Work station usage auto duration timers, fluorescent light.

11. The Work station shall be easily transportable through a standard 34 inch (864mm) wide door without disassembly.

12. Fluorescent lighting shall be externally mounted and provide a minimum of 90 (968) foot-candles (LUX) on work surface.

The ballast is to be electronic containing thermal protection with automatic reset.

PROCESS #9-1236 Rev 4 August/2019 Best Products. | Best Performance. | Best Protection. Page 3 of 4 NuAire, Inc. | 2100 Fernbrook Lane | Plymouth, MN 55447 | U.S.A | ph: 763.553.1270 | fx: 763.553.0459 | tf: 800.328.3352 | www.nuaire.com

13. Work station shall be easily converted to a free-standing console model with the addition of the optional base support stand.

14.* Work station work zone shall be all 16/18 gauge stainless steel.

15. Work station exterior shall be constructed of 16/18 gauge powdered coated (light gray) cold rolled steel.

16.* Work station shall have a permanent positive pressure plenum with quick release supply filter removal.

17.* Motor/blower shall be positioned so as to create an even filter loading, thereby prolonging the life of HEPA filters, automatically handling a 250% minimum increase in filter loading without reducing total air delivery by more than 10%.

18.* Work station shall be capable of front filter removal through the work zone without exterior disassembly.

19. The following optional equipment shall be available to support installation and user requirements:

• LED Lighting

• Additional Service Valves for Gas, Air, Vacuum

• Outlets (separate power cord)

• IV Bar with 6 Stainless Steel Hooks

• Base Support Stand (available as telescoping for work surface heights of 30 to 36 inches) (762 to 914mm) with or without Storage Shelves

• Adjustable Control for Support Stand or Storage Work station

• Storage Pull-Out Trays

• GFI Outlets

• Ergotron monitor/keyboard holders

• Polycarbonate side walls

• Laminate (white) work surface

PROCESS #9-1236 Rev 4 August/2019 Best Products. | Best Performance. | Best Protection. Page 4 of 4 NuAire, Inc. | 2100 Fernbrook Lane | Plymouth, MN 55447 | U.S.A | ph: 763.553.1270 | fx: 763.553.0459 | tf: 800.328.3352 | www.nuaire.com

AIREGARD™ ES Energy Saver Horizontal Laminar Flow Work station Models NU-240-330/336/430/436/530/536/630

Catalog Number

NU-240-330

NU-240-336

Nominal 3 foot (0.9m)

NU-240-430

NU-240-436

Nominal 4 foot (1.2m)

NU-240-530

NU-240-536

Nominal 5 foot (1.5m)

NU-240-630

Nominal 6 foot (1.8m) Performance Specifications

1. Product Protection

IEST-RP-CC002.4

ISO 14644-1

IEST-RP-CC002.4

ISO 14644-1

IEST-RP-CC002.4

ISO 14644-1

IEST-RP-CC002.4

ISO 14644-1

ISO 14644-1 ISO 5 ISO 5 ISO 5 ISO 5

Style of Work station Bench top/console w/base stand/storage Work station

Bench top/console w/base stand/storage Work station

Bench top/console w/base stand/storage Work station

Bench top/console w/base stand/storage Work station

Work Station Construction

16/18 GA, powder coated steel exterior;

16/18 GA type 304 SST work zone

16/18 GA, powder coated steel exterior;

16/18 GA type 304 SST work zone

16/18 GA, powder coated steel exterior;

16/18 GA type 304 SST work zone

16/18 GA, powder coated steel exterior;

16/18 GA type 304 SST work zone

Diffuser for Air Supply (Metal) Non-flammable Non-flammable Non-flammable Non-flammable HEPA Filter Seal Type:

Supply Filter-99.99% Eff. on 0.3 microns

HEPEX Seal- Neoprene, Spring-loaded

HEPEX Seal- Neoprene, Spring-loaded

HEPEX Seal- Neoprene, Spring-loaded

HEPEX Seal- Neoprene, Spring-loaded

Optional Services, (Total) Position:

Service Coupling (3/8 inch NPT) Gas Valve/Service Coupling (3/8inch NPT) Outlet

(3) Right/left sidewalls

(1) Right/left area under work surface

(1) Right/left area under work surface

Work Station Size Inches (mm):

Height (w/30” HEPA filter) Height (w/36” HEPA filter) Width Depth

55 3/8 (1406) 61 3/8 (1559) 38 (965) 32 (813)

50 (1270)

62 (1575)

74 (1880) 32 (813)

Work Zone Inches (mm):

Height (w/30” HEPA filter) Height (w/36” HEPA filter) Width Depth

35 7/8 (911) 19 5/8 (498)

47 7/8 (1216)

59 7/8 (1521)

71 7/8 (1826) 19 5/8 (498)

Airflow Volume at 90 fpm (.46 mps)

CFM (CMH)

(w/30” HEPA filter) (w/36” HEPA filter)

718 (1220) 852 (1448)

958 (1628) 1137 (1932)

1198 (2036) 1422 (2416)

1438 (2443)

Heat Rejected, BTU, Per Hour (w/30” HEPA filter) (w/36” HEPA filter)

Electrical:

Volts, AC 60 Hz +Amps: Blower/Lights (30”/36”) Rated Amps:

12 ft. Power Cord (one)

U.L./U.L.-C Listed 2.2/2.8 14 GA - 3 Wire, 15A

U.L./U.L.-C Listed 3.2/3.8 14 GA-3 Wire, 15A

U.L./U.L.-C Listed 4.2/4.8 14 GA-3 Wire, 15A

U.L./U.L.-C Listed 5.0 14 GA-3 Wire, 15A

Work Station Weights:*** Crated Shipping Weight (30”) Net Weight (30”) Crated Shipping Weight (36”) Net Weight (36”)

330 lbs. /150 kg.

290 lbs. /132 kg.

345 lbs. /156 kg.

305 lbs. /138 kg.

390 lbs. /177 kg.

340 lbs. /154 kg.

405 lbs. /184 kg.

355 lbs. /161 kg.

450 lbs. /204 kg.

400 lbs. /181 kg.

470 lbs. /213 kg.

420 lbs. /191 kg.

540 lbs. /245 kg.

470 lbs. /213 kg.

***Crated shipping weight does not include weight for accessories or options + Based on Work station with new filters running at 115VAC.

SECTION A-A

A

A

TITLE

CHKDDATEDFTM

TM

DRAWING

NUMBER

NU-240-600 HORIZONTAL

AIRFLOW WORKSTATION

1 1

BCD-19666 C

3/8/2018LS BP

34 1/2 [876]

26 1/2 [673]

TO

(AS SHOWN)

TELESCOPING

(DIM "A")

WORKZONE

HEIGHT

* (DIM "B")

* 74 [1880]

71 7/8 [1826]

WORKZONE

WIDTH

PERMANENT PLENUM

16 X 20 X 1

PREFILTERS

28 1/8 [716]

4 1/4 [109]

32 [814]*

INCH

MILLIMETER

30 7/8 [784]

19 5/8 [497]

WORKZONE

DEPTH

REV ECO DESCRIPTION DATE DFTM CHKD

C 13826 UPDATED DRAWING 9/11/2018 TH BP

1 3/8 [36]

* OVERALL DIMENSIONAL TOLERANCE ±1/4" [6.35]

ALL OTHER DIMENSIONS ±1/8" [3.17]

FLUORESCENT LIGHTS

MODEL DIMS

MODEL DIM "A" DIM "B" DIM "C" HEPA FILTER SIZE

NU-240-630 32 [813] 81 7/8 [2080] TO 89 7/8 [2282] 55 3/8 [1407] 70 [1778] X 30 [762] X 3 [76 ]

DC ECM

MOTOR / BLOWER

PAINTED EXTERIOR

PANELS

ALUMINUM DIFFUSER

(REMOVABLE) (2)

SST WORKZONE

SIDEWALLS

AND TOP PANEL

SST WORK SURFACE

W/ RIGID NON-SHEADING

PVC CORE

SILICONE

CORD PASSTHRU

BOTH SIDES HEPA FILTER (99.99% @ 0.3M)

SEE "MODEL DIMS" TABLE

POWER CORD

(NEMA 5-15P)

NU-200-136 (SHOWN)

OPTIONAL BASESTAND

IV BARS OPTIONAL

* DIM "C"

OUTLET

POWER CORD

(NEMA 5-15P)

DUPLEX OUTLET (OPTIONAL)

CONTROL PANEL

AEROMAX CONTROL SYSTEM

PRESSURE FLOW CONTROL SYSTEM

AIRFLOW CONTROL

LIGHT SWITCH

BLOWER SWITCH

SYSTEM DIAGNOSTICS

[660]

TO

[914]

OPTIONAL

AUTOLIFT

NSF APPROVED LEG

LEVELERS ± 3/4

ADJUSTMENT

File details come from the government source that posted it. Updated .