NuAire® | 2100 Fernbrook Lane | Plymouth, MN 55447 | USA | 763-553-1270 | www.nuaire.com
© Copyright 2018 NuAire, Inc. All Rights Reserved
20-1307-F-D-EN-6-0818
NU-543 LabGard® AIR Class II, Type A2 Biosafety Cabinet
Select from a variety of window access openings:
• 8” (203 mm) for lower operating cost (NSF/ANSI 49)
• 10” (254 mm) combing lower costs and improved ergonomics (NSF/ANSI 49, EN12469)
• 12” (305 mm) for greater ergonomic reach by 40%* (NSF/ANSI 49)
A recessed airfoil grill allows users to use the ergonomic armrest without blocking airflow compromising safety. A recessed work tray provides users greater flexibility when working with bottles or pipettes.
A shorter reach into the work zone allows users to bring work closer without sacrificing safety. These are just some of the user focused ergonomic designs NuAire® incorporates into all our laboratory products.
The LabGard® AIR model NU-543 a NSF/ANSI 49 and EN12469 listed Class II, Type A2 Biosafety Cabinet offering protection for Personnel, Product, and Environment. LabGard® AIR can be either exhausted back into the room or connected to a facility HVAC system to minimize cross-contamination of low to moderate risk biologicals in the absence of volatile toxic chemicals.
Class II, Type A2 Air In-flow 70% Recirculated vs. 30% Exhausted
HEPA Filtered Air
Contaminated Worksurface Air
Contaminated Room Air
Performance Features:
• 30% exhausted / 70% recirculated
• Dynamic air barrier of 105 fpm (0.53 m/s) inflow separating laboratory from work zone environment
• True Laminar Unidirectional Down-flow air of 60 fpm (0.30 m/s) for minimal cross contamination
• 300% Filter Load Capacity (±10 Year Average depending on laboratory environment)
• ISO Class 5 Air Quality
* As compared to an 8” (203 mm) access opening.
Downflow Inflow 12:00
60 105 24.0
FPM >>
o
Electronic Control System utilizes intelliflow, a digital thermistor airflow probe in the exhaust airflow to analyze, monitor and, control down-flow and inflow air.
Night setback mode is initiated by the window closure, will reduce motor / blower operational airflow to conserve energy while maintaining work zone sterility.
On/Off Outlets
Down-flow Air
Inflow Air
Air Temperature Time Up Key
On/Off Blower
On/Off Lights
On/Off UV Lights
Silence Enter Key Airflow Range Down KeyVisual Alarm
NU-543 CLASS II TYPE A2 BIOSAFETY CABINET
NuAire® | 2100 Fernbrook Lane | Plymouth, MN 55447 | USA | 763-553-1270 | www.nuaire.com
© Copyright 2018 NuAire, Inc. All Rights Reserved
20-1307-F-D-EN-6-0818
NU-543 LabGard® AIR Class II, Type A2 Biosafety Cabinet
*Specify model with appropriate letter suffix for electrical specifications. “NU-540-300E” for 230 VAC / 50 - 60 Hz
Model Width Size (nominal)
Electrical* (all sizes) Interior Dimensions
(W x D x H) [mm] Exterior Dimensions
(W x D x H) [mm] Net Weight
NU-543-300 3 ft [0.9 m]
115 VAC / 60 Hz E: 230 VAC / 50 Hz-60 Hz
34 3/8 x 25 3/4 x 28 1/2 [873 x 654 x 724]
41 5/8 x 31 7/16 x 60 7/8 [1057 x 799 x 1546]
410 lbs [186 kg]
NU-543-400 4 ft [1.2 m]
46 3/8 x 25 3/4 x 28 1/2 [1178 x 654 x 724]
53 5/8 x 31 7/16 x 60 7/8 [1362 x 799 x 1546]
480 lbs [218 kg]
NU-543-500 5 ft [1.5 m]
58 3/8 x 25 3/4 x 28 1/2 [1483 x 654 x 724]
65 5/8 x 31 7/16 x 60 7/8 [1669 x 799 x 1546]
570 lbs [258 kg]
NU-543-600 6 ft [1.8 m]
70 3/8 x 25 3/4 x 28 1/2 [1788 x 654 x 724]
77 5/8 x 31 7/16 x 60 7/8 [1972 x 799 x 1546]
640 lbs [290 kg]
Specifications
Efficiency Features:
• Ultra High Efficiency ECM Eon™ Motor
• Auto Filter Loading Compensation
• Motor Speed Controller
• Internal Exhaust Damper
• nitecare Night Setback
Superior Construction Features:
• Monolithic 16/18 Gauge, Type 304 Stainless Steel
Welded Pressure Tight Construction (Silicone Free)
• Removable Control Center
• Removable Stainless Steel Coved Work Tray
• Recessed and Removable Front Intake Grill
• Stainless Steel Spill Trough with Drain Valve
• Vibration Reducing Work Surface Support System
• External Non-Glare 100 lm/ft2 Fluorescent Lighting
Safety Features:
• Large HEPA Filters
• Airflow / Window Alarms with ring back function
• Security Password Protection
• HEPEX Zero Leak Airflow System
• Metal Diffuser over Supply HEPA Filter
• NSF/ANSI 49, EN12469, UL, CE Listed
Visit www2.nuaire.com/06380 to make a request online or scan the code.
Popular Accessories/Built-in Options
Prop Up Work Surface Smooth Interior with Back Wall Monitor Cutout and Cord Pass Through
Arm Rest Pads and
Elbow Rests
Ultraviolet (UV) LightMotorized Base StandTelescoping Leg Leveler or Castor Base Stand
Standard Features Single Ultra High Efficiency Motor FlowGard Electronic Control System IntelliFlow Airflow Control Motor Speed Control Internal Exhaust Dampener Monolithic Pressure Tight Design (Silicone
Free) HEPEX Zero Leak Airflow Front Filter Removal Permanent Plenum with Quick Release
Supply Filter 10” [254 mm] Access Opening (NSF/ANSI 49, EN12469)
Frame-less Sliding Tempered Glass View
Screen External Fluorescent Lighting Ergonomic PVC Armrest Removable Work Tray Work Surface Support System Metal Diffuser Over Supply Filter Spill Trough with Drain Valve One Service Valve Right Side Wall One Service Coupling Right Side Wall Two Service Couplings Left Side Wall
Power Cord NSF/ANSI 49 Standard Features
3” (76 mm) Metal Frame Supply HEPA Filter 99.99% @ 0.3 Microns
12” (305 mm) Metal Frame Exhaust HEPA Filter 99.99% @ 0.3 Microns
Two Duplex Outlets (Back Wall) EN12469 Standard Features
3” (76 mm) Metal Frame Supply H14 HEPA Filter 99.995% @ 0.025 MPPS
12” (305 mm) Metal Frame Supply H14 HEPA Filter 99.995% @ 0.025 MPPS
Two Outlets (Back Wall) Optional Features and Accessories
Glass Sidewalls 12” (305 mm) Access Opening (NSF/ANSI 49) 8” (203 mm) Access Opening (NSF/ANSI 49) Ultraviolet (UV) Light LED Lighting Additional Service Valves for Gas, Air, Vacuum IV Bar with 3 Locations, 6 Stainless Steel
Hooks Exhaust Transition Canopy Back Wall Monitor Cutout
Storage Pull-Out Trays Lay in Sorbent Exhaust Filter Decorative Side Panels Stainless Steel Armrest Prop-up Work Tray Stainless Steel Window Edge Protector NSF Approved Cord Pass Through (Right and/or Left Sidewall) Smooth Interior Plumbing Options:
- Soft Copper
- Black Pipe
- Rigid Copper Outlet Options:
- Two (2) GFI Duplex Outlets Back Wall (3
Amp Max)
- Two (2) GFI Duplex Outlets Back Wall (5
Amp Max)
- Two (2) Duplex Outlets Back Wall with
Dedicated Power Cord (15 Amp Max)
- Two (2) GFI Duplex Outlets Back Wall with
Dedicated Power Cord (15 Amp Max)
Two (2) Duplex Outlets Back Wall with Dedicated Power Cord (20 Amp Max)
- Two (2) GFI Duplex Outlets Back Wall with Dedicated Power Cord (20 Amp Max)
- Front Faring Outlets Seismic Brackets Base Stand
- Telescoping Base Stand
- Telescoping OSHPD OPM Rated*
- Telescoping Base Storage
- Motorized Base Stand Options
- Leg Levelers
- Casters
- 8” (203 mm) Shelf**
- 12” (305 mm) Shelf**
- Ergotron Monitor & Keyboard Mount *Leg Levelers Only **Telescoping Base Stand Only Custom Solutions
NuAire is a sponsor of the IFBA
NU-543 CLASS II TYPE A2 BIOSAFETY CABINET
PROCESS #9-1219 Rev 4 September/2017 Best Products. | Best Performance. | Best Protection. Page 1 of 4
NuAire, Inc. | 2100 Fernbrook Lane | Plymouth, MN 55447 | U.S.A | ph: 763.553.1270 | fx: 763.553.0459 | tf: 800.328.3352 | www.nuaire.com
PURCHASE SPECIFICATIONS: NUAIRE LabGard® ES ENERGY SAVER NU-543 BIOSAFETY CABINET
The intent herein is to provide a concise statement of requirements for a quality Class II Type A2 Laminar Airflow Biosafety Cabinet which may be used to augment your purchase request/order.
The LabGard® ES NU-543 meets the performance requirements of the NSF/ANSI 49. Your confidence is well placed in a Biosafety Cabinet that meets the NSF standard.
NuAire sales representatives will be pleased to explain the importance of the performance and control affected by each of the following requirements. The NuAire LabGard® ES NU-543 meets all of the requirements in the following SPECIFICATION.
1. Dimensions Inches (mm)
Overall Dimensions NU-543-300 NU-543-400 NU-543-500 NU-543-600 Width (W) 41 5/8 (1057) 53 5/8 (1362) 65 5/8 (1669) 77 5/8 (1972) Depth (D) (Armrest Removed) (Incl. Control Center)
31 7/16 (799) 31 7/16 (799) 31 7/16 (799) 31 7/16 (799)
Height (H) (Incl. Exhaust Grill)
60 7/8 (1546) 60 7/8 (1546) 60 7/8 (1546) 60 7/8 (1546)
Basestand, 30" W.S. 88 3/8 (2245) 88 3/8 (2245) 88 3/8 (2245) 88 3/8 (2245) Basestand, 36" W.S. 94 3/8 (2397) 94 3/8 (2397) 94 3/8 (2397) 94 3/8 (2397)
Interior Dimensions Width (W) 34 3/8 (873) 46 3/8 (1178) 58 3/8 (1483) 70 3/8 (1788) Depth (D)+ 25 3/4 (654) 25 3/4 (654) 25 3/4 (654) 25 3/4 (654) Height (H) 28 1/2 (724) 28 1/2 (724) 28 1/2 (724) 28 1/2 (724)
+ Measured at 10 inch (254) window height.
2. Cabinet shall provide airflows & Biosafety performance as specified.
a. Cabinet shall provide biological containment protection for both operator and product proven by an actual test, (e.g. test conducted by NSF) and routinely validated by NuAire.
* b. Cabinet shall be constructed from 16/18 gauge, Type 304 stainless steel forming a monolithic, sealed structure.
c. Cabinet shall be easily fumigated employing an established procedure such as that recommended by NIH or NSF.
d. Supply HEPA filter 99.99% efficient @ 0.3 microns shall be of full cabinet work zone width and depth.
* e. Supply HEPA filter shall be protected by a perforated metal diffuser covering the entire top of the work zone.
* f. Air Velocity from the supply filter shall average 55 to 65 FPM (.28 to .32 m/s) with no single point outside the 20% of average range measured in a horizontal plane defined by 4 inches (102mm) above the bottom edge of window.
*^ g. Work access opening shall be 10 inches (254 mm) high standard, optionally (8 inches (203 mm) or 12 inches (305mm). Average Inflow velocity shall nominally be 105 LFPM (.53m/s).
h. High capacity 11-1/2 inch (292mm) deep exhaust HEPA filter 99.99% efficient @ 0.3 microns
*Having all of these features is unique ONLY to NuAire cabinets.
^NU-543-300 work access opening shall be 10 inches (254mm) high standard, optionally 8 inches (203mm).
PROCESS #9-1219 Rev 4 September/2017 Best Products. | Best Performance. | Best Protection. Page 2 of 4
NuAire, Inc. | 2100 Fernbrook Lane | Plymouth, MN 55447 | U.S.A | ph: 763.553.1270 | fx: 763.553.0459 | tf: 800.328.3352 | www.nuaire.com
3.* The cabinet shall be ergonomically designed for maximum user comfort and adjustability to meet the requirements of the American Disabilities Act (ADA).
Standard non-metallic armrest/airfoil incorporating large 2 inch (51mm) forearm support area 1/2 inch (12mm) recessed front grill designed for armrest comfort while maintaining containment performance.
Maximum visibility into cabinet workzone shall be at least 23-3/4 inches (603mm) from front access airfoil to exterior light housing.
Cabinet shall have a centrally located instrument panel within the control center that is easily serviced with quick disconnects.
Cabinet shall have the capability of incorporating a user adjustable basestand or base storage cabinet as an option.
The cabinet shall have a smooth operating sliding window from full closure to full opening at 21 inches (533mm).
Cabinet shall have a large worktray (20-3/4 inch (527mm) depth) removable with coved corners for easy cleaning.
Cabinet shall have a 10 degree slope.
4.* The cabinet shall have all positive pressure plenums surrounded by a vacuum relative to the room (the LabGard® ES employs the HEPEX™ Zero Leak Airflow System).
5. Electrical power shall be supplied with a 12-foot (2.5m), 3-wire cord with molded plug. Electrical supply should be 115 VAC, 60 Hz (Current rating varies per cabinet size. Reference electrical requirements page 4) protected with thermal circuit breaker from distribution panel.
6. The cabinet shall use a DC ECM Motor with an optimally determined forward-curved fan for each model size/width to maximize both energy efficiency and filter loading capacity.
7. The cabinet shall have two internal electrical circuits; one for blower/lights and one for the duplex outlets. Each circuit shall be protected with a fuse located in the Control Center on the electronic module.
8.* The cabinet shall be listed by Underwriters Laboratories to meet the requirements of both the U.S. and Canada for electrical/mechanical integrity.
9.* Cabinet shall contain a FlowGard™ control system consisting of electronic modules that will perform the following functions:
Easy user interface via OLED (Organic Light Emitting Diode) display/function keys.
Language selectable user interface menu’s (English, Spanish, German, French, Chinese, Japanese) Control blower via solid state switch.
Control lights via solid state switch.
Control outlets via solid state switch.
Disable audible alarm switch with ring back function.
Control blower DC ECM motor with solid-state DC Motor Controller that provides automatic compensation (constant volume control) for both filter loading and line voltage variances.
Monitor and display airflow velocity performance via IntelliFlow™ monitor.
Display time of day and temperature.
10.* Cabinet shall contain the FlowGard™ control system that provides the following optional functional features (included with cabinet, but must be configured during certification):
Security password protection of cabinet use.
Night Care™ setback mode. Upon sliding window closure, blower will continue to operate at a lower rate to save energy and maintain interior clean air conditions ready for use upon sliding widow opening.
Cabinet usage displays filter life, blower hours and UV light hours.
Cabinet usage sync functions with blower, fluorescent light, outlets and accessory outlet.
Cabinet usage auto duration timers, Nite Care, fluorescent light, UV light and outlets.
Cabinet decontamination sequence control system
11. Balancing of cabinet workzone downflow (recycling flow) to exhaust flow shall be accomplished with an internal exhaust flow damper, externally adjustable with screwdriver
12. The cabinet shall be easily transportable through a standard 36 inch (914mm) wide door without disassembly.
PROCESS #9-1219 Rev 4 September/2017 Best Products. | Best Performance. | Best Protection. Page 3 of 4
NuAire, Inc. | 2100 Fernbrook Lane | Plymouth, MN 55447 | U.S.A | ph: 763.553.1270 | fx: 763.553.0459 | tf: 800.328.3352 | www.nuaire.com
13. Sound level shall be no more than 60 dbA measured 15 inches (381mm) above the work tray and 12 inches (305mm) in front of viewing window.
14. Fluorescent lighting shall be externally mounted and provide 90 (968) to 120 (1291) foot-candles (LUX) on work surface.
The ballast is to be electronic containing thermal protection with automatic reset.
15.* Cabinet shall come standard with two duplex outlets with drip proof covers on back wall; one gas valve/service coupling on right side wall; one service coupling on right side wall; two service couplings on left sidewall.
16. Cabinet shall be easily converted to a free-standing console model with the addition of the optional Base Support Stand.
17.* Cabinet work zone shall be all welded 16/18 gauge stainless steel (silicone free) and reinforced with stainless steel U channels to minimize vibration.
18. A 3/8 inch (10 mm) inch ball valve shall be provided in the drain trough beneath the work tray.
19.* Cabinet shall have a permanent positive pressure plenum with quick release supply filter removal.
20.* Motor/blower shall be positioned so as to create an even filter loading, thereby prolonging the life of HEPA filters, automatically handling a 250% minimum increase in filter loading without reducing total air delivery by more than 10%.
21.* Cabinet shall be capable of front filter removal without disassembly of the control panel and sliding window tracks/hardware.
22. The following optional equipment shall be available to support installation and user requirements:
8 inch (203mm) Access Opening @ 105fpm (.53 m/s) Inflow
12 inch (305mm) Access Opening @ 105fpm (.53mls) (-400, -500, -600)
Ultraviolet Light
LED Lighting
Additional Service Valves for Gas, Air, Vacuum
Additional Duplex Outlet
IV Bar with 6 Stainless Steel Hooks
Exhaust Transition Canopy
Base Support Stand (available as telescoping for work surface heights of 30 to 36 inches) (762 or 914mm) with or without Storage Shelves
Adjustable Control for Support Stand or Storage Cabinet
Sinks with Hot/Cold or DI Water Faucets
Storage Pull-Out Trays
Lay in Sorbent Exhaust Filter
Decorative Side Panels (hides plumbing fixture connections)
HEPA Filters 99.999% @ 0.3 Micron
Arm Rest (Stainless Steel)
GFI Outlets
*Having all of these features is unique ONLY to NuAire cabinets.
PROCESS #9-1219 Rev 4 September/2017 Best Products. | Best Performance. | Best Protection. Page 4 of 4
NuAire, Inc. | 2100 Fernbrook Lane | Plymouth, MN 55447 | U.S.A | ph: 763.553.1270 | fx: 763.553.0459 | tf: 800.328.3352 | www.nuaire.com
LabGard® ES Energy Saver Class II Type, A2 Laminar Flow Biosafety Cabinet Models NU-543-300/400/500/600
Catalog Number
Catalog Number NU-543-300 Nominal 3 foot (0.9m)
NU-543-400
Nominal 4 foot (1.2m)
NU-543-500
Nominal 5 foot (1.5m)
NU-543-600
Nominal 6 foot (1.8m)
Performance Specifications
1. Personal Protection
2. Product Protection
NSF/ANSI 49
NSF/ANSI 49 Class II, Type A2 Class II, Type A2 Class II, Type A2 Class II, Type A2
Style of Cabinet Bench top/console w/base stand/storage cabinet
Bench top/console w/base stand/storage cabinet
Bench top/console w/base stand/storage cabinet
Bench top/console w/base stand/storage cabinet
Cabinet Construction All welded stainless steel 16/18 gauge, Type 304 pressure tight design
All welded stainless steel 16/18 gauge, Type 304 pressure tight design
All welded stainless steel 16/18 gauge, Type 304 pressure tight design
All welded stainless steel 16/18 gauge, Type 304 pressure tight design
Diffuser for Air Supply (Metal) Non-flammable Non-flammable Non-flammable Non-flammable
HEPA Filter Seal Type:
Supply Filter-99.99% Eff. on 0.3 microns Exhaust Filter-99.99% Eff. on 0.3 microns
HEPEX Seal Neoprene, Spring-loaded
Fumigation : per NIH/NSF Procedures Yes Yes Yes Yes
Standard Services:
Service Coupling (3/8 inch NPT) Service Coupling (3/8 inch NPT) Gas Valve/Service Coupling (3/8inch NPT) Duplex Outlet
One Right Sidewall Two Left Sidewall
Two Backwall
One Right Sidewall Two Backwall
Optional Services:
Gas Cocks 3/8" NPT Ultraviolet Light Standard/Cup Sinks
Up to 3 ea. Sidewall One, Backwall Left or Right Work Surface
Left or Right Work Surface
Cabinet Size Inches (mm):
Height (Fully Assembled) Height (Minimum for Transport) Width Depth with Armrest Removed
60 7/8 (1546) 59 (1499) 41 5/8 (1057) 31 7/16 (799)
53 5/8 (1362)
65 5/8 (1669)
77 5/8 (1972) 31 7/16 (799)
Work Access Opening Inches (mm):
Standard Opening Height/Optional Standard Inflow Velocity
10 (254)/8(203) 105 FPM (.53 m/s)
10 (254)/8(203)/12(305)
Work Zone Inches (mm):
Height Width Depth measured at 10 inches (254mm) window height
28 1/2 (724) 34 3/8 (873) 25 3/4 (654)
46 3/8 (1178)
46 3/8 (1178)
70 3/8 (1788) 25 3/4 (654)
Viewing Window Inches (mm):
Standard is safety plate sliding glass
Fully closed to 21(533) open
Fully closed to 21(533) open
Fully closed to 21(533) open
Fully closed to 21(533) open
Required Exhaust CFM/CMH (opening) Standard/Optional:
Canopy Variable Flow Thimble (NU-911)
Canopy Fixed Flow Thimble (NU-907)
10(254)/8(203)/12 (305)
CFM (CMH)
276-501 (649-851) 255-450 (383-765) 341-554 (576-933) 320 (544) 269 (457) 387 (650)
10(254) /8(203) /12(305)
CFM (CMH)
363-588 (617-1000) 295-520 (502-884) 430-655 (731-1113) 426 (724) 359 (610) 483 (821)
10(254) /8(203) /12(305)
CFM (CMH)
451-676 (766-1149) 365-590 (621-1003) 531-760 (909-1291) 531 (902) 445 (756) 610 (1036)
10(254) /8(203) /12(305)
CFM (CMH)
538-763 (915-1297) 436-661 (741-1124) 634-865 (1088-1470) 634 (1077) 532 (904) 763 (1251)
Plant Duct Static Pressure Eng./Metric 0.05-0.1"/1.27-2.54mm H2O
0.05-0.1"/1.27-2.54mm H2O
0.05-0.1"/1.27-2.54mm H2O
0.05-0.1"/1.27-2.54mm H2O
Heat Rejected, BTU, Per Hour (opening) (non-vented) (vented)
10(254) /8(208) 903 / 826
10(254) /8(208) /12(305) 1140 / 1020 / 1256
10(254) /8(208) /12(305) 1768 / 1611 / 1884
10(254) /8(208) /12(305) 1884 / 1768 / 2041
Electrical:
Volts, AC 60 Hz +Amps: Blower/Lights (10/8/12 openings) Amps: Duplex Rated Amps:
12 ft. Power Cord (one)
U.L./U.L.-C Listed 2.3/2.1 14 gauge - 3 Wire, 15A
U.L./U.L.-C Listed 2.9/2.6/3.2 12 gauge - 3 Wire, 20A
U.L./U.L.-C Listed 4.5/4.1/4.8 12 gauge - 3 Wire, 20A
U.L./U.L.-C Listed 4.8/4.5/5.2 12 gauge - 3 Wire, 20A
Crated Shipping Weight:*** Net Weight
460 lbs. /209 kg.
410 lbs. /186 kg.
530 lbs. /240 kg.
480 lbs. /218 kg.
620 lbs. /281 kg.
570 lbs. /258 kg.
690 lbs. /313 kg.
640 lbs. /290 kg.
***Crated shipping weight does not include weight for accessories or options + Based on cabinet with new filters running at 115VAC.
TITLE
CHKDDATEDFTM
TM
SHEET
OF
DRAWING
NUMBER
NU-543-400 Class II, Type A2 Biosafety Cabinet
1 1
BCD-16373 D
10/15/2013LS BP
ROOM AIR
21[563] X 44[1118] X 3[76]
HEPA SUPPLY FILTER
ACCESSORY OUTLET
20 AMP
CABINET POWER CORD
(NEMA 5-20P)
SLIDING WINDOW
REV CO DESCRIPTION DATE DFTM CHKD
D 000098
UPDATED, ADDED CORD
PASS-THRU
8/30/2019 TH DSH
INCH
MILLIMETER
OVERALL DIMENSIONAL TOLERANCE ±1/4 [6.35]
ALL OTHER DIMESIONS ±1/8 [3.17]*
3/4 [19] ADJUSTMENT
DUPLEX OUTLET (2)
OPTIONAL
BASE STAND
(TELESCOPING)
PROTECTIVE
GRILL
CONTROL PANEL
OLED DISPLAY
FLOWGARD CONTROL SYSTEM
INTELLIFLOW MONITOR
LIGHT SWITCH
BLOWER SWITCH
OUTLET SWITCH
UV SWITCH (W/ OPTION)
SYSTEM DIAGNOSTICS
INTERNAL BALANCING
DAMPER ADJUSTMENT
46 3/8 [1178]
WORKZONE
WIDTH23 7/8
[608]
WORKZONE
VIEW HEIGHT
10 [254] STANDARD
8 [203] OPTIONAL
NSF APPROVED
LEG LEVELERS
53 1/2 [1358]
3 EQ SPCS
@ 6.792 [173] = 20 3/8 [518]
1 1/8 [28]
2 3/8 [60]
4 EQ SPCS
@ 5.594 [142] = 22 3/8 [568]
31 3/8 [797]
32 1/8 [817]
58 7/8 [1495]
[51]
30 3/8 [771]
25 3/4 [653]
MEDIAN
WORKZONE
DEPTH 28 1/4
[716]
WORKZONE
HEIGHT
OPTIONAL
IV BAR
DC ECM
MOTOR/BLOWER
PAPER CATCH
REMOVABLE PVC
ARMREST
REMOVABLE SST
WORKTRAY
FLUORESCENT LIGHT
*26 1/2 [673]
TO
*32 1/2 [825]
IN 2"
INCREMENTS
1/2 [13]
13 7/16 [342]
16 7/16 [418]
11 13/16 [301]
14 13/16 [377]
FLUSH MOUNT
FRONT PANEL
CORD PASS-THRU
DRAIN VALVE
SERVICE VALVE
3/8" FEM NPT
PLUGGED SERVICE
COUPLINGS (1) RS, (2) LS
** ADD 1/2" [13] TO OVERALL HEIGHT
WITH PROTECTIVE GRILL REMOVED
30 3/4 [781]
OPTIONAL
UV LIGHT
PERMANENT
PLENUM
20[508] X 22[559] X 11 1/2[292]
HEPA EXHAUST FILTER