Cryostat Isolation System
Purpose: To vibrational isolate a cryostat hanging inside a granite table and pit in a NIST A1 inertial slab for scanning probe measurements.
Critical dimensions.
1. Max OD of isolation system shall be 38.2” or less to fit inside existing pit ~ 40” ID in NIST A1 inertia slab.
2. Cryostat OD is 32.0”. Dewar height, bottom flange to top flange is 64.6”.
3. Top of granite table to bottom of the pit is 78.1”.
4. Top of cryostat is located at 2.6” below top of granite table in down position.
Cryostat Isolation System
Existing granite table NIST A1 inertia slab inside lab
Cryostat isolation system hangs cryostat inside pit and table in A1 slab
78.1”
Cryostat
40” ID