Salient Characteristics.doc
DOC document 96 KB Posted
- Solicitation number
- W911QX23Q0155
About this file
This document outlines the technical requirements for a Vector Network Analyzer (VNA) to be procured by the Army Research Laboratory through solicitation number W911QX23Q0155. The VNA must have four ports, operate between 10 MHz and 40 GHz with a dynamic range of at least 95 dB across the full frequency range. Additional requirements include trace noise levels, temperature stability, output power levels and linearity, third harmonic levels, and noise floor. The VNA must integrate with existing Lakeshore equipment and include two years of warranty, training, and on-site support.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Salient Characteristics - Amendment P0001.doc | DOC document | |
| Combination Synopsis-Solicitation for Commercial Purchases - Amendment P0001.docx | DOCX document | |
| Provisions and Clauses Full-Text Attachment.doc | DOC document | |
| Combination Synopsis-Solicitation for Commercial Purchases.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics
Commercial Availability Certification:
I have evaluated the commercial marketplace and determined that all of the salient characteristics documented below are available in products/services offered in the commercial marketplace.
Signature (typed name): Owen Vail Technical Point of Contact (TPOC) or Contracting Officer’s Representative (COR)
Technical Requirements for Vector Network Analyzer (VNA) The Advanced Electronics Branch at the Army Research Laboratory (ARL) in service of the Gordian TOPS program requires one (1) Vector Network Analyzer (VNA) with the following minimum characteristics:
· Four (4) ports
· Maximum operating frequency of at least 40 gigahertz (GHz)
· Minimum operation frequency of at most ten (10) megahertz (MHz)
· Dynamic Range of at least 95 decibels (dB) over the entire frequency range stated above
· Trace noise at zero (0) decibels-milliwatt (dBm) source power and zero (0) dB reflection of less than a hundredth (0.01) dB root mean square (RMS) at 10 kHz Intermediate Frequency Band Width (IFBW) for 40 GHz
· Trace noise at zero (0) dBm source power and zero (0) dB reflection of less than four thousandths (0.004) dB RMS at ten (10) kHz IFBW for ten (10) MHz
· Temperature instability of less than twenty-five thousandths (0.025) decibels per Kelvin (dB/K) at negative (-) ten (10) dBm source power over the entire frequency range stated
· Maximum output power of at least positive (+) two (2) dBm at 40 GHz, with typical operation of at least positive (+) ten (10) dBm over the frequency range between ten (10) MHz and ten (10) GHz
· Minimum output power of at most negative (-) 30 dBm
· Power level inaccuracy of at most two and a half (2.5) dB over the entire frequency range stated
· Power linearity of less than two (2) dB referenced to negative (-) ten (10) dBm over the entire frequency range stated
· Third harmonics at zero (0) dBm of at most negative (-) 17 dBc over the entire frequency range stated
· Typical noise floor of at most negative (-) 115 dBm over the entire frequency range stated
· Future (or included) options for pulse modulation.
· On-site support for upgrades and software updates, as well as email and phone access to applications experts on an as-needed basis for at least two (2) years
· Instrument fundamentals and operations starter shall be included
· Shall integrate with 2.4 millimeter (mm) SubMiniature version A (SMA) connections of an existing Government-owned Lakeshore vertical field probe station (CRX-VF) located at Adelphi Laboratory Center
· Intuitive user interface core software shall be provided with future options for time domain analysis and harmonic analysis
· Shall include at least two (2) years of warranty
· Shall include at least one (1) day of in-person training for standard users and functionalities of the software, as well as initial setup and standards testing with ground-signal-ground (GSG) probes.
Contracting Officer: