S02 - Attach 12 Narrative Set 1 of 3.pdf
PDF 5 MB Posted
- Attached to
- Replace Electrical Panel Board in OR Federal contract opportunity
- Solicitation number
- 36C24823R0073
About this file
This solicitation seeks replacement of electrical panel boards in operating rooms at the Miami VA Medical Center. The Department of Veterans Affairs, through VISN 8, requires replacing existing ground panel boards with Isolated Power Systems or Isolated Power Panels in 10 operating rooms to comply with current VA criteria. The contractor shall analyze the complete design package and perform all construction activities such as installation, replacement, wiring, control, calibration, testing, and maintenance of operations during construction according to the final design, narrative and specifications from the Architect Engineer firm. A site visit is scheduled for January 10, 2023 and RFIs are due January 17, 2023 and price proposals are due February 3, 2023 at 1:00PM. The agency is the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 8.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Amendment 0001 - 36C24823R0073.docx | DOCX document | |
| S02 - Attach 11 Set of Drawings 4 of 4.pdf | ||
| S02 - Attach 13 Narrative Set 2 of 3.pdf | ||
| S02 - Attach 5 Past Performance Form.pdf | ||
| S02 - Attach 9 Set of Drawings 2 of 4.pdf | ||
| S02 - SOLICITATION 36C24823R0073.pdf | ||
| S02 - Attach 7 Price Matrix.xlsx | XLSX spreadsheet | |
| S02 - Attach 14 Narrative Set 3 of 3.pdf | ||
| S02 - Attach 1 Site Plan.pdf | ||
| S02 - Attach 2 _ Wage Dtermination FL20220125 _9_29_22.pdf | ||
| S02 - Attach 8 Set of Drawings 1 of 4.pdf | ||
| S02 - Attach 10 Set of Drawings 3 of 4.pdf | ||
| S02 - Attach 15 Floor Plan.pdf | ||
| S02 - Attach 3 Limitations of Subcontracting .pdf | ||
| S02 - Attach 6 Experience Information Sheet.docx | DOCX document | |
| S02 - Attach 4 - EMR.docx | DOCX document |
Show all 16
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Isolated Power Planning Guide System Options for Isolated Power in Healthcare Facilities
32 3
Standards such as NFPA 99 and CSA Z32 require isolated power systems in all areas deemed “wet procedure locations” in healthcare facilities.
allows for critical systems to remain online in a latest standards and code requirements.
Bender provides advanced electrical safety technology to the healthcare industry, including:
to help reduce downtime and increase operational the system remains online customization and installation
PROTECT PATIENTS AND STAFF
With isolated power solutions from the leader in electrical safety.
SERVICES FOR ISOLATED POWER SYSTEMS
BENDER Services offer a wide range of benefits to ensure optimal performance of your Isolated Power Systems including:
Complete
System Centers for
The Joint documentation of Isolated Power Systems equipment factory trained and authorized technicians calculations of system impedence surfaces torques
Testing & Evaluation of your Isolated Power Systems includes:
BENDER’s Isolated Power System (IPS) Panel
BENDER’s LIM2010 Line Isolation Monitor
BENDER’s LT3000, An industry-leading testing instrument
4 5
Basic Configuration Includes required equipment for NFPA 99 compliance
Load Monitoring Includes capability of monitoring transformer load
Standard isolated power panel
System notes:
Minimum equipment required for compliance with NFPA 99 when panel is not installed
Panel configuration:
Sample part numbers:
Additional equipment:
d d i l t P
L1 L2
MK2000P-G1 remote indicator
Key system components:
Isolation transformer
MK2000CBM digital remote indicator
STW3 current transformer
Transformer load monitoring
L1 L2
Key additional components:System notes:
remote former load
Panel configuration:
Sample part numbers:
Additional equipment:
MK2000CBM remote indicator
6 7
Branch ground fault location Includes automatic indication of individual branches with ground faults
Branch load monitoring Adds load monitoring for individual branch loads
Fault location expansion kit
System notes:
Modular current transformer sizes allow for simple installation panel module systems / panels other rooms and central nurses'
Addition of ground fault location and digital remote station
MK2430 remote indicating station
Branch current monitoring expansion kit
Addition of branch load current monitoring
Key additional components:
System notes:
Automatically detects and with other equipment
ON
L1 L2
Key system components:
Included in panel beyond standard:
hardware
Sample part numbers:
Additional equipment:
ON
L1 L2
Included in panel beyond standard:
modules
Sample part numbers:
Additional equipment:
8 9
Central monitoring station Provides a central point for monitoring system alarms
Technician visualization & remote integration Includes advanced technical visualizations & communication
MK800RS (as shown) remote station
Addition of central monitoring station at nurses' desk
Communication expansion kit
Addition of technician monitoring station for all systems
Key additional components:
System notes:
from a centralized point
Key system components:
Included in panel beyond standard:
and hardware
Sample part numbers:
ON
L1 L2
System notes:
of systems point
CP907 HMI Station
ON
L1 L2
10 11
Complete systems for multiple rooms A complete solution for multiple operating rooms, centrally monitored
Sample part numbers:
Other equipment:
System notes:
personnel for all systems
L1 L2 L2 L1 L2
Modbus/TCP network
ON ON
ON
OR 1 OR 2 ICU 1 Nurse Desk Central Maintenance
14.3H.REV.0 | 11.2019 | © Bender Inc. All Rights Reserved.
Bender is located in 70 countries around the world
USA • Exton, PA +1 800.356.4266 • info@benderinc.com www.benderinc.com
México • Ciudad de México +55 7916.2799 • info@bender.com.mx www.bender.com.mx
Canada • Missisauga, ON +1 800.243.2438 • info@bender-ca.com www.bender-ca.com
Chile • Santiago de Chile +56 2.2933.4211 • info@bender-cl.com www.bender-cl.com
South America, Central America, Caribbean +1 484.288.7434 • info@bender-latinamerica.com www.bender-latinamerica.com
Spain • San Sebastián de los Reyes +34 9.1375.1202 • info@bender.es www.bender.es
Isolated Power Panels With LIM2010 Line Isolation Monitor
For Hospitals and Critical Care Areas
Technical Bulletin
NAE2092010 / 01.2012
Introduction
BENDER Isolated Power Panels are designed to provide isolated power to electrical systems in operating rooms and other critical areas. Designed in strict compliance with UL 1047, UL 1022, and UL 50, BENDER isolated power panels offer the most up-to-date technology for all isolated power distribution requirements.
Standard Features
Standard isolation power panels typically include:
Additional Features
Additional features available include:
Refer to the following pages for more information on these options.
Backbox
Front Trim surface-mounted panel boards, the front trim has the same dimensions as the enclosure.
Isolation Transformer
Isolation transformers are available with various primary and secondary single-phase volt-ages. See ordering information for available primary and secondary voltages.
Line Isolation Monitor of capacitive and resistive faults, including balanced, unbalanced, and hybrid faults. A self-test and self-calibration function is included.
BENDER EDS series module to create a ground fault location system. For more information, Loadcenter
Isolated Power Panels Isolated Power System with Line Isolation Monitor for hospitals and other critical care areas
Standard Isolated Power Panel
Features
Power Distribution: Loadcenter avail-able for either plug-in and bolt-on circuit
Mounting: Available for flush- or surface-mounted applications
Advanced technology:
isolation monitor features self-testing, self-calibration, and a wide variety of alarms including voltage, overload, and overtemperature monitoring
Outlets: Optional arrangement of recep-
Controller: Optional combination of PLC and remote / local controls for up to 12
Standards: Listed to UL 1047, the stan-dard for isolated power system equip-ment
Warranty: Industry-first 5 year warranty
1 - Stainless steel front trim
2 -
3 -
4 -
5 -
6 -
7 -
8 -
9 -
10 -
11 - Loadcenter
12 - Isolation transformer, single-phase
13 -
14 - LIM connector plate
15 -
Sample Outline: Standard Isolated Power Panel
Wiring Diagram: Standard Isolated Power Panel
IP Panels
A
B
C
C
A
D E
F
TOP
FRONT SIDE
A
B
C
C
A
D
F
E
TOP
FRONT SIDE
1 - Stainless steel front trim
2 -
3 -
4 -
5 -
6 -
7 -
8 -
9 -
10 -
11 - Loadcenter
12 - Isolation transformer, single-phase
13 -
14 - LIM connector plate
15 -
16 -
17 -
18 - Receptacle hat station
Isolated Power Panels with Receptacles and Ground Jacks
BENDER isolated power panels may have provisions for hospital grade power receptacles
Sample Outline: Isolated Power Panel with Optional Receptacles and Ground Jacks
IP Panels
1 - Stainless steel front trim
2 -
3 -
4 -
5 -
6 -
7 -
8 -
9 -
10 -
11 -
12 - Isolation transformer, single-phase
13 -
14 - LIM connector plate
15 - Control transformer
16 -
17 -
18 -
19 - Circuit contactors, 2P
20 -
Isolated Power Panels for X-Ray / Laser Equipment
Programmable Logic Controller (PLC) configured to a particular system‘s requirements. Control wiring from push-buttons, door switches, etc. are permanently wired into the input section. Signals or contact closures from the output section may be used to actuate contactors and to apply power to the
BENDER type XRM receptacle modules are recommended for use with this panel
Sample Outline: Isolated Power Panel with Circuit Control
IP Panels
A
B
C
D
F
E
C
A
TOP
FRONT SIDE
A complete ground fault location system for hospitals
L1
L2
Power Supply Class 2
GF Location Module EDS Series
MK2000 Series
>˚F
Line Isolation Monitor
LIM2010
MK2430
The BENDER Advantage remains online displays at both the location device and remote indicators
System Functions retro-fitting/upgrading available in many different shapes and types of ungrounded systems
Complete system including LIM, EDS ground fault location, and remote indication
Advantages:
location device and remote indicators
System functions:
Function a complete ground fault location system. After an alarm is device generates a test signal for a set period of time. Its the location of the ground fault. Current transformers placed of the fault is displayed on the EDS series via either a digital dis-play or an LED bar graph. Special remote indicators connected to the system may also display the location of the fault.
1 -w/ Vibration Mounts
2 -
3 -
4 -
5 -
6 - Sub-feed Lug
7 -
8 -
9 - Load Center
10 - Connector Plate, LIM
11 -
12 -
13 -
14 - EDS Units
15 -
Sample Outline: Panel w/ Load Monitoring and Integrated Fault Location System
IP Panels w/ Integrated Fault Location
Ground Fault Location Module an installed ground fault location system for ungrounded AC and DC systems. Once a ground fault is detected, the test pulse generated by the LIM2010 is scanned by the transformers may be connected to the device. A total of 90 EDS devices may be intercon-isolated power system can be identified within seconds.
EDS461-D
Module Types
Type Display Type Outputs EDS461-D LCD display 1 DPDT contact EDS461-L LED indication 1 DPDT contact
Ordering Information
BENDER’s complete Isolated Power Panels are comprise of four assembly types:
NOTE -
1 - Panel Interior kVA rating
2 - Panel Interior Primary voltage rating
3 - Panel Interior Secondary voltage rating
4 - Loadcenter/Panelboard Manufacturer
SB: Square D, Bolt-On
CB: Cutler-Hammer, Bolt-On
Option 1A:
5 - Quantity of ground jacks (0-8)
6 - Type of receptacle
7 - Quantity of receptacles
Option 1B:
8 - Type of control and indication individual receptacle modules with in-use lamps station located remotely
9 - Amperage per circuit*
10 - TOTAL quantity of controlled circuits (01-12)
11 - Quantity of SIMULTANEOUSLY Active circuits*
H1: One H4: Four
I P
1 5 A- A 2 3
S P 4 5
S R 6 7
P N- A 2
Option 1A Option 1B
0 5 H 2
STEP 1: Interior
Base Information (required):
Select EITHER Option 1A,1B OR 1C (leave blank if not required):
Note*: For Option 1B, the amperage per circuit and number of simultaneously hot circuits must be rated properly in accordance with the secondary vol-
IP Panels
LM-
10 11 12 13
OPTION 1: Interior Options
Option 1C Option 2
Option 1C:
12 - Fault Location System (EDS) Type
X M
0 5 B A 2 3
Backbox / Front Trim Combinations for IP Panels kVA Backbox Size (H x W x D)
Front Trim Size (H x W) Mounting Supports
Option 1A Supports Option 1B
Supports
EDS
Backbox Part Number
Front Trim Part Number
3, 5 43" x 24" x 6" 45" x 26" Flush No No No B432406 T4526
7.5, 10 43" x 24" x 8" 45" x 26" Flush No No No B432408 T4526
15 51" x 30" x 12" 53" x 32" Flush No Yes No B513012 T5332
25 51" x 30" x 14" 53" x 32" Flush No Yes No B513014 T5332
3, 5, 7.5, 10 48" x 24" x 8" 50" x 26" Flush Yes No No B482408 T5026R
3, 5, 7.5, 10 48“ x 24“ x 8“ 50“ x 26“ Flush No No Yes B482408 T5026E
3, 5 43" x 24" x 6" 43" x 24" Surface No No No B432406S T4324
7.5, 10 43" x 24" x 8" 43" x 24" Surface No No No B432408S T4324
15 51" x 30" x 12" 51" x 30" Surface No Yes No B513012S T5130
25 51" x 30" x 14" 51" x 30" Surface No Yes No B513014S T5130
3, 5, 7.5, 10 48" x 24" x 8" 48" x 24" Surface Yes No No B482408S T4824R
3, 5, 7.5, 10 48“ x 24“ x 8“ 50“ x 26“ Surface No No Yes B482408S T4824E
STEP 2: Transformer Contact your local BENDER sales office for additional or custom configurations.
2 - Primary voltage of transformer
3 - Secondary voltage of transformer1 - kVA of transformer
OPTION 2: Additional Monitoring Features
13 - Transformer Load Monitoring
[ ]: Feature not required
*Note:
IP Panels
STEPS 3 & 4: Backbox and Front Trim
Dual Isolated Power Panels Dual Voltage and Dual System Isolated Power Panels for hospitals and other critical care areas
Dual Output Voltage Isolated Power Panel
Introduction
BENDER’s dual voltage isolated power panels provide two different output voltages
Ideally suited for:
receptacles
Standard Features: Dual Output Voltage Panel
Dual output voltage isolated power panels provide two separate voltage outputs using a single isolation transformer. A standard dual output voltage panel consists of the following:
Standard Features: Dual System Panel
Dual system panels provide two separate voltage outputs from two isolation transformers.
enclosure. A standard dual system panel consists of the following:
Features
Power Distribution: Loadcenters available for either plug-in or bolt-on cir-
Mounting: Available for flush- or surface-mounted applications
Advanced technology:
isolation monitor features self-testing, self-calibration, and a wide variety of alarms including voltage, overload, and overtemperature monitoring
Standards: Listed to UL 1047, the standard for isolated power system equipment
Warranty: Industry-first 5 year warranty
Dual System Isolated Power Panel
File details come from the government source that posted it. Updated .