This document outlines requirements for a high volume sequencing system procurement. The Army Contracting Command at Aberdeen Proving Ground seeks to acquire one NextSeq 2000 sequencing system from Illumina, along with a silver support plan. The system must be capable of producing up to 330 megabases of data from a 48-hour paired-end run with 150 base pair reads. It must fit on a standard laboratory benchtop and include reagent kits for various sequencing applications including whole genome, amplicon, RNA, and metagenomics. The platform should have a track record of producing quality publications for organisms with read lengths at or below 150 base pairs. Responses to solicitation number W911QX-24-Q-0047 are due from offerors by the date specified.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Synopsis-RFI (Sole Source using 6.302-1) (HV NGS).doc | DOC document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Requirements Document 2 January 2024 Commercial Availability Certification:
I have evaluated the commercial marketplace and determined that all of the salient characteristics documented below are available in products/services offered in the commercial marketplace.
Signature (typed name): Randi Pullen Technical Point of Contract (TPOC) The Contractor shall provide the following llumina NextSeq2000:
| CLIN Number |
| Quantity |
| Description |
| 0001 |
| One (1) |
| NextSeq 2000 Sequencing System |
| 0002 |
| One (1) |
| NextSeq 2000 Silver Support Plan |
Government’s minimum technical requirements:
· Produce up to 330 Megabase (Mb) of data from a 48 hour 2 by 150 base pair (bp) run.
· The sequencer should fit on a standard laboratory benchtop.
· The manufacturer shall provide compatible reagent kits for flexible run formats (whole genome sequencing, amplicon sequencing, library sequencing, Ribonucleic Acid (RNA) sequencing, transcriptomics, metagenomics).
· The sequencer shall be capable of automated paired-end sequencing.
· The sequencing platform shall have an excellent track record of producing quality publications, especially concerning organisms with low (less than (<) 35 percent (%)) guanine and cytosine (GC) content.
· The sequencer shall have adjustable read length from 50 to 300 bp.
· The sequencer shall produce high quality data with Quotient factor scores (Q-scores) greater than or equal to (>=) 30 at two (2) by 150 bp read length.