Required Minimum Specifications.pdf
PDF 136 KB Posted
- Attached to
- Blood Chemistry Analyzer
- Solicitation number
- 12505B21Q0142
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Required Minimum Specifications Veterinary Blood Chemistry Analyzer
• The analyzer shall measure the following parameters: ALB, ALKP, ALT, AMYL, AST, BUN, Ca, CHOL, CK, Cl, CREA, CRP, FRU, GGT, GLU, K, LAC, LDH, LIPA, Mg, Na, NH3, PHBR, PHOS, SDMA, TBIL, TP, TRIG, TT4, UCRE, UPRO, URIC.
• The analyzer shall utilize dry slide technology (no need for rotors) that enables us to run individual or multiple parameters on individual samples.
• The analyzer shall measure the following species: Avian, bovine, caprine, equine, lagomorphs, murine, mustelids, ovine, and porcine.
• The analyzer shall process serum, plasma, whole blood, and urine.
• The analyzer shall perform processing at a speed of a minimum of 144 samples per hour timeframe and allow for storage and printing of results data. Flexibility of results reports shall include chronological order and per patient ID.
• The analyzer shall be a countertop unit not to exceed the dimensions of: 25.4 cm wide x
37.6 cm deep x 35.6 cm high.
• The analyzer shall operate standard 120-volt AC power at less than 20 amps.
• Warranty to be included with the unit shall be a minimum of 3 years with the option to purchase an extension of the warranty at the conclusion of the initial period. This shall include full replacement of original equipment manufacturer’s parts with labor performed by original equipment manufacturers technicians and a backup machine provided by the manufacturer to use while any warranty work is being conducted..
The quote shall include installation and training for USDA personnel of operation of the unit.
| The analyzer shall measure the following parameters: ALB, ALKP, ALT, AMYL, AST, BUN, Ca, CHOL, CK, Cl, CREA, CRP, FRU, GGT, GLU, K, LAC, LDH, LIPA, Mg, Na, NH3, PHBR, PHOS, SDMA, TBIL, TP, TRIG, TT4, UCRE, UPRO, URIC. |
| The analyzer shall utilize dry slide technology (no need for rotors) that enables us to run individual or multiple parameters on individual samples. |
| The analyzer shall measure the following species: Avian, bovine, caprine, equine, lagomorphs, murine, mustelids, ovine, and porcine. |
| The analyzer shall process serum, plasma, whole blood, and urine. |
| The analyzer shall perform processing at a speed of a minimum of 144 samples per hour timeframe and allow for storage and printing of results data. Flexibility of results reports shall include chronological order and per patient ID. |
| The analyzer shall be a countertop unit not to exceed the dimensions of: 25.4 cm wide x 37.6 cm deep x 35.6 cm high. |
| The analyzer shall operate standard 120-volt AC power at less than 20 amps. |
| Warranty to be included with the unit shall be a minimum of 3 years with the option to purchase an extension of the warranty at the conclusion of the initial period. This shall include full replacement of original equipment manufacturer’s parts wit... |
File details come from the government source that posted it. Updated .