HU0001-17-Q-0004
Pipette calibration BPA - AMENDMENT #1
This amendment #1 is provided to answer questions received from interested vendors. Please acknowledge this amendment #1 in your response. All other terms and conditions remain the same.
Questions:
• What are the brands and models of the pipettes?
Answer: Multiple Brands
• Is there a total of 300-400 pipettes at each location or is this the total of both USU and AFRRI?
Answer: There is not an actual count; however we may go over 400 pipettes easily between both locations
• Are As-found and As-calibrated data required OR Just As Found OR Just As Left?
Answer: Both As-found and As-calibrated data are required
• How many samples are to be recorded and Which Levels are recorded – Low, Mid and High OR Low and High?
Answer: Low, Mid, and High data should be recorded
• Are pipettes to be calibrated against ISO8655 specifications or Manufacturer’s specifications and test points (accuracy and repeat-ability standards)?
Answer: The pipettes should be calibrated using the most stringent specifications for the pipet, whether it be those set by the manufacturer or ISO8655.
• Is there room for two QCI technicians working together?
Answer: In our shop there will be enough room
• If the procurement process is complete before the end of January, 2017 – can we begin calibrations in January, 2017?
Answer: The exact dates of performance will be provided to the awardee by TSB (Technical Service Branch)
• If performing the work on-site, will the technicians have internet or WiFi access?
Answer: Yes
HU0001-17-Q-0004
Pipette calibration BPA - AMENDMENT #1
• Ratio of multichannel and single channel pipettes?
Answer: Can vary
• How many channels are in the multi-channel pipettes?
Answer: Different multi-channels pipettes