Question_and_Answers_2.13.2017.pdf

PDF 67 KB Posted

Attached to
Stationary X-Ray & Interlock System Federal contract opportunity
Solicitation number
FA8125-17-Q-0038
Issued by
Department of the Air Force Materiel Command Lifecycle Management Center Tinker Air Force Base

About this file

Questions & Answers 2/13/2017

View the file

Other files for this federal contract opportunity

Other files attached to Stationary X-Ray & Interlock System, newest first.
File Type Posted
FA812517Q00380002_Amendment_13Feb17.pdf PDF
FA812517Q0038_Solicitation.pdf PDF
SOW.pdf PDF
Attachment_D[1].pdf PDF
Attachment_E_Electrical[1].pdf PDF
Redacted_Brandname[1].pdf PDF
FA812517Q0038_Amendment.pdf PDF
Item_Description.pdf PDF
Appendix_C_Safety-Specification.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

FA8125-17-Q-0038

Question: "The specifications of the room alarm seem to be very undefined. Does this device have to comply with the US Air Force requirement for monitored beacon lamps? Does it have to be programmable? Does it have to have removable plugs at the enclosure?”

Answer:

They have to be of high intensity , flash or rotate to be IAW our AFIs and safety procedures.

The warning beacon lamp (amber or yellow) beacon lamps must at least be 20 seconds after all the doors are closed. The red beacon lamp must be on during the X-ray transmission signifying danger/caution x-ray are on. No, it doesn’t have to be programmable but it is highly desired to have it programmable delay to meet all our possible mission needs. As for removable plugs, no, it does not have to be removable/portable. This will be a stationary system. However, we want the capability to replace burnt out bulbs when they do occur (note the safety interlock system is design such that if a light bulb burns out a safety shut down will occur and x-ray system will not operate until the beacon light is replaced.

POC: Linsey Laird (405) 739-9117 linsey.laird@us.af.mil

February 13, 2017 Questions & Answers: Stationary X-Ray & Interlock System