Tech_Exhibit_1_SDS.pdf

PDF 594 KB Posted

Attached to
Air Force Primary Standards Laboratory Federal contract opportunity
Solicitation number
FA2263-20-R-0002
Issued by
Department of the Air Force Materiel Command Lifecycle Management Center Wright Patterson Air Force Base

About this file

This technical exhibit outlines key performance standards for a service contract providing calibration and maintenance services to the Air Force Primary Standards Laboratory (AFPSL). Key requirements include maintaining calibration accuracy for AFPSL equipment within thresholds of 1% major defects and 2% minor defects per month. Other standards cover areas such as turnaround times for services, not to exceed 3% defects per month; database maintenance with less than 3% defects; and inventory accuracy requirements of less than 2% defective items. The contractor must also meet reporting deadlines with no more than two defective reports submitted per month. Performance will be monitored through random sampling of workloads across several functional areas.

DRAFT Tech Exhibit 1

View the file

Other files for this federal contract opportunity

Other files attached to Air Force Primary Standards Laboratory, newest first.
File Type Posted
AFPSL_Recompete_PWS.docx DOCX document
Tech_Exhibit_2a_Workload.pdf PDF
Tech_Exhibit_2b_Turnaround_Time_Exceptions.pdf PDF
DRAFT_Solicitation_-_FA226320R0002.pdf PDF
AFPSL_Recompete_Section_M.docx DOCX document
Draft_Data_Requirements.pdf PDF
AFPSL_Recompete_QASP.pdf PDF
AFPSL_Recompete_Section_L.docx DOCX document
Tech_Exhibit_7__Accreditation_Areas.pdf PDF
Tech_Exhibit_5b_GFE_List.xlsx XLSX spreadsheet
Tech_Exhibit_5a_Facilities_Lease_Agreement.docx DOCX document
Tech_Exhibit_6_GFE_Usage_Fee.doc DOC document
Tech_Exhibit_4_Data_Requirements_Cross_Reference.docx DOCX document
Show all 13

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Technical Exhibit 1 (SDS).xls

ITEM # REQUIRED SERVICE

PWS para.

STANDARD

Performance Objective

PERFORMANCE REQUIRED

THRESHOLD

METHOD OF

SURVEILLANCE

APPLIES TO:

LAB T.O.

1.00 ACCURACY

1.01 ACCURACY OF CALIBRATIONS

1.4, 1.5, 5.1.7.3, 5.1.8.8, 5.1.8.8.1, 5.1.8.10, 5.1.8.11, 5.1.8.12, 5.1.8.13, 5.1.8.14, 5.1.8.15, 5.1.8.16, 5.1.8.17, 5.1.8.18, 5.1.8.19, 5.1.9.4, 5.1.9.5, 5.1.12

ENSURE ALL AFPSL CALIBRATIONS ARE

ACCURATE.

MAJOR DEFECTS: 1. THE ITEM IS UNSAFE OR

HAZARDOUS TO USE, OR 2. THE ITEM DOES NOT

MEET CALIBRATION ACCURACIES FOR ALL

FUNCTIONS CERTIFIED, 3. ALSO AS DEFINED IN

T.O. 00-20-14 AND PWS PARAGRAPH 2.1.19.

MINOR DEFECTS: ANY DEFECT THAT IS NOT

MAJOR. ALSO MAJOR OR MINOR DEFECTS AS

DEFINED IAW T.O. 00-20-14, AND THIS PWS

TO ALLOW CONTRACTOR TO

EXCEED THIS ELEMENT,

ONE MAJOR DEFECT IS

ALLOWED PER MONTH OF

QUALITY PRODUCTION.

TWO MINOR DEFECTS

ALLOWED PER MONTH OF

QULAITY PRODUCTION.

RANDOM

SAMPLING

LAB

1.02 ACCURACY OF T.O. DEVELOPMENT

5.1.8.15, 5.1.8.17.1, 5.2.1.1, 5.2.1.2, 5.2.3, 5.2.4

ALL T.O DEVELOPMENT, REVISIONS, AND

REWRITES ARE ACCURATE.

MAJOR DEFECTS: THOSE DEFECTS THAT DO

NOT ACHIEVE THE REQUIRED CALIBRATION

PARAMETERS OF THE ITEM TO BE CALIBRATED.

ALSO, AS DEFINED IN PWS PARAGRAPH 2.1.19.

MINOR DEFECTS: THOSE DEFECTS THAT DO NOT

AFFECT THE REQUIRED CALIBRATION

PARAMETERS.

NEW AND REWRITE 1 or 10%

(WHICHEVER IS GREATER)

REJECTS ALLOWED.

REVISIONS/CHANGES 1 or 2% of

SAMPLE (WHICHEVER IS

GREATER) REJECTS ALLOWED.

REJECT = T.O. WITH ONE

MAJOR OR MORE THAN TWO

MINOR DEFECTS.

LOT SIZE IS MONTHLY

REVIEWED

NEW & REWRITE:

ONE HUNDRED

PERCENT

INSPECTION

REVISION/CHANG

E: ONE HUNDRED

PERCENT

INSPECTION,

RANDOM

SAMPLING AFTER

6 MONTHS

ACCEPTABLE

PERFORMANCE

T.O.

Service Delivery Summary (SDS) Technical Exhibit 1

DRAFT

1.03 ACCURACY OF RADIAC

CALIBRATIONS

1.4, 5.1.7.3, 5.1.8.8, 5.1.8.8.1, 5.1.8.10, 5.1.8.11, 5.1.8.12, 5.1.8.13, 5.1.8.14, 5.1.8.15, 5.1.8.16, 5.1.8.17, 5.1.8.18, 5.1.8.19, 5.1.9.4, 5.1.9.5, 5.1.12

ENSURE ALL RADIAC CALS ARE ACCURATE.

MAJOR DEFECTS: 1. THE ITEM IS UNSAFE OR

HAZARDOUS TO USE, OR 2. THE ITEM DOES NOT

MEET CAL ACCURACIES FOR ALL FUNCTIONS

CERTIFIED, 3 ALSO AS DEFINED IN T.O. 00-20-14

AND PWS PARAGRAPH 2.2.19.

MINOR DEFECTS: ANY DEFECT THAT IS NOT A

MAJOR. ALSO MAJOR OR MINOR DEFECTS AS

DEFINED IN T.O. 00-20-14 AND THIS PWS.

1% OR LESS MAJOR DEFECTS

OF LOT ALLOWED

2% OR LESS MINOR DEFECTS

OF LOT ALLOWED

** LOT SIZE IS MONTHLY

PRODUCTION MINUS BATCHED

ITEMS.

(TOTAL ITEMS IN THE

BATCHES).

FOR BATCHED ITEMS 1 MAJOR

DEFECT IS ALLOWED OR 2

MINOR DEFECTS.

**LOT SIZE IS MONTHLY

PRODUCTION OF BATCHES.

RANDOM

SAMPLING

RADIAC LAB

ONLY

2.00 FACILITY CONTROL

5.1.4, 5.1.5, 5.1.6, 5.1.9.4.3

MUST MAINTAIN FACILITY TO OPERATING

SPECIFICATIONS IAW T.O. 00-20-14, AND THIS

PWS

ENVIRONMENTAL SYSTEMS

WILL BE WITHIN TOLERANCE

90% OF TIME. ALL OUT OF

TOLERANCE CONDITIONS WILL

BE ANNOTATED AND

CORRECTIVE ACTIONS

DOCUMENTED

RANDOM

SAMPLING

LAB

3.00 MAINTAIN AFPSL MEASUREMENT

AREA CAPABILITIES

5.1.8.8.1, 5.1.12

ENSURE REGULAR PROCEDURES FOR

MAINTAINING MEASUREMENT AREA ASSURANCE

ARE FOLLOWED AND ACCURATE RECORDS

DOCUMENTING THE RESULTS OF THESE

PROCEDURES ARE AVAILABLE.

ONE OR 25% OF SAMPLE,

(WHICHEVER IS GREATER),

DEFECTS ALLOWED PER

MONTH.

LOT SIZE IS ALL AFPSL

MEASUREMENT AREAS

RANDOM

SAMPLING

LAB

4.00 SERVICE CUSTOMERS WITHIN

ALLOTTED TIME

4.01 SERVICE CUSTOMERS WITHIN

ALLOTTED TURN-AROUND-TIME

5.1.8.1, 5.1.8.2, 5.1.8.3, 5.1.8.4, 5.1.8.4.1, 5.1.8.4.2, 5.1.9.1, 5.1.9.4.1, 5.1.8.7.1, 5.1.13.2

ENSURE WORK FLOW TIMES DO NOT EXCEED

THOSE STANDARDS DEFINED IN THIS PWS

3% DEFECTS ALLOWED PER

MONTH. IF THRESHOLD IS

EXCEEDED, 0% DEFECTS

ALLOWED 15 DAYS BEYOND

TURN-AROUND-TIME; WRITTEN

JUSTIFICATION REQUIRED FOR

ALL ITEMS EXCEEDING TURN-

AROUND-TIME BY 10 DAYS OR

MORE. LOT SIZE IS MONTHLY

PRODUCTION.

ONE HUNDRED

PERCENT

INSPECTION

LAB

4.02 SERVICE CUSTOMERS WITHIN

ALLOTTED TURN-AROUND-TIME

5.1.8.1, 5.1.8.2, 5.1.8.3, 5.1.8.4, 5.1.8.4.1, 5.1.8.4.2, 5.1.9.1, 5.1.9.4.1, 5.1.8.7.1, 5.1.13.2

ENSURE WORK FLOW TIMES DO NOT EXCEED

STANDARDS DEFINED IN THIS PWS

2 "BATCHED ITEM" DEFECTS

ALLOWED PER MONTH AND

0.75% "NON-BATCHED ITEM"

DEFECTS ALLOWED PER

MONTH. IF THRESHOLD IS

EXCEEDED, 0% DEFECTS

ALLOWED 5 DAYS BEYOND

TURN-AROUND-TIME; WRITTEN

JUSTIFICATION REQUIRED FOR

ALL ITEMS EXCEEDING TURN-

AROUND-TIME. LOT SIZE IS

MONTHLY PRODUCTION.

ONE HUNDRED

PERCENT

INSPECTION

RADIAC LAB

ONLY

4.03 SERVICE TO CUSTOMERS WITHIN

ALLOTED TURN-AROUND-TIME FOR

T.O. DEVELOPMENT

5.2.1.2.4, 5.2.1.4, 5.2.1.5.2, 5.2.1.6

ENSURE WORKFLOW TIMES DO NOT EXCEED

THOSE STANDARDS DEFINED IN THIS PWS

10% NEW T.O. DEFECTS

ALLOWED PER MONTH; 10%

REWRITE DEFECTS ALLOWED

PER MONTH; 1%

REVISION/CHANGE DEFECTS

ALLOWED PER MONTH. IF

THRESHOLD IS EXCEEDED, 0%

DEFECTS ALLOWED 5 DAYS

BEYOND TURN-AROUND-TIME;

WRITTEN JUSTIFICATION

REQUIRED FOR ALL ITEMS

EXCEEDING TURN-AROUND-

TIME LOT SIZE IS MONTHLY

ONE HUNDRED

PERCENT

INSPECTION

PWS 5.2.7

RANDOM

SAMPLING

T.O.

4.04 SERVICE CUSTOMERS WITHIN

ALLOTTED TURN-AROUND-TIME

5.1.8.1

ENSURE WORK FLOW TIMES DO NOT EXCEED

THOSE STANDARDS DEFINED IN THIS PWS

1% DEFECTS ALLOWED PER

MONTH. IF THRESHOLD IS

EXCEEDED, 0% DEFECTS

ALLOWED FIVE DAYS BEYOND

TURN-AROUND-TIME; WRITTEN

JUSTIFICATION REQUIRED FOR

ALL ITEMS EXCEEDING TURN-

AROUND-TIME. LOT SIZE IS

MONTHLY PRODUCTION

ONE HUNDRED

PERCENT

INSPECTION

LAB

4.05 SERVICE CUSTOMERS WITHIN

ALOTTED TURN-AROUND-TIME

ENSURE WORK FLOW TIMES DO NOT EXCEED

THOSE STANDARDS DEFINED IN THIS PWS

3% DEFECTS ALLOWED PER

MONTH. IF THRESHOLD IS

EXCEEDED, 0% DEFECTS

ALLOWED 15 DAYS BEYOND

TURN-AROUND-TIME; WRITTEN

JUSTIFICATION REQUIRED FOR

ALL ITEMS EXCEEDING TURN-

AROUND-TIME BY 10 DAYS OR

MORE. LOT SIZE IS MONTHLY

PRODUCTION.

ONE HUNDRED

PERCENT

INSPECTION

LAB (FMS)

5.00 DATABASE MAINTENANCE

5.01 MAINTAIN PMEL AUTOMATED

MANAGEMENT SYSTEM (LABMET)

5.1.8.7.4, 5.1.8.17

ENSURE ALL LABMET ENTRIES ARE ACCURATE

FOR ALL TMDE SUPPORTED, AND ALL DATA

LISTINGS ARE PRODUCED AND DISTRIBUTED IAW

THIS PWS

3% OR LESS DEFECTS OF

MONTHLY PRODUCTION

ALLOWED

RANDOM

SAMPLING

LAB

6.00 MAINTAIN PUBLICATION T.O. & CPIN

LIBRARIES

5.1.8.9, 5.2.5

MUST BE MAINTAINED IAW T.O.s 00-5-1 AND 00-5-

3; AND AFI 33-360

5.2.5 THREE

NONCONFORMATIES PER

AUDIT 5.1.8.9 AUDITED AS

PART OF 1.01 & 1.02 THREE

NONCONFORMATIES PER

MONTH

RANDOM

SAMPLING

T.O. & LAB

7.00 LOGISTICS

7.01 RECEIPT PROCESSING

5.1.8.1, 5.1.9.1

100% OF ALL ASSETS ARE ENTERED INTO

COMPUTER RECORDS.

(LABMET ENTRIES NOT LATER THAN ONE WORK

DAY AFTER RECEIPT)

EIGHT DEFECTS ARE

ALLOWED. LOT SIZE IS

MONTHLY RECEIPTS

RANDOM

SAMPLING

LAB

7.02 STORE AND LOCATE EQUIPMENT

5.1.7.2.3, 5.1.8.4.5, 5.1.10.5, 5.1.10.6, 5.1.10.7

100% OF ALL ASSETS ARE ASSIGNED

WAREHOUSE LOCATIONS, AND STORED WITHIN

THREE WORKDAYS AFTER RECEIPT

TWO DEFECTS ARE ALLOWED.

LOT SIZE IS MONTHLY

RECEIPTS

RANDOM

SAMPLING

LAB

7.03 PACKAGING

5.1.11, 5.2.7.1

ALL ITEMS SHIPPED SHALL BE PACKAGED IAW

TRANSPORTATION PACKAGING ORDER (TPO) OR

SPECIAL PACKAGING INSTRUCTION (SPI). (AFMAN

24-204, AFI 24-602, CFR TITLE 49, AND MIL-STD-

2073-1)

1% DEFECTS ARE ALLOWED.

LOT SIZE IS TOTAL SHIPPED

PER MONTH

T.O. TWO DEFECTS ALLOWED

RANDOM

SAMPLING

T.O. USE OF

CUSTOMER

COMPLAINTS

LAB & T.O.

8.00 PROPERTY CONTROL

4.1.2, 5.1.10.1.1, 5.1.10.9.1, 5.1.10.9.2, 5.1.10.9.3, 5.1.10.9.4, 5.1.10.10, 5.1.10.11, 5.1.10.22

PROPERTY CONTROL SYSTEM IS ADMINISTERED

IAW FAR 45.5

ALL PROPERTY WILL BE

MAINTAINED TO

CONTRACTOR/GOVERNMENT

AGREED CONDITION. LOT SIZE

IS TOTAL ACCOUNTABLE

PROPERTY.

SYSTEM SURVEYS

DOD 4161.2-M, OR

RANDOM

SAMPLING

DCMA FUNCTION

LAB

9.00 INVENTORY ACCURACY

1.2.3.4.1, 3.1.1, 3.1.2, 3.1.3, 3.1.4, 5.1.10.3, 5.1.10.8, 5.1.10.9.1, 5.1.10.9.2

100% OF ALL GOVERNMENT FURNISHED

EQUIPMENT MUST BE INVENTORIED.

2% DEFECTS OF ITEMS

AUDITED ARE ALLOWED.

RANDOM

SAMPLING

LAB

10.00 RESPOND TO REPORTS REQUIRED

IN TECHNICAL EXHIBIT 4

RESPOND WITH ACCURATE SUBMISSIONS AND

WITHIN TIMES SPECIFIED FOR THESE REPORTS.

PERFORMANCE REQUIREMENT

IS TWO DEFECTIVE REPORTS.

LOT SIZE IS NUMBER OF

REPORTS PER MONTH

NO DEFECTS OF CDRL A010

ONE HUNDRED

PERCENT

INSPECTION

LAB & T.O.

Sheet1

File details come from the government source that posted it. Updated .