Attachment One.docx
DOCX document 15 KB Posted
- Contract number
- 26-493ISP-FOREN-P-96051
- Issued by
- Illinois
About this file
I appreciate your request, but I'm unable to access the actual content of the document "Attachment One.docx" that you've referenced. While you've provided the GovTribe ID and filename, the document itself has not been included in a format I can read.
To provide the comprehensive summary you've requested, please share the document content by either:
- Pasting the text directly into our conversation
- Providing the document in a text-readable format
- Sharing the key details you'd like summarized
Once you provide the document content, I'll deliver a professional two-paragraph summary following your specified format.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Standard Terms and Conditions Pipettes.pdf | ||
| Standard IL Certification Pipettes.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
1. All completed calibrations of pipettes will be accompanied by a uniquely identifiable ISO 17025 compliant calibration certificate.
2. Stickers must be issued for an affixed to each pipette calibrated listing the expiration date (month/year), date of service (date/month/year), and vendor name.
3. Vendor must maintain ISO 17025 accreditation during this contract.
4. Calibrations of pipettes must be done on-site.
5. Preventative maintenance must be done as part of the calibration check. Services will include: replacement of worn seals, a-rings, shafts, and replacement of pistons. Method used must be a minimum of 4 weighings (at 3 volume settings, if an adjustable pipette). Adjustment to manufacturer specifications, calibration certificates, pipette sticker as described above, thorough pipette cleaning, visual and pressurized leak test, reminder notice for scheduled service, and 12 month warranty of labor and parts.
1. All certificates must be provided in either electronic or paper format. Certificates must include, at a minimum: Name/address of calibration vendor, the date of calibration (day/month/year), next due date (month/year), documented uncertainty, 'as found' and 'as left/calibrated’, status (pass/fail), manufacturer, size, printed name and signature of individual conducting calibration, what balance was used (identified by model and serial number), environmental factors (temperature, humidity, barometric pressure, air density, and Z-factor), method used, and data indicating the performance of the pipette.