This document is a Statement of Work (SOW) for a Large Scale Fuel Cell Test Stand. The purpose is to replace an aging test stand at the NIST Neutron Imaging Facility, which is used to test low polymer electrolyte membrane hydrogen fuel cells.
The key requirements are: 1) The test stand must be less than or equal to 48" wide x 112" long x 90" tall. 2) It must have a programmable load bank supporting dual mode operation, up to 1500 Amps at 10V. 3) The stand must work on one of the available US AC power standards (480V 3-phase, 208V 3-phase, or 120V single phase). 4) It must be able to control a single fuel cell or a stack of up to 6 cells. 5) It must support operation up to at least 120C with pressurized gas flows, humidification, and a coolant loop. 6) It must have mass flow controllers for hydrogen, nitrogen, and air with specific flow rate and turndown requirements. 7) It must have rapid response humidification capabilities.
This SOW is attached to a Federal Contract Opportunity for a Fuel Cell Test Stand, solicitation number 1333ND24QNB680494, issued by the Department of Commerce National Institute of Standards and Technology.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| QandA 1333ND24QNB680494 NIST Fuel Cell Test System Questions.xlsx | XLSX spreadsheet | |
| Combined Synopsis Solicitation 1333ND24QNB680494.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK/REQUIREMENTS DOCUMENT
Title: Large Scale Fuel Cell Test Stand Requesting Lab: 682.03
I. BACKGROUND INFORMATION
Purpose/Objective:
The NIST Neutron Imaging Facility performs testing of low polymer electrolyte membrane hydrogen fuel cells (PEMFCs) by imaging the water production in these PEMFCs using neutron imaging. Currently the facility must replace its aging large scale PEMFC test stand. NIST supplies this test stand with U.S. power, stand cooling water, deionized water, hydrogen gas (100 psi), nitrogen gas (100 psi) and air (100 psi). This test stand can control a single large PEMFC or a PEMFC stack by suppling it humidified hydrogen and air and dissipating the resulting electrical load.
In addition, the stand provides water coolant to the PEMFC to control the cell temperature. The water in this water coolant loop has been replace by heavy water by NIST because neutrons are used to image the cell. Only heavy water cell coolant may be used and will be supplied by NIST upon delivery of the test stand. Neutron imaging experiments run in an enclosed, shielded, high radiation enclosure. The test stand sits outside the enclosure in a low radiation area. Inside the enclosure, single cells and/or short stacks are supported by the test stand using long heated lines and long load cables that penetrate the walls of the enclosure through holes.
II. SCOPE
The Contractor shall deliver a fuel cell test stand, inclusive of FOB Destination delivery, installation, warranty and training, and option line items.
III. MINIMUM REQUIREMENTS
The Contractor shall provide a system that meets all technical specifications identified below. All items must be new. Used or remanufactured equipment will not be considered for award.
Experimental, prototype, or custom items will not be considered. The use of “gray market” components are not authorized for sale in the U.S. by the Contractor is not acceptable. All line items shall be shipped in the original manufacturer’s packaging and include all original documentation and software, when applicable.
Line Item 0001:
Description: Large Scale Fuel Cell Test Stand Quantity: 1
A. Technical Specifications
1) Stand dimensions must be less than or equal to 48” wide x 112” long x 90” tall to meet available lab footprint.
2) Programmable load bank
a. Supports dual mode operation at 10% and 100% of capacity
b. Supports at least 1500 Amps with Zero Volt operation across the stack
c. Voltage range up to 10 V
3) Stand electrical requirements.
a. Must work on one of the available US AC electrical power standards (60 Hz frequency) which include 480 V (3-phase), 208 V (3-phase), and/or 120 V (single phase).
4) Stand can control either a single cell or a multicell stack (max cells is 6)
5) Stand supports operation up to at least 120 C
a. Pressurized gas flows
b. Humidification
c. Coolant loop
6) Gas flows controlled by mass flow controllers
a. Anode flows:
7) Hydrogen max flows 80 slpm, 100:1 turndown, 2 MFC’s minimum for finer flow control
8) ii) Nitrogen min max flows 80 slpm, 100:1 turndown, 2 MFC’s minimum for finer flow control
a. Cathode flows:
i. Air max flows 200 slpm, 100:1 turndown, 2 MFC’s minimum for finer flow control
b. Both anode and cathode mass flow controller must be customer changeable to allow for lower flow rates.
c. Automatic back pressure control 1-4 Bara
d. Reactant gas flow controller can be swapped out by users to better support single cell or multicell flow rates
e. Incorporation of Customer-Supplied Heated Hoses (Mfg. Clayborn)
9) Humidification
a. Rapid response <10 s (through use of two humidifiers supplying each electrode)
b. Supports up to 120 C operation and gas temperatures up to 130 C
c. Can be completely bypassed for dry gas operation
10) Fuel cell stack coolant loop
a. 0.2-5 liters per minute
b. Supports operation at 130 Celcius
c. Built in deionized water filter
d. NIST will supply heavy water for coolant loop operation
e. Nitrogen purge port to remove coolant from loop
f. Liquid coolant refill port(s)
11) All stand thermocouples are T-type
12) Includes Cell Voltage Monitor
13) Expandable additional inputs that can be logged
a. T-type thermocouple inputs
b. Voltage inputs
14) Fuel cell stack high frequency resistance measurement
a. Resistance range: 3 milli-Ohm to 3000 Ohm.
b. Frequency: 1 kHz.
c. Current: 100 milli-Ampere to 10 micro-Ampere.
15) Electrochemical Impedance Spectroscopy
a. Customer-Supplied Gamry reference 3000
b. Integration in stand software by manufacturer
16) Test stand must be fully automated
a. NIST supplies control computer to meet internal IT requirements.
b. Manufacturer supplied standalone software that runs natively on the Windows 10 operating system and be able be transitioned to Windows 12.
c. NIST in house data acquisition software must be able to programmatically start and stop test stand fuel cell automation as well as receive notification when automated stand tasks have finished.
17) NIST supplies hydrogen, nitrogen and air gas using ½” Swagelok fittings.
18) All internal stand plumbing must be Swagelok fittings.
19) Stand exhaust heat exchangers are cooled via NIST house coolant loop.
20) Stand must have emergency stop/interlock with NIST facility 20 V DC inputs.
21) Installation, Training, Support and Warranty
a. One year or more warranty
b. On site installation and training
c. Call in customer support
d. Options for additional customer support beyond initial period
IV. DELIVERABLES
Description Quantity or Format Due Date
Large Scale Fuel Cell Test Stand
1 14 months ARO
Standards of Acceptance: The NIST TPOC or COR shall review (list deliverable) and respond with an acceptance or request for revision email to the Contractor Point of Contact (POC) within 14 days/weeks of receipt of deliverable. This section applies to deliverables which must be reviewed and approved for acceptances (draft plans, drawings, etc.)
V. PLACE OF PERFORMANCE
NIST
VI. PERIOD OF PERFORMANCE/LEAD TIME
14 months after receipt of order
VII. DELIVERY TERMS
Delivery shall be F.O.B Destination and shall occur in accordance with the delivery due dates provided in the above table.
FOB Destination means: The contractor shall pack and mark the shipment in conformance with carrier requirements, deliver the shipment in good order and condition to the point of delivery specified in the purchase order, be responsible for any loss of and/or damage to the goods occurring before receipt and acceptance of the shipment by the consignee at the delivery point specified in the purchase order; and pay all charges to the specified point of delivery. The contractor shall deliver all Line Items to:
National Institute of Standards and Technology Shipping and Receiving 100 Bureau Drive, Building 301 Gaithersburg, MD 20899 C/O/ POC/COR David L Jacobson, Building 235, Room K133
VIII. INSPECTION & ACCEPTANCE
In addition to the inspection and acceptance terms articulated in 52.212-4, the Government reserves the right to perform such performance tests and evaluations as defined below to verify specified system performance. Such tests and evaluations, if performed, shall be conducted within the environment that the system is to be operated. The Contractor has the right to be present during the tests and evaluations, if performed, at the Contractor’s expense.
Performance Tests:
1. Vendor will demonstrate performance during installation and training.
NIST may choose at its discretion to forego this part of acceptance testing.
A visual inspection of the equipment will be performed by the NIST TPOC to identify surface defects or any form of indication that any equipment was damaged during transport to NIST. The Government shall have sole discretion to require repair or replacement of damaged and/or nonconforming supplies at no cost to the Government. The Government at any time prior to acceptance shall reject the equipment due to defects and/or nonconformance. The vendor is responsible for latent defects discovered any time after final inspection. However, the extent of its liability shall be prorated over the useful life of the equipment.
Ownership of the equipment shall transfer to NIST upon acceptance by the Government.
The Government will test, inspect, and accept or reject the equipment within (insert working days it will take to inspect, accept or reject) of the receipt of the equipment unless otherwise indicated above. The Government reserves the right to conduct quality assurance testing to confirm that a given instrument(s) meets the manufacturer’s and/or the Government’s performance specifications. It is anticipated that the equipment will meet all manufacturer’s specifications and/or the Government’s performance specifications identified in the most recent operations and maintenance manual for each piece of equipment and/or in this document.
IX. WARRANTY
The contractor shall warranty the entire system for a period of a minimum of (enter required warranty term) after receipt of the equipment and shall be in accordance with terms in FAR 52.212-
4. Warranty shall commence upon acceptance of the system by the Government and at a minimum shall include the following:
X. PAYMENT SCHEDULE (This section to be revised as needed)
Advance payment is not authorized. The Contractor must invoice in arrears according to the payment schedule. The Contractor shall be paid, in accordance with Net 30-day payment terms, upon receipt and acceptance of a proper invoice, in accordance with the following schedule:
1. 100% after installation and acceptance by the TPOC of fully installed system, AND
2. After the successful completion of the testing requirements set forth in this document under section set forth in this document, AND
3. After successful demonstration by the instrumentation that it performs IAW the technical requirements set forth in this document AND
4. After receiving an invoice submitted properly, IAW the purchase order terms and conditions.
NOTE: Partial shipments and partial invoices will not be accepted, unless otherwise requested and accepted by the Contracting Officer prior to award offer. Proposed payment schedules shall be submitted with vendor’s response to the RFQ for consideration.