Attachment 1 - Specifications.pdf
PDF 605 KB Posted
- Solicitation number
- 12805B24Q0368
About this file
This document is an attachment to a federal contract opportunity for a custom biosafety cabinet compatible with a BD FACSMelody cell sorter. The key details are:
The U.S. Department of Agriculture (USDA), Agricultural Research Service (ARS) requires a biosafety cabinet that meets specific size, power, and configuration requirements to be used with a BD FACSMelody cell sorter located at their facility in Manhattan, Kansas. The solicitation number is 12805B24Q0368 and the NAICS code is 334516 with a 1,000 employee size standard. Offerors must provide a quotation that addresses the technical specifications, shipping/delivery costs, and pricing, valid for at least 90 days. Award will be made to the responsible offeror whose quote conforms to the specifications and is most advantageous to the government considering technical, past performance, and price factors. Quotes are due by 12:00 pm Central Time on July 2, 2024 and should be submitted via email to the contracting officer.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment 2 - Buy American Certificate 12805B24Q0368.docx | DOCX document | |
| Clauses and Provisions - 12805B24Q0368.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
United States Department of Agriculture
Research, Education, and Economics Agricultural Research Service
Attachment 1 - Specifications Custom Baker Biosafety Cabinet for BD FACSMelody
Solicitation# 12805B24Q0368
Background and Scope The USDA ARS Foreign Animal Disease Research Unit (FADRU) requires a biosafety cabinet (BSC) that is compatible for use with BD FACSMelody cell sorter.
Applicable Documents and Standards
Specifications:
BSC must meet the specifications laid out in the attached document BGA-BD404XD-A2-A-B.pdf Briefly, the salient characteristics of the BSC are as follows:
1. Workspace dimensions must be 27 7/8in (D) x 27 7/16in (H) x 46in (L).
2. Work Surface dimensions must be 23 15/16in (D) x 44in (L).
3. 4” diameter cable ports should be present on the right (3) and left (2) hand side of the BSC.
4. A duplex receptacle with self-resetting circuit breaker should be present on the right-hand side wall for 100, 115 or 220V.
Testing and Verification Qualification via a customer site certification test will be the verification standard.
Government Furnished Property FADRU has already purchased the BD FACSMelody cell sorter and it is located in B3.144.
Data Requirements Required documentation must include certification of calibration of air flow.
Packaging and Shipping All packaging, shipping, and shipping-insurance costs must be included with purchase price.
Delivery: 60 Days ARO
USDA ARS NBAF
1200 Kretschmer Drive Manhattan, KS 66502
±1/32FRACTIONS
CHECKED BY
DESIGNED BY
FINISH: 125
DECIMALS
.XXX
.XX
.X ±.02
±.01
MACHINED
±.005 THIRD ANGLE
DATE
DATE
PROJECTION
NO 4
ANGULARSHEET METAL
±.010
±.03
±.06
D
SQUARE FOOTAGE
SCALE
DWG NOSIZE
TITLE
SHEET
REV
A ORIGINAL RELEASE - - - -
REV DESCRIPTION DATE DRN APP-BY APP-DATE
STERILGARD, MODEL BD404XD / BD404XD-INT
FOR BD FACSMELODY CELL SORTER
CLASS II TYPE A2
BGA-BD404XD-A2-A B
UNLESS OTHERWISE SPECIFIED
DO NOT SCALE DRAWING
REMOVE ALL BURRS AND SHARP EDGES
U.S. CUSTOMARY TOLERANCES: (INCHES)
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 8
A
B
C
D
A
B
C
AStroshane
L. McCarthy
09/22/2017
9/22/2017 1 OF 1
ALL ITEMS MUST BE
RoHS COMPLIANT
THIS DRAWING AND ITS CONTENTS
ARE THE PROPERTY OF THE
BAKER COMPANY. THE
INFORMATION CONTAINED HERE
ON MAY NOT BE USED OR COPIED
IN ANY MANNER WITHOUT THE
WRITTEN PERMISSION OF BAKER.
161 Gatehouse Road Sanford, ME 04073 Ph: 207-324-8773 / web: www.bakerco.com
±1' - 0°
D
61 15/16"
1572.8MM
79 7/16"
2017.3MM
SHIPPING
HEIGHT
34 5/16"
871.6MM
STANDARD UNIT WEIGHT WITH CHANNEL STAND = 616 LBS [279 KG]
NOTES:
1. ELECTRICAL REQUIREMENTS:
100 VAC, 15 AMP, 50/60 HZ, SINGLE PHASE
115 VAC, 15 AMP, 60 HZ, SINGLE PHASE
220 VAC, 13 AMP, 50/60 HZ, SINGLE PHASE
2. CONSTRUCTION:
WORK AREA - 16 GA[1.5 MM] 304 SS, #4 FINISH
WORKSURFACE - 16 GA[1.5 MM] 316L SS, #4 FINISH
ACCESS PANELS - 16[1.5 MM] & 18 GA.[1.2 MM] C.R.S. WITH A BAKER
PERMA WHITE POLYESTER HYBRID POWDER COAT FINISH
3. FILTER, MOTOR, BLOWER AND ALL ELECTRICAL COMPONENTS
ARE FRONT ACCESSIBLE.
21 5/8"
549.2MM
VIEWING
AREA
2 7/8"
73.1MM
4 9/16"
116.6MM
TO CENTER LINE
OF VALVE
61 3/4"
1568MM
36 5/8"
930.3MM
MAXIMUM
1/2" BALL VALVE DRAIN
CABINET VENT TO ROOM
EXHAUST FILTER PROTECTOR
ACCESS REQUIRED FOR
CABINET CERTIFICATION
INTERNAL DAMPER
CONTROL
(W/LOCKING DEVICE)
6"[152.4MM] MINIMUM CLEARANCE
TO THE CEILING RECOMMENDED
FOR MAINTENANCE. 2 3/4"[69.9MM]
MINIMUM CLEARANCE REQUIRED
FOR PROPER CABINET OPERATION
CABINET
CONTROLS
53 11/16"
1363.8MM
46"
1168.4MM
WORKSPACE
LENGTH
DOP PORT FOR CABINET
CERTIFICATION
44"
1116.8MM
WORKSURFACE
LENGTH
SUPPLY DIFFUSER
RIGHT HAND SIDEWALL
DUPLEX RECEPTACLE WITH SELF-
RESETTING CIRCUIT BREAKER
100V: 5A TOTAL
115V: 5A TOTAL, GFCI PROTECTED
220V: 3A TOTAL
AIR BYPASS
ARM REST
FLUORESCENT
LIGHT
4" DIA CABLE PORT
(3) RIGHT SIDE
(2) LEFT SIDE
23 15/16"
608MM
WORK SURFACE
DEPTH
1 1/8"
28.2MM
27 7/8"
708.4MM
WORK SPACE
DEPTH
10"
253.8MM
ACCESS
OPENING
27 7/16"
697.4MM
WORK SPACE
HEIGHT
1 5/8"
41.3MM
MAX ADJUST
LEVELER
29"
736.6MM
24 1/2"
622.3MM
MINIMUM
POWER CORD:
20 FT[ 6M] EXTERNAL LENGTH
100V: 15 AMP NEMA 5-15 PLUG
115V: 15 AMP NEMA 5-15 PLUG
220V: LISTED PLUG FOR DESTINATION COUNTRY
CABINET PRESSURE
MONITOR
AMS AIRFLOW
MONITOR
AMS CONTROLS
BLOWER AND FILTER FOR
AEROSOL MANAGEMENT
SYSTEM (AMS)
NOTE: REFER TO THE LATEST VERSION OF PRODUCT
SPECIFICATION SHEET FOR THE EXHAUST
REQUIREMENTS OF THIS MODEL
BGA-BD404XD-A2-A-B.pdf
| Attachment 1 - Specifications.pdf |
| BGA-BD404XD-A2-A-B.pdf |
File details come from the government source that posted it. Updated .