This Agency Specifications document outlines the technical requirements for an autoclave procurement by the Alabama Department of Public Health (ADPH), Bureau of Clinical Laboratory located at 757 Museum Drive, Mobile, Alabama 36608. The specification details the equipment requirements, performance standards, and delivery conditions necessary for the autoclave system. The procurement specifies that delivery must be accomplished via a truck equipped with a liftgate, as the facility does not have a loading dock available. The document establishes the functional and technical parameters that qualified bidders must meet to be considered for award of this contract.
The specifications document provides detailed performance criteria, dimensional requirements, and operational standards for the autoclave equipment being procured. The document serves as the binding technical reference that all submitted bids must comply with to be deemed responsive. All bidders must ensure their proposed autoclave meets or exceeds the specifications outlined in this document, and delivery logistics must accommodate the site-specific constraint of the lack of loading dock facilities at the Bureau of Clinical Laboratory location.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Additional Terms and Conditions.docx | DOCX document | |
| ITB Standard Boilerplate.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Autoclave Technical Specification
General
One (1) autoclave that sterilizes media and equipment to be used in the laboratory. The equipment also disinfects waste, allowing it to be disposed of properly. The autoclaves should have the following specifications:
• A microcomputer-controlled steam sterilizer that can offer either prevacuum or gravity displacement cycles for sterilization.
• Internal chamber dimensions 21"x21"x38".
• Internal volume of 9.7 cu FT
• Full autoclave dimensions for each unit: 30 x 74 x 68.125" with a load height of 38.375".
• Design Operating Temperature Gravity/Vacuum Cycles: 230°–275°F (110°–135°C)
Liquid Cycles: 219°–275°F (104°–135°C) Optional Low Temp Cycle: 169°–212°F (76°– 100°C).
• Steam pressure requirements: 40-50 psig. (2.76-3.45 bar). Flow requirements are 120lb/hr.
• Autoclaves must be designed for connections to a facility (house) steam supply.
• Power supply: 100-240V, 50/60Hz
• Single Manual Door
• With Interior Rack & Two Shelves
• Exports cycle records to USB storage
• 8.4" touch-screen control interface with 10 selectable pre-programmed cycles
• Thermal printer that documents cycle performance
• NetCOM ethernet interface
• A minimum one-year warranty from the date of installation.
• Disposal of the previous autoclave from the laboratory into storage for salvage pickup.
Storage area is onsite, behind the laboratory at 757 Museum Drive, Mobile, AL, 36608.
• On-site training and instructional materials to be included with installation.
• Delivery, installation, and training to take place during the laboratory's business hours.
• Location does not have a loading dock. The delivery must arrive on a truck with a liftgate.
• Equipment provided must be equivalent or equal to the specifications listed above.
• All items must be received within 30 days after receiving the purchase order.