36C25918R0391-012.pdf
PDF 626 KB Posted
- Attached to
- Radiology Rooms 1 & 3 Federal contract opportunity
- Solicitation number
- 36C25918R0391
About this file
36C25918R0391 RAD LAB 1 DWGS.pdf
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C25918C0232-000.docx | DOCX document | |
| 36C25918R0391-00001000.docx | DOCX document | |
| 36C25918R0391-00001002.docx | DOCX document | |
| 36C25918R0391-00001001.pdf | ||
| 36C25918R0391-00001004.docx | DOCX document | |
| 36C25918R0391-00001003.pdf | ||
| 36C25918R0391-014.pdf | ||
| 36C25918R0391-011.pdf | ||
| 36C25918R0391-016.pdf | ||
| 36C25918R0391-018.docx | DOCX document | |
| 36C25918R0391-010.docx | DOCX document | |
| 36C25918R0391-013.pdf | ||
| 36C25918R0391-017.docx | DOCX document | |
| 36C25918R0391-015.pdf | ||
| 36C25918R0391-009.docx | DOCX document |
Show all 15
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
PLANNING BY
DR400FINALV4
01/12/2017
DAVE HENRY
DRAWN BY:
FILE NAME:
DATE:
1/4"=1'-0"
SHEET:
SCALE:FACILITY EQUIPMENT
RADIOGRAPHIC SYSTEM SUITE
DIGITAL HIGH FREQUENCY
TYPICAL ROOM LAYOUT
(NOT SITE SPECIFIC)
DR 400
Brendan Conley RT(R), MRT | Agfa HealthCare Manager, DR Implementation Western Zone M 214-548-7214
Jim Dungee | Agfa HealthCare Senior Project Manager, DR Implementation Eastern Zone T 856-722-5221 | M 856-430-5086
DAVE HENRY
FILE NAME:
DATE:
DRAWN BY:
SHEET:
SCALE:
REVISION DESCRIPTIONNUMBER
SHEET
REV REP DATE
01/12/2017
DR400FINALV4
RECEIPT OF FINAL DRAWINGS
NAME: DATE:
COULD RESULT IN ADDITIONAL EXPENSES BEING INCURRED.
ALSO, I AM AWARE THAT ANY CHANGES MADE AFTER THIS DATE
THE EQUIPMENT OPERATION, SERVICEABILITY, OR SAFETY GUIDELINES.
DRAWINGS, DETAILS, AND SPECIFICATIONS, MAY ENCROACH UPON
PLANS. IT IS UNDERSTOOD THAT ANY DEVIATION FROM THESE
THIS SIGNATURE REPRESENTS RECEIPT OF THIS SET OF FINAL
REVISION HISTORY
TYPICAL ROOM LAYOUT
(NOT SITE SPECIFIC)
COVER SHEET.
PROPOSED ROOM LAYOUT.
PROPOSED EQUIPMENT LAYOUT.
E3 E4
E2
B1 D1
E1
SHEET
A3 A2
NUMBER
A1
SHEET DESCRIPTION
A4
E5
ELECTRICAL ELEVATIONS.
D2
SHEET LEGEND
SHEET LEGEND, RECEIPT OF DRAWINGS, REVISIONS HISTORY,
RECOMMENDED VENDORS, ELECTRONIC TRANSFER AGREEMENT.
GENERAL NOTES, ARCHITECTURAL NOTES AND SUPPORT NOTES,
SUPPLEMENTAL NOTES.
EQUIPMENT DETAILS. *PENDING
EQUIPMENT DETAILS. *PENDING
ELECTRICAL LAYOUT PLAN VIEW, TERMINAL SCHEDULE, ELECTRICAL
TAG DESCRIPTIONS.
ELECTRICAL BLOCK DIAGRAM, INCOMING WIRE SIZE NOTES,
CONDUIT & WIRING SCHEDULE.
CONDUIT LINE DIAGRAM.
TYPICAL TROUGH ELEVATION, JUNCTION BOX COVERPLATES DETAIL,
WALL DUCT PART LIST.
E6 DOOR SWITCH / WARNING LIGHT DETAIL AND NOTES.
herewith, regardless of its format or the means of its transmission. There is no guarantee or representation to the user as to the accuracy, currency, suitability or reliability of this data for any purpose. User accepts the data "as is", and assumes all risks associated with its use. By acceptance of this data, the user agrees not to transmit this data or provide access or any part of it to
All sheets included shall be considered a complete set of specification plans and may not be separated for any purpose. Any information contained in this set shall NOT be applied to other projects.
Gamma Tech, Inc. makes no representation or warranties, express or implied, with respect to the reuse of the data provided another party unless the user shall include with the data a copy of this disclaimer. Gamma Tech, Inc. assumes no responsibility for actual or consequential damage incurred as a result of any user's reliance on this data. Furthermore, electronic files on this media contain confidential and proprietary information that cannot be reproduced or divulged, in whole or in part, without written
ELECTRONIC FILE TRANSFER AGREEMENT
authorization from Gamma Tech, Inc.
E7 ELECTRICAL NOTES.
NONE
SCALE:
SHEET:
NONE
DRAWN BY:
FILE NAME:
DATE:
DAVE HENRY
NO ATTENTION HAS BEEN GIVEN TO SPECIFIC BUILDING STRUCTURAL INTEGRITY. A
9. THE EQUIPMENT REPRESENTATIVES MUST APPROVE DEVIATION FROM THESE PLANS IN
PREVAILING SEISMIC REQUIREMENTS. SEISMIC ENGINEERING DRAWINGS TO BE DONE BY
CUSTOMER CONSULTS WITH A PROFESSIONAL STRUCTURAL ENGINEER FAMILIAR WITH ALL
RESPECTIVE CONTRACTORS OF CHOICE. IT IS STRONGLY RECOMMENDED THAT THE
AND COMPLIANCE SHALL BE THE RESPONSIBILITY OF THE FACILITY (END USER) OR THEIR
AFFECTED BY THE SCOPE OF THIS WORK. ALL EXPENSES ASSOCIATED WITH SEISMIC TESTING
NO ATTENTION HAS BEEN GIVEN TO SPECIFIC SEISMIC ANCHORAGE CONSIDERATIONS
SUPPORT FOR THE DIAGNOSTIC IMAGING EQUIPMENT DESCRIBED IN THE 'EQUIPMENT
THE FACILITY SUPPORT STRUCTURE (e.g. WALLS, FLOORS, CEILINGS) CAN PROVIDE ADEQUATE
PROFESSIONAL STRUCTURAL ENGINEER SHALL INSPECT FIELD CONDITIONS TO ENSURE THAT
SCHEDULE'. STRUCTURAL ENGINEERING DRAWINGS TO BE DONE BY OTHERS.
WRITING PRIOR TO ANY CHANGES.
OTHERS.
ELECTRICAL AND MECHANICAL DESIGNERS SHALL BE REQUIRED TO REVIEW THESE
AFFECTED BY THE SCOPE OF THIS WORK. MECHANICAL ENGINEERING DRAWINGS TO BE DONE
8. NO ATTENTION HAS BEEN GIVEN TO SPECIFIC PLUMBING OR HVAC CONSIDERATIONS
SINGLE SHEET OF PLATE GLASS ON BOTH SIDES TO HELP PREVENT SCRATCHING.
BE TILTED 7° TO HELP ELIMINATE GLARE. IF TILTED 7° VIEWING WINDOW IS ACRYLIC, PROVIDE A
7. VIEWING WINDOW TO BE 1/16" LEAD EQUIVALENT. RECOMMEND LARGE VIEWING WINDOWS TO
ELECTRICAL NOTES. GROUNDING MUST CONFORM TO THE CURRENT CODE REQUIREMENTS.
EQUIPMENT REQUIRES VERY LARGE CURRENTS FOR EXTREMELY SHORT TIMES. REFER TO
6. THE SUPPLY VOLTAGE MUST REMAIN RELATIVELY STABLE. ELECTRICAL POWER FOR X-RAY
MANAGER SHALL CONSULT WITH FACILITY REGARDING DESIRED CONVENIENCES.
RESPECTIVE ROOM (i.e., STORAGE, CABINETS, SINKS AND COUNTERS, ETC.) PROJECT
5. SOME CONVENIENCES MAY BE REQUIRED BY CODE OR DESIRED BY CUSTOMER IN EACH
4. ALL WORK, EXCEPT INSTALLATION OF DIAGNOSTIC IMAGING EQUIPMENT, IS TO BE DONE BY
3. ARCHITECT AND ENGINEERS SHALL USE EXISTING ELECTRICAL AND STRUCTURAL ITEMS
CUSTOMER AND APPROVED BY PREVAILING CODES. LIGHTING FIXTURES, VENTS, ETC. TO BE
OPERATORS. RECOMMEND 2' x 4' LAY-IN CEILING PANELS OF MATERIAL ACCEPTABLE TO
2. ALL SUPPLY DIFFUSERS SHALL NOT CAUSE DIRECT AIR CURRENTS ON PATIENTS OR
BUILDING INTO RESPECTIVE ROOMS. RECOMMEND COVERING OR REMOVING THRESHOLDS.
AND PASSAGEWAY SIZES ARE ADEQUATE FOR TRANSPORTING EQUIPMENT FROM EXTERIOR OF
FINISHED SURFACES UNLESS OTHERWISE NOTED. ARCHITECT SHALL VERIFY THAT ALL DOOR
CONSTRUCTION PARTICIPANTS TO ADHERE TO THESE DISTANCES. ALL DIMENSIONS ARE FROM
DIMENSIONAL RELATIONSHIPS HAVE BEEN SUPPLIED. IT IS THE RESPONSIBILITY OF ALL
CONSTRUCTION PARTICIPANTS SHALL VERIFY AND ADJUST DIMENSIONS. CRITICAL
GAMMA TECH'S FLOOR PLAN WAS SCALED FROM PRINTS DRAWN BY OTHERS. ALL
INTO A WORKING SET OF CONSTRUCTION DRAWINGS OBSERVING ALL PREVAILING CODES.
TEAM SHALL EXTRAPOLATE PERTINENT INFORMATION FROM THIS LAYOUT AND INCORPORATE
TO CONFORM DETAILS TO THE ACTUAL EQUIPMENT EXPECTED TO BE INSTALLED. THE DESIGN
WITH SPECIFICATION AND CONCEPTUAL INFORMATION ONLY. EVERY EFFORT HAS BEEN MADE
GAMMA TECH'S DRAWINGS ARE INTENDED TO PROVIDE THE CONSTRUCTION PARTICIPANTS
RELYING ON THIS MATERIAL ASSUMES ALL RISKS AND LIABILITY ARISING FROM SUCH USE OR
BUILDING CODES AND REGULATIONS THAT HAVE JURISDICTION. THOSE MAKING USE OF OR
OR PROJECT MANAGER WHO SHALL ENSURE THAT ALL CONSTRUCTION COMPLIES WITH
GAMMA TECH'S DRAWINGS ARE FOR THE BENEFIT OF PURCHASER'S ARCHITECT OF CHOICE
FOR CONSTRUCTION. THE EQUIPMENT REPRESENTATIVES OR THEIR DESIGNER ARE NOT
1. GAMMA TECH'S SET OF DRAWINGS ARE FOR THE PURPOSE OF A LAYOUT PLAN ONLY - NOT
ARCHITECTURAL NOTES
THOSE OTHER THAN THE EQUIPMENT REPRESENTATIVES.
RESPONSIBLE OR LIABLE FOR ANY ERRORS OR OMISSIONS.
HELP MINIMIZE CONFLICTING CONSTRUCTION EFFORTS. ELECTRICAL ENGINEERING DRAWINGS
DRAWINGS TO HELP THEM GAIN AN UNDERSTANDING OF THE CONCEPT OF THIS WORK AND TO
BY OTHERS.
WHEN FEASIBLE.
FLUSH WITH FINISHED CEILING.
TO BE DONE BY OTHERS.
RELIANCE.
PURCHASE ORDER. ADDITIONALLY, A SITE INSPECTION SHALL BE CONDUCTED BY THE
EQUIPMENT REPRESENTATIVES TO CONFIRM THE SITE'S READINESS PRIOR TO X-RAY
EQUIPMENT DELIVERY. THIS INSPECTION SHALL BE PERFORMED WHEN NOTIFIED OF THE
CONSTRUCTION'S COMPLETION.
THE EQUIPMENT REPRESENTATIVES OR THEIR PLANNER ASSUME NO RESPONSIBILITY
AND ADDITIONAL TESTS REQUIRED TO MEET THE SATISFACTION OF THE HEALTH
REPLACEMENT OF ALL DEFECTIVE WORK (INCLUDING WORK SUBSEQUENTLY AFFECTED)
OF THE CONTRACTOR. THEREFORE, THE CONTRACTOR IS RESPONSIBLE FOR REPAIRS OR
14. THE DECISION OF ACCEPTABILITY BY A HEALTH PHYSICIST SHALL BE THE RESPONSIBILITY
THE SHIELDING SHALL BE TESTED, AT THE OWNER'S EXPENSE, BY A CERTIFIED HEALTH
AFTER X-RAY EQUIPMENT HAS BEEN INSTALLED AND PLACED IN OPERATING CONDITION,
13. SHIELDING TESTING (LEAD OR EQUIVALENT):
THESE MUST BE SPECIFIED BY A QUALIFIED RADIATION PHYSICIST.
12. RADIATION PROTECTION REQUIREMENTS:
10. ANY AIR CONDITIONING REQUIREMENTS. (HVAC - HEATING, VENTILATION, AIR CONDITIONING
ENTRANCE DOORS OF X-RAY ROOM TO BE COORDINATED WITH FACILITY, EQUIPMENT
9. INSTALLATION OF OPTIONAL WARNING LIGHT AND/OR DOOR INTERLOCK SWITCH AT
SPRINKLER HEADS, HALON NOZZLES, ETC. SHALL NOT EXTEND MORE THAN 1/2" BELOW
7. ANY CEILING MOUNTED LIGHT FIXTURES, DIFFUSERS, VENTS, SMOKE/FIRE DETECTORS,
SERVICE, A MINIMUM NUMBER OF OUTLETS WILL BE DEPICTED.
OUTLETS REQUIRED BY EQUIPMENT WILL BE SPECIFIED. FOR EASE IN EQUIPMENT
6. FOR THE MOST PART CONVENIENCE OUTLETS ARE TO BE SPECIFIED BY OTHERS.
5. INSTALLATION OF WALL AND CEILING SUPPORT AND SUSPENSION FRAME WORK AS
4. INTERCONNECTING WIRE AND CABLE AS SPECIFIED.
3. INSTALLATION OF LOAD CENTERS, BRANCH PANELS, CIRCUIT BREAKERS, JUNCTION BOXES,
2. INSTALLATION OF PROPER SPECIFIED POWER.
FINISHED BEFORE EQUIPMENT INSTALLATION CAN BEGIN.
1. ANY ROOM REMODELING AND CONSTRUCTION - FLOOR CEILING AND WALLS MUST BE
PROVIDE THE FOLLOWING (IF REQUIRED) PRIOR TO X-RAY EQUIPMENT DELIVERY AND
AT THEIR EXPENSE, THE CUSTOMER AND/OR THE CONTRACTOR WILL BE EXPECTED TO
X-RAY EQUIPMENT OR SITE PLANNING DOCUMENTS. NO WORK MAY BE PERFORMED OR
MATERIALS FURNISHED AT THEIR EXPENSE UNLESS AUTHORIZED BY A SIGNED, FORMAL
FOR ANY CONSTRUCTION COSTS, WHETHER OR NOT RELATED TO THE INSTALLATION OF
11. ALL ANCILLARY AND ACCESSORY ITEMS.
AND HUMIDIFICATION)
PHYSICIST.
INSTALLERS, AND DESIGN TEAM.
8. ANY PLUMBING REQUIREMENTS.
THE FINISHED CEILING.
15. LIABILITY:
PHYSICIST.
CONDUIT AND CABLE TRAYS AS SPECIFIED.
GENERAL NOTES
INSTALLATION:
SPECIFIED.
5. ALL SURFACE MOUNTED SUPPORTS ARE TO BE PAINTED THE SAME COLOR AS THE
ADJACENT SURFACE OR CONSULT WITH FACILITY REGARDING ACCENT COLORS.
( CONTRACTOR FURNISHED - CONTRACTOR INSTALLED )
2. ALL SUPPORTS STRUCTURES ARE TO BE PROVIDED BY THE FACILITY OR THEIR
3. WALL SUPPORTS ARE TO BE HIDDEN WITHIN WALLS WHENEVER POSSIBLE.
THE EQUIPMENT REPRESENTATIVE OR THEIR DESIGNER IS IN NO WAY RESPONSIBLE FOR THE
SUPPORT STRUCTURE FOR ANY OVERHEAD CEILING MOUNTED EQUIPMENT / DEVICES OR WALL
MOUNTED EQUIPMENT / DEVICES. THIS IS THE RESPONSIBILITY OF THE CONSTRUCTION
1. ALL UNITS THAT ARE WALL MOUNTED OR CEILING SUSPENDED ARE TO BE PROVIDED WITH
GIVEN (i.e.,BLOCK). CONSULTATION WITH A STRUCTURAL ENGINEER IS STRONGLY
4. WALL SUPPORTS MAY NOT BE REQUIRED IF WALL MATERIAL IS ADEQUATE FOR THE LOADS
SUPPORTS STRUCTURES WHERE NECESSARY.
SUPPORTS:
CONTRACTOR.
RECOMMENDED.
SUPPORT NOTES
PARTICIPANTS.
01/12/2017
DR400FINALV4
ALL ITEMS LISTED BELOW ARE TO BE PROVIDED/DESIGNED BY THE ARCHITECT. THE ARCHITECT SHALL
4. TYPICALLY THE MINIMUM RECOMMENDED FINISHED CEILING HEIGHT IS 8'-4" A.F.F. UNLESS STATED
3. THE FLOOR IN THE X-RAY AND DARKROOM MUST BE ESSENTIALLY LEVEL.TOLERANCE IS 1/8" OVER A
2. IT IS RECOMMENDED THAT ALL DIAGNOSTIC ROOMS SHOULD HAVE TILE OR VINYL FLOORING.
CONTROLS SHOULD BE LOCATED NEAR THE ROOM ENTRANCES AND IN THE CONTROL AREAS.
THE ROOM. THE CANNED LIGHT FIXTURES SHOULD BE PROVIDED WITH DIMMER CONTROLS. THE DIMMER WELL AS FLORESCENT LIGHT FIXTURES. THE CANNED LIGHT FIXTURES ARE TO BE DISTRIBUTED AROUND
1. ALL DIAGNOSTIC ROOMS INCLUDING X-RAY, MAMMO, R/F ETC... SHALL HAVE CANNED LIGHT FIXTURES AS
SUPPLEMENTAL NOTES
OTHERWISE IN THIS DRAWING SET.
CONSULT WITH THE CUSTOMER FOR ANY ADDITIONAL INFORMATION.
10'-0" SPAN.
TYPICAL ROOM LAYOUT
(NOT SITE SPECIFIC)
1/4"=1'-0"
SCALE:
DAVE HENRY
FILE NAME:
DRAWN BY:
DATE:
SYSTEM. AREA UNDER COUNTER MUST REMAIN CLEAR FOR TRIPLET BOX. WORKING DRAWINGS BY ARCHITECT.
PHYSICIST REGARDING SHIELDING REQUIREMENTS. CONSULT CUSTOMER FOR WINDOW SIZE. WORKING
CONSTRUCT CONTROL BOOTH PARTITION FROM FLOOR TO CEILING WHERE SHOWN. PROVIDE ONE LEADED
PROVIDE ONE 2"T x 16"H x 24"W WOOD SUPPORT BOARD HIDDEN IN WALL WHERE SHOWN. LOCATE BOARD
PROVIDE A 40" HIGH STANDING COUNTER WITH ONE 3" DIA. CABLE ACCESS HOLE THROUGH COUNTER WHERE
SHOWN. BRACE COUNTER TO SUPPORT 300 POUNDS STATIC LOAD, COUNTER WILL BE USED TO SUPPORT DR
WINDOW AND FRAME. LOCATE WINDOW CENTER LINE 5'-0" A.F.F. CONSULT WITH A CERTIFIED RADIOLOGICAL
CENTER LINE 84" A.F.F. BOARD SHALL SUPPORT 300 POUNDS PULL FORCE. REFER TO SUPPORT NOTES.
PAINT BOARD THE SAME COLOR AS FINISHED WALL.
CONSTRUCTION NOTES
(CONTRACTOR FURNISHED - CONTRACTOR INSTALLED)
DRAWINGS BY ARCHITECT.
EQUIPMENT VENDOR IS NOT RESPONSIBLE FOR DAMAGES. CONSULT WITH EQUIPMENT VENDOR FOR
PROVIDE WALL AND FLOOR PROTECTION FOR INSTALLATION AND DELIVERY OF X-RAY EQUIPMENT. THE
FURTHER INFORMATION.
01/12/2017
DR400FINALV4
X-RAY TABLE BASE AND WALL STAND ANCHOR HOLE LOCATION. REFER TO DETAIL SHEET <D2> FOR FURTHER
INFORMATION. THE FLOOR MUST SUPPORT THE TABLE WEIGHT AND MAXIMUM PATIENT WEIGHT.
TYPICAL ROOM LAYOUT
(NOT SITE SPECIFIC)
'-6
'-0
'-6
7'-7 3/8"
C
C
16'-0"
5'
-0
5"
'-0
'-5
/1
6"
2'-0"
'-1
/1
6"
'-0
3'- 0"
'-0
'-0
45°
Ø 3"
C
3'-0"
3'-6"
THE FLOOR IN THE X-RAY ROOM MUST BE LEVEL TO WITHIN ±1/8" OVER A 10'-0" SPAN. THE MINIMUM
RECOMMENDED FINISHED CEILING HEIGHT IS 8'-4" A.F.F.
JUNCTION BOX <F>. CONSULT WITH EQUIPMENT INSTALLER FOR EXACT SIZE AND LOCATION.
PROVIDE SMALL SHELF ON CONTROL BOOTH WALL TO SUPPORT THE SCU. SHELF TO BE LOCATED UNDER
DRAWN BY:
DATE: SHEET:
1/4"=1'-0"
SCALE:
01/12/2017
DR400FINALV4
*AGFA DR 400
86.60 47.44 89.57 1389
31.46 9.37 88.39 346
23.50 17.50 14.50 29
TUBESTAND LEFT HAND LOAD VERSION SHOWN
AGFA DR 400
AGFA DR 400
VERTICAL BUCKY STAND
STANDARD
TRIPLET BOX
CONTAINS PC, UPS AND DR SYSTEM COMPONENTS
ELEVATING 4-WAY FLOAT TOP INTEGRATED TABLE
INTEGRATED INSIDE TABLE
W x D x HW x D x H
WEIGHT (LBS.)WEIGHT (LBS.)
DIMENSIONS (IN.)DIMENSIONS (IN.)
EQUIPMENT SCHEDULEEQUIPMENT SCHEDULE
EQUIPMENT IDENTIFICATION NUMBEREQUIPMENT IDENTIFICATION NUMBER
AGFA DR 400
(LOCATED IN TABLE BASE)
BTU / HOURBTU / HOUR
DESCRIPTIONDESCRIPTION
SHF GENERATOR CABINET
AGFA DR 400
MONITOR, KEYBOARD AND MOUSE
GENERATOR CONTROL AND DR WORKSTATION PC TYPICAL
'-6
C
7'-7 3/8"
C
40" S.I.D.
72" S.I.D.
TYPICAL ROOM LAYOUT
(NOT SITE SPECIFIC)
50° to 95° (F)
30% to 90% non condensing
Above specification apply to all areas where the Radiographic
(Control area, Procedure room and Equipment room if required.)
equipment is to be located.
Ambient Operation Temperature:
Humidity:
*AGFA DR 400
TUBESTAND RIGHT HAND LOAD VERSION AVAILABLE.
ELEVATING 4-WAY FLOAT TOP INTEGRATED TABLE
NEC CLEARANCE
FOR DISCONNECT <A>.
EC SHALL MAINTAIN
NOTE: EC SHALL ENSURE THAT LUGS ARE LARGE ENOUGH TO FACILITATE OVERSIZED INCOMING CABLES
REQUIRED BY X-RAY EQUIPMENT. REFER TO "INCOMING WIRE SIZE" NOTES.
FLUSH MOUNT IN WALL WHERE SHOWN. LOCATE ENCLOSURE CENTER LINE 5'-6" A.F.F.
ENCLOSURE SHALL CONTAIN ONE XXAMP, X-POLE THERMOMAGNETIC CIRCUIT BREAKER WITH SHUNT TRIP DEVICE.
NEMA 1, XXX VAC, XXX PHASE, FRONT OPERATED, CIRCUIT BREAKER ENCLOSURE WITH GROUND ACCESSORIES.
FACILITY TO DETERMINE SOURCE LOCATION.
DAVE HENRY
DATE:
FILE NAME:
DRAWN BY:
SHEET:
1/4"=1'-0"
SCALE:
01/12/2017
DR400FINALV4
TYPICAL ROOM LAYOUT
(NOT SITE SPECIFIC)
TE
5'-0" 11'-0"
C
'-0
'-6
'-6
'-0
2'-0"2'-0"3'-4"
8"8"8"10"8"
C C C C
C
C
C C C
( CONTRACTOR FURNISHED - CONTRACTOR INSTALLED )
EC SHALL PROVIDE ONE NORMALLY OPEN SWITCH WITH RED MUSHROOM HEAD AND AN "EMERGENCY OFF" NAME PLATE.SWITCH WILL BE USED FOR SHUNT TRIP DEVICE LOCATED ON CIRCUIT BREAKER <A>.
( CONTRACTOR FURNISHED - CONTRACTOR INSTALLED )
DOOR IS OPEN. (INSTALL ONLY IF REQUIRED BY PREVAILING CODE OR REQUESTED BY CUSTOMER).
EC SHALL FURNISH AND INSTALL ONE FACILITY GRADE, THREE PRONG, 120VAC, 15AMP, DUPLEX RECEPTACLE.
OF CONNECTOR (e.g.. RJ-45/CAT-5, FIBER OPTICS, BNC/COAX, ETC.). REFER TO DICOM NOTES.
NOTE: ETHERNET CONNECTION SHOULD HAVE INTERNET ACCESS WITH VPN CAPABILITY.
THE TELEPHONE NUMBER MUST BE PROVIDED TO THE TECHNICIAN WHO INSTALLS THE SYSTEM.
ELECTRICAL TAG DESCRIPTIONS
FLUSH MOUNT IN WALL WHERE SHOWN 12" A.F.F. PROVIDES POWER FOR THE DR SYSTEM COMPONENTS.
FLUSH MOUNT IN WALL WHERE SHOWN AT NORMAL HEIGHT. PROVIDES POWER FOR SERVICE AND HOUSE
EC SHALL FURNISH AND INSTALL TWO FACILITY GRADE, THREE PRONG, 120VAC, 20AMP, QUAD RECEPTACLE.
BETWEEN THE FACILITY, THE CONSTRUCTION PARTICIPANTS AND THE EQUIPMENT INSTALLERS. REFER TO WIRING LOCATION TO BE DETERMINED BY THE FACILITY. THE X-RAY ON WARNING LAMP CIRCUIT SHALL BE COORDINATED 4" x 4" x 2" JUNCTION BOX FOR THE "X-RAY IN USE" WARNING SIGN. FLUSH MOUNT IN WALL OR CEILING. EXACT
4" x 2" x 2" DEEP SWITCH BOX. FLUSH MOUNT IN WALL WHERE SHOWN. LOCATE BOX CENTER LINE AT 6'-0" A.F.F.
6" x 6" x 4" DEEP JUNCTION BOX. FLUSH MOUNT IN WALL WHERE SHOWN. LOCATE BOX CENTER LINE 6" A.F.F.
8" x 8" x 4" DEEP JUNCTION BOX. FLUSH MOUNT IN WALL WHERE SHOWN. LOCATE BOX CENTER LINE 12" A.F.F.
SHALL HAVE NORMALLY OPEN CONTACTS RATED AT 120VAC, 15AMPS. SWITCH SHOULD OPEN CONTACTS WHEN PROVIDE ONE 1-POLE, SINGLE THROW, MOMENTARY, EXPOSURE INTERLOCK SWITCH IN DOOR FRAME. SWITCH
PROVIDE A 4-PORT ETHERNET CONNECTION WHERE SHOWN. EC SHALL CONSULT WITH END USER REGARDING TYPE
PROVIDE ALL HARDWARE NECESSARY INCLUDING COVER PLATES, CONNECTORS AND INTERCONNECTING CABLES.
(TELEPHONE LINE WHICH DOES NOT RUN THROUGH ANY FACILITY PBX, SWITCHBOARD OR A SHARED ACCESS LINE.
PERFORMED FROM A REMOTE LOCATION. THE RJ11 LINE MUST BE A DEDICATED DIRECT IN DIAL (DID) VOICE-GRADE PROVIDE A SINGLE TELEPHONE LINE WHERE SHOWN FOR CERTAIN SERVICE AND MAINTENANCE ACTIVITIES TO BE
AT THE TIME OF EQUIPMENT INSTALLATION.
CABLE ACCESS FOR THE CONTROL CONSOLE.
CABLE ACCESS FOR THE TABLE.
CABLE ACCESS FOR THE WALL STAND.
PROVIDE A COVER PLATE AS SHOWN IN COVER PLATE DETAILS.
PROVIDE A BLANK COVER PLATE ON FLOOR TRENCH. EQUIPMENT INSTALLER TO PROVIDE CABLE ACCESS HOLE
KEEPING. REQUIRES HOUSE POWER, LOCATION NOT CRITICAL.
DEDICATED OUTLET, POWER CAN BE SUPPLIED FROM <A1>.
PROVIDE A COVER PLATE AS SHOWN IN COVER PLATE DETAILS.
E
T
W
ITEM
DIAGRAM.
F
N
G
D
C
B
DESCRIPTION
A
ITEM
TERMINAL SCHEDULE
DESCRIPTION
10" x 3 1/2" DEEP FLUSH MOUNTED VERTICAL RISER. FLUSH MOUNT IN WALL AT FLOOR LEVEL AND TERMINATE ABOVE THE FINISHED CEILING TO ALLOW FOR CONDUIT CONNECTIONS. PROVIDE DIVIDERS FOR LOW AND HIGH
VOLTAGE. REFER TO TROUGH NOTES AND TROUGH ELEVATION.
10" x 3 1/2" DEEP FLUSH MOUNTED FLOOR TRENCH. FLUSH MOUNT IN FLOOR WHERE SHOWN AND CONNECT
TO VERTICAL RISER. PROVIDE 3 COMPARTMENTS FOR LOW AND HIGH VOLTAGE.
REFER TO TROUGH NOTES AND TROUGH ELEVATION.
4" x 4" x 4" DEEP JUNCTION BOX. FLUSH MOUNT IN WALL WHERE SHOWN. LOCATE BOX ABOVE WINDOW OR
CABLE ACCESS FOR THE SCU. NOTE: BOX OPENS OUT INTO X-RAY ROOM NOT INTO CONTROL AREA.
CENTER LINE 88" A.F.F. PROVIDE A BLANK OVERSIZE COVER PLATE.
FLUSH MOUNT IN WALL WHERE SHOWN 88" A.F.F. NEXT TO JUNCTION BOX <F>. PROVIDES POWER TO SCU.
EC SHALL FURNISH AND INSTALL ONE FACILITY GRADE, THREE PRONG, 120VAC, 15AMP, SINGLE RECEPTACLE.
DEDICATED OUTLET, REQUIRES HOUSE POWER.
4'-10"
NONE
SCALE:
POWER SOURCE
MAIN
INCOMING
ELECTRICAL BLOCK DIAGRAM
G
F
D
W
RUN #
INCOMING POWER
JUNCTION BOXES/COVERS
CONDUIT CONNECTIONS
CABLE ACCESS POINTS
LEGEND
OR RACEWAY
TROUGH (WALL DUCT)
N
C
A B
LEAVE PULL STRINGS.
2-#14AWG PLUS GROUND.
FOR SHUNT TRIP SOLENOID.
MAXIMUM CONDUIT LENGTH IS 24'-0".
TAILS IN (G). EC SHALL IDENTIFY WIRES AS 'X-RAY WARNING SIGN' 'L', 'N', 'G' .
CIRCUIT PER EQUIPMENT SUPPLIER RECOMMENDATIONS AND LEAVE 10'-0"
2-#14AWG PLUS GROUND. EC TO CONNECT "X-RAY IN USE" WARNING LIGHT
EC SHALL CONNECT TO (A) AND (B). PROVIDE POWER IF NECESSARY
EC SHALL CONNECT TO (N) AND LEAVE 10'-0" TAILS AT <G>.
IDENTIFY WIRES AS DOOR INTERLOCK SWITCH.
2-#14AWG PLUS GROUND IF REQUIRED.1/2"N
F
VERTICAL
RISER CONDUIT
C VERTICAL
D
W
B A
VERTICAL
RISER
RISER
RISER
VERTICAL
CONDUIT
1/2"
CONDUIT
1/2"
DAVE HENRY
DATE:
DRAWN BY:
FILE NAME:
( CONTRACTOR FURNISHED - CONTRACTOR INSTALLED )
CONDUIT & WIRING SCHEDULE
FROM
MAIN
POWER
A
TO
A
SIZE
VARIES
VARIES
REMARKS
1 1/2"
CONDUIT
VERTICAL RISER
(WALL DUCT)
VERTICAL
RISER
2 1/2"
CONDUIT
D
2 1/2"
CONDUIT
FLOOR TRENCH
LEAVE PULL STRINGS.
MAXIMUM CONDUIT LENGTH IS 24'-0".
LEAVE PULL STRINGS.
SOURCE
GENERATOR
KW RATING
RECOMMENDED
LINE VOLTAGE
SUPPLY TRANSFORMER
KVA RATING
MAXIMUM MOMENTARY
CURRENT
RECOMMENDED MAIN
CIRCUIT BREAKER <A>
MAXIMUM WIRE LINE
IMPEDANCE
HFe 601
65KW
480VAC/3-PHASE/60Hz
VOLTAGE REGULATION
82KVA
THREE HOT WIRES ONE NEUTRAL AND ONE EARTH GROUND
TRANSFORMER: Y WITH BALANCED NEUTRAL
UNDER LOAD ±5%.
EC SHALL CALCULATE INCOMING WIRE SIZE FROM BUILDING SOURCE (TRANSFORMER)
TO MAIN DISCONNECT <A> WITH VOLTAGE REGULATION OF ±5% AT MAXIMUM LOAD.
EC SHALL CALCULATE INCOMING WIRE SIZE BASED OFF OF INFORMATION
CONDUIT SIZE DEPENDS ON INCOMING WIRE SIZE.
ENSURE CONNECTION COMPLIES WITH ALL STATE AND LOCAL CODES.
EC TO CONNECT AT <A> AND AT GENERATOR ITEM <2> SHEET B1.
WIRE SIZE TO BE DETERMINED BY CALCULATED 50' WIRE LENGTH.
PROVIDED IN CHART AND LENGTH OF WIRE AS REQUIRED AT SITE.
HFe 401
40KW
480VAC/3-PHASE/60Hz
50KVA
UNDER LOAD ±5%.
VOLTAGE REGULATION
HFe 501
50KW
480VAC/3-PHASE/60Hz
60KVA
UNDER LOAD ±5%.
VOLTAGE REGULATION
HFe 801
80KW
480VAC/3-PHASE/60Hz
VOLTAGE REGULATION
100KVA
UNDER LOAD ±5%.
01/12/2017
DR400FINALV4
DATE:
DRAWN BY:
SHEET:
SCALE:
01/12/2017
DR400FINALV4
01/12/2017
DR400FINALV4
TYPICAL ROOM LAYOUT
(NOT SITE SPECIFIC)
LINE (TYPICAL).
FINISHED FLOOR
LINE (TYPICAL).
FINISHED CEILING
'-0
6" C
'-0
C
'-0
" C
'-6
C
'-0
C
'-0
"C TE
'-0
" C
CONTROL BOOTH WALL SHOWN UNFOLDED
FOLD FOLD
COUNTER
1/4"=1'-0"
'-4
C
DATE:
DRAWN BY:
FILE NAME:
SHEET:
NONE
SCALE:
01/12/2017
DR400FINALV4
01/12/2017
DR400FINALV4
TYPICAL ROOM LAYOUT
(NOT SITE SPECIFIC)
SHORTEST ROUTE POSSIBLE. NOTE: DO NOT FOLLOW WALL LINE
NOTE TO ELECTRICAL CONTRACTOR: THE CONDUIT ROUTING
IS SHOWN AS A VISUAL REFERENCE ONLY (NOT TO SCALE). THE
UNLESS IT IS THE SHORTEST ROUTE POSSIBLE.
EC SHALL ROUTE CONDUITS FROM POINT TO POINT USING THE
POWER SOURCE
MAIN
INCOMING
FOR APPROPRIATE PART NUMBERS.
PLACEMENT AND TROUGH CLEARANCES.
10" X 3 1/2" FLUSH MOUNTED WALL
5'-0" LONG DIVIDER STRIPS FOR
10" X 3 1/2" DUCT (3 COMPARTMENTS).
10" X 3 1/2" END CLOSURES.
DUCT (5'-0" LENGTHS, INCLUDES COVERS)
WALL DUCT PARTS LIST
CONDITIONS REQUIRE SURFACE MOUNTED DUCT THEN CONTACT THE ELECTRICAL SUPPLIER
THE PROJECT REQUIREMENTS. PLEASE HAVE THE DRAWING FILE NAME (BOTTOM RIGHT CORNER
A COPY OF THESE PLANS HAVE BEEN FORWARDED TO THE SUPPLIER TO ENSURE ACCURACY OF
ON PRICING AND AVAILABILITY. SUFFICIENT LEAD TIME MUST BE GIVEN ON ALL TROUGH ORDERS.
NOTE: CONTACT CHRIS NAJDA (440) 835-3500 (ELECTRICAL SALES CO.) FOR FURTHER INFORMATION
NOTE: PROVIDE TWO DIVIDER STRIPS WITHIN TROUGH FOR THE SEPARATION OF HIGH TENSION
CROSSOVER TUNNELS BEHIND COVER PLATES AND MULTIPLE CONDUIT CONNECTION POINTS.
X-RAY CABLES, POWER AND LOW VOLTAGE CABLES WHERE REQUIRED. PROVIDE APPROPRIATE
CLEAR FOR REMOVAL AND INSERTION OF PRE TERMINATED CABLES.
NOTE: ALL TROUGH SHALL BE FLUSH WITH FINISHED SURFACES UNLESS OTHERWISE SPECIFIED
AND PROVIDED WITH REMOVABLE, OVERSIZED, COVERS THE FULL LENGTH. COVERS SHALL BE
NOTE: ALL PART NUMBERS LISTED ARE FOR FLUSH MOUNTED APPLICATIONS, IF THE FIELD
NOTE: IF IT IS NECESSARY THAT THE TROUGH BE SURFACE MOUNTED, WHEN NOT INDICATED ON
THESE PLANNING DRAWINGS, ALL EQUIPMENT PLACEMENT AND TRAVEL CLEARANCE MUST BE
VERIFIED TO PREVENT EQUIPMENT AND TROUGH INTERFERENCE AND POSSIBLE EQUIPMENT
DAMAGE. CONTACT GAMMA TECHNOLOGIES, INC. FOR PROPER VERIFICATION OF EQUIPMENT
COUPLERS DO NOT HAVE TO BE ORDERED UNLESS CUSTOMER WILL BE MAKING A LARGE
NOTE: COUPLERS ARE INCLUDED WITH ALL WALL DUCT STRAIGHTS AND FITTINGS. ADDITIONAL
JUNCTION BOX COVERPLATES
6"
(C)
COVERPLATE
6"
2" DIAMETER CHASE
NIPPLE WITH PLASTIC
(CENTERED)
BUSHING AND LOCKNUT.
OF PLAN).
NUMBER OF FIELD CUTS.
PART# ESW3560D
PART NUMBER
PART# ESW103560F
PART# ESW1035ECF
CONNECTION POINTS.
TYPICAL CONDUIT CONNECTIONS
LOCATED ABOVE FINISHED CEILING.
EC TO DETERMINE ACTUAL CONDUIT
ALL FLUSH MOUNTED DEVICES SHALL HAVE OVERSIZED COVERPLATES.
NOTE: COVERPLATES DEPICTED ARE FOR SURFACE MOUNTED DEVICES.
COVERPLATE DETAILS AND TROUGH NOTES
BUSHING AND LOCKNUT.
8"
(D)
COVERPLATE
8"
2" DIAMETER CHASE
NIPPLE WITH PLASTIC
(CENTERED)
DESCRIPTION QTY
'-0
W
A L
L D
U C
T S
E C
T
IO
N C
U T
T O
L E
N G
T H
FINISHED FLOOR
(WALL DUCT)
VERTICAL RISER
MOUNTED TROUGH, HIDES TROUGH AND
DRYWALL JUNCTION. COVERPLATES TO
PRE TERMINATED EQUIPMENT CABLES.
OVERSIZE COVERPLATES FOR FLUSH
REMAIN UNOBSTRUCTED FOR EASY
REMOVAL. ALLOWS INSTALLATION OF
PART# ESW103560F
FINISHED CEILING
'-0
W
A L
L D
U C
T S
E C
T
IO
N
NOT TO SCALE
PART# ESW3560D
DIVIDERS FOR LOW
AND HIGH VOLTAGE.
TYPICAL TROUGH ELEVATION
TO EQUIPMENT INSTALLER FOR CABLE INSTALLATION.
CONDUIT CONNECTION POINTS MUST BE ACCESSIBLE
END CAP PART# ESW1035ECF
SHEET:
NONE
SCALE:
DAVE HENRY
DATE:
DRAWN BY:
FILE NAME:
01/12/2017
DR400FINALV4
01/12/2017
DR400FINALV4
NOTE: CONTACT CHRIS NAJDA (440) 835-3500 (ELECTRICAL SALES CO.) FOR FURTHER INFORMATION.
TYPICAL ROOM LAYOUT
(NOT SITE SPECIFIC)
ADAPTER
TRENCH TO WALL DUCT
PART# ESTWA1035
3 COMPARTMENTS
TRENCH DUCT
PART# EST1035C72
END CAP
TRENCH DUCT
PART# EST1035EC
TRENCH TO WALL DUCT
ADAPTERPART# ESTWA1035 1
10" X 3 1/2" FLUSH MOUNTED FLOOR TRENCH
DUCT, 3-COMPARTMENTS. (6'-0" LENGTHS)PART# EST1035C72 1
TRENCH DUCT
END CAPPART# EST1035EC 2
4'-0" CUT TO LENGTH
PART# EST10C12
10" X 3 1/2" FLUSH MOUNTED FLOOR TRENCH
12" LONG SECTIONAL COVER PLATE. 1
COVER PLATE
12" SECTIONAL
PART# EST10C12
TRENCH DUCT
12"
THE MAIN CONSIDERATION WHEN LEAVING WIRES IN JUNCTION BOXES IS TO ENSURE THEY ARE NOT
LIVE AND THAT THEIR ASSOCIATED SWITCHING MEANS ARE LOCKED OUT (OR EQUIVALENT).
THIS STANDARD SHALL BE APPLIED TO ALL OPEN ENDED WIRE AND CABLE MENTIONED IN THIS DRAWING
SET. OBVIOUSLY, GROUND WIRE OR CABLE BEING THE EXCEPTION.
IT IS THE ELECTRICAL CONTRACTOR'S RESPONSIBILITY TO ENSURE THAT WIRES ARE NOT ENERGIZED
UNTIL THE EQUIPMENT INSTALLERS ARE READY FOR TESTING.
THE ELECTRICAL CONTRACTOR SHALL INFORM THE EQUIPMENT INSTALLER(S) AS TO THE LOCATION AND MEANS OF ACTIVATING ALL SWITCHING DEVICES ASSOCIATED WITH THE OPEN ENDED WIRE AND CABLE
MENTIONED IN THIS DRAWING SET.
TO (J) GENERATOR CABINET.
WARNING LIGHT WIRING DIAGRAM DETAIL.
WIRES SHALL NOT BE LIVE. REFER TO
X-RAY IN USE WARNING SIGN NOTES:
MAXIMUM 20 WATT BULBS RECOMMENDED FOR WARNING LIGHT.
CONTRACTOR SHALL CONSULT WITH FACILITY REGARDING EXACT WORDING, STYLE AND LOCATION.
OPTIONAL X-RAY IN USE WARNING SIGN TO BE INSTALLED ONLY IF REQUIRED BY PREVAILING CODE OR
REQUESTED BY CUSTOMER.
DOOR SWITCH NOTES:
OPTIONAL DOOR SWITCH TO BE SQUARE D "CLASS 9007, TYPE A0" OR EQUIVALENT TO BE LOCATED ON HINGE
SIDE OF DOOR JAMB.
OPTIONAL DOOR SWITCH TO BE INSTALLED ONLY IF REQUIRED BY PREVAILING CODE OR REQUESTED BY
CUSTOMER.
IN CASE OF A DOUBLE JAMB, ONE DOOR SWITCH SHOULD BE LOCATED ON EACH SIDE. SEE TROUGH /
RACEWAY LAYOUT FOR DOOR SWITCH / WARNING LIGHT LOCATION IN ROOM.
SEE POINT TO POINT WIRING DESCRIPTION FOR WIRE SIZES TO BE USED IN CONDUITS.
USING 'W' AS A JUNCTION BOX FOR THE DOOR INTERLOCK SWITCH WIRES ELLIMINATES AN ADDITIONAL CONDUIT RUN TO THE SYSTEM CABINET PULL BOX. ALTHOUGH THIS DECREASES CONSTRUCTION COSTS, IT
IS NOT REQUIRED.
WARNING SIGN NOTES:
OPTIONAL WALL SWITCH (SW) MAY BE CONTRACTOR INSTALLED AT THE REQUEST OF THE CUSTOMER OR EQUIPMENT SUPPLIER. PURPOSE OF WALL SWITCH IS TO REMOVE ALL VOLTAGES FROM GENERATOR CABINET DURING SERVICE. EC SHALL CONSULT WITH CUSTOMER AND EQUIPMENT SUPPLIER REGARDING
WALL SWITCH INSTALLATION, TYPE AND EXACT LOCATION.
DOOR SWITCH / WARNING LIGHT DETAIL
JB
X-ray On Warning Sign Circuit Options
Housing
c. Aluminum, etc.
Note: Recommend 20 watts or less
a. 120vac * a. red on white * Facility panel * Rotor prep *
b. 277vac b. white on red Systems supply System ON
c. 120/277vac c. black on white Always ON
d. 26vac d. white on black
e. DC e. red on black, etc.
Refer to detail number
The engineer of record shall consult with the facility and equipment supplier to coordinate the x-ray on warning sign circuit combinations. Majority picks marked by an asterisk.
a. White * LED Single * Wall * Surface * X-RAY IN USE *
b. Black LED Double Ceiling Flush X-RAY ON, etc.
Refer to detail number WLD01.0310 for typical x-ray on warning light circuit.
Example "X-RAY IN USE" Warning Sign Circuit Options
Many potential factors influence choice of x-ray on warning sign. E.g. Matching existing; efficiency; aesthetics; price;
quality; number of options; etc.
Gamma Tech never knows the exact warning sign type, style, lamp voltage, power source, method or means of control prior to drawing set issuance. For this reason more than one detail may be depicted in this drawing set.
It is our hope and intent to encourage the construction participants and the equipment suppliers / installers to coordinate efforts. For this reason we have listed some potential options and their caveats below. (not to be construed as all inclusive). Important decisions to be shared are numbered. Sample options are listed alphabetically. For your convenience, the majority pick is designated by the word 'typical' in parentheses.
The facility or their designee selects an x-ray on warning sign.
1) Lamp voltage? - usually one of four options.
a. 120vac (typical)
b. 120/277vac
c. 277vac
d. Low voltage ac
e. Low voltage dc (battery) Note: The system installer should be informed of the lamp voltage as soon as possible.
The facility determines at which point in time the warning sign will illuminate.
2) When to illuminate? - Engineer to coordinat with the facility and equipment supplier.
a. Warning lamp turns on upon rotor prep / turns off upon x-ray termination. (typical)
b. Warning lamp illuminates / extinguishes when equipment is switched on / off.
c. Warning lamp illuminates 24/7.
Note: the equipment being installed may or may not be capable of interfacing with a warning light or may have just one option. The equipment supplier shall coordinate with the facility and the construction participants regarding interface options.
Voltage
DAVID HENRY
warning light circuit.for typical WLD02.0696
FACILITY:
120VAC TO 24VAC
FILE NAME:
DATE:
DRAWN BY:
SHEET:
SCALE:
Faces
Power source
Lamp
Panel color
Surface
ON / OFF
Mounting Wording
TO SUPPLY POWER TO (W).
LAMP VOLTAGE CAN VARY.
REFER TO ARCHITECT'S
DRAWINGS REGARDING HOW
( CONTRACTOR FURNISHED - CONTRACTOR INSTALLED )
1/2" CONDUIT
SW
N
W
WALL SWITCH
TO BE LABELED
"WARNING LIGHT"
OPTIONAL
1/2" CONDUIT
JB
1/2" CONDUIT
FLOOR LINE
CEILING LINE
A typical 'X-RAY IN USE' warning sign circuit is depicted below.
Due to many varibles, the x-ray on warning sign circuit must be coordinated between the equipment supplier, electrical engineer and the facility.
The electrical contractor shall consult with the facility and the equipment supplier prior to installing any x-ray on warning sign.
TYPICAL WARNING LIGHT CIRCUIT
X-RAY ON WARNING LIGHT
N
L
G
20A
120VAC
POWER
IN USE
X-RAY
N L
G
W
JUNCTION BOX
In the example above, 120vac house power is supplied to the warning light.
The circuit is not complete until the anode, in the x-ray tube, begins to turn.
When the exposure button is pressed, the rotor prep signal is activated.
At this time the "x-ray in use" light illuminates.
It turns off when x-ray generation is terminated.
OUT
IN
G
ROTOR PREP
DETAIL NOT TO SCALE
CONTACTS
RELAY
WITH N.O.
CABINET
SYSTEMS
WLD01.0214
G
NONE
DR400FINALV4
01/12/2017
TYPICAL ROOM LAYOUT
(NOT SITE SPECIFIC)
CB PANEL
POWER BY EC
DATE:
DRAWN BY:
FILE NAME:
SHEET:
NONE
SCALE:
ELECTRICAL CONTRACTOR SHALL FURNISH AND INSTALL ALL CONDUITS, WIRE, WIREWAYS,
DISTRIBUTION TRANSFORMERS, JUNCTION BOXES, ENCLOSURES, CIRCUIT BREAKERS, ETC.
ALL WIRES TO BE THHN TYPE, STRANDED COPPER UNLESS OTHERWISE SPECIFIED. THE
ELECTRICAL CONTRACTOR SHALL SUPPLY LUG PAIRS OF #8 OR GREATER FOR ALL CABLES
CONDUIT AND DUCT RUNS ARE SHOWN SCHEMATICALLY. ACTUAL BUILDING CONDITIONS WILL
DETERMINE CONDUIT AND DUCT ROUTES. CONDUIT TURNS TO HAVE LARGE SWEEPING BENDS
WITH MINIMUM RADIUS AS SPECIFIED IN NEC ART. 346-10. IF LARGE BENDS IN CONDUIT ARE
NOT POSSIBLE, USE SERVICE ENTRANCE ELBOW CONNECTORS TO AVOID DAMAGE TO CABLE.
ELECTRICAL CONTRACTOR TO PROVIDE 10'-0" TAILS FOR ALL WIRES UNLESS OTHERWISE
SPECIFIED AND IDENTIFY BOTH ENDS OF ALL WIRES. ELECTRICAL CONTRACTOR SHALL MAKE
EVERY EFFORT TO ENSURE THAT POINT TO POINT CONDUIT RUNS ARE AS SHORT AS FIELD
WORK SHALL COMPLY WITH REGULATORY AGENCIES OF JURISDICTION. ALL SUCH AGENCIES
SUPERSEDE THESE DRAWINGS AND MUST BE ADHERED TO. THE ELECTRICAL CONTRACTOR
SHALL BE RESPONSIBLE FOR SCRUTINIZING THESE DRAWINGS AND SHALL MAKE CORRECTIONS
NOTE: CONDUITS MAY BE INCREASED IN SIZE BUT NEVER DECREASED WITHOUT CONSENT
RESPONSIBLE FOR INSTALLING THE PATIENT GROUNDING NETWORK PER NATIONAL ELECTRIC
ALL CONDUIT SHALL BE RIGID TYPE UNLESS SPECIFIED OTHERWISE IN THIS DRAWING SET. ALL
Supplying and installing network connections between raceways for containment and routing of networking cables, as determined by site to, the customer's responsibility is the purchase and installation of conduits and / or transceivers, cabling, repeaters, routers, modems, etc. which are consistent with the responsibility. Customer must specify, supply and install proper receptacles, AUI diagnostic imaging equipment and other node connections shall be the customer's
DICOM, PACS, LANS/WANS, Ethernet, Modems site needs and sound network implementation practices. Also included, but not limited
ALL FLUSH MOUNTED ELECTRICAL BOXES AND WIRE TROUGHS SHALL HAVE OVERSIZED COVER
PLATES. ALL FLUSH MOUNTED ELECTRICAL BOXES AND WIRE TROUGHS EXPOSED TO A WET
THE ELECTRICAL CONTRACTOR SHALL INSTALL, PROVIDE POWER TO, AND CONNECT THE
DOOR INTERLOCK SWITCH, EMERGENCY OFF SWITCH AND120VAC DUPLEX OUTLETS FOR
THE ELECTRICAL CONTRACTOR SHALL COORDINATE WITH THE FACILITY AND THE EQUIPMENT
SUPPLIER / INSTALLERS REGARDING SPECIFIC INSTALLATION DETAILS. REFER TO ASSOCIATED
FROM THE EQUIPMENT REPRESENTATIVES. ELECTRICAL CONTRACTOR SHALL BE
CODE ARTICLE 517 AND NFPA 70.
CONDUIT RUNS EXPOSED TO WET ENVIRONMENTS SHALL BE MADE LIQUID TIGHT (e.g.
Networking Notes:
BLOCK DIAGRAMS AND DETAILS.
conditions or required by prevailing codes.
ENVIRONMENT SHALL BE MADE LIQUID TIGHT (e.g. IN FLOOR, ETC.).
FOLLOWING, IF SPECIFIED FOR THIS PROJECT.
SERVICE AND HOUSEKEEPING DESCRIBED IN 'GENERAL NOTES' ITEM #6.
UNDERGROUND, ETC.).
LISTED WITHIN THE SCOPE OF THIS WORK.
WIRING NOTES
LEAVE PULL WIRE IN EMPTY CONDUITS.
ACCORDING TO PREVAILING CODES.
CONDITIONS ALLOW. NEVER LEAVE 'HOT' WIRES CURLED IN PULL BOXES.
IF APPLICABLE, INSTALL THE EQUIPMENT GROUNDING (BONDING) CONDUCTOR (SIZED PER TABLE 250.122) IN A METAL RACEWAY SYSTEM OR A CABLE ARMOR OR SHEATH ASSEMBLY, EACH OF WHICH MUST QUALIFY
CODES. THIS GROUNDING SYSTEM IS REQUIRED IN AREAS WHERE ELECTRICALLY SUSCEPTIBLE PATIENTS MIGHT BE EXAMINED OR TREATED UNDER PRESENT, FUTURE OR EMERGENCY CONDITIONS. CONSULT WITH
SYSTEM. IF SUCH A GROUNDING SYSTEM IS REQUIRED, THE RESPONSIBLE CONSTRUCTION PARTICIPANTS
IF APPLICABLE, USE AN INSULATED COPPER EQUIPMENT GROUNDING (BONDING) CONDUCTOR TO GROUND
- ISOLATED GROUND (IG) RECEPTACLES, IF APPLICABLE.
REQUIREMENTS OR SCHEMATIC.
INTO THEIR WORKING DRAWINGS.
COURTESY TO OUR CUSTOMERS.
THE ELECTRICAL SPECIFICATIONS IN THESE PLANS ARE DEDICATED TO THE
ALL ADDITIONAL EQUIPMENT SUPPLIERS SHALL NOT USE ANY ELECTRICAL
DEVICES CALLED OUT IN THESE PLANS AS A PART OF THEIR ELECTRICAL
ALL ADDITIONAL EQUIPMENT SUPPLIERS SHALL ISSUE A SEPARATE DETAILED
SPECIFICATIONS PACKAGE TO THE ARCHITECT OR TO THE CUSTOMER. THE
ADDITIONAL EQUIPMENT DEPICTED IN THESE PLANS HAVE BEEN SHOWN AS A
ARCHITECT SHALL EXTRAPOLATE THIS INFORMATION AND INCORPORATE IT
EQUIPMENT PROVIDED BY THE X-RAY EQUIPMENT DEALER.
EQUIPMENT SUPPLIERS:
SPECIAL NOTE TO ALL ANCILLARY
ON EQUIPMENT THEY DO NOT PROVIDE.
X-RAY EQUIPMENT DEALER IS NOT RESPONSIBLE FOR FURNISHING INFORMATION
ENERGIZED AND ARE SUBJECT TO PERSONAL CONTACT.
AS AN EFFECTIVE GROUND-FAULT CURRENT PATH PER 250.118 [517.13(A)].
THE GROUNDING TERMINALS OF ALL RECEPTACLES AND ALL NONCURRENT CARRYING CONDUCTIVE SURFACES OF FIXED ELECTRIC EQUIPMENT OPERATING AT OVER 100V, IF THEY ARE LIKELY TO BECOME
- EQUIPMENT GROUNDING.
- REDUNDANT GROUND ADHERENCE.
PATIENT SAFETY GROUNDING CONSIDERATIONS
(CONTRACTOR FURNISHED - CONTRACTOR INSTALLED)
A SPECIAL GROUNDING SYSTEM IS REQUIRED IN ALL DIAGNOSTIC ROOMS BY SOME STATE AND LOCAL
THE GOVERNING ELECTRICAL CODES AND CONFER WITH APPROPRIATE HEALTHCARE FACILITY
PERSONNEL AND CONSULTING ENGINEERS TO DETERMINE THE AREAS REQUIRING THIS GROUNDING
SHALL DESIGN AND INSTALL GROUNDING SYSTEM TO INCLUDE ALL METAL OBJECTS AND FIXTURES PER
NEC ARTICLE 517 CONSIDERATIONS
(CONTRACTOR FURNISHED - CONTRACTOR INSTALLED)
THE PROJECT ENGINEER OF RECORD IS RESPONSIBLE FOR SHOWING COMPLIANCE WITH NEC 517.
- FACILITY GRADE RECEPTACLES IN PATIENT CARE AREAS.
SECTION II - WIRING AND PROTECTION
NEC 517.
IF ANY OF THE ABOVE POWER SPECIFICATIONS REGARDING VOLTAGE REGULATION OR
DISCONNECT WITH 5% VOLTAGE REGULATION UNDER INTERMITTENT NO LOAD TO FULL LOAD
2. VOLTAGE REGULATION: XXX KVA, XXX PHASE, XXX VAC , 60HZ ±0.5% INCOMING LINE TO MAIN
ELECTRICAL CONTRACTOR SHALL USE A DRANETZ POWER LINE ANALYZER OR EQUIVALENT.
B. POWER LINES SHALL BE MEASURED FOR TRANSIENTS WITH THE SYSTEM IN STANDBY. THE
1. TRANSIENTS: NO TRANSIENTS SHALL OCCUR THAT EXCEED 30% OF NOMINAL PEAK LINE
ELECTRICAL EQUIPMENT BE CONNECTED TO THESE POWER LINES NOW OR IN THE FUTURE.
NEAREST FACILITY POWER SOURCE. UNDER NO CIRCUMSTANCES SHALL ANY OTHER
POWER LINES FOR EQUIPMENT LISTED SHALL BE DEDICATED LINES RUN DIRECTLY FROM
PROBLEM MAY BE EMPLOYED TO MAINTAIN LINE VOLTAGE WITHIN 5% OF NOMINAL UNDER
A. A REGULATION TRANSFORMER OR LINE CONDITIONER DEPENDING ON THE SEVERITY OF THE
TRANSIENTS CANNOT BE MET THE FOLLOWING CORRECTIVE ACTION CAN BE TAKEN.
A. TRANSIENTS, FOR DISCUSSION PURPOSES, SHALL BE DEFINED AS IMPULSES FROM 0.5 TO 800
INTERMITTENT NO LOAD TO FULL LOAD CONDITIONS.
CONDITIONS REQUIRED.
LINE REQUIREMENTS:
( CONTRACTOR FURNISHED - CONTRACTOR INSTALLED )
IMPORTANT POWER NOTES
MICROSECONDS.
VOLTAGE.
01/12/2017
DR400FINALV4
01/12/2017
DR400FINALV4