36C25918R0391-012.pdf

PDF 626 KB Posted

Attached to
Radiology Rooms 1 & 3 Federal contract opportunity
Solicitation number
36C25918R0391
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 19

About this file

36C25918R0391 RAD LAB 1 DWGS.pdf

View the file

Other files for this federal contract opportunity

Other files attached to Radiology Rooms 1 & 3, newest first.
File Type Posted
36C25918C0232-000.docx DOCX document
36C25918R0391-00001000.docx DOCX document
36C25918R0391-00001002.docx DOCX document
36C25918R0391-00001001.pdf PDF
36C25918R0391-00001004.docx DOCX document
36C25918R0391-00001003.pdf PDF
36C25918R0391-014.pdf PDF
36C25918R0391-011.pdf PDF
36C25918R0391-016.pdf PDF
36C25918R0391-018.docx DOCX document
36C25918R0391-010.docx DOCX document
36C25918R0391-013.pdf PDF
36C25918R0391-017.docx DOCX document
36C25918R0391-015.pdf PDF
36C25918R0391-009.docx DOCX document
Show all 15

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

PLANNING BY

DR400FINALV4

01/12/2017

DAVE HENRY

DRAWN BY:

FILE NAME:

DATE:

1/4"=1'-0"

SHEET:

SCALE:FACILITY EQUIPMENT

RADIOGRAPHIC SYSTEM SUITE

DIGITAL HIGH FREQUENCY

TYPICAL ROOM LAYOUT

(NOT SITE SPECIFIC)

DR 400

Brendan Conley RT(R), MRT | Agfa HealthCare Manager, DR Implementation Western Zone M 214-548-7214

Jim Dungee | Agfa HealthCare Senior Project Manager, DR Implementation Eastern Zone T 856-722-5221 | M 856-430-5086

DAVE HENRY

FILE NAME:

DATE:

DRAWN BY:

SHEET:

SCALE:

REVISION DESCRIPTIONNUMBER

SHEET

REV REP DATE

01/12/2017

DR400FINALV4

RECEIPT OF FINAL DRAWINGS

NAME: DATE:

COULD RESULT IN ADDITIONAL EXPENSES BEING INCURRED.

ALSO, I AM AWARE THAT ANY CHANGES MADE AFTER THIS DATE

THE EQUIPMENT OPERATION, SERVICEABILITY, OR SAFETY GUIDELINES.

DRAWINGS, DETAILS, AND SPECIFICATIONS, MAY ENCROACH UPON

PLANS. IT IS UNDERSTOOD THAT ANY DEVIATION FROM THESE

THIS SIGNATURE REPRESENTS RECEIPT OF THIS SET OF FINAL

REVISION HISTORY

TYPICAL ROOM LAYOUT

(NOT SITE SPECIFIC)

COVER SHEET.

PROPOSED ROOM LAYOUT.

PROPOSED EQUIPMENT LAYOUT.

E3 E4

E2

B1 D1

E1

SHEET

A3 A2

NUMBER

A1

SHEET DESCRIPTION

A4

E5

ELECTRICAL ELEVATIONS.

D2

SHEET LEGEND

SHEET LEGEND, RECEIPT OF DRAWINGS, REVISIONS HISTORY,

RECOMMENDED VENDORS, ELECTRONIC TRANSFER AGREEMENT.

GENERAL NOTES, ARCHITECTURAL NOTES AND SUPPORT NOTES,

SUPPLEMENTAL NOTES.

EQUIPMENT DETAILS. *PENDING

EQUIPMENT DETAILS. *PENDING

ELECTRICAL LAYOUT PLAN VIEW, TERMINAL SCHEDULE, ELECTRICAL

TAG DESCRIPTIONS.

ELECTRICAL BLOCK DIAGRAM, INCOMING WIRE SIZE NOTES,

CONDUIT & WIRING SCHEDULE.

CONDUIT LINE DIAGRAM.

TYPICAL TROUGH ELEVATION, JUNCTION BOX COVERPLATES DETAIL,

WALL DUCT PART LIST.

E6 DOOR SWITCH / WARNING LIGHT DETAIL AND NOTES.

herewith, regardless of its format or the means of its transmission. There is no guarantee or representation to the user as to the accuracy, currency, suitability or reliability of this data for any purpose. User accepts the data "as is", and assumes all risks associated with its use. By acceptance of this data, the user agrees not to transmit this data or provide access or any part of it to

All sheets included shall be considered a complete set of specification plans and may not be separated for any purpose. Any information contained in this set shall NOT be applied to other projects.

Gamma Tech, Inc. makes no representation or warranties, express or implied, with respect to the reuse of the data provided another party unless the user shall include with the data a copy of this disclaimer. Gamma Tech, Inc. assumes no responsibility for actual or consequential damage incurred as a result of any user's reliance on this data. Furthermore, electronic files on this media contain confidential and proprietary information that cannot be reproduced or divulged, in whole or in part, without written

ELECTRONIC FILE TRANSFER AGREEMENT

authorization from Gamma Tech, Inc.

E7 ELECTRICAL NOTES.

NONE

SCALE:

SHEET:

NONE

DRAWN BY:

FILE NAME:

DATE:

DAVE HENRY

NO ATTENTION HAS BEEN GIVEN TO SPECIFIC BUILDING STRUCTURAL INTEGRITY. A

9. THE EQUIPMENT REPRESENTATIVES MUST APPROVE DEVIATION FROM THESE PLANS IN

PREVAILING SEISMIC REQUIREMENTS. SEISMIC ENGINEERING DRAWINGS TO BE DONE BY

CUSTOMER CONSULTS WITH A PROFESSIONAL STRUCTURAL ENGINEER FAMILIAR WITH ALL

RESPECTIVE CONTRACTORS OF CHOICE. IT IS STRONGLY RECOMMENDED THAT THE

AND COMPLIANCE SHALL BE THE RESPONSIBILITY OF THE FACILITY (END USER) OR THEIR

AFFECTED BY THE SCOPE OF THIS WORK. ALL EXPENSES ASSOCIATED WITH SEISMIC TESTING

NO ATTENTION HAS BEEN GIVEN TO SPECIFIC SEISMIC ANCHORAGE CONSIDERATIONS

SUPPORT FOR THE DIAGNOSTIC IMAGING EQUIPMENT DESCRIBED IN THE 'EQUIPMENT

THE FACILITY SUPPORT STRUCTURE (e.g. WALLS, FLOORS, CEILINGS) CAN PROVIDE ADEQUATE

PROFESSIONAL STRUCTURAL ENGINEER SHALL INSPECT FIELD CONDITIONS TO ENSURE THAT

SCHEDULE'. STRUCTURAL ENGINEERING DRAWINGS TO BE DONE BY OTHERS.

WRITING PRIOR TO ANY CHANGES.

OTHERS.

ELECTRICAL AND MECHANICAL DESIGNERS SHALL BE REQUIRED TO REVIEW THESE

AFFECTED BY THE SCOPE OF THIS WORK. MECHANICAL ENGINEERING DRAWINGS TO BE DONE

8. NO ATTENTION HAS BEEN GIVEN TO SPECIFIC PLUMBING OR HVAC CONSIDERATIONS

SINGLE SHEET OF PLATE GLASS ON BOTH SIDES TO HELP PREVENT SCRATCHING.

BE TILTED 7° TO HELP ELIMINATE GLARE. IF TILTED 7° VIEWING WINDOW IS ACRYLIC, PROVIDE A

7. VIEWING WINDOW TO BE 1/16" LEAD EQUIVALENT. RECOMMEND LARGE VIEWING WINDOWS TO

ELECTRICAL NOTES. GROUNDING MUST CONFORM TO THE CURRENT CODE REQUIREMENTS.

EQUIPMENT REQUIRES VERY LARGE CURRENTS FOR EXTREMELY SHORT TIMES. REFER TO

6. THE SUPPLY VOLTAGE MUST REMAIN RELATIVELY STABLE. ELECTRICAL POWER FOR X-RAY

MANAGER SHALL CONSULT WITH FACILITY REGARDING DESIRED CONVENIENCES.

RESPECTIVE ROOM (i.e., STORAGE, CABINETS, SINKS AND COUNTERS, ETC.) PROJECT

5. SOME CONVENIENCES MAY BE REQUIRED BY CODE OR DESIRED BY CUSTOMER IN EACH

4. ALL WORK, EXCEPT INSTALLATION OF DIAGNOSTIC IMAGING EQUIPMENT, IS TO BE DONE BY

3. ARCHITECT AND ENGINEERS SHALL USE EXISTING ELECTRICAL AND STRUCTURAL ITEMS

CUSTOMER AND APPROVED BY PREVAILING CODES. LIGHTING FIXTURES, VENTS, ETC. TO BE

OPERATORS. RECOMMEND 2' x 4' LAY-IN CEILING PANELS OF MATERIAL ACCEPTABLE TO

2. ALL SUPPLY DIFFUSERS SHALL NOT CAUSE DIRECT AIR CURRENTS ON PATIENTS OR

BUILDING INTO RESPECTIVE ROOMS. RECOMMEND COVERING OR REMOVING THRESHOLDS.

AND PASSAGEWAY SIZES ARE ADEQUATE FOR TRANSPORTING EQUIPMENT FROM EXTERIOR OF

FINISHED SURFACES UNLESS OTHERWISE NOTED. ARCHITECT SHALL VERIFY THAT ALL DOOR

CONSTRUCTION PARTICIPANTS TO ADHERE TO THESE DISTANCES. ALL DIMENSIONS ARE FROM

DIMENSIONAL RELATIONSHIPS HAVE BEEN SUPPLIED. IT IS THE RESPONSIBILITY OF ALL

CONSTRUCTION PARTICIPANTS SHALL VERIFY AND ADJUST DIMENSIONS. CRITICAL

GAMMA TECH'S FLOOR PLAN WAS SCALED FROM PRINTS DRAWN BY OTHERS. ALL

INTO A WORKING SET OF CONSTRUCTION DRAWINGS OBSERVING ALL PREVAILING CODES.

TEAM SHALL EXTRAPOLATE PERTINENT INFORMATION FROM THIS LAYOUT AND INCORPORATE

TO CONFORM DETAILS TO THE ACTUAL EQUIPMENT EXPECTED TO BE INSTALLED. THE DESIGN

WITH SPECIFICATION AND CONCEPTUAL INFORMATION ONLY. EVERY EFFORT HAS BEEN MADE

GAMMA TECH'S DRAWINGS ARE INTENDED TO PROVIDE THE CONSTRUCTION PARTICIPANTS

RELYING ON THIS MATERIAL ASSUMES ALL RISKS AND LIABILITY ARISING FROM SUCH USE OR

BUILDING CODES AND REGULATIONS THAT HAVE JURISDICTION. THOSE MAKING USE OF OR

OR PROJECT MANAGER WHO SHALL ENSURE THAT ALL CONSTRUCTION COMPLIES WITH

GAMMA TECH'S DRAWINGS ARE FOR THE BENEFIT OF PURCHASER'S ARCHITECT OF CHOICE

FOR CONSTRUCTION. THE EQUIPMENT REPRESENTATIVES OR THEIR DESIGNER ARE NOT

1. GAMMA TECH'S SET OF DRAWINGS ARE FOR THE PURPOSE OF A LAYOUT PLAN ONLY - NOT

ARCHITECTURAL NOTES

THOSE OTHER THAN THE EQUIPMENT REPRESENTATIVES.

RESPONSIBLE OR LIABLE FOR ANY ERRORS OR OMISSIONS.

HELP MINIMIZE CONFLICTING CONSTRUCTION EFFORTS. ELECTRICAL ENGINEERING DRAWINGS

DRAWINGS TO HELP THEM GAIN AN UNDERSTANDING OF THE CONCEPT OF THIS WORK AND TO

BY OTHERS.

WHEN FEASIBLE.

FLUSH WITH FINISHED CEILING.

TO BE DONE BY OTHERS.

RELIANCE.

PURCHASE ORDER. ADDITIONALLY, A SITE INSPECTION SHALL BE CONDUCTED BY THE

EQUIPMENT REPRESENTATIVES TO CONFIRM THE SITE'S READINESS PRIOR TO X-RAY

EQUIPMENT DELIVERY. THIS INSPECTION SHALL BE PERFORMED WHEN NOTIFIED OF THE

CONSTRUCTION'S COMPLETION.

THE EQUIPMENT REPRESENTATIVES OR THEIR PLANNER ASSUME NO RESPONSIBILITY

AND ADDITIONAL TESTS REQUIRED TO MEET THE SATISFACTION OF THE HEALTH

REPLACEMENT OF ALL DEFECTIVE WORK (INCLUDING WORK SUBSEQUENTLY AFFECTED)

OF THE CONTRACTOR. THEREFORE, THE CONTRACTOR IS RESPONSIBLE FOR REPAIRS OR

14. THE DECISION OF ACCEPTABILITY BY A HEALTH PHYSICIST SHALL BE THE RESPONSIBILITY

THE SHIELDING SHALL BE TESTED, AT THE OWNER'S EXPENSE, BY A CERTIFIED HEALTH

AFTER X-RAY EQUIPMENT HAS BEEN INSTALLED AND PLACED IN OPERATING CONDITION,

13. SHIELDING TESTING (LEAD OR EQUIVALENT):

THESE MUST BE SPECIFIED BY A QUALIFIED RADIATION PHYSICIST.

12. RADIATION PROTECTION REQUIREMENTS:

10. ANY AIR CONDITIONING REQUIREMENTS. (HVAC - HEATING, VENTILATION, AIR CONDITIONING

ENTRANCE DOORS OF X-RAY ROOM TO BE COORDINATED WITH FACILITY, EQUIPMENT

9. INSTALLATION OF OPTIONAL WARNING LIGHT AND/OR DOOR INTERLOCK SWITCH AT

SPRINKLER HEADS, HALON NOZZLES, ETC. SHALL NOT EXTEND MORE THAN 1/2" BELOW

7. ANY CEILING MOUNTED LIGHT FIXTURES, DIFFUSERS, VENTS, SMOKE/FIRE DETECTORS,

SERVICE, A MINIMUM NUMBER OF OUTLETS WILL BE DEPICTED.

OUTLETS REQUIRED BY EQUIPMENT WILL BE SPECIFIED. FOR EASE IN EQUIPMENT

6. FOR THE MOST PART CONVENIENCE OUTLETS ARE TO BE SPECIFIED BY OTHERS.

5. INSTALLATION OF WALL AND CEILING SUPPORT AND SUSPENSION FRAME WORK AS

4. INTERCONNECTING WIRE AND CABLE AS SPECIFIED.

3. INSTALLATION OF LOAD CENTERS, BRANCH PANELS, CIRCUIT BREAKERS, JUNCTION BOXES,

2. INSTALLATION OF PROPER SPECIFIED POWER.

FINISHED BEFORE EQUIPMENT INSTALLATION CAN BEGIN.

1. ANY ROOM REMODELING AND CONSTRUCTION - FLOOR CEILING AND WALLS MUST BE

PROVIDE THE FOLLOWING (IF REQUIRED) PRIOR TO X-RAY EQUIPMENT DELIVERY AND

AT THEIR EXPENSE, THE CUSTOMER AND/OR THE CONTRACTOR WILL BE EXPECTED TO

X-RAY EQUIPMENT OR SITE PLANNING DOCUMENTS. NO WORK MAY BE PERFORMED OR

MATERIALS FURNISHED AT THEIR EXPENSE UNLESS AUTHORIZED BY A SIGNED, FORMAL

FOR ANY CONSTRUCTION COSTS, WHETHER OR NOT RELATED TO THE INSTALLATION OF

11. ALL ANCILLARY AND ACCESSORY ITEMS.

AND HUMIDIFICATION)

PHYSICIST.

INSTALLERS, AND DESIGN TEAM.

8. ANY PLUMBING REQUIREMENTS.

THE FINISHED CEILING.

15. LIABILITY:

PHYSICIST.

CONDUIT AND CABLE TRAYS AS SPECIFIED.

GENERAL NOTES

INSTALLATION:

SPECIFIED.

5. ALL SURFACE MOUNTED SUPPORTS ARE TO BE PAINTED THE SAME COLOR AS THE

ADJACENT SURFACE OR CONSULT WITH FACILITY REGARDING ACCENT COLORS.

( CONTRACTOR FURNISHED - CONTRACTOR INSTALLED )

2. ALL SUPPORTS STRUCTURES ARE TO BE PROVIDED BY THE FACILITY OR THEIR

3. WALL SUPPORTS ARE TO BE HIDDEN WITHIN WALLS WHENEVER POSSIBLE.

THE EQUIPMENT REPRESENTATIVE OR THEIR DESIGNER IS IN NO WAY RESPONSIBLE FOR THE

SUPPORT STRUCTURE FOR ANY OVERHEAD CEILING MOUNTED EQUIPMENT / DEVICES OR WALL

MOUNTED EQUIPMENT / DEVICES. THIS IS THE RESPONSIBILITY OF THE CONSTRUCTION

1. ALL UNITS THAT ARE WALL MOUNTED OR CEILING SUSPENDED ARE TO BE PROVIDED WITH

GIVEN (i.e.,BLOCK). CONSULTATION WITH A STRUCTURAL ENGINEER IS STRONGLY

4. WALL SUPPORTS MAY NOT BE REQUIRED IF WALL MATERIAL IS ADEQUATE FOR THE LOADS

SUPPORTS STRUCTURES WHERE NECESSARY.

SUPPORTS:

CONTRACTOR.

RECOMMENDED.

SUPPORT NOTES

PARTICIPANTS.

01/12/2017

DR400FINALV4

ALL ITEMS LISTED BELOW ARE TO BE PROVIDED/DESIGNED BY THE ARCHITECT. THE ARCHITECT SHALL

4. TYPICALLY THE MINIMUM RECOMMENDED FINISHED CEILING HEIGHT IS 8'-4" A.F.F. UNLESS STATED

3. THE FLOOR IN THE X-RAY AND DARKROOM MUST BE ESSENTIALLY LEVEL.TOLERANCE IS 1/8" OVER A

2. IT IS RECOMMENDED THAT ALL DIAGNOSTIC ROOMS SHOULD HAVE TILE OR VINYL FLOORING.

CONTROLS SHOULD BE LOCATED NEAR THE ROOM ENTRANCES AND IN THE CONTROL AREAS.

THE ROOM. THE CANNED LIGHT FIXTURES SHOULD BE PROVIDED WITH DIMMER CONTROLS. THE DIMMER WELL AS FLORESCENT LIGHT FIXTURES. THE CANNED LIGHT FIXTURES ARE TO BE DISTRIBUTED AROUND

1. ALL DIAGNOSTIC ROOMS INCLUDING X-RAY, MAMMO, R/F ETC... SHALL HAVE CANNED LIGHT FIXTURES AS

SUPPLEMENTAL NOTES

OTHERWISE IN THIS DRAWING SET.

CONSULT WITH THE CUSTOMER FOR ANY ADDITIONAL INFORMATION.

10'-0" SPAN.

TYPICAL ROOM LAYOUT

(NOT SITE SPECIFIC)

1/4"=1'-0"

SCALE:

DAVE HENRY

FILE NAME:

DRAWN BY:

DATE:

SYSTEM. AREA UNDER COUNTER MUST REMAIN CLEAR FOR TRIPLET BOX. WORKING DRAWINGS BY ARCHITECT.

PHYSICIST REGARDING SHIELDING REQUIREMENTS. CONSULT CUSTOMER FOR WINDOW SIZE. WORKING

CONSTRUCT CONTROL BOOTH PARTITION FROM FLOOR TO CEILING WHERE SHOWN. PROVIDE ONE LEADED

PROVIDE ONE 2"T x 16"H x 24"W WOOD SUPPORT BOARD HIDDEN IN WALL WHERE SHOWN. LOCATE BOARD

PROVIDE A 40" HIGH STANDING COUNTER WITH ONE 3" DIA. CABLE ACCESS HOLE THROUGH COUNTER WHERE

SHOWN. BRACE COUNTER TO SUPPORT 300 POUNDS STATIC LOAD, COUNTER WILL BE USED TO SUPPORT DR

WINDOW AND FRAME. LOCATE WINDOW CENTER LINE 5'-0" A.F.F. CONSULT WITH A CERTIFIED RADIOLOGICAL

CENTER LINE 84" A.F.F. BOARD SHALL SUPPORT 300 POUNDS PULL FORCE. REFER TO SUPPORT NOTES.

PAINT BOARD THE SAME COLOR AS FINISHED WALL.

CONSTRUCTION NOTES

(CONTRACTOR FURNISHED - CONTRACTOR INSTALLED)

DRAWINGS BY ARCHITECT.

EQUIPMENT VENDOR IS NOT RESPONSIBLE FOR DAMAGES. CONSULT WITH EQUIPMENT VENDOR FOR

PROVIDE WALL AND FLOOR PROTECTION FOR INSTALLATION AND DELIVERY OF X-RAY EQUIPMENT. THE

FURTHER INFORMATION.

01/12/2017

DR400FINALV4

X-RAY TABLE BASE AND WALL STAND ANCHOR HOLE LOCATION. REFER TO DETAIL SHEET <D2> FOR FURTHER

INFORMATION. THE FLOOR MUST SUPPORT THE TABLE WEIGHT AND MAXIMUM PATIENT WEIGHT.

TYPICAL ROOM LAYOUT

(NOT SITE SPECIFIC)

'-6

'-0

'-6

7'-7 3/8"

C

C

16'-0"

5'

-0

5"

'-0

'-5

/1

6"

2'-0"

'-1

/1

6"

'-0

3'- 0"

'-0

'-0

45°

Ø 3"

C

3'-0"

3'-6"

THE FLOOR IN THE X-RAY ROOM MUST BE LEVEL TO WITHIN ±1/8" OVER A 10'-0" SPAN. THE MINIMUM

RECOMMENDED FINISHED CEILING HEIGHT IS 8'-4" A.F.F.

JUNCTION BOX <F>. CONSULT WITH EQUIPMENT INSTALLER FOR EXACT SIZE AND LOCATION.

PROVIDE SMALL SHELF ON CONTROL BOOTH WALL TO SUPPORT THE SCU. SHELF TO BE LOCATED UNDER

DRAWN BY:

DATE: SHEET:

1/4"=1'-0"

SCALE:

01/12/2017

DR400FINALV4

*AGFA DR 400

86.60 47.44 89.57 1389

31.46 9.37 88.39 346

23.50 17.50 14.50 29

TUBESTAND LEFT HAND LOAD VERSION SHOWN

AGFA DR 400

AGFA DR 400

VERTICAL BUCKY STAND

STANDARD

TRIPLET BOX

CONTAINS PC, UPS AND DR SYSTEM COMPONENTS

ELEVATING 4-WAY FLOAT TOP INTEGRATED TABLE

INTEGRATED INSIDE TABLE

W x D x HW x D x H

WEIGHT (LBS.)WEIGHT (LBS.)

DIMENSIONS (IN.)DIMENSIONS (IN.)

EQUIPMENT SCHEDULEEQUIPMENT SCHEDULE

EQUIPMENT IDENTIFICATION NUMBEREQUIPMENT IDENTIFICATION NUMBER

AGFA DR 400

(LOCATED IN TABLE BASE)

BTU / HOURBTU / HOUR

DESCRIPTIONDESCRIPTION

SHF GENERATOR CABINET

AGFA DR 400

MONITOR, KEYBOARD AND MOUSE

GENERATOR CONTROL AND DR WORKSTATION PC TYPICAL

'-6

C

7'-7 3/8"

C

40" S.I.D.

72" S.I.D.

TYPICAL ROOM LAYOUT

(NOT SITE SPECIFIC)

50° to 95° (F)

30% to 90% non condensing

Above specification apply to all areas where the Radiographic

(Control area, Procedure room and Equipment room if required.)

equipment is to be located.

Ambient Operation Temperature:

Humidity:

*AGFA DR 400

TUBESTAND RIGHT HAND LOAD VERSION AVAILABLE.

ELEVATING 4-WAY FLOAT TOP INTEGRATED TABLE

NEC CLEARANCE

FOR DISCONNECT <A>.

EC SHALL MAINTAIN

NOTE: EC SHALL ENSURE THAT LUGS ARE LARGE ENOUGH TO FACILITATE OVERSIZED INCOMING CABLES

REQUIRED BY X-RAY EQUIPMENT. REFER TO "INCOMING WIRE SIZE" NOTES.

FLUSH MOUNT IN WALL WHERE SHOWN. LOCATE ENCLOSURE CENTER LINE 5'-6" A.F.F.

ENCLOSURE SHALL CONTAIN ONE XXAMP, X-POLE THERMOMAGNETIC CIRCUIT BREAKER WITH SHUNT TRIP DEVICE.

NEMA 1, XXX VAC, XXX PHASE, FRONT OPERATED, CIRCUIT BREAKER ENCLOSURE WITH GROUND ACCESSORIES.

FACILITY TO DETERMINE SOURCE LOCATION.

DAVE HENRY

DATE:

FILE NAME:

DRAWN BY:

SHEET:

1/4"=1'-0"

SCALE:

01/12/2017

DR400FINALV4

TYPICAL ROOM LAYOUT

(NOT SITE SPECIFIC)

TE

5'-0" 11'-0"

C

'-0

'-6

'-6

'-0

2'-0"2'-0"3'-4"

8"8"8"10"8"

C C C C

C

C

C C C

( CONTRACTOR FURNISHED - CONTRACTOR INSTALLED )

EC SHALL PROVIDE ONE NORMALLY OPEN SWITCH WITH RED MUSHROOM HEAD AND AN "EMERGENCY OFF" NAME PLATE.SWITCH WILL BE USED FOR SHUNT TRIP DEVICE LOCATED ON CIRCUIT BREAKER <A>.

( CONTRACTOR FURNISHED - CONTRACTOR INSTALLED )

DOOR IS OPEN. (INSTALL ONLY IF REQUIRED BY PREVAILING CODE OR REQUESTED BY CUSTOMER).

EC SHALL FURNISH AND INSTALL ONE FACILITY GRADE, THREE PRONG, 120VAC, 15AMP, DUPLEX RECEPTACLE.

OF CONNECTOR (e.g.. RJ-45/CAT-5, FIBER OPTICS, BNC/COAX, ETC.). REFER TO DICOM NOTES.

NOTE: ETHERNET CONNECTION SHOULD HAVE INTERNET ACCESS WITH VPN CAPABILITY.

THE TELEPHONE NUMBER MUST BE PROVIDED TO THE TECHNICIAN WHO INSTALLS THE SYSTEM.

ELECTRICAL TAG DESCRIPTIONS

FLUSH MOUNT IN WALL WHERE SHOWN 12" A.F.F. PROVIDES POWER FOR THE DR SYSTEM COMPONENTS.

FLUSH MOUNT IN WALL WHERE SHOWN AT NORMAL HEIGHT. PROVIDES POWER FOR SERVICE AND HOUSE

EC SHALL FURNISH AND INSTALL TWO FACILITY GRADE, THREE PRONG, 120VAC, 20AMP, QUAD RECEPTACLE.

BETWEEN THE FACILITY, THE CONSTRUCTION PARTICIPANTS AND THE EQUIPMENT INSTALLERS. REFER TO WIRING LOCATION TO BE DETERMINED BY THE FACILITY. THE X-RAY ON WARNING LAMP CIRCUIT SHALL BE COORDINATED 4" x 4" x 2" JUNCTION BOX FOR THE "X-RAY IN USE" WARNING SIGN. FLUSH MOUNT IN WALL OR CEILING. EXACT

4" x 2" x 2" DEEP SWITCH BOX. FLUSH MOUNT IN WALL WHERE SHOWN. LOCATE BOX CENTER LINE AT 6'-0" A.F.F.

6" x 6" x 4" DEEP JUNCTION BOX. FLUSH MOUNT IN WALL WHERE SHOWN. LOCATE BOX CENTER LINE 6" A.F.F.

8" x 8" x 4" DEEP JUNCTION BOX. FLUSH MOUNT IN WALL WHERE SHOWN. LOCATE BOX CENTER LINE 12" A.F.F.

SHALL HAVE NORMALLY OPEN CONTACTS RATED AT 120VAC, 15AMPS. SWITCH SHOULD OPEN CONTACTS WHEN PROVIDE ONE 1-POLE, SINGLE THROW, MOMENTARY, EXPOSURE INTERLOCK SWITCH IN DOOR FRAME. SWITCH

PROVIDE A 4-PORT ETHERNET CONNECTION WHERE SHOWN. EC SHALL CONSULT WITH END USER REGARDING TYPE

PROVIDE ALL HARDWARE NECESSARY INCLUDING COVER PLATES, CONNECTORS AND INTERCONNECTING CABLES.

(TELEPHONE LINE WHICH DOES NOT RUN THROUGH ANY FACILITY PBX, SWITCHBOARD OR A SHARED ACCESS LINE.

PERFORMED FROM A REMOTE LOCATION. THE RJ11 LINE MUST BE A DEDICATED DIRECT IN DIAL (DID) VOICE-GRADE PROVIDE A SINGLE TELEPHONE LINE WHERE SHOWN FOR CERTAIN SERVICE AND MAINTENANCE ACTIVITIES TO BE

AT THE TIME OF EQUIPMENT INSTALLATION.

CABLE ACCESS FOR THE CONTROL CONSOLE.

CABLE ACCESS FOR THE TABLE.

CABLE ACCESS FOR THE WALL STAND.

PROVIDE A COVER PLATE AS SHOWN IN COVER PLATE DETAILS.

PROVIDE A BLANK COVER PLATE ON FLOOR TRENCH. EQUIPMENT INSTALLER TO PROVIDE CABLE ACCESS HOLE

KEEPING. REQUIRES HOUSE POWER, LOCATION NOT CRITICAL.

DEDICATED OUTLET, POWER CAN BE SUPPLIED FROM <A1>.

PROVIDE A COVER PLATE AS SHOWN IN COVER PLATE DETAILS.

E

T

W

ITEM

DIAGRAM.

F

N

G

D

C

B

DESCRIPTION

A

ITEM

TERMINAL SCHEDULE

DESCRIPTION

10" x 3 1/2" DEEP FLUSH MOUNTED VERTICAL RISER. FLUSH MOUNT IN WALL AT FLOOR LEVEL AND TERMINATE ABOVE THE FINISHED CEILING TO ALLOW FOR CONDUIT CONNECTIONS. PROVIDE DIVIDERS FOR LOW AND HIGH

VOLTAGE. REFER TO TROUGH NOTES AND TROUGH ELEVATION.

10" x 3 1/2" DEEP FLUSH MOUNTED FLOOR TRENCH. FLUSH MOUNT IN FLOOR WHERE SHOWN AND CONNECT

TO VERTICAL RISER. PROVIDE 3 COMPARTMENTS FOR LOW AND HIGH VOLTAGE.

REFER TO TROUGH NOTES AND TROUGH ELEVATION.

4" x 4" x 4" DEEP JUNCTION BOX. FLUSH MOUNT IN WALL WHERE SHOWN. LOCATE BOX ABOVE WINDOW OR

CABLE ACCESS FOR THE SCU. NOTE: BOX OPENS OUT INTO X-RAY ROOM NOT INTO CONTROL AREA.

CENTER LINE 88" A.F.F. PROVIDE A BLANK OVERSIZE COVER PLATE.

FLUSH MOUNT IN WALL WHERE SHOWN 88" A.F.F. NEXT TO JUNCTION BOX <F>. PROVIDES POWER TO SCU.

EC SHALL FURNISH AND INSTALL ONE FACILITY GRADE, THREE PRONG, 120VAC, 15AMP, SINGLE RECEPTACLE.

DEDICATED OUTLET, REQUIRES HOUSE POWER.

4'-10"

NONE

SCALE:

POWER SOURCE

MAIN

INCOMING

ELECTRICAL BLOCK DIAGRAM

G

F

D

W

RUN #

INCOMING POWER

JUNCTION BOXES/COVERS

CONDUIT CONNECTIONS

CABLE ACCESS POINTS

LEGEND

OR RACEWAY

TROUGH (WALL DUCT)

N

C

A B

LEAVE PULL STRINGS.

2-#14AWG PLUS GROUND.

FOR SHUNT TRIP SOLENOID.

MAXIMUM CONDUIT LENGTH IS 24'-0".

TAILS IN (G). EC SHALL IDENTIFY WIRES AS 'X-RAY WARNING SIGN' 'L', 'N', 'G' .

CIRCUIT PER EQUIPMENT SUPPLIER RECOMMENDATIONS AND LEAVE 10'-0"

2-#14AWG PLUS GROUND. EC TO CONNECT "X-RAY IN USE" WARNING LIGHT

EC SHALL CONNECT TO (A) AND (B). PROVIDE POWER IF NECESSARY

EC SHALL CONNECT TO (N) AND LEAVE 10'-0" TAILS AT <G>.

IDENTIFY WIRES AS DOOR INTERLOCK SWITCH.

2-#14AWG PLUS GROUND IF REQUIRED.1/2"N

F

VERTICAL

RISER CONDUIT

C VERTICAL

D

W

B A

VERTICAL

RISER

RISER

RISER

VERTICAL

CONDUIT

1/2"

CONDUIT

1/2"

DAVE HENRY

DATE:

DRAWN BY:

FILE NAME:

( CONTRACTOR FURNISHED - CONTRACTOR INSTALLED )

CONDUIT & WIRING SCHEDULE

FROM

MAIN

POWER

A

TO

A

SIZE

VARIES

VARIES

REMARKS

1 1/2"

CONDUIT

VERTICAL RISER

(WALL DUCT)

VERTICAL

RISER

2 1/2"

CONDUIT

D

2 1/2"

CONDUIT

FLOOR TRENCH

LEAVE PULL STRINGS.

MAXIMUM CONDUIT LENGTH IS 24'-0".

LEAVE PULL STRINGS.

SOURCE

GENERATOR

KW RATING

RECOMMENDED

LINE VOLTAGE

SUPPLY TRANSFORMER

KVA RATING

MAXIMUM MOMENTARY

CURRENT

RECOMMENDED MAIN

CIRCUIT BREAKER <A>

MAXIMUM WIRE LINE

IMPEDANCE

HFe 601

65KW

480VAC/3-PHASE/60Hz

VOLTAGE REGULATION

82KVA

THREE HOT WIRES ONE NEUTRAL AND ONE EARTH GROUND

TRANSFORMER: Y WITH BALANCED NEUTRAL

UNDER LOAD ±5%.

EC SHALL CALCULATE INCOMING WIRE SIZE FROM BUILDING SOURCE (TRANSFORMER)

TO MAIN DISCONNECT <A> WITH VOLTAGE REGULATION OF ±5% AT MAXIMUM LOAD.

EC SHALL CALCULATE INCOMING WIRE SIZE BASED OFF OF INFORMATION

CONDUIT SIZE DEPENDS ON INCOMING WIRE SIZE.

ENSURE CONNECTION COMPLIES WITH ALL STATE AND LOCAL CODES.

EC TO CONNECT AT <A> AND AT GENERATOR ITEM <2> SHEET B1.

WIRE SIZE TO BE DETERMINED BY CALCULATED 50' WIRE LENGTH.

PROVIDED IN CHART AND LENGTH OF WIRE AS REQUIRED AT SITE.

HFe 401

40KW

480VAC/3-PHASE/60Hz

50KVA

UNDER LOAD ±5%.

VOLTAGE REGULATION

HFe 501

50KW

480VAC/3-PHASE/60Hz

60KVA

UNDER LOAD ±5%.

VOLTAGE REGULATION

HFe 801

80KW

480VAC/3-PHASE/60Hz

VOLTAGE REGULATION

100KVA

UNDER LOAD ±5%.

01/12/2017

DR400FINALV4

DATE:

DRAWN BY:

SHEET:

SCALE:

01/12/2017

DR400FINALV4

01/12/2017

DR400FINALV4

TYPICAL ROOM LAYOUT

(NOT SITE SPECIFIC)

LINE (TYPICAL).

FINISHED FLOOR

LINE (TYPICAL).

FINISHED CEILING

'-0

6" C

'-0

C

'-0

" C

'-6

C

'-0

C

'-0

"C TE

'-0

" C

CONTROL BOOTH WALL SHOWN UNFOLDED

FOLD FOLD

COUNTER

1/4"=1'-0"

'-4

C

DATE:

DRAWN BY:

FILE NAME:

SHEET:

NONE

SCALE:

01/12/2017

DR400FINALV4

01/12/2017

DR400FINALV4

TYPICAL ROOM LAYOUT

(NOT SITE SPECIFIC)

SHORTEST ROUTE POSSIBLE. NOTE: DO NOT FOLLOW WALL LINE

NOTE TO ELECTRICAL CONTRACTOR: THE CONDUIT ROUTING

IS SHOWN AS A VISUAL REFERENCE ONLY (NOT TO SCALE). THE

UNLESS IT IS THE SHORTEST ROUTE POSSIBLE.

EC SHALL ROUTE CONDUITS FROM POINT TO POINT USING THE

POWER SOURCE

MAIN

INCOMING

FOR APPROPRIATE PART NUMBERS.

PLACEMENT AND TROUGH CLEARANCES.

10" X 3 1/2" FLUSH MOUNTED WALL

5'-0" LONG DIVIDER STRIPS FOR

10" X 3 1/2" DUCT (3 COMPARTMENTS).

10" X 3 1/2" END CLOSURES.

DUCT (5'-0" LENGTHS, INCLUDES COVERS)

WALL DUCT PARTS LIST

CONDITIONS REQUIRE SURFACE MOUNTED DUCT THEN CONTACT THE ELECTRICAL SUPPLIER

THE PROJECT REQUIREMENTS. PLEASE HAVE THE DRAWING FILE NAME (BOTTOM RIGHT CORNER

A COPY OF THESE PLANS HAVE BEEN FORWARDED TO THE SUPPLIER TO ENSURE ACCURACY OF

ON PRICING AND AVAILABILITY. SUFFICIENT LEAD TIME MUST BE GIVEN ON ALL TROUGH ORDERS.

NOTE: CONTACT CHRIS NAJDA (440) 835-3500 (ELECTRICAL SALES CO.) FOR FURTHER INFORMATION

NOTE: PROVIDE TWO DIVIDER STRIPS WITHIN TROUGH FOR THE SEPARATION OF HIGH TENSION

CROSSOVER TUNNELS BEHIND COVER PLATES AND MULTIPLE CONDUIT CONNECTION POINTS.

X-RAY CABLES, POWER AND LOW VOLTAGE CABLES WHERE REQUIRED. PROVIDE APPROPRIATE

CLEAR FOR REMOVAL AND INSERTION OF PRE TERMINATED CABLES.

NOTE: ALL TROUGH SHALL BE FLUSH WITH FINISHED SURFACES UNLESS OTHERWISE SPECIFIED

AND PROVIDED WITH REMOVABLE, OVERSIZED, COVERS THE FULL LENGTH. COVERS SHALL BE

NOTE: ALL PART NUMBERS LISTED ARE FOR FLUSH MOUNTED APPLICATIONS, IF THE FIELD

NOTE: IF IT IS NECESSARY THAT THE TROUGH BE SURFACE MOUNTED, WHEN NOT INDICATED ON

THESE PLANNING DRAWINGS, ALL EQUIPMENT PLACEMENT AND TRAVEL CLEARANCE MUST BE

VERIFIED TO PREVENT EQUIPMENT AND TROUGH INTERFERENCE AND POSSIBLE EQUIPMENT

DAMAGE. CONTACT GAMMA TECHNOLOGIES, INC. FOR PROPER VERIFICATION OF EQUIPMENT

COUPLERS DO NOT HAVE TO BE ORDERED UNLESS CUSTOMER WILL BE MAKING A LARGE

NOTE: COUPLERS ARE INCLUDED WITH ALL WALL DUCT STRAIGHTS AND FITTINGS. ADDITIONAL

JUNCTION BOX COVERPLATES

6"

(C)

COVERPLATE

6"

2" DIAMETER CHASE

NIPPLE WITH PLASTIC

(CENTERED)

BUSHING AND LOCKNUT.

OF PLAN).

NUMBER OF FIELD CUTS.

PART# ESW3560D

PART NUMBER

PART# ESW103560F

PART# ESW1035ECF

CONNECTION POINTS.

TYPICAL CONDUIT CONNECTIONS

LOCATED ABOVE FINISHED CEILING.

EC TO DETERMINE ACTUAL CONDUIT

ALL FLUSH MOUNTED DEVICES SHALL HAVE OVERSIZED COVERPLATES.

NOTE: COVERPLATES DEPICTED ARE FOR SURFACE MOUNTED DEVICES.

COVERPLATE DETAILS AND TROUGH NOTES

BUSHING AND LOCKNUT.

8"

(D)

COVERPLATE

8"

2" DIAMETER CHASE

NIPPLE WITH PLASTIC

(CENTERED)

DESCRIPTION QTY

'-0

W

A L

L D

U C

T S

E C

T

IO

N C

U T

T O

L E

N G

T H

FINISHED FLOOR

(WALL DUCT)

VERTICAL RISER

MOUNTED TROUGH, HIDES TROUGH AND

DRYWALL JUNCTION. COVERPLATES TO

PRE TERMINATED EQUIPMENT CABLES.

OVERSIZE COVERPLATES FOR FLUSH

REMAIN UNOBSTRUCTED FOR EASY

REMOVAL. ALLOWS INSTALLATION OF

PART# ESW103560F

FINISHED CEILING

'-0

W

A L

L D

U C

T S

E C

T

IO

N

NOT TO SCALE

PART# ESW3560D

DIVIDERS FOR LOW

AND HIGH VOLTAGE.

TYPICAL TROUGH ELEVATION

TO EQUIPMENT INSTALLER FOR CABLE INSTALLATION.

CONDUIT CONNECTION POINTS MUST BE ACCESSIBLE

END CAP PART# ESW1035ECF

SHEET:

NONE

SCALE:

DAVE HENRY

DATE:

DRAWN BY:

FILE NAME:

01/12/2017

DR400FINALV4

01/12/2017

DR400FINALV4

NOTE: CONTACT CHRIS NAJDA (440) 835-3500 (ELECTRICAL SALES CO.) FOR FURTHER INFORMATION.

TYPICAL ROOM LAYOUT

(NOT SITE SPECIFIC)

ADAPTER

TRENCH TO WALL DUCT

PART# ESTWA1035

3 COMPARTMENTS

TRENCH DUCT

PART# EST1035C72

END CAP

TRENCH DUCT

PART# EST1035EC

TRENCH TO WALL DUCT

ADAPTERPART# ESTWA1035 1

10" X 3 1/2" FLUSH MOUNTED FLOOR TRENCH

DUCT, 3-COMPARTMENTS. (6'-0" LENGTHS)PART# EST1035C72 1

TRENCH DUCT

END CAPPART# EST1035EC 2

4'-0" CUT TO LENGTH

PART# EST10C12

10" X 3 1/2" FLUSH MOUNTED FLOOR TRENCH

12" LONG SECTIONAL COVER PLATE. 1

COVER PLATE

12" SECTIONAL

PART# EST10C12

TRENCH DUCT

12"

THE MAIN CONSIDERATION WHEN LEAVING WIRES IN JUNCTION BOXES IS TO ENSURE THEY ARE NOT

LIVE AND THAT THEIR ASSOCIATED SWITCHING MEANS ARE LOCKED OUT (OR EQUIVALENT).

THIS STANDARD SHALL BE APPLIED TO ALL OPEN ENDED WIRE AND CABLE MENTIONED IN THIS DRAWING

SET. OBVIOUSLY, GROUND WIRE OR CABLE BEING THE EXCEPTION.

IT IS THE ELECTRICAL CONTRACTOR'S RESPONSIBILITY TO ENSURE THAT WIRES ARE NOT ENERGIZED

UNTIL THE EQUIPMENT INSTALLERS ARE READY FOR TESTING.

THE ELECTRICAL CONTRACTOR SHALL INFORM THE EQUIPMENT INSTALLER(S) AS TO THE LOCATION AND MEANS OF ACTIVATING ALL SWITCHING DEVICES ASSOCIATED WITH THE OPEN ENDED WIRE AND CABLE

MENTIONED IN THIS DRAWING SET.

TO (J) GENERATOR CABINET.

WARNING LIGHT WIRING DIAGRAM DETAIL.

WIRES SHALL NOT BE LIVE. REFER TO

X-RAY IN USE WARNING SIGN NOTES:

MAXIMUM 20 WATT BULBS RECOMMENDED FOR WARNING LIGHT.

CONTRACTOR SHALL CONSULT WITH FACILITY REGARDING EXACT WORDING, STYLE AND LOCATION.

OPTIONAL X-RAY IN USE WARNING SIGN TO BE INSTALLED ONLY IF REQUIRED BY PREVAILING CODE OR

REQUESTED BY CUSTOMER.

DOOR SWITCH NOTES:

OPTIONAL DOOR SWITCH TO BE SQUARE D "CLASS 9007, TYPE A0" OR EQUIVALENT TO BE LOCATED ON HINGE

SIDE OF DOOR JAMB.

OPTIONAL DOOR SWITCH TO BE INSTALLED ONLY IF REQUIRED BY PREVAILING CODE OR REQUESTED BY

CUSTOMER.

IN CASE OF A DOUBLE JAMB, ONE DOOR SWITCH SHOULD BE LOCATED ON EACH SIDE. SEE TROUGH /

RACEWAY LAYOUT FOR DOOR SWITCH / WARNING LIGHT LOCATION IN ROOM.

SEE POINT TO POINT WIRING DESCRIPTION FOR WIRE SIZES TO BE USED IN CONDUITS.

USING 'W' AS A JUNCTION BOX FOR THE DOOR INTERLOCK SWITCH WIRES ELLIMINATES AN ADDITIONAL CONDUIT RUN TO THE SYSTEM CABINET PULL BOX. ALTHOUGH THIS DECREASES CONSTRUCTION COSTS, IT

IS NOT REQUIRED.

WARNING SIGN NOTES:

OPTIONAL WALL SWITCH (SW) MAY BE CONTRACTOR INSTALLED AT THE REQUEST OF THE CUSTOMER OR EQUIPMENT SUPPLIER. PURPOSE OF WALL SWITCH IS TO REMOVE ALL VOLTAGES FROM GENERATOR CABINET DURING SERVICE. EC SHALL CONSULT WITH CUSTOMER AND EQUIPMENT SUPPLIER REGARDING

WALL SWITCH INSTALLATION, TYPE AND EXACT LOCATION.

DOOR SWITCH / WARNING LIGHT DETAIL

JB

X-ray On Warning Sign Circuit Options

Housing

c. Aluminum, etc.

Note: Recommend 20 watts or less

a. 120vac * a. red on white * Facility panel * Rotor prep *

b. 277vac b. white on red Systems supply System ON

c. 120/277vac c. black on white Always ON

d. 26vac d. white on black

e. DC e. red on black, etc.

Refer to detail number

The engineer of record shall consult with the facility and equipment supplier to coordinate the x-ray on warning sign circuit combinations. Majority picks marked by an asterisk.

a. White * LED Single * Wall * Surface * X-RAY IN USE *

b. Black LED Double Ceiling Flush X-RAY ON, etc.

Refer to detail number WLD01.0310 for typical x-ray on warning light circuit.

Example "X-RAY IN USE" Warning Sign Circuit Options

Many potential factors influence choice of x-ray on warning sign. E.g. Matching existing; efficiency; aesthetics; price;

quality; number of options; etc.

Gamma Tech never knows the exact warning sign type, style, lamp voltage, power source, method or means of control prior to drawing set issuance. For this reason more than one detail may be depicted in this drawing set.

It is our hope and intent to encourage the construction participants and the equipment suppliers / installers to coordinate efforts. For this reason we have listed some potential options and their caveats below. (not to be construed as all inclusive). Important decisions to be shared are numbered. Sample options are listed alphabetically. For your convenience, the majority pick is designated by the word 'typical' in parentheses.

The facility or their designee selects an x-ray on warning sign.

1) Lamp voltage? - usually one of four options.

a. 120vac (typical)

b. 120/277vac

c. 277vac

d. Low voltage ac

e. Low voltage dc (battery) Note: The system installer should be informed of the lamp voltage as soon as possible.

The facility determines at which point in time the warning sign will illuminate.

2) When to illuminate? - Engineer to coordinat with the facility and equipment supplier.

a. Warning lamp turns on upon rotor prep / turns off upon x-ray termination. (typical)

b. Warning lamp illuminates / extinguishes when equipment is switched on / off.

c. Warning lamp illuminates 24/7.

Note: the equipment being installed may or may not be capable of interfacing with a warning light or may have just one option. The equipment supplier shall coordinate with the facility and the construction participants regarding interface options.

Voltage

DAVID HENRY

warning light circuit.for typical WLD02.0696

FACILITY:

120VAC TO 24VAC

FILE NAME:

DATE:

DRAWN BY:

SHEET:

SCALE:

Faces

Power source

Lamp

Panel color

Surface

ON / OFF

Mounting Wording

TO SUPPLY POWER TO (W).

LAMP VOLTAGE CAN VARY.

REFER TO ARCHITECT'S

DRAWINGS REGARDING HOW

( CONTRACTOR FURNISHED - CONTRACTOR INSTALLED )

1/2" CONDUIT

SW

N

W

WALL SWITCH

TO BE LABELED

"WARNING LIGHT"

OPTIONAL

1/2" CONDUIT

JB

1/2" CONDUIT

FLOOR LINE

CEILING LINE

A typical 'X-RAY IN USE' warning sign circuit is depicted below.

Due to many varibles, the x-ray on warning sign circuit must be coordinated between the equipment supplier, electrical engineer and the facility.

The electrical contractor shall consult with the facility and the equipment supplier prior to installing any x-ray on warning sign.

TYPICAL WARNING LIGHT CIRCUIT

X-RAY ON WARNING LIGHT

N

L

G

20A

120VAC

POWER

IN USE

X-RAY

N L

G

W

JUNCTION BOX

In the example above, 120vac house power is supplied to the warning light.

The circuit is not complete until the anode, in the x-ray tube, begins to turn.

When the exposure button is pressed, the rotor prep signal is activated.

At this time the "x-ray in use" light illuminates.

It turns off when x-ray generation is terminated.

OUT

IN

G

ROTOR PREP

DETAIL NOT TO SCALE

CONTACTS

RELAY

WITH N.O.

CABINET

SYSTEMS

WLD01.0214

G

NONE

DR400FINALV4

01/12/2017

TYPICAL ROOM LAYOUT

(NOT SITE SPECIFIC)

CB PANEL

POWER BY EC

DATE:

DRAWN BY:

FILE NAME:

SHEET:

NONE

SCALE:

ELECTRICAL CONTRACTOR SHALL FURNISH AND INSTALL ALL CONDUITS, WIRE, WIREWAYS,

DISTRIBUTION TRANSFORMERS, JUNCTION BOXES, ENCLOSURES, CIRCUIT BREAKERS, ETC.

ALL WIRES TO BE THHN TYPE, STRANDED COPPER UNLESS OTHERWISE SPECIFIED. THE

ELECTRICAL CONTRACTOR SHALL SUPPLY LUG PAIRS OF #8 OR GREATER FOR ALL CABLES

CONDUIT AND DUCT RUNS ARE SHOWN SCHEMATICALLY. ACTUAL BUILDING CONDITIONS WILL

DETERMINE CONDUIT AND DUCT ROUTES. CONDUIT TURNS TO HAVE LARGE SWEEPING BENDS

WITH MINIMUM RADIUS AS SPECIFIED IN NEC ART. 346-10. IF LARGE BENDS IN CONDUIT ARE

NOT POSSIBLE, USE SERVICE ENTRANCE ELBOW CONNECTORS TO AVOID DAMAGE TO CABLE.

ELECTRICAL CONTRACTOR TO PROVIDE 10'-0" TAILS FOR ALL WIRES UNLESS OTHERWISE

SPECIFIED AND IDENTIFY BOTH ENDS OF ALL WIRES. ELECTRICAL CONTRACTOR SHALL MAKE

EVERY EFFORT TO ENSURE THAT POINT TO POINT CONDUIT RUNS ARE AS SHORT AS FIELD

WORK SHALL COMPLY WITH REGULATORY AGENCIES OF JURISDICTION. ALL SUCH AGENCIES

SUPERSEDE THESE DRAWINGS AND MUST BE ADHERED TO. THE ELECTRICAL CONTRACTOR

SHALL BE RESPONSIBLE FOR SCRUTINIZING THESE DRAWINGS AND SHALL MAKE CORRECTIONS

NOTE: CONDUITS MAY BE INCREASED IN SIZE BUT NEVER DECREASED WITHOUT CONSENT

RESPONSIBLE FOR INSTALLING THE PATIENT GROUNDING NETWORK PER NATIONAL ELECTRIC

ALL CONDUIT SHALL BE RIGID TYPE UNLESS SPECIFIED OTHERWISE IN THIS DRAWING SET. ALL

Supplying and installing network connections between raceways for containment and routing of networking cables, as determined by site to, the customer's responsibility is the purchase and installation of conduits and / or transceivers, cabling, repeaters, routers, modems, etc. which are consistent with the responsibility. Customer must specify, supply and install proper receptacles, AUI diagnostic imaging equipment and other node connections shall be the customer's

DICOM, PACS, LANS/WANS, Ethernet, Modems site needs and sound network implementation practices. Also included, but not limited

ALL FLUSH MOUNTED ELECTRICAL BOXES AND WIRE TROUGHS SHALL HAVE OVERSIZED COVER

PLATES. ALL FLUSH MOUNTED ELECTRICAL BOXES AND WIRE TROUGHS EXPOSED TO A WET

THE ELECTRICAL CONTRACTOR SHALL INSTALL, PROVIDE POWER TO, AND CONNECT THE

DOOR INTERLOCK SWITCH, EMERGENCY OFF SWITCH AND120VAC DUPLEX OUTLETS FOR

THE ELECTRICAL CONTRACTOR SHALL COORDINATE WITH THE FACILITY AND THE EQUIPMENT

SUPPLIER / INSTALLERS REGARDING SPECIFIC INSTALLATION DETAILS. REFER TO ASSOCIATED

FROM THE EQUIPMENT REPRESENTATIVES. ELECTRICAL CONTRACTOR SHALL BE

CODE ARTICLE 517 AND NFPA 70.

CONDUIT RUNS EXPOSED TO WET ENVIRONMENTS SHALL BE MADE LIQUID TIGHT (e.g.

Networking Notes:

BLOCK DIAGRAMS AND DETAILS.

conditions or required by prevailing codes.

ENVIRONMENT SHALL BE MADE LIQUID TIGHT (e.g. IN FLOOR, ETC.).

FOLLOWING, IF SPECIFIED FOR THIS PROJECT.

SERVICE AND HOUSEKEEPING DESCRIBED IN 'GENERAL NOTES' ITEM #6.

UNDERGROUND, ETC.).

LISTED WITHIN THE SCOPE OF THIS WORK.

WIRING NOTES

LEAVE PULL WIRE IN EMPTY CONDUITS.

ACCORDING TO PREVAILING CODES.

CONDITIONS ALLOW. NEVER LEAVE 'HOT' WIRES CURLED IN PULL BOXES.

IF APPLICABLE, INSTALL THE EQUIPMENT GROUNDING (BONDING) CONDUCTOR (SIZED PER TABLE 250.122) IN A METAL RACEWAY SYSTEM OR A CABLE ARMOR OR SHEATH ASSEMBLY, EACH OF WHICH MUST QUALIFY

CODES. THIS GROUNDING SYSTEM IS REQUIRED IN AREAS WHERE ELECTRICALLY SUSCEPTIBLE PATIENTS MIGHT BE EXAMINED OR TREATED UNDER PRESENT, FUTURE OR EMERGENCY CONDITIONS. CONSULT WITH

SYSTEM. IF SUCH A GROUNDING SYSTEM IS REQUIRED, THE RESPONSIBLE CONSTRUCTION PARTICIPANTS

IF APPLICABLE, USE AN INSULATED COPPER EQUIPMENT GROUNDING (BONDING) CONDUCTOR TO GROUND

- ISOLATED GROUND (IG) RECEPTACLES, IF APPLICABLE.

REQUIREMENTS OR SCHEMATIC.

INTO THEIR WORKING DRAWINGS.

COURTESY TO OUR CUSTOMERS.

THE ELECTRICAL SPECIFICATIONS IN THESE PLANS ARE DEDICATED TO THE

ALL ADDITIONAL EQUIPMENT SUPPLIERS SHALL NOT USE ANY ELECTRICAL

DEVICES CALLED OUT IN THESE PLANS AS A PART OF THEIR ELECTRICAL

ALL ADDITIONAL EQUIPMENT SUPPLIERS SHALL ISSUE A SEPARATE DETAILED

SPECIFICATIONS PACKAGE TO THE ARCHITECT OR TO THE CUSTOMER. THE

ADDITIONAL EQUIPMENT DEPICTED IN THESE PLANS HAVE BEEN SHOWN AS A

ARCHITECT SHALL EXTRAPOLATE THIS INFORMATION AND INCORPORATE IT

EQUIPMENT PROVIDED BY THE X-RAY EQUIPMENT DEALER.

EQUIPMENT SUPPLIERS:

SPECIAL NOTE TO ALL ANCILLARY

ON EQUIPMENT THEY DO NOT PROVIDE.

X-RAY EQUIPMENT DEALER IS NOT RESPONSIBLE FOR FURNISHING INFORMATION

ENERGIZED AND ARE SUBJECT TO PERSONAL CONTACT.

AS AN EFFECTIVE GROUND-FAULT CURRENT PATH PER 250.118 [517.13(A)].

THE GROUNDING TERMINALS OF ALL RECEPTACLES AND ALL NONCURRENT CARRYING CONDUCTIVE SURFACES OF FIXED ELECTRIC EQUIPMENT OPERATING AT OVER 100V, IF THEY ARE LIKELY TO BECOME

- EQUIPMENT GROUNDING.

- REDUNDANT GROUND ADHERENCE.

PATIENT SAFETY GROUNDING CONSIDERATIONS

(CONTRACTOR FURNISHED - CONTRACTOR INSTALLED)

A SPECIAL GROUNDING SYSTEM IS REQUIRED IN ALL DIAGNOSTIC ROOMS BY SOME STATE AND LOCAL

THE GOVERNING ELECTRICAL CODES AND CONFER WITH APPROPRIATE HEALTHCARE FACILITY

PERSONNEL AND CONSULTING ENGINEERS TO DETERMINE THE AREAS REQUIRING THIS GROUNDING

SHALL DESIGN AND INSTALL GROUNDING SYSTEM TO INCLUDE ALL METAL OBJECTS AND FIXTURES PER

NEC ARTICLE 517 CONSIDERATIONS

(CONTRACTOR FURNISHED - CONTRACTOR INSTALLED)

THE PROJECT ENGINEER OF RECORD IS RESPONSIBLE FOR SHOWING COMPLIANCE WITH NEC 517.

- FACILITY GRADE RECEPTACLES IN PATIENT CARE AREAS.

SECTION II - WIRING AND PROTECTION

NEC 517.

IF ANY OF THE ABOVE POWER SPECIFICATIONS REGARDING VOLTAGE REGULATION OR

DISCONNECT WITH 5% VOLTAGE REGULATION UNDER INTERMITTENT NO LOAD TO FULL LOAD

2. VOLTAGE REGULATION: XXX KVA, XXX PHASE, XXX VAC , 60HZ ±0.5% INCOMING LINE TO MAIN

ELECTRICAL CONTRACTOR SHALL USE A DRANETZ POWER LINE ANALYZER OR EQUIVALENT.

B. POWER LINES SHALL BE MEASURED FOR TRANSIENTS WITH THE SYSTEM IN STANDBY. THE

1. TRANSIENTS: NO TRANSIENTS SHALL OCCUR THAT EXCEED 30% OF NOMINAL PEAK LINE

ELECTRICAL EQUIPMENT BE CONNECTED TO THESE POWER LINES NOW OR IN THE FUTURE.

NEAREST FACILITY POWER SOURCE. UNDER NO CIRCUMSTANCES SHALL ANY OTHER

POWER LINES FOR EQUIPMENT LISTED SHALL BE DEDICATED LINES RUN DIRECTLY FROM

PROBLEM MAY BE EMPLOYED TO MAINTAIN LINE VOLTAGE WITHIN 5% OF NOMINAL UNDER

A. A REGULATION TRANSFORMER OR LINE CONDITIONER DEPENDING ON THE SEVERITY OF THE

TRANSIENTS CANNOT BE MET THE FOLLOWING CORRECTIVE ACTION CAN BE TAKEN.

A. TRANSIENTS, FOR DISCUSSION PURPOSES, SHALL BE DEFINED AS IMPULSES FROM 0.5 TO 800

INTERMITTENT NO LOAD TO FULL LOAD CONDITIONS.

CONDITIONS REQUIRED.

LINE REQUIREMENTS:

( CONTRACTOR FURNISHED - CONTRACTOR INSTALLED )

IMPORTANT POWER NOTES

MICROSECONDS.

VOLTAGE.

01/12/2017

DR400FINALV4

01/12/2017

DR400FINALV4