553-CSI-006 RFI answers 1.docx
DOCX document 26 KB Posted
- Attached to
- Z1DA--Site Prep for PET/MRI 553-CSI-006
- Solicitation number
- 36C25020B0041
About this file
This document contains questions and answers related to a solicitation for site preparation work for a PET/MRI equipment upgrade project at the John D. Dingle VA Medical Center in Detroit, Michigan. The contractor will be required to renovate approximately 2,200 square feet of existing space to accommodate new MRI, PET, and patient support rooms. Work includes additions such as uptake/dosing rooms, a hot lab, and storage areas. The contractor must complete interior design elements and provide all necessary labor, materials, and tools to complete the renovations according to drawings and specifications. The North American Industry Classification code for this project is 2389900 and the estimated value is between $1,000,000 and $5,000,000. The solicitation will be issued on June 11, 2020 and the full bid documents will be available on beta.SAM.gov with responses due by the specified date in the solicitation. The contract will be a firm fixed-price awarded within 120 days of bid receipt.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C25020B0041 0009.docx | DOCX document | |
| 36C25020B0041 0008.docx | DOCX document | |
| 36C25020B0041 0007.docx | DOCX document | |
| 19.1010 - 2020.07.16 - Addendum 1 Sheet AS302.pdf | ||
| 36C25020B0041 0006.docx | DOCX document | |
| 36C25020B0041 0005.docx | DOCX document | |
| 36C25020B0041 0004.docx | DOCX document | |
| 553-CSI-006 RFI answers.docx | DOCX document | |
| 553-CSI-006 Attendance Log .pdf | ||
| 36C25020B0041 0003.docx | DOCX document | |
| 553-CSI-006 - Addendum 1 Sheet AS302.pdf | ||
| 553-CSI-006 - Drawing Set.pdf | ||
| 36C25020B0041 0001.docx | DOCX document | |
| 36C25020B0041.docx | DOCX document | |
| Project 553-CSI-006 - Spec Vol 1 Div00 14.pdf | ||
| Wage Determination M120200101.txt | TXT text file | |
| Project 553-CSI-006 - Bid Documents Drawing Set.pdf | ||
| Contractor EMR Certification Form 2020.docx | DOCX document | |
| Project 553-CSI-006 - Spec Vol 2 Div21 28.pdf |
Show all 19
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
36C25020B0041 – Project 553-CSI-006 MRI/PET Equipment - Detroit VAMC
1. Door assembly 1680 including door, frame, and all hardware shall be provided by shielding vendor. See attached Revised Sheet AS302.
Answer: Per Window Schedule, window ‘R’ shall be provided by shielding vendor.
2. Not all drawings listed on page G1000 are in bid set Answer: All Drawings on GI000 will be provided to bidders.
3. Please identify existing RF shielding material and provide specification:
Answer: RF Shielding is typically copper. Per AS101, contractor’s shielding vendor shall perform testing of existing and new RF shielding to determine that it meets requirements of new equipment.
4. Has an RF shielding report be conducted of existing shielding to verify integrity?
Answer: Per AS101, contractor’s shielding vendor shall perform testing of existing and new RF shielding to determine that it meets requirements of new equipment
5. Is magnetic shielding required for Siemens new equipment.
Answer: No Magnetic shielding is required. Radiation (lead) and RF shielding is required per contract documents.
6. Siemens drwg. M501 calls for new quench vent. Drwg.AS 200 note 10 calls existing to remain please clarify.
Answer: It is the understanding of the AE that portions of the quench vent shall be existing to remain. See MEP and Vendor drawings for additional information.
7. Many of the drawings that are listed on Drawing G1000 have NOT been provided on BetaSam. The only drawings that were distributed are the Guidon Architectural Drawings. None of the other drawings; Fire Protection, Plumbing, Mechanical and Electrical Drawings have been made available (28 drawings missing). The Siemen Drawings have been distributed. The Medical Physical Consultants Inc. drawings were not posted as well.
Answer: Drawings listed on GI000 will be provided to bidders.
8. We have assumed and we need confirmation that the Siemen Drawings and Medical Physics Consultant Drawings are for reference only and that the Guidon Drawings only describe the work of this contract.
Answer: The Siemens drawings and Medical Physics Consultant Drawings are part of the contract documents.
9. We cannot find any specification for the lockers as shown on Drawing AQ101, Item 102.
Answer: Lockers referenced on AQ101 shall be JSN #A1030, Metal lockers. 2-person 72"h x 12"w x 18"d. Users shall provide locking mechanism.
10. We cannot find any specification for the Waste Disposal Unit on Drawing AQ101, Item 103 Answer: Waste disposal unit shall be provided by VA.
11. In Specification Section 27 52 23, Nurse Call and Code Blue Systems a reference is made to Specification Section 27 11 00. This specification section is not included in the documents although it is listed in the Specification Index.
Answer: Disregard specification Section 27 11 00.
12. We have submitted RFis # I though #4 previously and they have not been answered. We request that this project' s due date be postponed until the RFis are answered.
Answer: date will be revised as needed.
BID Opening Call in information:
Thursday August 27, 2020 at 1:00PM, Following the BID submission information in the solicitation:
There is NO visitor access to Dominos Farm. Contracting personnel will be waiting by the entrance to Lobby M to receive dropped off bids. You may drop off your bid, observing social distancing, and call in from your vehicles.
Join Skype Meeting Trouble Joining? Try Skype Web App Join by phone
| 844-825-8490,67259051# (South) | English (United States) |
| 844-352-6288,67259051# (South) | English (United States) |
| 844-355-6288,67259051# (South) | English (United States) |
Find a local number Conference ID: 67259051
File details come from the government source that posted it. Updated .