Salient Characteristics

Control Room:

· Climate controlled space with AC/heating with thermometer separate from procedure room.

· Phillips XPER Hemodynamic Workstation with two monitors, 1 for hemodynamics and 1 for recording case details.

· Phillips monitor for live x-ray recording

· Excelera cardiology PACS viewing station (can substitute with WITT)

· Monitor for IVUS/FFR

· Network hook up for transferring FFR images from OCT/TVC systems

· Power and data outlets to support 3 additional VA computers and one printer/copier/fax

· Speaker for intercom for procedure room.

· 2 phones, several phone lines

· Emergency lighting for power outages

· Hard coat flooring non-carpet

· Lead Apron storage rack mounted to wall (8-10 Aprons).

Lab/Procedure Room:

· Phillips X-ray System (Philips Clarity Allura or equivalent system)

· The system must be compatible with our current FFR, IVUS, TVC, OTC and Xcelera

· (4) monitors on boom: 1 hemodynamic 1 IVUS, live x-ray, 1 freeze frame

· 4 pressure channels and cardiac output

· CO2 monitoring

· Acist power injector with outlet

· Suction

· Oxygen (cylinders can be used but will need sufficient location to store)

· Code cart space with defibrillator with outlet.

· Code blue/emergency alert button, wired to main building.

· Data outlets to support two additional computers

· Hard coat seamless flooring

· AC/Heat and climate controlled room with thermostat

· Electrical Outlets all on emergency backup generator.

· Emergency lighting in case of power outage

Power Requirements The contractor will be required to provide and install necessary power connectors, wiring distribution, and any ancillary items needed to connect the unit to the power source.

The government is prepared to provide a single electrical power source as required (3/N/PE AC 480V service fused at 200 amperes 60Hz ±1.0Hz)

Maximum Voltage Variation The contractor will analyze the power source to ensure it meets with power requirements for the unit. The government will be responsible for ensuring the quality (voltage, amperes, and voltage variations) to meet the requirements of the unit.

Mobile Grounding The contractor shall be responsible for properly grounding the unit. The contractor shall purchase all necessary grounding material and provide the labor for grounding.

Telephone and Data Service Requirements The mobile unit is supplied with four (4) telephone connections.

The contractor shall purchase and install four (4) Hubbell all weather telephone connections.

Data Service The contractor shall provide twelve (12) data line connections that utilize RJ-45 outlets.

The contractor shall provide the data connection cables for use with the data line connections. The additional connections will provide data flow for:

· XPER Hemodynamic Workstation

· Xcelera PACS viewing workstation

· Network hook up for their Mobile OCT/TVC system

· Pharmacy Pyxis

· GE Telligence Nurse Call System

Code Blue and Code Red Notification To facilitate the “Code Blue” system; the contractor will provide a duty station with connection to the GE Telligence Nurse Call System. This will require a data line.

Sanitary Sink The mobile unit must provide sanitary sink for us by system operators and patients as needed.

Water The contractor shall be required to make all necessary water and sewage connections as part of the setup process. The government will provide a suitable water source and a suitable sewage disposal point.

X-Ray System Requirements The mobile unit must provide Philips Clarity Allura or equivalent x-ray system installed.

· Poly-G Floor mounted G-arm geometry

• Field of view: 25cm diagonal square

• Support of the cardiac workflow through the Xper platform: a unique workflow

• Concept that allows personalization of user settings

• DoseWise for low X-ray dose with excellent image quality

• Functionality integration in one intuitive user interface at the tableside

• 100KW generator

• AD5 cardiac table

• 10” digital flat panel detector

• MRC 0508 grid switch X-ray tube

• (3) high resolution procedure room monitors with Ceiling-mounted suspension

• FD-10 digital system with DICOM output

Power Consumption

• General system : 480 VAC; 60 Hz; 3-phase

• X-ray generator: Supplied by system

Siting Requirements

• Room Size Needed 7.6m x 5.3m (24.9 ft x 17.4 ft)

• Minimum Room Height Needed: 2.9 m (9.5 ft.)

Supplementals

• 3D-RA, XperCT, 3D roadmap, XperGuide, StentBoost, 3D-CA

• Exposure control

• Last image hold: Yes

• Virtual collimation using lih: Yes

• Auto-adjustable copper filtration: Yes

• Dose monitoring: Yes

Imaging Features

• 512 x 512 60 fps

• 1024 x 1024 15 and 30 fps

• Other 2048 x 2048, 0.5 to 6 fps

X-ray generator

• Power rating (kw): @ 100 kvp 100

• Radiographic (ma): 1-1,250

• Radiographic (kv): 40-125

• Radiographic timer: 2 ms-4 sec

• Fluoroscopic (ma): 60 mA nominal for pulsed fluoro

• Fluoroscopic (kv): 40-125

• Pulsed fluoroscopy: Grid switched

• Cine camera range: 15, 30, and 60 fps optional

• Full in-room control: Yes

X-Ray Tube

• Maximum Number of X-ray tubes: 1

• Heat storage capacity: 5400000 hu

• Heat dissipation rate: 910000 hu/min

• Focal spot size: 0.5/0.8 mm (MRC)

• Grid pulsed: Yes

• Target angle: 9 degrees

• Cooling method: Recirculating oil

Detector

• Generator: aSi detector with CsI scintillator

• Dimensions: 40cm x 30cm (15.7 in. x 11.8 in.)

• Pixel Size: 154mm x 154mm

• Bit depth: 14

• Max frame rate: 60 fps

Image storage

• 1024 x 1024, Images: 50000

• Additional Storage: up to 100,000 images

• Digital imaging system: Integrated

Patient Table

• Maximum patient load: 250 kg (550 lbs) with additional 50 kg (110 lbs) for CPR

• Movement: 4-way floating, Xper table

• Dimensions (l x w): 319 cm x 50 cm (125.5 in. x 20 in.)

• Vertical: 79-107 cm (31-42 in.)

• Lateral: +/-18 cm (7.1 in.)

• Longitudinal: 125 cm (49.2 in.)

Decking

· Existing decking extends from the loading dock along the right side of the trailer. The purpose is to provide a level surface to bring patients and equipment through the roll door. There is a 98 inch opening that starts at the 27 foot line that allows for the lift gate to drop down smoothly with access to the roll door.