This presolicitation notice is for a project to construct a Hybrid Operating Room at the Hunter Holmes McGuire VA Medical Center in Richmond, Virginia. The project has an estimated value between $1,000,000 and $2,000,000, and will be solicited as a 100% Service Disabled Veteran Owned Small Business set-aside under NAICS code 236220. Interested vendors must be registered in both the System for Award Management and VetBiz databases. The solicitation is expected to be released on or around March 2, 2020, with proposals due at least 30 days later. The site visit location, date, and points of contact will be provided in the solicitation. Prospective contractors should monitor the website for any amendments or notices.
36C24620R0046 0007 RFI Response Attachment 1 - Steris Drawings.pdf
View the file
Other files
Show 40 more
Latest 50 of 79 listed here. See the rest on GovTribe.
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
HARMONY iQ3600 INSTALLATION GUIDELINES
VA MEDICAL CENTER OF RICHMOND
RICHMOND, VIRGINIA
INDEX
PAGE # SHEET TITLE ROOM #
1 TITLE PAGE, INDEX -
2 PROPOSED OPERATING ROOM LAYOUT -
3 MOUNTING ASSEMBLY FOR A1&A2 (iQ3600) ATOM DETAILS -
4 PROPOSED OPERATING ROOM CONDUIT RISER DIAGRAM -
5 PROPOSED CONTROL ROOM CONDUIT RISER DIAGRAM -
6 TYPICAL OPERATING ROOM INSTALLATION DETAIL -
7 TYPICAL BOSE CEILING SPEAKER (G1-G4) DETAILS -
8 CALL & CONNECT HUB DETAIL -
ABBREVIATIONS
A.F.F. = ABOVE FINISHED FLOOR
B.C.P. = BLANK COVER PLATE
C. = CONDUIT
C.P. = COVER PLATE
C.K.T. = CIRCUIT
E.C. = ELECTRICAL CONTRACTOR
I.G. = ISOLATED GROUND (TECHNICAL GROUND)
N.S.E. = NURSE STATION EQUIPMENT
GENERAL NOTES
1. SUPPLY AND INSTALLATION OF ALL CONDUIT, POWER CIRCUITS, BACK BOXES, ELECTRICAL DEVICES, ELECTRICAL WIRING IS TO BE
PROVIDED BY THE CLIENTS ELECTRICAL CONTRACTOR UNLESS NOTED OTHERWISE
2. ONE WIRE HANGER FOR CABLE COILS REQUIRED FOR EACH MOUNT
3. CLIENT CONTRACTOR TO PROVIDE BLANK COVER PLATES FOR ALL JUNCTIONS BOXES WHERE NECESSARY
4. ALL CABLE RUNS OUTSIDE INTEGRATED ROOMS ARE THE RESPONSIBILITY OF HOSPITAL/CONTRACTOR; ALL CABLE RUNS SHOULD
BE CONTINUOUS EXCEPT WHERE NOTED
5. 24x24" ACCESS MUST BE PROVIDED WITHIN 36" OF ALL JUNCTION BOXES
6. CABLE RUNS THRU PLENUM SPACE MUST BE PLENUM RATED OR SHALL BE INSTALLED WITH CONDUIT
7. CABLES NOT RUN IN CONDUIT MUST HAVE J-HOOKS EVERY 4' FOR SUPPORT (NOT BY STERIS)
8. WHERE J-HOOKS ARE USED IN CEILING, CONDUIT IS REQUIRED IN ALL CONCEALED LOCATIONS (SIZED PER SCHEDULES)
9. EVERY RACK REQUIRES 4 LAN CONNECTIONS (DATA PORTS) ADJACENT TO T1 BOX; EACH OR REQUIRES 2 NETWORK PORTS AT RACK
10. ALL POWER FOR INTEGRATION EQUIPMENT SHOULD COME FROM A SINGLE ELECTRICAL PANEL IN THE OR IT SERVES
11. CONFORMANCE TO NATIONAL, STATE AND LOCAL CODES FOR ELECTRICAL CIRCUITS AND CONDUIT IS THE RESPONSIBILITY OF THE
HOSPITAL OR THEIR CONTRACTED ELECTRICIAN
12. THE EQUIPMENT AND ROUGH-IN REQUIREMENTS SHOWN ARE FOR BUDGETARY PLANNING PURPOSES ONLY; EQUIPMENT IS SUBJECT TO CHANGE AND DIFFERENT OPTIONS ARE AVAILABLE TO MEET SPECIFIC CUSTOMER DESIGN CRITERIA; POST-ORDER DRAWINGS WILL REFLECT
EXACT EQUIPMENT SPECIFICATIONS, AND REQUIRES CUSTOMER SIGNED APPROVAL
13. USE OF ISOLATED GROUND POWER TO BE DETERMINED BY ELECTRICAL CONTRACTOR
REVISION HISTORY
REVISION # DESCRIPTION PLANNER/DATE
1 UPDATED C1 MOUNT; E1 & E2 MADE FUTURE; MOVED NURSE
DESK
JF/30SEP2019
36-01 iQ3600 TITLE PAGE, INDEXPlanning & Design Group
STERIS
STERIS Corporation
5960 Heisley Road
Mentor, OH 44060 USA
STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD
Dr
Ck
Date
Scale
H- B-
800-548-4873 NA
DATE
LL
KS
C3R1-INT
11224
PAGE: OF 36-08
VA MEDICAL CENTER OF RICHMOND
RICHMOND, VIRGINA
Ken Jones Text Box
HYBRID OR
BID QUESTION RFI'S
ATTACHMENT 1
A1 A2
E4 E5
SHEET NOTES
1. INTEGRATED SYSTEM (A1) AND VIDEO CONFERENCING BOX (A2); REQUIRES DEDICATED
EMERGENCY BACK-UP 120V ON 20 AMP POWER FROM THE O.R. IT SERVES (NOT BY STERIS)
2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS
3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE RIGHT
HALF OF THE BOX
4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE
AROUND ALL SIDES OF THE BOX
5. AUDIO REQUIREMENT; HOSPITAL GRADE HUBBELL USB CHARGER RECEPTACLE DUPLEX
(ACCEPTABLE MODEL NUMBERS: #SB8300LA, #USB8300R, #USB8300W OR SIMILAR; NOT BY
STERIS)
6. TOUCH PANEL REQUIRES DUPLEX OUTLET 120 VAC ON 20 AMP SERVICE (STANDARD OR
EMERGENCY)
LIGHT CONDUIT SCHEDULE NOTES
1. 1/2" CONDUIT FOR AC POWER TO K1 & K2 BY E.C.
2.LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET
3.CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO
REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE SPECIFIED
SURGICAL LIGHT WALL CONTROL SYSTEM
4.DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN DOUBLE
RUNS ARE REQUIRED (LONG RUNS)
***ALL LIGHT MOUNTS (B) REQUIRE 2 J-HOOKS EACH, 8"
APART -OR- EQUIVALENT SUSPENSION POINTS ABOVE THE
STRUCTURAL MOUNTING PLATE
E SERIES LIGHT CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
K1 - K2 1 3/4" ELECTRICAL
CONTRACTOR
K1 - B1* 1 3/4" ELECTRICAL
CONTRACTOR
K1 - B1 1 3/4" ELECTRICAL
CONTRACTOR
K2 - B1* 1 3/4" ELECTRICAL
CONTRACTOR
CONDUIT SCHEDULE
FROM - TO QUANTITY SIZE RESPONSIBILITY
A1 - A2 1 1" ELECTRICAL CONTRACTOR
T1 - A1 2 2" ELECTRICAL CONTRACTOR
T1 - A2 1 2" ELECTRICAL CONTRACTOR
T1 - B1 1 1 1/2" ELECTRICAL CONTRACTOR
T1 - B2 1 1 1/2" ELECTRICAL CONTRACTOR
T1 - C1 1 1 1/2" ELECTRICAL CONTRACTOR
T1 - C2 1 1 1/2" ELECTRICAL CONTRACTOR
T1 - C3 1 1 1/2" ELECTRICAL CONTRACTOR
T1 - CORRIDOR 1 1 1/2" ELECTRICAL CONTRACTOR
T1 - E1 1 1" ELECTRICAL CONTRACTOR
T1 - E2 1 1" ELECTRICAL CONTRACTOR
T1 - E3 1 1" ELECTRICAL CONTRACTOR
T1 - E4 1 1" ELECTRICAL CONTRACTOR
T1 - E5 1 1" ELECTRICAL CONTRACTOR
T1 - F1 1 1 1/2" ELECTRICAL CONTRACTOR
T1 - G1 1 3/4" ELECTRICAL CONTRACTOR
T1 - G3 1 3/4" ELECTRICAL CONTRACTOR
T1 - G2 1 3/4" ELECTRICAL CONTRACTOR
T1 - G4 1 3/4" ELECTRICAL CONTRACTOR
T1 - K1 1 1" ELECTRICAL CONTRACTOR
T1 - N1 2 2" ELECTRICAL CONTRACTOR
T1 - U2 1 1 1/4" ELECTRICAL CONTRACTOR
T1 - V2 2 1 1/2" ELECTRICAL CONTRACTOR
T1 - T2 1 2" ELECTRICAL CONTRACTOR
T2 - CR1 1 1 1/4" ELECTRICAL CONTRACTOR
T2 - CR2 1 1 1/4" ELECTRICAL CONTRACTOR
T2 - CR3 1 1 1/4" ELECTRICAL CONTRACTOR
T2 - CRQ1 1 1" ELECTRICAL CONTRACTOR
T2 - CRTP1 1 1 1/4" ELECTRICAL CONTRACTOR
HOSPITAL
NETWORK 1 1" ELECTRICAL CONTRACTOR
N1 - U1 1 1 1/4" ELECTRICAL CONTRACTOR
N1 - TP1 1 1 1/4" ELECTRICAL CONTRACTOR
C1*** MONITOR CARRIER STERIS
C2*** EQUIPMENT BOOM STERIS
C3*** ANESTHESIA BOOM STERIS
CR1
WALL MOUNTED HUBBELL - RACO 698 4 GANG
JUNCTION BOX FOR GUEST PORTS (2x) ELECTRICAL CONTRACTOR
CR2 WALL MOUNTED HUBBELL - RACO 698 4 GANG
JUNCTION BOX FOR MACLAB ELECTRICAL CONTRACTOR
CR3
WALL MOUNTED HUBBELL - RACO 698 4 GANG
JUNCTION BOX FOR PACS PC (2x) ELECTRICAL CONTRACTOR
CRQ1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR
CRTP1
TOUCH PANEL STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR
E1
FUTURE WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR WALL
MOUNTED DISPLAY
ELECTRICAL CONTRACTOR
E2
FUTURE WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR WALL
MOUNTED DISPLAY
ELECTRICAL CONTRACTOR
E3
WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR WALL
MOUNTED DISPLAY
ELECTRICAL CONTRACTOR
E4
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING FOR SIEMENS LIVE & REF
(DVI)
ELECTRICAL CONTRACTOR
E5
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING FOR 3RD PARTY DVI HD
DISPLAY OR DESTINATION
ELECTRICAL CONTRACTOR
F1
WALL MOUNTED HD PTZ CAMERA STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR WALL
MOUNTED HD PTZ CAMERA
ELECTRICAL CONTRACTOR
G1 CEILING MOUNTED SPEAKER STERIS
G3 CEILING MOUNTED SPEAKER STERIS
G2 CEILING MOUNTED SPEAKER STERIS
G4 CEILING MOUNTED SPEAKER STERIS
K1 SURGICAL LIGHT CONTROLLER STERIS
K2 SURGICAL REMOTE POWER MODULE STERIS
N1 WALL MOUNTED 12x12x4" JUNCTION BOX ELECTRICAL CONTRACTOR
T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE
CEILING ELECTRICAL CONTRACTOR
T2 12" H x 12" W X 6" D JUNCTION BOX ABOVE
CEILING ELECTRICAL CONTRACTOR
TP1
TOUCH PANEL STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR
U1 UNIVERSAL PORT - WALL MOUNTED 4 11/16"
JUNCTION BOX WITH SINGLE GANG MUD RING ELECTRICAL CONTRACTOR
U2 UNIVERSAL PORT - WALL MOUNTED 4 11/16"
JUNCTION BOX WITH SINGLE GANG MUD RING ELECTRICAL CONTRACTOR
V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER
DISTRIBUTION ELECTRICAL CONTRACTOR
E3
CRTP1 CR3
CRTP1
T1 V2
G4 G3
SPEAKER PLACEMENT
(G1-G4) FOR REFERENCE
ONLY; SITE VERIFY
G1 G2
F1
K1 K2
NURSE
DESK
T2
TP1
N1 TP1
CR1
U1
U2
CR2
CRQ1
B1
B2
C1
C2
C3
E2 E1
INTEGRATION EQUIPMENT SCHEDULE
LEGEND DESCRIPTION RESPONSIBILITY
A1 STERIS iQ3600 INTEGRATION SYSTEM STERIS
A2 VIDEO CONFERENCE BOX STERIS
B1*** SURGICAL LIGHT MOUNT STERIS
B2*** SURGICAL LIGHT MOUNT STERIS
36-02 iQ3600
LAYOUT
PROPOSED OPERATING ROOMPlanning & Design Group
STERIS
STERIS Corporation
5960 Heisley Road
Mentor, OH 44060 USA
STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD
Dr
Ck
Date
Scale
H- B-
800-548-4873 NA
DATE
LL
KS
C3R1-INT
11224
PAGE: OF 36-08
VA MEDICAL CENTER OF RICHMOND
RICHMOND, VIRGINA
AutoCAD SHX Text
EQUIP. ROOM
AutoCAD SHX Text
CONTROL ROOM
AutoCAD SHX Text
HYBRID OR
REQUIRES DEDICATED 120V ON
20 AMP
CIRCUIT FROM THE O.R. IT
SERVES
(ON EMERGENCY BACK-UP
POWER;NOT BY STERIS)
HEAT LOADS:
iQ3600:
1200 BTU/Hr TO BE MAINTAINED AT 78 DEGREES F. OR LESS
POWER MODULE: (INCLUDING POWERING MONITORS):
450 BTU/Hr TO BE MAINTAINED AT 78 DEGREES F. OR LESS
4 LAN CONNECTIONS REQUIRED ADJACENT TO T1
.65
(ABOVE WALL)
42.00
3.00
29.50
3.91
(INSIDE WALL)
1.95
2.81
2X Ø 1.50
2X 1.71
2X 1.96
2X 2.91
14.18
(14.33 WITH ADJUSTABLE L-BRACKET)
0.5" TO 1.50" - ADJUSTABLE
(DRYWALL THICKNESS)
4X Ø1.002
2X Ø 2.00
ELEVATION A
iQ3600 ELECTRICAL BOX
FINISHED CEILING
iQ3600
STERIS
INTEGRATION/VIDEO
CONFERENCING BOX
17.55
FINISHED FLOOR
36-03 iQ3600
A1,A2 (iQ3600) ATOM DETAILS MOUNTING ASSEMBLY FORPlanning & Design Group
STERIS
STERIS Corporation
5960 Heisley Road
Mentor, OH 44060 USA
STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD
Dr
Ck
Date
Scale
H- B-
800-548-4873 NA
DATE
LL
KS
C3R1-INT
11224
PAGE: OF 36-08
RICHMOND, VIRGINA
B2B1 C2C1
SURGICAL
LIGHT MOUNT
SURGICAL
LIGHT MOUNT
SURGICAL
LIGHT HEAD
(3x)
RAD SHIELD
ANESTHESIA
BOOM
LARGE FORMAT
MONITOR
/EQUIPMENT
BOOM
UNIVERSAL
PORT
VITALS
JUNCTION BOX SCHEDULE
J-BOX # SIZE TYPE MOUNTING
N1 12"H X 12"W X 4"D PULL BOX IN WALL
T1 24"H X 24"W X 8"D NEMA 1, SCREW
COVER ABOVE CEILING
JUNCTION BOX SCHEDULE NOTES
1. ALL BOXES ARE PROVIDED BY THE CLIENTS CONTRACTOR UNLESS NOTED OTHERWISE
2. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS
OTHERWISE NOTED/DETAILED
3. ALL BOXES AND COVER PLATES ARE PROVIDED BY THE CLIENTS CONTRACTOR UNLESS
NOTED OTHERWISE
4. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS
OTHERWISE NOTED/DETAILED
5. BOXES/OUTLETS/RECEPTACLES LOCATED ABOVE THE CEILING MUST BE WITHIN 18
INCHES OF THE CEILING AND BE ACCESSIBLE FROM THE SPACE BELOW; CLIENT/G.C.
TO PROVIDE SUITABLE CEILING ACCESS PANELS AT DEVICE LOCATIONS
6. T1 JUNCTION BOX - 24" x 24" x 8" (PROVIDED AND MOUNTED VERTICALLY NOT BY
STERIS); ALL FIELD CONDUITS AND 2" & 1" CONDUIT TO (A1) STERIS INTEGRATION
RACK ATTACH TO JUNCTION BOX (T1); CEILING ACCESS REQUIRED FOR SERVICE
ACCESS; ALL CONDUIT TO ENTER T1 BOX ANYWHERE EXCEPT FOR THE BACK SIDE; T1
BOX SHOULD NOT BE HIGHER THAN 12' ABOVE FINISHED FLOOR;
NOT TO BE LOCATED DIRECTLY ABOVE NURSE DESKS OR ANY OTHER FIXED EQUIPMENT
CONDUIT NOTES
1. IF A DIRECT CONNECTION IS NEEDED BETWEEN TWO PIECES OF
EQUIPMENT, CONDUIT MUST BE RUN
2. EXACT CONDUIT PATH DETERMINED BY ELECTRICAL CONTRACTOR
3. CONDUIT SHOULD NOT EXCEED 50'
4. MUST BE A PULL BOX FOR EVERY 270° OF CONDUIT BEND
5. ALL CONDUITS SHOULD BE PROVIDED WITH PULL STRINGS AND MUST
HAVE INSULATED BUSHINGS ON ALL OPEN ENDS
6. USE OF FLEXIBLE CONDUIT IS NOT ACCEPTABLE
7. STERIS WILL NOT BE RESPONSIBLE FOR FIRE STOPPING IN ANY FORM
ABBREVIATIONS
A.F.F. = ABOVE FINISHED FLOOR
B.C.P. = BLANK COVER PLATE
C. = CONDUIT
C.P. = COVER PLATE
C.K.T. = CIRCUIT
E.C. = ELECTRICAL CONTRACTOR
I.G. = ISOLATED GROUND (TECHNICAL GROUND)
N.S.E. = NURSE STATION EQUIPMENT
RISER LEGEND
SYMBOL DESCRIPTION MOUNTING/COMMENT
AP CEILING ACCESS PANEL FOR REQUIRED ACCESS TO BOXES
B1 EQUIPMENT LABEL DENOTED CEILING MOUNT
N1 CONDUIT OR CABLE RUN JUNCTION BOX REFERENCE
CONDUIT STUB OUT CONDUIT STUB OUT WITH BUSHING
G1 CEILING SPEAKER CEILING, FLUSH, RECEPTACLE HEIGHT
DUPLEX I.G. POWER RECEPTACLE WALL, FLUSH, RECEPTACLE HEIGHT
QUAD I.G. POWER RECEPTACLE WALL, FLUSH, RECEPTACLE HEIGHT
DT DATA/TELEPHONE BOX WALL, FLUSH, RECEPTACLE HEIGHT
T1
NURSE
DESK
NURSE
DESK
U1
UNIVERSAL
PORT
TOUCH
PANEL
1 1/4"C.
TP1
P W R
C K T
I.G.
P W R
C K T
I.G.
NURSE
DESK
(2)2"C.
(N1)12x12
JUNCTION
BOX
1 1/2"C.
1 1/2"C.
1 1/2"C. 1 1/2"C.1"C.
K1
3/4"C.
1/2"C.
I N C O M I N G
H O S P
P W R
REMOTE
POWER
MODULE
SURGICAL
LIGHT
CONTROLLER
K2 (2)2"C.
A1 iQ3600
STERIS
INTEGRATION
BOX
A2 iQ3600
STERIS
VIDEO
CONFERENCING
BOX
SEE MOUNTING
ASSEMBLY PAGE
FOR POWER
REQUIREMENTS
1"C.
2"C.
(F1)HD PTZ
CAMERA
I.G.
P W R
C K T
3/4"C.
G1
3/4"C. STUB
WITHIN 18"
OF SPEAKER
G3
3/4"C.
(E2)FUTURE
WALL
DISPLAY
1"C.
I.G.
P W R
C K T
(E3)WALL
DISPLAY
1"C.
1 1/2"C.
3/4"C.
G2
3/4"C. STUB
WITHIN 18"
OF SPEAKER
G4
3/4"C.
E4
I.G.
P W R
C K T
I.G.
P W R
C K T
SIEMENS
LIVE &
REF
1"C.
(V2)
LEVITON
BOX
P W R
C K T
(2)1 1/2"C.
V2 REQUIRES DEDICATED
EMERGENCY BACK-UP 120V
ON 20 AMP POWER (NOT
BY STERIS)
UNIVERSAL
PORT
1 1/4"C.
U2
SURGICAL
DISPLAY
ENDO
CAMERA
HEADLIGHT
CAM
UNIVERSAL
PORTS (2x)
IN-LIGHT
CAMERA
C3
1 1/2"C.
(E1)FUTURE
WALL
DISPLAY
1"C.
E5
3RD PARTY
DVI
1"C.
1 1/4"C.
LARGE
FORMAT
MONITOR
SHEET NOTES
1. INTEGRATED SYSTEM (A1) AND VIDEO CONFERENCING BOX (A2); REQUIRES DEDICATED
EMERGENCY BACK-UP 120V ON 20 AMP POWER FROM THE O.R. IT SERVES (NOT BY STERIS)
2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS
3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE RIGHT
HALF OF THE BOX
4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE
AROUND ALL SIDES OF THE BOX
5. AUDIO REQUIREMENT; HOSPITAL GRADE HUBBELL USB CHARGER RECEPTACLE DUPLEX
(ACCEPTABLE MODEL NUMBERS: #SB8300LA, #USB8300R, #USB8300W OR SIMILAR; NOT BY
STERIS)
6. TOUCH PANEL REQUIRES DUPLEX OUTLET 120 VAC ON 20 AMP SERVICE (STANDARD OR
EMERGENCY)
K2 SURGICAL REMOTE POWER MODULE STERIS
N1 WALL MOUNTED 12x12x4" JUNCTION BOX ELECTRICAL CONTRACTOR
T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE
CEILING ELECTRICAL CONTRACTOR
T2 12" H x 12" W X 6" D JUNCTION BOX ABOVE
CEILING ELECTRICAL CONTRACTOR
TP1
TOUCH PANEL STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR
U1 UNIVERSAL PORT - WALL MOUNTED 4 11/16"
JUNCTION BOX WITH SINGLE GANG MUD RING ELECTRICAL CONTRACTOR
U2 UNIVERSAL PORT - WALL MOUNTED 4 11/16"
JUNCTION BOX WITH SINGLE GANG MUD RING ELECTRICAL CONTRACTOR
V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER
DISTRIBUTION ELECTRICAL CONTRACTOR
INTEGRATION EQUIPMENT SCHEDULE
LEGEND DESCRIPTION RESPONSIBILITY
A1 STERIS iQ3600 INTEGRATION SYSTEM STERIS
A2 VIDEO CONFERENCE BOX STERIS
B1*** SURGICAL LIGHT MOUNT STERIS
B2*** SURGICAL LIGHT MOUNT STERIS
C1*** MONITOR CARRIER STERIS
C2*** EQUIPMENT BOOM STERIS
C3*** ANESTHESIA BOOM STERIS
CR1
WALL MOUNTED HUBBELL - RACO 698 4 GANG
JUNCTION BOX FOR GUEST PORTS (2x) ELECTRICAL CONTRACTOR
CR2 WALL MOUNTED HUBBELL - RACO 698 4 GANG
JUNCTION BOX FOR MACLAB ELECTRICAL CONTRACTOR
CR3
WALL MOUNTED HUBBELL - RACO 698 4 GANG
JUNCTION BOX FOR PACS PC (2x) ELECTRICAL CONTRACTOR
CRQ1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR
CRTP1
TOUCH PANEL STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR
E1
FUTURE WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR WALL
MOUNTED DISPLAY
ELECTRICAL CONTRACTOR
E2
FUTURE WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR WALL
MOUNTED DISPLAY
ELECTRICAL CONTRACTOR
E3
WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR WALL
MOUNTED DISPLAY
ELECTRICAL CONTRACTOR
E4
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING FOR SIEMENS LIVE & REF
(DVI)
ELECTRICAL CONTRACTOR
E5
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING FOR 3RD PARTY DVI HD
DISPLAY OR DESTINATION
ELECTRICAL CONTRACTOR
F1
WALL MOUNTED HD PTZ CAMERA STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR WALL
MOUNTED HD PTZ CAMERA
ELECTRICAL CONTRACTOR
G1 CEILING MOUNTED SPEAKER STERIS
G3 CEILING MOUNTED SPEAKER STERIS
G2 CEILING MOUNTED SPEAKER STERIS
G4 CEILING MOUNTED SPEAKER STERIS
K1 SURGICAL LIGHT CONTROLLER STERIS
36-04 iQ3600
CONDUIT RISER DIAGRAM
PROPOSED OPERATING ROOMPlanning & Design Group
STERIS
STERIS Corporation
5960 Heisley Road
Mentor, OH 44060 USA
STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD
Dr
Ck
Date
Scale
H- B-
800-548-4873 NA
DATE
LL
KS
C3R1-INT
11224
PAGE: OF 36-08
JUNCTION BOX SCHEDULE
J-BOX # SIZE TYPE MOUNTING
T2 12"H X 12"W X 6"D PULL BOX ABOVE CEILING
JUNCTION BOX SCHEDULE NOTES
1. ALL BOXES ARE PROVIDED BY THE CLIENTS CONTRACTOR UNLESS NOTED OTHERWISE
2. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS
OTHERWISE NOTED/DETAILED
3. ALL BOXES AND COVER PLATES ARE PROVIDED BY THE CLIENTS CONTRACTOR
UNLESS NOTED OTHERWISE
4. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS
OTHERWISE NOTED/DETAILED
5. BOXES/OUTLETS/RECEPTACLES LOCATED ABOVE THE CEILING MUST BE WITHIN 18"
OF THE CEILING AND BE ACCESSIBLE FROM THE SPACE BELOW; CLIENT/G.C. TO
PROVIDE SUITABLE CEILING ACCESS PANELS AT DEVICE LOCATIONS
6. T2 JUNCTION BOX - 12" x 12" x 6" (MOUNTED VERTICALLY; NOT BY STERIS);
ALL FIELD CONDUITS AND 2" CONDUIT TO 24x24x8" (T1); CEILING ACCESS
REQUIRED FOR SERVICE ACCESS; ALL CONDUIT TO ENTER T2 BOX ANYWHERE EXCEPT
FOR THE BACK SIDE; T2 BOX SHOULD NOT BE HIGHER THAN 12' ABOVE FINISHED
FLOOR
*NOT TO BE LOCATED DIRECTLY ABOVE DESKS OR ANY OTHER FIXED EQUIPMENT*
CONDUIT NOTES
1. IF A DIRECT CONNECTION IS NEEDED BETWEEN TWO PIECES OF EQUIPMENT,
CONDUIT MUST BE RUN
2. EXACT CONDUIT PATH DETERMINED BY ELECTRICAL CONTRACTOR
3. CONDUIT SHOULD NOT EXCEED 50'
4. MUST BE A PULL BOX FOR EVERY 270° OF CONDUIT BEND
5. ALL CONDUITS SHOULD BE PROVIDED WITH PULL STRINGS AND MUST HAVE
INSULATED BUSHINGS ON ALL OPEN ENDS
6. USE OF FLEXIBLE CONDUIT IS NOT ACCEPTABLE
7. STERIS WILL NOT BE RESPONSIBLE FOR FIRE STOPPING IN ANY FORM
ABBREVIATIONS
A.F.F. = ABOVE FINISHED FLOOR
B.C.P. = BLANK COVER PLATE
C. = CONDUIT
C.P. = COVER PLATE
C.K.T. = CIRCUIT
E.C. = ELECTRICAL CONTRACTOR
I.G. = ISOLATED GROUND (TECHNICAL GROUND)
N.S.E. = NURSE STATION EQUIPMENT
RISER LEGEND
SYMBOL DESCRIPTION MOUNTING/COMMENT
AP CEILING ACCESS PANEL FOR REQUIRED ACCESS TO BOXES
B1 EQUIPMENT LABEL DENOTED CEILING MOUNT
N1 CONDUIT OR CABLE RUN JUNCTION BOX REFERENCE
CONDUIT STUB OUT CONDUIT STUB OUT WITH BUSHING
G1 CEILING SPEAKER CEILING, FLUSH, RECEPTACLE HEIGHT
DUPLEX I.G. POWER RECEPTACLE WALL, FLUSH, RECEPTACLE HEIGHT
QUAD I.G. POWER RECEPTACLE WALL, FLUSH, RECEPTACLE HEIGHT
DT DATA/TELEPHONE BOX WALL, FLUSH, RECEPTACLE HEIGHT
I.G.
I.G.
T2
CRQ1
MUSIC
AUDIO
INPUT
TOUCH PANEL
PACS PC
(2x)
CR3
1 1/4"C.
1 1/4"C.
CRTP1
P W R
C K T
P W R
C K T
I.G. I.G.
CONTROL
ROOM
DESK
CONTROL
ROOM
DESK
CONTROL
ROOM
DESK
1 1/4"C.
CR2
CONTROL
ROOM
DESK
TO T1
2"C.
MACLAB
1 1/4"C.
GUEST PORT
DVI (2x)
CR1
CONTROL
ROOM
DESK
1 1/4"C.
SHEET NOTES
1. INTEGRATED SYSTEM (A1) AND VIDEO CONFERENCING BOX (A2); REQUIRES DEDICATED
EMERGENCY BACK-UP 120V ON 20 AMP POWER FROM THE O.R. IT SERVES (NOT BY STERIS)
2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS
3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE RIGHT
HALF OF THE BOX
4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE
AROUND ALL SIDES OF THE BOX
5. AUDIO REQUIREMENT; HOSPITAL GRADE HUBBELL USB CHARGER RECEPTACLE DUPLEX
(ACCEPTABLE MODEL NUMBERS: #SB8300LA, #USB8300R, #USB8300W OR SIMILAR; NOT BY
STERIS)
6. TOUCH PANEL REQUIRES DUPLEX OUTLET 120 VAC ON 20 AMP SERVICE (STANDARD OR
EMERGENCY)
K2 SURGICAL REMOTE POWER MODULE STERIS
N1 WALL MOUNTED 12x12x4" JUNCTION BOX ELECTRICAL CONTRACTOR
T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE
CEILING ELECTRICAL CONTRACTOR
T2 12" H x 12" W X 6" D JUNCTION BOX ABOVE
CEILING ELECTRICAL CONTRACTOR
TP1
TOUCH PANEL STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR
U1 UNIVERSAL PORT - WALL MOUNTED 4 11/16"
JUNCTION BOX WITH SINGLE GANG MUD RING ELECTRICAL CONTRACTOR
U2 UNIVERSAL PORT - WALL MOUNTED 4 11/16"
JUNCTION BOX WITH SINGLE GANG MUD RING ELECTRICAL CONTRACTOR
V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER
DISTRIBUTION ELECTRICAL CONTRACTOR
INTEGRATION EQUIPMENT SCHEDULE
LEGEND DESCRIPTION RESPONSIBILITY
A1 STERIS iQ3600 INTEGRATION SYSTEM STERIS
A2 VIDEO CONFERENCE BOX STERIS
B1*** SURGICAL LIGHT MOUNT STERIS
B2*** SURGICAL LIGHT MOUNT STERIS
C1*** MONITOR CARRIER STERIS
C2*** EQUIPMENT BOOM STERIS
C3*** ANESTHESIA BOOM STERIS
CR1
WALL MOUNTED HUBBELL - RACO 698 4 GANG
JUNCTION BOX FOR GUEST PORTS (2x) ELECTRICAL CONTRACTOR
CR2 WALL MOUNTED HUBBELL - RACO 698 4 GANG
JUNCTION BOX FOR MACLAB ELECTRICAL CONTRACTOR
CR3
WALL MOUNTED HUBBELL - RACO 698 4 GANG
JUNCTION BOX FOR PACS PC (2x) ELECTRICAL CONTRACTOR
CRQ1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR
CRTP1
TOUCH PANEL STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR
E1
FUTURE WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR WALL
MOUNTED DISPLAY
ELECTRICAL CONTRACTOR
E2
FUTURE WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR WALL
MOUNTED DISPLAY
ELECTRICAL CONTRACTOR
E3
WALL MOUNTED DISPLAY STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR WALL
MOUNTED DISPLAY
ELECTRICAL CONTRACTOR
E4
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING FOR SIEMENS LIVE & REF
(DVI)
ELECTRICAL CONTRACTOR
E5
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING FOR 3RD PARTY DVI HD
DISPLAY OR DESTINATION
ELECTRICAL CONTRACTOR
F1
WALL MOUNTED HD PTZ CAMERA STERIS
WALL MOUNTED 4 11/16" JUNCTION BOX WITH
SINGLE GANG MUD RING REQUIRED FOR WALL
MOUNTED HD PTZ CAMERA
ELECTRICAL CONTRACTOR
G1 CEILING MOUNTED SPEAKER STERIS
G3 CEILING MOUNTED SPEAKER STERIS
G2 CEILING MOUNTED SPEAKER STERIS
G4 CEILING MOUNTED SPEAKER STERIS
K1 SURGICAL LIGHT CONTROLLER STERIS
36-05 iQ3600
CONDUIT RISER DIAGRAM
PROPOSED CONTROL ROOMPlanning & Design Group
STERIS
STERIS Corporation
5960 Heisley Road
Mentor, OH 44060 USA
STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD
Dr
Ck
Date
Scale
H- B-
800-548-4873 NA
DATE
LL
KS
C3R1-INT
11224
PAGE: OF 36-08
CUSTOMER
VERIFY
CUSTOMER
VERIFY
3"3"
A B O V E
F I N I S H E D
F L O O R
I F
U S E D
F O R
X R A Y
V I E W I N G
S T A T I O N
C U S T O M E R
V E R I F Y
A L L
A B O V E
F I N I S H E D
F L O O R
E1,E2,E3
F1
TYPICAL WALL DISPLAY INSTALLATION (OPTIONAL)
*REFERENCE SHEET NOTES #1,#2,#3
TYPICAL WALL MOUNTED HD PTZ CAMERA
INSTALLATION (OPTIONAL)
*REFERENCE SHEET NOTES #3,#4
1558"
33 8
POWER (SEE
SHEET NOTES
FOR DETAILS)
TYPICAL LEVITON BOX FOR STERIS
POWER DISTRIBUTION INSTALLATION
*REFERENCE SHEET NOTES #5,#6
V2
1034"13 1/16"
6"
1012"
12 1/16"
2"
60" ABOVE
FINISHED
FLOOR
2"17 1/2"2"
13 1/2"
6"
SURGICAL
REMOTE POWER
MODULE
ROUGH OPENING
SURGICAL LIGHT
TOUCH PANEL
CONTROL UNIT
ROUGH OPENING
6"
E SERIES SURGICAL LIGHT CONTROLLER
AND REMOTE POWER MODULE (OPTION)
K2
K1
EQUIPMENT SCHEDULE
LEGEND DESCRIPTION RESPONSIBILITY
E1,E2,E3 WALL MOUNTED DISPLAY STERIS
F1 HD PTZ WALL MOUNTED CAMERA STERIS
K1 SURGICAL LIGHT CONTROLLER STERIS
K2 SURGICAL REMOTE POWER MODULE STERIS
V2 LEVITON BOX FOR STERIS POWER DISTRIBUTION ELECTRICAL CONTRACTOR
SHEET NOTES
1. WALL DISPLAY BACKING TO BE PROVIDED AND INSTALLED BY CONTRACTOR; WALL DISPLAY BACKING
(NOT BY STERIS) MUST SUPPORT A 250LBS LOAD; WALL BRACKET NOT INSTALLED BY STERIS
2. WALL DISPLAY REQUIRES 4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING
3. WALL DISPLAY & PTZ CAMERA REQUIRE DUPLEX OUTLET 120 VAC ON 20 AMP SERVICE (STANDARD OR
EMERGENCY)
4. HD PTZ CAMERA MOUNTED DIRECTLY TO FRONT OF A/V PLATE; EXACT LOCATION OF H.D. CAMERA TO
BE DETERMINED BY CUSTOMER; CONTRACTOR TO PROVIDE PROPER BACKING FOR H.D. CAMERA; (1)
DUPLEX OUTLET REQUIRED AT EACH PTZ CAMERA LOCATION
5. V2 REQUIRES DEDICATED EMERGENCY BACK-UP 120V ON 20 AMP POWER FROM THE O.R. IT SERVES
(NOT BY STERIS)
6. V2 LEVITON BOX (NOT BY STERIS) WITH VENTED DOOR FOR STERIS POWER DISTRIBUTION; V2 BOX PRODUCT NUMBER: 47605-42N (REQUIRES MINIMUM 2" SERVICE CLEARANCE); V2 DOOR (NOT BY STERIS) PRODUCT NUMBER: 47605-42S; LEVITON AC DUAL POWER MODULE (NOT BY STERIS - BOLTED
TO BOTTOM OF LEVITON BOX - LOCATED INSIDE OF WALL) PRODUCT NUMBER: 47605-0DP
36-06 iQ3600
TYPICAL OPERATING ROOM
INSTALLATION DETAILS
Planning & Design Group
STERIS
STERIS Corporation
5960 Heisley Road
Mentor, OH 44060 USA
STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD
Dr
Ck
Date
Scale
H- B-
800-548-4873 NA
DATE
LL
KS
C3R1-INT
11224
PAGE: OF 36-08
EQUIPMENT SCHEDULE
LEGEND DESCRIPTION RESPONSIBILITY
G1,G2,G3,G4 BOSE CEILING SPEAKER (OPTIONAL) STERIS
SHEET NOTES
1. CONTRACTOR TO INSTALL SEISMIC SAFETY SUPPORT WIRE FOR SPEAKER ANCHORING TO
THE BUILDING STRUCTURE
2. CONTRACTOR TO CUT ROUND HOLE IN CEILING AT THE SPEAKER LOCATION ACCORDING
TO THE SPEAKER INSTALLATION TEMPLATE
3. TOP CLEARANCE FOR SPEAKERS IS 9"; ROUGH-IN PAN TO BE INSTALLED WHERE
OPENING IS NOT OBSTRUCTED ABOVE CEILING
T1 24"x24"x8"
G1 G2
G4 G3
1/2" FLEXIBLE CONDUIT WITH
CONNECTOR ON SPEAKER END
(NOT BY STERIS)(TYPICAL AT
EACH SPEAKER LOCATION)
SPEAKER (INSTALLED BY STERIS)
BOSE ROUGH-IN PAN PRE-SHIPPED
BY STERIS (NOT BY STERIS)
3/4" CONDUIT (NON-FLEX)
(NOT BY STERIS)
3/4" CONDUIT (NON-FLEX)
(NOT BY STERIS)
1/2" FLEXIBLE CONDUIT WITH
CONNECTOR ON SPEAKER END
(NOT BY STERIS)(TYPICAL AT
EACH SPEAKER LOCATION)
SPEAKER (INSTALLED BY STERIS)
BOSE ROUGH-IN PAN PRE-SHIPPED
BY STERIS (NOT BY STERIS)
3/4" CONDUIT (NON-FLEX)
(NOT BY STERIS)
R O U G H I N
H O L E
R E Q U I R E D
6.94"
MOUNTING EARS
FINISHED CEILING
SIDE
VIEW
FRONT VIEW
CEILING SECTION
VIEW
FREESPACE DS 16F ROUGH-IN PAN
26.00 [660.mm] m m
Ø 8.40 [213.4mm] BOSE ROUGH-IN PAN
PRE-SHIPPED BY STERIS
(NOT INSTALLED BY
STERIS)
36-07 iQ3600 Planning & Design Group
STERIS
STERIS Corporation
5960 Heisley Road
Mentor, OH 44060 USA
STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD
Dr
Ck
Date
Scale
H- B-
800-548-4873 NA
DATE
LL
KS
C3R1-INT
11224
PAGE: OF 36-08
VA MEDICAL CENTER OF RICHMOND
RICHMOND, VIRGINA
TYPICAL BOSE CEILING SPEAKER
(G1-G4) INSTALLATION DETAILS
36-08 iQ3600
CALL AND CONNECT HUB DETAIL
WITH ELECTRICAL REQUIREMENTS
Planning & Design Group
STERIS
STERIS Corporation
5960 Heisley Road
Mentor, OH 44060 USA
STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD
Dr
Ck
Date
Scale
H- B-
800-548-4873 NA
DATE
LL
KS
C3R1-INT
11224
PAGE: OF 36-08
VA MEDICAL CENTER OF RICHMOND
RICHMOND, VIRGINA
CALL AND CONNECT HUB - ELEVATION DETAIL PAGE
49.2"
23.4"
FINISHED CEILING
3" MINIMUM
REQUIRED
26.0"
FINISHED CEILING
DESCRIPTION RESPONSIBILITY
CALL AND CONNECT HUB STERIS
SHEET NOTES
1. CALL AND CONNECT HUB REQUIRES (1) DEDICATED 120 VAC ON 20 AMP
CIRCUIT
2. CALL AND CONNECT HUB REQUIRES (2) DATA DROPS - (1) FOR
CONTROL FOR ROOM-TO-ROOM AND (1) FOR CODEC
3. CALL AND CONNECT HUB MIDDLE ATLANTIC MODEL# DWR2426 24U WALL
MOUNTED AV RACK
4. CALL AND CONNECT HUB REQUIRES REINFORCED WALL BEHIND TO
SUPPORT 150LBS (NOT BY STERIS)
5. CALL AND CONNECT REQUIRES A MINIMUM OF 2" CLEARANCE REQUIRED
AROUND FRONT AND SIDES FOR PROPER VENTILATION
6. 8x8x4" JUNCTION BOX LOCATED IN WALL BEHIND THE CALL & CONNECT
HUB RACK WITH DUPLEX POWER
2" C.
TO CORRIDOR
CALL & CONNECT HUB FRONT ELEVATION CALL & CONNECT HUB SIDE ELEVATION
CORRIDOR
CALL AND
CONNECT HUB
1 1/2" CONDUITT1 AT
HYBRID
OR
*CONDUIT SIZE
DEPENDANT ON SIZE OF
SYSTEM
| Sheets and Views |
| 11224C3R1-INT-PUBLISHED-36-01 TITLE |
| 11224C3R1-INT-PUBLISHED-36-02 LAYOUT |
| 11224C3R1-INT-PUBLISHED-36-03 3600 DETAIL |
| 11224C3R1-INT-PUBLISHED-36-04 RISER DETAIL |
| 11224C3R1-INT-PUBLISHED-36-05 RISER DETAIL |
| 11224C3R1-INT-PUBLISHED-36-06 INSTALLTION DETAIL |
| 11224C3R1-INT-PUBLISHED-36-07 SPEAKER DETAIL |
| 11224C3R1-INT-PUBLISHED-36-08 CALLCONNECT |