This presolicitation notice is for a project to construct a Hybrid Operating Room at the Hunter Holmes McGuire VA Medical Center in Richmond, Virginia. The project has an estimated value between $1,000,000 and $2,000,000, and will be solicited as a 100% Service Disabled Veteran Owned Small Business set-aside under NAICS code 236220. Interested vendors must be registered in both the System for Award Management and VetBiz databases. The solicitation is expected to be released on or around March 2, 2020, with proposals due at least 30 days later. The site visit location, date, and points of contact will be provided in the solicitation. Prospective contractors should monitor the website for any amendments or notices.

36C24620R0046 0007 RFI Response Attachment 1 - Steris Drawings.pdf

View the file

Other files

File Type Posted
36C24620R0046-0012001.pdf PDF
36C24620R0046-0012000.docx DOCX document
36C24620R0046-0011003.pdf PDF
36C24620R0046-0011001.pdf PDF
36C24620R0046-0011000.docx DOCX document
36C24620R0046-0011002.pdf PDF
36C24620R0046-0010001.pdf PDF
36C24620R0046-0010003.pdf PDF
36C24620R0046-0010000.docx DOCX document
36C24620R0046-0010002.pdf PDF
36C24620R0046-0009000.docx DOCX document
36C24620R0046-0009008.pdf PDF
36C24620R0046-0009005.pdf PDF
36C24620R0046-0009016.pdf PDF
36C24620R0046-0009015.pdf PDF
36C24620R0046-0009014.pdf PDF
36C24620R0046-0009019.pdf PDF
36C24620R0046-0009009.pdf PDF
36C24620R0046-0009002.pdf PDF
36C24620R0046-0009012.pdf PDF
36C24620R0046-0009006.pdf PDF
36C24620R0046-0009021.pdf PDF
36C24620R0046-0009022.pdf PDF
36C24620R0046-0009007.pdf PDF
36C24620R0046-0009020.pdf PDF
36C24620R0046-0009004.pdf PDF
36C24620R0046-0009018.pdf PDF
36C24620R0046-0009011.pdf PDF
36C24620R0046-0009003.pdf PDF
36C24620R0046-0009017.pdf PDF
36C24620R0046-0009001.pdf PDF
36C24620R0046-0009013.pdf PDF
36C24620R0046-0009010.pdf PDF
36C24620R0046-0008006.pdf PDF
36C24620R0046-0008007.pdf PDF
36C24620R0046-0008000.docx DOCX document
36C24620R0046-0008002.pdf PDF
36C24620R0046-0008004.pdf PDF
36C24620R0046-0008003.pdf PDF
36C24620R0046-0008005.pdf PDF
36C24620R0046-0008001.pdf PDF
36C24620R0046-0007004.pdf PDF
36C24620R0046-0007002.pdf PDF
36C24620R0046-0007005.pdf PDF
36C24620R0046-0007001.pdf PDF
36C24620R0046-0007000.docx DOCX document
36C24620R0046-0007006.pdf PDF
36C24620R0046-0007007.pdf PDF
36C24620R0046-0006000.docx DOCX document
36C24620R0046-0006001.pdf PDF
Show 40 more

Latest 50 of 79 listed here. See the rest on GovTribe.

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

HARMONY iQ3600 INSTALLATION GUIDELINES

VA MEDICAL CENTER OF RICHMOND

RICHMOND, VIRGINIA

INDEX

PAGE # SHEET TITLE ROOM #

1 TITLE PAGE, INDEX -

2 PROPOSED OPERATING ROOM LAYOUT -

3 MOUNTING ASSEMBLY FOR A1&A2 (iQ3600) ATOM DETAILS -

4 PROPOSED OPERATING ROOM CONDUIT RISER DIAGRAM -

5 PROPOSED CONTROL ROOM CONDUIT RISER DIAGRAM -

6 TYPICAL OPERATING ROOM INSTALLATION DETAIL -

7 TYPICAL BOSE CEILING SPEAKER (G1-G4) DETAILS -

8 CALL & CONNECT HUB DETAIL -

ABBREVIATIONS

A.F.F. = ABOVE FINISHED FLOOR

B.C.P. = BLANK COVER PLATE

C. = CONDUIT

C.P. = COVER PLATE

C.K.T. = CIRCUIT

E.C. = ELECTRICAL CONTRACTOR

I.G. = ISOLATED GROUND (TECHNICAL GROUND)

N.S.E. = NURSE STATION EQUIPMENT

GENERAL NOTES

1. SUPPLY AND INSTALLATION OF ALL CONDUIT, POWER CIRCUITS, BACK BOXES, ELECTRICAL DEVICES, ELECTRICAL WIRING IS TO BE

PROVIDED BY THE CLIENTS ELECTRICAL CONTRACTOR UNLESS NOTED OTHERWISE

2. ONE WIRE HANGER FOR CABLE COILS REQUIRED FOR EACH MOUNT

3. CLIENT CONTRACTOR TO PROVIDE BLANK COVER PLATES FOR ALL JUNCTIONS BOXES WHERE NECESSARY

4. ALL CABLE RUNS OUTSIDE INTEGRATED ROOMS ARE THE RESPONSIBILITY OF HOSPITAL/CONTRACTOR; ALL CABLE RUNS SHOULD

BE CONTINUOUS EXCEPT WHERE NOTED

5. 24x24" ACCESS MUST BE PROVIDED WITHIN 36" OF ALL JUNCTION BOXES

6. CABLE RUNS THRU PLENUM SPACE MUST BE PLENUM RATED OR SHALL BE INSTALLED WITH CONDUIT

7. CABLES NOT RUN IN CONDUIT MUST HAVE J-HOOKS EVERY 4' FOR SUPPORT (NOT BY STERIS)

8. WHERE J-HOOKS ARE USED IN CEILING, CONDUIT IS REQUIRED IN ALL CONCEALED LOCATIONS (SIZED PER SCHEDULES)

9. EVERY RACK REQUIRES 4 LAN CONNECTIONS (DATA PORTS) ADJACENT TO T1 BOX; EACH OR REQUIRES 2 NETWORK PORTS AT RACK

10. ALL POWER FOR INTEGRATION EQUIPMENT SHOULD COME FROM A SINGLE ELECTRICAL PANEL IN THE OR IT SERVES

11. CONFORMANCE TO NATIONAL, STATE AND LOCAL CODES FOR ELECTRICAL CIRCUITS AND CONDUIT IS THE RESPONSIBILITY OF THE

HOSPITAL OR THEIR CONTRACTED ELECTRICIAN

12. THE EQUIPMENT AND ROUGH-IN REQUIREMENTS SHOWN ARE FOR BUDGETARY PLANNING PURPOSES ONLY; EQUIPMENT IS SUBJECT TO CHANGE AND DIFFERENT OPTIONS ARE AVAILABLE TO MEET SPECIFIC CUSTOMER DESIGN CRITERIA; POST-ORDER DRAWINGS WILL REFLECT

EXACT EQUIPMENT SPECIFICATIONS, AND REQUIRES CUSTOMER SIGNED APPROVAL

13. USE OF ISOLATED GROUND POWER TO BE DETERMINED BY ELECTRICAL CONTRACTOR

REVISION HISTORY

REVISION # DESCRIPTION PLANNER/DATE

1 UPDATED C1 MOUNT; E1 & E2 MADE FUTURE; MOVED NURSE

DESK

JF/30SEP2019

36-01 iQ3600 TITLE PAGE, INDEXPlanning & Design Group

STERIS

STERIS Corporation

5960 Heisley Road

Mentor, OH 44060 USA

STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD

Dr

Ck

Date

Scale

H- B-

800-548-4873 NA

DATE

LL

KS

C3R1-INT

11224

PAGE: OF 36-08

VA MEDICAL CENTER OF RICHMOND

RICHMOND, VIRGINA

Ken Jones Text Box

HYBRID OR

BID QUESTION RFI'S

ATTACHMENT 1

A1 A2

E4 E5

SHEET NOTES

1. INTEGRATED SYSTEM (A1) AND VIDEO CONFERENCING BOX (A2); REQUIRES DEDICATED

EMERGENCY BACK-UP 120V ON 20 AMP POWER FROM THE O.R. IT SERVES (NOT BY STERIS)

2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS

3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE RIGHT

HALF OF THE BOX

4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE

AROUND ALL SIDES OF THE BOX

5. AUDIO REQUIREMENT; HOSPITAL GRADE HUBBELL USB CHARGER RECEPTACLE DUPLEX

(ACCEPTABLE MODEL NUMBERS: #SB8300LA, #USB8300R, #USB8300W OR SIMILAR; NOT BY

STERIS)

6. TOUCH PANEL REQUIRES DUPLEX OUTLET 120 VAC ON 20 AMP SERVICE (STANDARD OR

EMERGENCY)

LIGHT CONDUIT SCHEDULE NOTES

1. 1/2" CONDUIT FOR AC POWER TO K1 & K2 BY E.C.

2.LIGHTING CONDUIT RUNS NOT TO EXCEED 25 FEET

3.CONDUIT REQUIREMENTS FOR LIGHTS AND CONNECT POINT ALSO

REFERENCED WITHIN THE EQUIPMENT DRAWINGS FOR THE SPECIFIED

SURGICAL LIGHT WALL CONTROL SYSTEM

4.DC SUPPLY LIGHTHEAD CABLES REQUIRE 1" CONDUIT WHEN DOUBLE

RUNS ARE REQUIRED (LONG RUNS)

***ALL LIGHT MOUNTS (B) REQUIRE 2 J-HOOKS EACH, 8"

APART -OR- EQUIVALENT SUSPENSION POINTS ABOVE THE

STRUCTURAL MOUNTING PLATE

E SERIES LIGHT CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

K1 - K2 1 3/4" ELECTRICAL

CONTRACTOR

K1 - B1* 1 3/4" ELECTRICAL

CONTRACTOR

K1 - B1 1 3/4" ELECTRICAL

CONTRACTOR

K2 - B1* 1 3/4" ELECTRICAL

CONTRACTOR

CONDUIT SCHEDULE

FROM - TO QUANTITY SIZE RESPONSIBILITY

A1 - A2 1 1" ELECTRICAL CONTRACTOR

T1 - A1 2 2" ELECTRICAL CONTRACTOR

T1 - A2 1 2" ELECTRICAL CONTRACTOR

T1 - B1 1 1 1/2" ELECTRICAL CONTRACTOR

T1 - B2 1 1 1/2" ELECTRICAL CONTRACTOR

T1 - C1 1 1 1/2" ELECTRICAL CONTRACTOR

T1 - C2 1 1 1/2" ELECTRICAL CONTRACTOR

T1 - C3 1 1 1/2" ELECTRICAL CONTRACTOR

T1 - CORRIDOR 1 1 1/2" ELECTRICAL CONTRACTOR

T1 - E1 1 1" ELECTRICAL CONTRACTOR

T1 - E2 1 1" ELECTRICAL CONTRACTOR

T1 - E3 1 1" ELECTRICAL CONTRACTOR

T1 - E4 1 1" ELECTRICAL CONTRACTOR

T1 - E5 1 1" ELECTRICAL CONTRACTOR

T1 - F1 1 1 1/2" ELECTRICAL CONTRACTOR

T1 - G1 1 3/4" ELECTRICAL CONTRACTOR

T1 - G3 1 3/4" ELECTRICAL CONTRACTOR

T1 - G2 1 3/4" ELECTRICAL CONTRACTOR

T1 - G4 1 3/4" ELECTRICAL CONTRACTOR

T1 - K1 1 1" ELECTRICAL CONTRACTOR

T1 - N1 2 2" ELECTRICAL CONTRACTOR

T1 - U2 1 1 1/4" ELECTRICAL CONTRACTOR

T1 - V2 2 1 1/2" ELECTRICAL CONTRACTOR

T1 - T2 1 2" ELECTRICAL CONTRACTOR

T2 - CR1 1 1 1/4" ELECTRICAL CONTRACTOR

T2 - CR2 1 1 1/4" ELECTRICAL CONTRACTOR

T2 - CR3 1 1 1/4" ELECTRICAL CONTRACTOR

T2 - CRQ1 1 1" ELECTRICAL CONTRACTOR

T2 - CRTP1 1 1 1/4" ELECTRICAL CONTRACTOR

HOSPITAL

NETWORK 1 1" ELECTRICAL CONTRACTOR

N1 - U1 1 1 1/4" ELECTRICAL CONTRACTOR

N1 - TP1 1 1 1/4" ELECTRICAL CONTRACTOR

C1*** MONITOR CARRIER STERIS

C2*** EQUIPMENT BOOM STERIS

C3*** ANESTHESIA BOOM STERIS

CR1

WALL MOUNTED HUBBELL - RACO 698 4 GANG

JUNCTION BOX FOR GUEST PORTS (2x) ELECTRICAL CONTRACTOR

CR2 WALL MOUNTED HUBBELL - RACO 698 4 GANG

JUNCTION BOX FOR MACLAB ELECTRICAL CONTRACTOR

CR3

WALL MOUNTED HUBBELL - RACO 698 4 GANG

JUNCTION BOX FOR PACS PC (2x) ELECTRICAL CONTRACTOR

CRQ1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR

CRTP1

TOUCH PANEL STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR

E1

FUTURE WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR WALL

MOUNTED DISPLAY

ELECTRICAL CONTRACTOR

E2

FUTURE WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR WALL

MOUNTED DISPLAY

ELECTRICAL CONTRACTOR

E3

WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR WALL

MOUNTED DISPLAY

ELECTRICAL CONTRACTOR

E4

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING FOR SIEMENS LIVE & REF

(DVI)

ELECTRICAL CONTRACTOR

E5

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING FOR 3RD PARTY DVI HD

DISPLAY OR DESTINATION

ELECTRICAL CONTRACTOR

F1

WALL MOUNTED HD PTZ CAMERA STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR WALL

MOUNTED HD PTZ CAMERA

ELECTRICAL CONTRACTOR

G1 CEILING MOUNTED SPEAKER STERIS

G3 CEILING MOUNTED SPEAKER STERIS

G2 CEILING MOUNTED SPEAKER STERIS

G4 CEILING MOUNTED SPEAKER STERIS

K1 SURGICAL LIGHT CONTROLLER STERIS

K2 SURGICAL REMOTE POWER MODULE STERIS

N1 WALL MOUNTED 12x12x4" JUNCTION BOX ELECTRICAL CONTRACTOR

T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE

CEILING ELECTRICAL CONTRACTOR

T2 12" H x 12" W X 6" D JUNCTION BOX ABOVE

CEILING ELECTRICAL CONTRACTOR

TP1

TOUCH PANEL STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR

U1 UNIVERSAL PORT - WALL MOUNTED 4 11/16"

JUNCTION BOX WITH SINGLE GANG MUD RING ELECTRICAL CONTRACTOR

U2 UNIVERSAL PORT - WALL MOUNTED 4 11/16"

JUNCTION BOX WITH SINGLE GANG MUD RING ELECTRICAL CONTRACTOR

V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER

DISTRIBUTION ELECTRICAL CONTRACTOR

E3

CRTP1 CR3

CRTP1

T1 V2

G4 G3

SPEAKER PLACEMENT

(G1-G4) FOR REFERENCE

ONLY; SITE VERIFY

G1 G2

F1

K1 K2

NURSE

DESK

T2

TP1

N1 TP1

CR1

U1

U2

CR2

CRQ1

B1

B2

C1

C2

C3

E2 E1

INTEGRATION EQUIPMENT SCHEDULE

LEGEND DESCRIPTION RESPONSIBILITY

A1 STERIS iQ3600 INTEGRATION SYSTEM STERIS

A2 VIDEO CONFERENCE BOX STERIS

B1*** SURGICAL LIGHT MOUNT STERIS

B2*** SURGICAL LIGHT MOUNT STERIS

36-02 iQ3600

LAYOUT

PROPOSED OPERATING ROOMPlanning & Design Group

STERIS

STERIS Corporation

5960 Heisley Road

Mentor, OH 44060 USA

STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD

Dr

Ck

Date

Scale

H- B-

800-548-4873 NA

DATE

LL

KS

C3R1-INT

11224

PAGE: OF 36-08

VA MEDICAL CENTER OF RICHMOND

RICHMOND, VIRGINA

AutoCAD SHX Text

EQUIP. ROOM

AutoCAD SHX Text

CONTROL ROOM

AutoCAD SHX Text

HYBRID OR

REQUIRES DEDICATED 120V ON

20 AMP

CIRCUIT FROM THE O.R. IT

SERVES

(ON EMERGENCY BACK-UP

POWER;NOT BY STERIS)

HEAT LOADS:

iQ3600:

1200 BTU/Hr TO BE MAINTAINED AT 78 DEGREES F. OR LESS

POWER MODULE: (INCLUDING POWERING MONITORS):

450 BTU/Hr TO BE MAINTAINED AT 78 DEGREES F. OR LESS

4 LAN CONNECTIONS REQUIRED ADJACENT TO T1

.65

(ABOVE WALL)

42.00

3.00

29.50

3.91

(INSIDE WALL)

1.95

2.81

2X Ø 1.50

2X 1.71

2X 1.96

2X 2.91

14.18

(14.33 WITH ADJUSTABLE L-BRACKET)

0.5" TO 1.50" - ADJUSTABLE

(DRYWALL THICKNESS)

4X Ø1.002

2X Ø 2.00

ELEVATION A

iQ3600 ELECTRICAL BOX

FINISHED CEILING

iQ3600

STERIS

INTEGRATION/VIDEO

CONFERENCING BOX

17.55

FINISHED FLOOR

36-03 iQ3600

A1,A2 (iQ3600) ATOM DETAILS MOUNTING ASSEMBLY FORPlanning & Design Group

STERIS

STERIS Corporation

5960 Heisley Road

Mentor, OH 44060 USA

STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD

Dr

Ck

Date

Scale

H- B-

800-548-4873 NA

DATE

LL

KS

C3R1-INT

11224

PAGE: OF 36-08

RICHMOND, VIRGINA

B2B1 C2C1

SURGICAL

LIGHT MOUNT

SURGICAL

LIGHT MOUNT

SURGICAL

LIGHT HEAD

(3x)

RAD SHIELD

ANESTHESIA

BOOM

LARGE FORMAT

MONITOR

/EQUIPMENT

BOOM

UNIVERSAL

PORT

VITALS

JUNCTION BOX SCHEDULE

J-BOX # SIZE TYPE MOUNTING

N1 12"H X 12"W X 4"D PULL BOX IN WALL

T1 24"H X 24"W X 8"D NEMA 1, SCREW

COVER ABOVE CEILING

JUNCTION BOX SCHEDULE NOTES

1. ALL BOXES ARE PROVIDED BY THE CLIENTS CONTRACTOR UNLESS NOTED OTHERWISE

2. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS

OTHERWISE NOTED/DETAILED

3. ALL BOXES AND COVER PLATES ARE PROVIDED BY THE CLIENTS CONTRACTOR UNLESS

NOTED OTHERWISE

4. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS

OTHERWISE NOTED/DETAILED

5. BOXES/OUTLETS/RECEPTACLES LOCATED ABOVE THE CEILING MUST BE WITHIN 18

INCHES OF THE CEILING AND BE ACCESSIBLE FROM THE SPACE BELOW; CLIENT/G.C.

TO PROVIDE SUITABLE CEILING ACCESS PANELS AT DEVICE LOCATIONS

6. T1 JUNCTION BOX - 24" x 24" x 8" (PROVIDED AND MOUNTED VERTICALLY NOT BY

STERIS); ALL FIELD CONDUITS AND 2" & 1" CONDUIT TO (A1) STERIS INTEGRATION

RACK ATTACH TO JUNCTION BOX (T1); CEILING ACCESS REQUIRED FOR SERVICE

ACCESS; ALL CONDUIT TO ENTER T1 BOX ANYWHERE EXCEPT FOR THE BACK SIDE; T1

BOX SHOULD NOT BE HIGHER THAN 12' ABOVE FINISHED FLOOR;

NOT TO BE LOCATED DIRECTLY ABOVE NURSE DESKS OR ANY OTHER FIXED EQUIPMENT

CONDUIT NOTES

1. IF A DIRECT CONNECTION IS NEEDED BETWEEN TWO PIECES OF

EQUIPMENT, CONDUIT MUST BE RUN

2. EXACT CONDUIT PATH DETERMINED BY ELECTRICAL CONTRACTOR

3. CONDUIT SHOULD NOT EXCEED 50'

4. MUST BE A PULL BOX FOR EVERY 270° OF CONDUIT BEND

5. ALL CONDUITS SHOULD BE PROVIDED WITH PULL STRINGS AND MUST

HAVE INSULATED BUSHINGS ON ALL OPEN ENDS

6. USE OF FLEXIBLE CONDUIT IS NOT ACCEPTABLE

7. STERIS WILL NOT BE RESPONSIBLE FOR FIRE STOPPING IN ANY FORM

ABBREVIATIONS

A.F.F. = ABOVE FINISHED FLOOR

B.C.P. = BLANK COVER PLATE

C. = CONDUIT

C.P. = COVER PLATE

C.K.T. = CIRCUIT

E.C. = ELECTRICAL CONTRACTOR

I.G. = ISOLATED GROUND (TECHNICAL GROUND)

N.S.E. = NURSE STATION EQUIPMENT

RISER LEGEND

SYMBOL DESCRIPTION MOUNTING/COMMENT

AP CEILING ACCESS PANEL FOR REQUIRED ACCESS TO BOXES

B1 EQUIPMENT LABEL DENOTED CEILING MOUNT

N1 CONDUIT OR CABLE RUN JUNCTION BOX REFERENCE

CONDUIT STUB OUT CONDUIT STUB OUT WITH BUSHING

G1 CEILING SPEAKER CEILING, FLUSH, RECEPTACLE HEIGHT

DUPLEX I.G. POWER RECEPTACLE WALL, FLUSH, RECEPTACLE HEIGHT

QUAD I.G. POWER RECEPTACLE WALL, FLUSH, RECEPTACLE HEIGHT

DT DATA/TELEPHONE BOX WALL, FLUSH, RECEPTACLE HEIGHT

T1

NURSE

DESK

NURSE

DESK

U1

UNIVERSAL

PORT

TOUCH

PANEL

1 1/4"C.

TP1

P W R

C K T

I.G.

P W R

C K T

I.G.

NURSE

DESK

(2)2"C.

(N1)12x12

JUNCTION

BOX

1 1/2"C.

1 1/2"C.

1 1/2"C. 1 1/2"C.1"C.

K1

3/4"C.

1/2"C.

I N C O M I N G

H O S P

P W R

REMOTE

POWER

MODULE

SURGICAL

LIGHT

CONTROLLER

K2 (2)2"C.

A1 iQ3600

STERIS

INTEGRATION

BOX

A2 iQ3600

STERIS

VIDEO

CONFERENCING

BOX

SEE MOUNTING

ASSEMBLY PAGE

FOR POWER

REQUIREMENTS

1"C.

2"C.

(F1)HD PTZ

CAMERA

I.G.

P W R

C K T

3/4"C.

G1

3/4"C. STUB

WITHIN 18"

OF SPEAKER

G3

3/4"C.

(E2)FUTURE

WALL

DISPLAY

1"C.

I.G.

P W R

C K T

(E3)WALL

DISPLAY

1"C.

1 1/2"C.

3/4"C.

G2

3/4"C. STUB

WITHIN 18"

OF SPEAKER

G4

3/4"C.

E4

I.G.

P W R

C K T

I.G.

P W R

C K T

SIEMENS

LIVE &

REF

1"C.

(V2)

LEVITON

BOX

P W R

C K T

(2)1 1/2"C.

V2 REQUIRES DEDICATED

EMERGENCY BACK-UP 120V

ON 20 AMP POWER (NOT

BY STERIS)

UNIVERSAL

PORT

1 1/4"C.

U2

SURGICAL

DISPLAY

ENDO

CAMERA

HEADLIGHT

CAM

UNIVERSAL

PORTS (2x)

IN-LIGHT

CAMERA

C3

1 1/2"C.

(E1)FUTURE

WALL

DISPLAY

1"C.

E5

3RD PARTY

DVI

1"C.

1 1/4"C.

LARGE

FORMAT

MONITOR

SHEET NOTES

1. INTEGRATED SYSTEM (A1) AND VIDEO CONFERENCING BOX (A2); REQUIRES DEDICATED

EMERGENCY BACK-UP 120V ON 20 AMP POWER FROM THE O.R. IT SERVES (NOT BY STERIS)

2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS

3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE RIGHT

HALF OF THE BOX

4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE

AROUND ALL SIDES OF THE BOX

5. AUDIO REQUIREMENT; HOSPITAL GRADE HUBBELL USB CHARGER RECEPTACLE DUPLEX

(ACCEPTABLE MODEL NUMBERS: #SB8300LA, #USB8300R, #USB8300W OR SIMILAR; NOT BY

STERIS)

6. TOUCH PANEL REQUIRES DUPLEX OUTLET 120 VAC ON 20 AMP SERVICE (STANDARD OR

EMERGENCY)

K2 SURGICAL REMOTE POWER MODULE STERIS

N1 WALL MOUNTED 12x12x4" JUNCTION BOX ELECTRICAL CONTRACTOR

T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE

CEILING ELECTRICAL CONTRACTOR

T2 12" H x 12" W X 6" D JUNCTION BOX ABOVE

CEILING ELECTRICAL CONTRACTOR

TP1

TOUCH PANEL STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR

U1 UNIVERSAL PORT - WALL MOUNTED 4 11/16"

JUNCTION BOX WITH SINGLE GANG MUD RING ELECTRICAL CONTRACTOR

U2 UNIVERSAL PORT - WALL MOUNTED 4 11/16"

JUNCTION BOX WITH SINGLE GANG MUD RING ELECTRICAL CONTRACTOR

V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER

DISTRIBUTION ELECTRICAL CONTRACTOR

INTEGRATION EQUIPMENT SCHEDULE

LEGEND DESCRIPTION RESPONSIBILITY

A1 STERIS iQ3600 INTEGRATION SYSTEM STERIS

A2 VIDEO CONFERENCE BOX STERIS

B1*** SURGICAL LIGHT MOUNT STERIS

B2*** SURGICAL LIGHT MOUNT STERIS

C1*** MONITOR CARRIER STERIS

C2*** EQUIPMENT BOOM STERIS

C3*** ANESTHESIA BOOM STERIS

CR1

WALL MOUNTED HUBBELL - RACO 698 4 GANG

JUNCTION BOX FOR GUEST PORTS (2x) ELECTRICAL CONTRACTOR

CR2 WALL MOUNTED HUBBELL - RACO 698 4 GANG

JUNCTION BOX FOR MACLAB ELECTRICAL CONTRACTOR

CR3

WALL MOUNTED HUBBELL - RACO 698 4 GANG

JUNCTION BOX FOR PACS PC (2x) ELECTRICAL CONTRACTOR

CRQ1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR

CRTP1

TOUCH PANEL STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR

E1

FUTURE WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR WALL

MOUNTED DISPLAY

ELECTRICAL CONTRACTOR

E2

FUTURE WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR WALL

MOUNTED DISPLAY

ELECTRICAL CONTRACTOR

E3

WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR WALL

MOUNTED DISPLAY

ELECTRICAL CONTRACTOR

E4

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING FOR SIEMENS LIVE & REF

(DVI)

ELECTRICAL CONTRACTOR

E5

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING FOR 3RD PARTY DVI HD

DISPLAY OR DESTINATION

ELECTRICAL CONTRACTOR

F1

WALL MOUNTED HD PTZ CAMERA STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR WALL

MOUNTED HD PTZ CAMERA

ELECTRICAL CONTRACTOR

G1 CEILING MOUNTED SPEAKER STERIS

G3 CEILING MOUNTED SPEAKER STERIS

G2 CEILING MOUNTED SPEAKER STERIS

G4 CEILING MOUNTED SPEAKER STERIS

K1 SURGICAL LIGHT CONTROLLER STERIS

36-04 iQ3600

CONDUIT RISER DIAGRAM

PROPOSED OPERATING ROOMPlanning & Design Group

STERIS

STERIS Corporation

5960 Heisley Road

Mentor, OH 44060 USA

STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD

Dr

Ck

Date

Scale

H- B-

800-548-4873 NA

DATE

LL

KS

C3R1-INT

11224

PAGE: OF 36-08

JUNCTION BOX SCHEDULE

J-BOX # SIZE TYPE MOUNTING

T2 12"H X 12"W X 6"D PULL BOX ABOVE CEILING

JUNCTION BOX SCHEDULE NOTES

1. ALL BOXES ARE PROVIDED BY THE CLIENTS CONTRACTOR UNLESS NOTED OTHERWISE

2. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS

OTHERWISE NOTED/DETAILED

3. ALL BOXES AND COVER PLATES ARE PROVIDED BY THE CLIENTS CONTRACTOR

UNLESS NOTED OTHERWISE

4. MOUNTING REQUIREMENTS SHOWN ARE THE DEFAULT REQUIREMENTS UNLESS

OTHERWISE NOTED/DETAILED

5. BOXES/OUTLETS/RECEPTACLES LOCATED ABOVE THE CEILING MUST BE WITHIN 18"

OF THE CEILING AND BE ACCESSIBLE FROM THE SPACE BELOW; CLIENT/G.C. TO

PROVIDE SUITABLE CEILING ACCESS PANELS AT DEVICE LOCATIONS

6. T2 JUNCTION BOX - 12" x 12" x 6" (MOUNTED VERTICALLY; NOT BY STERIS);

ALL FIELD CONDUITS AND 2" CONDUIT TO 24x24x8" (T1); CEILING ACCESS

REQUIRED FOR SERVICE ACCESS; ALL CONDUIT TO ENTER T2 BOX ANYWHERE EXCEPT

FOR THE BACK SIDE; T2 BOX SHOULD NOT BE HIGHER THAN 12' ABOVE FINISHED

FLOOR

*NOT TO BE LOCATED DIRECTLY ABOVE DESKS OR ANY OTHER FIXED EQUIPMENT*

CONDUIT NOTES

1. IF A DIRECT CONNECTION IS NEEDED BETWEEN TWO PIECES OF EQUIPMENT,

CONDUIT MUST BE RUN

2. EXACT CONDUIT PATH DETERMINED BY ELECTRICAL CONTRACTOR

3. CONDUIT SHOULD NOT EXCEED 50'

4. MUST BE A PULL BOX FOR EVERY 270° OF CONDUIT BEND

5. ALL CONDUITS SHOULD BE PROVIDED WITH PULL STRINGS AND MUST HAVE

INSULATED BUSHINGS ON ALL OPEN ENDS

6. USE OF FLEXIBLE CONDUIT IS NOT ACCEPTABLE

7. STERIS WILL NOT BE RESPONSIBLE FOR FIRE STOPPING IN ANY FORM

ABBREVIATIONS

A.F.F. = ABOVE FINISHED FLOOR

B.C.P. = BLANK COVER PLATE

C. = CONDUIT

C.P. = COVER PLATE

C.K.T. = CIRCUIT

E.C. = ELECTRICAL CONTRACTOR

I.G. = ISOLATED GROUND (TECHNICAL GROUND)

N.S.E. = NURSE STATION EQUIPMENT

RISER LEGEND

SYMBOL DESCRIPTION MOUNTING/COMMENT

AP CEILING ACCESS PANEL FOR REQUIRED ACCESS TO BOXES

B1 EQUIPMENT LABEL DENOTED CEILING MOUNT

N1 CONDUIT OR CABLE RUN JUNCTION BOX REFERENCE

CONDUIT STUB OUT CONDUIT STUB OUT WITH BUSHING

G1 CEILING SPEAKER CEILING, FLUSH, RECEPTACLE HEIGHT

DUPLEX I.G. POWER RECEPTACLE WALL, FLUSH, RECEPTACLE HEIGHT

QUAD I.G. POWER RECEPTACLE WALL, FLUSH, RECEPTACLE HEIGHT

DT DATA/TELEPHONE BOX WALL, FLUSH, RECEPTACLE HEIGHT

I.G.

I.G.

T2

CRQ1

MUSIC

AUDIO

INPUT

TOUCH PANEL

PACS PC

(2x)

CR3

1 1/4"C.

1 1/4"C.

CRTP1

P W R

C K T

P W R

C K T

I.G. I.G.

CONTROL

ROOM

DESK

CONTROL

ROOM

DESK

CONTROL

ROOM

DESK

1 1/4"C.

CR2

CONTROL

ROOM

DESK

TO T1

2"C.

MACLAB

1 1/4"C.

GUEST PORT

DVI (2x)

CR1

CONTROL

ROOM

DESK

1 1/4"C.

SHEET NOTES

1. INTEGRATED SYSTEM (A1) AND VIDEO CONFERENCING BOX (A2); REQUIRES DEDICATED

EMERGENCY BACK-UP 120V ON 20 AMP POWER FROM THE O.R. IT SERVES (NOT BY STERIS)

2. HUBBELL - RACO 698 4 GANG JUNCTION BOXES USED FOR FIBER SOLUTIONS

3. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE CONDUIT TO ENTER ON THE RIGHT

HALF OF THE BOX

4. HUBBELL - RACO 698 4 GANG JUNCTION BOXES REQUIRE 2" MINIMUM SERVICE CLEARANCE

AROUND ALL SIDES OF THE BOX

5. AUDIO REQUIREMENT; HOSPITAL GRADE HUBBELL USB CHARGER RECEPTACLE DUPLEX

(ACCEPTABLE MODEL NUMBERS: #SB8300LA, #USB8300R, #USB8300W OR SIMILAR; NOT BY

STERIS)

6. TOUCH PANEL REQUIRES DUPLEX OUTLET 120 VAC ON 20 AMP SERVICE (STANDARD OR

EMERGENCY)

K2 SURGICAL REMOTE POWER MODULE STERIS

N1 WALL MOUNTED 12x12x4" JUNCTION BOX ELECTRICAL CONTRACTOR

T1 24" H x 24" W X 8" D JUNCTION BOX ABOVE

CEILING ELECTRICAL CONTRACTOR

T2 12" H x 12" W X 6" D JUNCTION BOX ABOVE

CEILING ELECTRICAL CONTRACTOR

TP1

TOUCH PANEL STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR

U1 UNIVERSAL PORT - WALL MOUNTED 4 11/16"

JUNCTION BOX WITH SINGLE GANG MUD RING ELECTRICAL CONTRACTOR

U2 UNIVERSAL PORT - WALL MOUNTED 4 11/16"

JUNCTION BOX WITH SINGLE GANG MUD RING ELECTRICAL CONTRACTOR

V2 WALL MOUNTED LEVITON BOX FOR STERIS POWER

DISTRIBUTION ELECTRICAL CONTRACTOR

INTEGRATION EQUIPMENT SCHEDULE

LEGEND DESCRIPTION RESPONSIBILITY

A1 STERIS iQ3600 INTEGRATION SYSTEM STERIS

A2 VIDEO CONFERENCE BOX STERIS

B1*** SURGICAL LIGHT MOUNT STERIS

B2*** SURGICAL LIGHT MOUNT STERIS

C1*** MONITOR CARRIER STERIS

C2*** EQUIPMENT BOOM STERIS

C3*** ANESTHESIA BOOM STERIS

CR1

WALL MOUNTED HUBBELL - RACO 698 4 GANG

JUNCTION BOX FOR GUEST PORTS (2x) ELECTRICAL CONTRACTOR

CR2 WALL MOUNTED HUBBELL - RACO 698 4 GANG

JUNCTION BOX FOR MACLAB ELECTRICAL CONTRACTOR

CR3

WALL MOUNTED HUBBELL - RACO 698 4 GANG

JUNCTION BOX FOR PACS PC (2x) ELECTRICAL CONTRACTOR

CRQ1 WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING USED FOR AUDIO ELECTRICAL CONTRACTOR

CRTP1

TOUCH PANEL STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR TOUCH PANEL ELECTRICAL CONTRACTOR

E1

FUTURE WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR WALL

MOUNTED DISPLAY

ELECTRICAL CONTRACTOR

E2

FUTURE WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR WALL

MOUNTED DISPLAY

ELECTRICAL CONTRACTOR

E3

WALL MOUNTED DISPLAY STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR WALL

MOUNTED DISPLAY

ELECTRICAL CONTRACTOR

E4

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING FOR SIEMENS LIVE & REF

(DVI)

ELECTRICAL CONTRACTOR

E5

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING FOR 3RD PARTY DVI HD

DISPLAY OR DESTINATION

ELECTRICAL CONTRACTOR

F1

WALL MOUNTED HD PTZ CAMERA STERIS

WALL MOUNTED 4 11/16" JUNCTION BOX WITH

SINGLE GANG MUD RING REQUIRED FOR WALL

MOUNTED HD PTZ CAMERA

ELECTRICAL CONTRACTOR

G1 CEILING MOUNTED SPEAKER STERIS

G3 CEILING MOUNTED SPEAKER STERIS

G2 CEILING MOUNTED SPEAKER STERIS

G4 CEILING MOUNTED SPEAKER STERIS

K1 SURGICAL LIGHT CONTROLLER STERIS

36-05 iQ3600

CONDUIT RISER DIAGRAM

PROPOSED CONTROL ROOMPlanning & Design Group

STERIS

STERIS Corporation

5960 Heisley Road

Mentor, OH 44060 USA

STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD

Dr

Ck

Date

Scale

H- B-

800-548-4873 NA

DATE

LL

KS

C3R1-INT

11224

PAGE: OF 36-08

CUSTOMER

VERIFY

CUSTOMER

VERIFY

3"3"

A B O V E

F I N I S H E D

F L O O R

I F

U S E D

F O R

X R A Y

V I E W I N G

S T A T I O N

C U S T O M E R

V E R I F Y

A L L

A B O V E

F I N I S H E D

F L O O R

E1,E2,E3

F1

TYPICAL WALL DISPLAY INSTALLATION (OPTIONAL)

*REFERENCE SHEET NOTES #1,#2,#3

TYPICAL WALL MOUNTED HD PTZ CAMERA

INSTALLATION (OPTIONAL)

*REFERENCE SHEET NOTES #3,#4

1558"

33 8

POWER (SEE

SHEET NOTES

FOR DETAILS)

TYPICAL LEVITON BOX FOR STERIS

POWER DISTRIBUTION INSTALLATION

*REFERENCE SHEET NOTES #5,#6

V2

1034"13 1/16"

6"

1012"

12 1/16"

2"

60" ABOVE

FINISHED

FLOOR

2"17 1/2"2"

13 1/2"

6"

SURGICAL

REMOTE POWER

MODULE

ROUGH OPENING

SURGICAL LIGHT

TOUCH PANEL

CONTROL UNIT

ROUGH OPENING

6"

E SERIES SURGICAL LIGHT CONTROLLER

AND REMOTE POWER MODULE (OPTION)

K2

K1

EQUIPMENT SCHEDULE

LEGEND DESCRIPTION RESPONSIBILITY

E1,E2,E3 WALL MOUNTED DISPLAY STERIS

F1 HD PTZ WALL MOUNTED CAMERA STERIS

K1 SURGICAL LIGHT CONTROLLER STERIS

K2 SURGICAL REMOTE POWER MODULE STERIS

V2 LEVITON BOX FOR STERIS POWER DISTRIBUTION ELECTRICAL CONTRACTOR

SHEET NOTES

1. WALL DISPLAY BACKING TO BE PROVIDED AND INSTALLED BY CONTRACTOR; WALL DISPLAY BACKING

(NOT BY STERIS) MUST SUPPORT A 250LBS LOAD; WALL BRACKET NOT INSTALLED BY STERIS

2. WALL DISPLAY REQUIRES 4 11/16" JUNCTION BOX WITH SINGLE GANG MUD RING

3. WALL DISPLAY & PTZ CAMERA REQUIRE DUPLEX OUTLET 120 VAC ON 20 AMP SERVICE (STANDARD OR

EMERGENCY)

4. HD PTZ CAMERA MOUNTED DIRECTLY TO FRONT OF A/V PLATE; EXACT LOCATION OF H.D. CAMERA TO

BE DETERMINED BY CUSTOMER; CONTRACTOR TO PROVIDE PROPER BACKING FOR H.D. CAMERA; (1)

DUPLEX OUTLET REQUIRED AT EACH PTZ CAMERA LOCATION

5. V2 REQUIRES DEDICATED EMERGENCY BACK-UP 120V ON 20 AMP POWER FROM THE O.R. IT SERVES

(NOT BY STERIS)

6. V2 LEVITON BOX (NOT BY STERIS) WITH VENTED DOOR FOR STERIS POWER DISTRIBUTION; V2 BOX PRODUCT NUMBER: 47605-42N (REQUIRES MINIMUM 2" SERVICE CLEARANCE); V2 DOOR (NOT BY STERIS) PRODUCT NUMBER: 47605-42S; LEVITON AC DUAL POWER MODULE (NOT BY STERIS - BOLTED

TO BOTTOM OF LEVITON BOX - LOCATED INSIDE OF WALL) PRODUCT NUMBER: 47605-0DP

36-06 iQ3600

TYPICAL OPERATING ROOM

INSTALLATION DETAILS

Planning & Design Group

STERIS

STERIS Corporation

5960 Heisley Road

Mentor, OH 44060 USA

STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD

Dr

Ck

Date

Scale

H- B-

800-548-4873 NA

DATE

LL

KS

C3R1-INT

11224

PAGE: OF 36-08

EQUIPMENT SCHEDULE

LEGEND DESCRIPTION RESPONSIBILITY

G1,G2,G3,G4 BOSE CEILING SPEAKER (OPTIONAL) STERIS

SHEET NOTES

1. CONTRACTOR TO INSTALL SEISMIC SAFETY SUPPORT WIRE FOR SPEAKER ANCHORING TO

THE BUILDING STRUCTURE

2. CONTRACTOR TO CUT ROUND HOLE IN CEILING AT THE SPEAKER LOCATION ACCORDING

TO THE SPEAKER INSTALLATION TEMPLATE

3. TOP CLEARANCE FOR SPEAKERS IS 9"; ROUGH-IN PAN TO BE INSTALLED WHERE

OPENING IS NOT OBSTRUCTED ABOVE CEILING

T1 24"x24"x8"

G1 G2

G4 G3

1/2" FLEXIBLE CONDUIT WITH

CONNECTOR ON SPEAKER END

(NOT BY STERIS)(TYPICAL AT

EACH SPEAKER LOCATION)

SPEAKER (INSTALLED BY STERIS)

BOSE ROUGH-IN PAN PRE-SHIPPED

BY STERIS (NOT BY STERIS)

3/4" CONDUIT (NON-FLEX)

(NOT BY STERIS)

3/4" CONDUIT (NON-FLEX)

(NOT BY STERIS)

1/2" FLEXIBLE CONDUIT WITH

CONNECTOR ON SPEAKER END

(NOT BY STERIS)(TYPICAL AT

EACH SPEAKER LOCATION)

SPEAKER (INSTALLED BY STERIS)

BOSE ROUGH-IN PAN PRE-SHIPPED

BY STERIS (NOT BY STERIS)

3/4" CONDUIT (NON-FLEX)

(NOT BY STERIS)

R O U G H I N

H O L E

R E Q U I R E D

6.94"

MOUNTING EARS

FINISHED CEILING

SIDE

VIEW

FRONT VIEW

CEILING SECTION

VIEW

FREESPACE DS 16F ROUGH-IN PAN

26.00 [660.mm] m m

Ø 8.40 [213.4mm] BOSE ROUGH-IN PAN

PRE-SHIPPED BY STERIS

(NOT INSTALLED BY

STERIS)

36-07 iQ3600 Planning & Design Group

STERIS

STERIS Corporation

5960 Heisley Road

Mentor, OH 44060 USA

STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD

Dr

Ck

Date

Scale

H- B-

800-548-4873 NA

DATE

LL

KS

C3R1-INT

11224

PAGE: OF 36-08

VA MEDICAL CENTER OF RICHMOND

RICHMOND, VIRGINA

TYPICAL BOSE CEILING SPEAKER

(G1-G4) INSTALLATION DETAILS

36-08 iQ3600

CALL AND CONNECT HUB DETAIL

WITH ELECTRICAL REQUIREMENTS

Planning & Design Group

STERIS

STERIS Corporation

5960 Heisley Road

Mentor, OH 44060 USA

STERIS DRAWINGS DO NOT SUPERSEDE THE DRAWINGS FROM THE ARCHITECT OF RECORD

Dr

Ck

Date

Scale

H- B-

800-548-4873 NA

DATE

LL

KS

C3R1-INT

11224

PAGE: OF 36-08

VA MEDICAL CENTER OF RICHMOND

RICHMOND, VIRGINA

CALL AND CONNECT HUB - ELEVATION DETAIL PAGE

49.2"

23.4"

FINISHED CEILING

3" MINIMUM

REQUIRED

26.0"

FINISHED CEILING

DESCRIPTION RESPONSIBILITY

CALL AND CONNECT HUB STERIS

SHEET NOTES

1. CALL AND CONNECT HUB REQUIRES (1) DEDICATED 120 VAC ON 20 AMP

CIRCUIT

2. CALL AND CONNECT HUB REQUIRES (2) DATA DROPS - (1) FOR

CONTROL FOR ROOM-TO-ROOM AND (1) FOR CODEC

3. CALL AND CONNECT HUB MIDDLE ATLANTIC MODEL# DWR2426 24U WALL

MOUNTED AV RACK

4. CALL AND CONNECT HUB REQUIRES REINFORCED WALL BEHIND TO

SUPPORT 150LBS (NOT BY STERIS)

5. CALL AND CONNECT REQUIRES A MINIMUM OF 2" CLEARANCE REQUIRED

AROUND FRONT AND SIDES FOR PROPER VENTILATION

6. 8x8x4" JUNCTION BOX LOCATED IN WALL BEHIND THE CALL & CONNECT

HUB RACK WITH DUPLEX POWER

2" C.

TO CORRIDOR

CALL & CONNECT HUB FRONT ELEVATION CALL & CONNECT HUB SIDE ELEVATION

CORRIDOR

CALL AND

CONNECT HUB

1 1/2" CONDUITT1 AT

HYBRID

OR

*CONDUIT SIZE

DEPENDANT ON SIZE OF

SYSTEM

Sheets and Views
11224C3R1-INT-PUBLISHED-36-01 TITLE
11224C3R1-INT-PUBLISHED-36-02 LAYOUT
11224C3R1-INT-PUBLISHED-36-03 3600 DETAIL
11224C3R1-INT-PUBLISHED-36-04 RISER DETAIL
11224C3R1-INT-PUBLISHED-36-05 RISER DETAIL
11224C3R1-INT-PUBLISHED-36-06 INSTALLTION DETAIL
11224C3R1-INT-PUBLISHED-36-07 SPEAKER DETAIL
11224C3R1-INT-PUBLISHED-36-08 CALLCONNECT