Physicist Recommendations.pdf
PDF 195 KB Posted
- Attached to
- Y1DA--Relocate & Upgrade ANGIO Suite Federal contract opportunity
- Solicitation number
- 36C24221B0010
About this file
This document contains a federal contract opportunity solicitation and related shielding calculations for an interventional radiology suite renovation project. The solicitation is for relocating and upgrading the angiography suite at the VA New Jersey Health Care System facility in East Orange, New Jersey. The estimated cost range for the project is $2 million to $5 million. The scope of work involves general construction, structural alterations, asbestos abatement, HVAC, plumbing and electrical work to renovate space on the first floor for the new angiography suite. The shielding calculations provide recommendations for lead shielding requirements on walls, windows, doors and other penetrations for the interventional radiology suite based on the room layout and projected workloads. The solicitation response due date will be included in the solicitation document to be posted on SAM.gov on or around December 4, 2020. The project has a 365 calendar day completion deadline and requires bonding. The opportunity is set aside for Service-Disabled Veteran Owned Small Businesses.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C24221B0010 0004.docx | DOCX document | |
| 36C24221B0010 0003.docx | DOCX document | |
| Site Visit Sign-in Sheet.pdf | ||
| RFI-3_ST9 1ST FLOOR FRAMING (WINGS).pdf | ||
| RFI-13_A34 LONGITUDINAL SECTION X-X.pdf | ||
| Rev.1 _ Scope Clarification.pdf | ||
| RFI-3_A-6 RED CLINIC A-LEVEL.pdf | ||
| Interior Signage Standard.pdf | ||
| 36C24221B0010 0002.docx | DOCX document | |
| Door Hardware Specs.pdf | ||
| 36C24221B0010 0001.docx | DOCX document | |
| PLUMB Drawing.pdf | ||
| FP Drawing.pdf | ||
| ISSUE FOR BID Spec.pdf | ||
| Arch-Abate Drawing.pdf | ||
| ELEC Drawing.pdf | ||
| MECH Drawing.pdf | ||
| Struct Drawing.pdf | ||
| 36C24221B0010_1.docx | DOCX document |
Show all 19
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
RE:
4. All penetrations in lead walls must be covered by an equal amount of lead.
5. Lead equivalent glass windows must meet the same requirements as the lead walls.
6. Door frames must meet the same lead requirements as the walls.
7. All wall openings, such as outlet boxes, door knobs, locks, etc., must have the same lead equivalent as surrounding wall.
8. Please ensure the operator has both visual and audible communication with the patient during treatment.
Sincerely, Jason S. Tavel, PhD, DABR
IR Suite Shielding Calculations for:
VA New Jersey Health Care System
385 East Tremont Ave referenced project. Please note the following:
East Orange, NJ 07018
Att. Jacob Serfass
3. The lead on all walls must extend to a 7' height from the finished floor, except where noted.
Astarita Associates, Inc.
MEDICAL PHYSICS CONSULTANTS
414 Route 111, Smithtown, NY 11787
Phone (631) 265-2950
Jacob.Serfass@va.gov
April 23, 2020
Fax (631) 265-2962
1. Kindly reference the attached diagram ( page 3).
Dear Mr. Serfass:
As per your request, enclosed are the calculations for the shielding needed at the above
VA New Jersey Health Care System material may be used if preferred.
385 East Tremont Ave
East Orange, NJ 07018
These calculations are based upon the room layouts, occupancies, and dimensions provided. Any changes in parameters may result in a need for additional shielding. Should this occur, please contact
Astarita Associates immediately.
2. Please reference the Install column for the minimum shielding recommendations. Additional
Unit: GE Innova
Workload: 50 Patients/wk at 100 kVp
Exist Exist
Wall D(ft) U T P/S "Pb. "Con mR/wk R/NR
1 8 1.00 1.000 S 0 0.00 808.24 R
2 11 1.00 0.200 S 0 0.00 85.50 NR
3 11 1.00 0.050 S 0 0.00 21.38 NR
4 15 1.00 0.025 S 0 0.00 5.75 NR
5 15 1.00 0.025 S 0 0.00 5.75 NR
6 9 1.00 1.000 S 0 10.00 0.00 NR
7 9 1.00 1.000 S 0 10.00 0.00 NR
Allowable limit is 2mR/wk in Non-Restricted Areas and 100mR/wk in Restricted Areas.
T : Occupancy Factors based on NCRP 147.
P: Primary Beam S: Secondary Beam
Below - Occupied
Note: No lead needed where >0.5" concrete/brick is existing
Note: No lead needed where >0.5" concrete/brick is existing
R- Restricted Area NR-Non Restricted Area
Outside
Above - Occupied 0
"Pb.Area
Control 1/16
Shielding calculations for: VA New Jersey Health Care System
Install
Scrub Room, Hall & Doors 1/32
1/32
1/32
1/32
Outside
Prep Room & Doors
April 23, 2020
Room: IR Suite
= 1/16"Pb
= 1/32"Pb
Prep Room
Scrub Room
Hall
Outside
Outside