Physicist Recommendations.pdf

PDF 195 KB Posted

Attached to
Y1DA--Relocate & Upgrade ANGIO Suite Federal contract opportunity
Solicitation number
36C24221B0010
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 2

About this file

This document contains a federal contract opportunity solicitation and related shielding calculations for an interventional radiology suite renovation project. The solicitation is for relocating and upgrading the angiography suite at the VA New Jersey Health Care System facility in East Orange, New Jersey. The estimated cost range for the project is $2 million to $5 million. The scope of work involves general construction, structural alterations, asbestos abatement, HVAC, plumbing and electrical work to renovate space on the first floor for the new angiography suite. The shielding calculations provide recommendations for lead shielding requirements on walls, windows, doors and other penetrations for the interventional radiology suite based on the room layout and projected workloads. The solicitation response due date will be included in the solicitation document to be posted on SAM.gov on or around December 4, 2020. The project has a 365 calendar day completion deadline and requires bonding. The opportunity is set aside for Service-Disabled Veteran Owned Small Businesses.

View the file

Other files for this federal contract opportunity

Other files attached to Y1DA--Relocate & Upgrade ANGIO Suite, newest first.
File Type Posted
36C24221B0010 0004.docx DOCX document
36C24221B0010 0003.docx DOCX document
Site Visit Sign-in Sheet.pdf PDF
RFI-3_ST9 1ST FLOOR FRAMING (WINGS).pdf PDF
RFI-13_A34 LONGITUDINAL SECTION X-X.pdf PDF
Rev.1 _ Scope Clarification.pdf PDF
RFI-3_A-6 RED CLINIC A-LEVEL.pdf PDF
Interior Signage Standard.pdf PDF
36C24221B0010 0002.docx DOCX document
Door Hardware Specs.pdf PDF
36C24221B0010 0001.docx DOCX document
PLUMB Drawing.pdf PDF
FP Drawing.pdf PDF
ISSUE FOR BID Spec.pdf PDF
Arch-Abate Drawing.pdf PDF
ELEC Drawing.pdf PDF
MECH Drawing.pdf PDF
Struct Drawing.pdf PDF
36C24221B0010_1.docx DOCX document
Show all 19

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

RE:

4. All penetrations in lead walls must be covered by an equal amount of lead.

5. Lead equivalent glass windows must meet the same requirements as the lead walls.

6. Door frames must meet the same lead requirements as the walls.

7. All wall openings, such as outlet boxes, door knobs, locks, etc., must have the same lead equivalent as surrounding wall.

8. Please ensure the operator has both visual and audible communication with the patient during treatment.

Sincerely, Jason S. Tavel, PhD, DABR

IR Suite Shielding Calculations for:

VA New Jersey Health Care System

385 East Tremont Ave referenced project. Please note the following:

East Orange, NJ 07018

Att. Jacob Serfass

3. The lead on all walls must extend to a 7' height from the finished floor, except where noted.

Astarita Associates, Inc.

MEDICAL PHYSICS CONSULTANTS

414 Route 111, Smithtown, NY 11787

Phone (631) 265-2950

Jacob.Serfass@va.gov

April 23, 2020

Fax (631) 265-2962

1. Kindly reference the attached diagram ( page 3).

Dear Mr. Serfass:

As per your request, enclosed are the calculations for the shielding needed at the above

VA New Jersey Health Care System material may be used if preferred.

385 East Tremont Ave

East Orange, NJ 07018

These calculations are based upon the room layouts, occupancies, and dimensions provided. Any changes in parameters may result in a need for additional shielding. Should this occur, please contact

Astarita Associates immediately.

2. Please reference the Install column for the minimum shielding recommendations. Additional

Unit: GE Innova

Workload: 50 Patients/wk at 100 kVp

Exist Exist

Wall D(ft) U T P/S "Pb. "Con mR/wk R/NR

1 8 1.00 1.000 S 0 0.00 808.24 R

2 11 1.00 0.200 S 0 0.00 85.50 NR

3 11 1.00 0.050 S 0 0.00 21.38 NR

4 15 1.00 0.025 S 0 0.00 5.75 NR

5 15 1.00 0.025 S 0 0.00 5.75 NR

6 9 1.00 1.000 S 0 10.00 0.00 NR

7 9 1.00 1.000 S 0 10.00 0.00 NR

Allowable limit is 2mR/wk in Non-Restricted Areas and 100mR/wk in Restricted Areas.

T : Occupancy Factors based on NCRP 147.

P: Primary Beam S: Secondary Beam

Below - Occupied

Note: No lead needed where >0.5" concrete/brick is existing

Note: No lead needed where >0.5" concrete/brick is existing

R- Restricted Area NR-Non Restricted Area

Outside

Above - Occupied 0

"Pb.Area

Control 1/16

Shielding calculations for: VA New Jersey Health Care System

Install

Scrub Room, Hall & Doors 1/32

1/32

1/32

1/32

Outside

Prep Room & Doors

April 23, 2020

Room: IR Suite

= 1/16"Pb

= 1/32"Pb

Prep Room

Scrub Room

Hall

Outside

Outside