2. Audio Booth Minimum Specifications 20240826.pdf
PDF 153 KB Posted
- Attached to
- Audiometric Booth
- Solicitation number
- HT941024Q2080
- Issued by
-
Defense Health Agency
About this file
This document contains the minimum specifications for an audiometric booth to be installed at the United States Naval Hospital in Yokosuka, Japan. The booth must meet specific requirements, including the ability to test multiple people simultaneously, soundproof construction, high-performance acoustic doors, and various technical features such as a master jack panel, LED lighting, and ventilation system. The document also outlines the technical equipment to be provided by the government, including audiometers, headphones, and computer interface cables. The booth must pass a certification test to ensure it meets the required sound pressure level standards before government acceptance. The installation and certification must be completed by April 15, 2025.
The related federal contract opportunity is a Request for Quote (RFQ) for the provision and installation of the audiometric booth. This is a 100% set-aside for small businesses under NAICS code 334510, with a size standard of 1,250 employees. The solicitation number is HT941024Q2080 and is being issued by the Defense Health Agency.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Audio Booth Minimum Specifications 26 Aug 24
1. List of all specific booth requirements:
• 6 person booth
• Capable of testing multiple people simultaneously
• Soundproof booth to isolate outside noise that passes annual booth certification; see attached document of booth certification requirements.
• High performance acoustic doors; 32” w x 73 ¾” h clear opening door c/w 22 x 20 window in door
• Fully isolated floors
• Privacy curtains
• Silenced ventilation system; fan exhaust and static intake ceiling silencers
• Plug and Play electrical; CSA approved electrical; 120 v, 12 A, 60 Hz
• Master Jack Panel; plate with twenty four ¼” phone jacks c/w six individual stations of 4 plug jack plates
• LED ceiling lighting
• One 3” pass thru
• 4” acoustic floor with RD-3 green isolators
• Standard carpeting
• One quad outlet
• Light and fan switches
• Door stop
• Headset and hand switch hooks at each station
• Stools for each station
• Set up options to include left hinge standard, right hinge standard, left hinge reverse, right hinge revers
• Floor options; 7” rise- 4” floo/rd3 isolators, 4” rise- 2” floor/rd2 isolators, 1 9/16” rise- low profile option
• Collar for HVAC hook up
• Vent for exhaust
2. List of all tech that will be installed: The hardware supplies to be provided by the Government; the wire covers shall be provided by the contractor. There will be six patient testing stations.
• Benson Medical CCA-200 automatic/semi-manual audiometers
• Benson Medical BAS-200 (Bio-Acoustic Simulator)
• External Workstation USB Computer Interface cable that connects to booth to provide power to the internal audiometers and connection to the workstation (computer) for DOEHRS-HC function (software to test hearing)
• Telephonic TDH-39 headphones
• Patient Response Switch
• Daisy Chain- Expansion and Power VGA style cables; six stations
• Wire Covers- not provided by NMCFHPC and requesting vendor to supply enough to connect six work stations.
3. Specifics on booth certification; see the attached ‘Booth-Cert-Fillable-Form’ document that is used for booth certification. The form covers the standards (Max Sound Pressure Level allowed (dB) for each test frequency that the booth must meet for medical surveillance testing. The instructions are found on page two of how the booth certification must be completed. Government acceptance will not occur until the booth passes the certification test. The contractor is responsible for ensuring the booth certification including all supplies, tools, equipment and installation.
4. Installation and certification shall be complete NLT 15 April 2025.
5. Additional Information:
Current space measurements where booth shall be installed:
a. 98 3/8” by 112 9/32” (WxL) with an additional 47 ¾” for an open door
b. Height: 99 1/8”
Door option that is required for the space, Left hinge standard.