Plant Growth Chamber Specifications

Control System Control system is a robust industrial grade control system that is designed specifically for plant growth chambers and rooms. Its intuitive color touch screen interface and limitless scalability make it a reliable, flexible and user‐friendly system.

Programming: Capable of storing up to 16 real‐time user‐entered programs each containing up to 48 lines with one minute resolution. Each program defines a single 24‐hour period. Programs are then entered into a “Schedule” (a sequence of up to 8 program entries) to create a single day, multi day, or seasonal schedule. Programs are entered via a high‐resolution color touch‐screen display.

Process Control Mode: All chamber process transitions occur in STEP mode, with the exception of temperature, which can be set to either STEP or global ramp (gRAMP) mode.

Data Logging and Graphing Local Data: The data available at the local display screen includes up to seven chamber variables such as temperature, humidity, lighting and CO2. Parameters are all logged automatically every 30 seconds and can be viewed in the form of Trend graphs. Trend graphs can be viewed, with up to five days of history, directly on the local display screen for both ‘set‐point’ and ‘actual’ values. Zoom and Pan Functions provide additional visualization tools allowing the User to precisely and accurately view and access the data. A cursor can be used to display the value and the setpoint at the cursor location.

The data available remotely is more substantial and includes every input and output parameter being monitored and controlled. Data is automatically logged and log rate can be set in multiples of 6 seconds regardless of the number of parameters. This provides unprecedented data resolution for the User while it also benefits service personnel. Log data is accessed in several ways:

∙ A network‐connected PC (Note: Requires optional products, sold separately) ∙ An internet‐enabled PC (Note: Requires optional products, sold separately)

Trend data can also be downloaded through the USB port of the chamber touch screen. Chamber data archived on the touch screen is logged every 30 seconds with up to fifteen days of history.

Alarms Audible and color‐coded visual alarms are provided. User‐settable absolute alarms for all controlled processes and User‐settable process tracking alarms for temperature are available for the User to determine how tightly they want to monitor the experiment. In addition, there are also “built in” alarms available for chamber component monitoring. Alarm history is logged to daily text files easily retrievable from any Ethernet connected computer or through the local control screen (last 15 alarm events).

User‐settable absolute alarms for monitoring sensors. User‐settable absolute alarms are available to determine an alarm condition for auxiliary sensors used to monitor chamber conditions.

User‐settable alarm delays are provided for all process sensors and monitoring sensors. An alarm delay of up to 540 minutes can be enabled that will bypass the user settable alarm limits. Alarm delays allow the user the ability to temporarily interact with the chamber without causing an alarm condition. Once the alarm delay has expired the process and monitoring alarms will no longer be bypassed.

Security: Password‐protected levels are reserved for users, administrators and service personnel. The unit is capable of storing up to 50 unique names and passwords for user and administrator profiles.

Start‐Up Delay: Chambers can be programmed with varying restart times to reduce start‐up loads after a power failure.

Communications: Ethernet connectivity is delivered with every controller. Controller, mask and Host IP address are easily settable from the Status screen. The controller is shipped communications‐ready.

Auxiliary and Central Alarm Contacts: Alarm contacts for connection to a building management system or (optional) auto dialer. The chamber local display can be mirrored on company’s central system (sold separately) allowing remote support, training and access to all the local chamber functions remotely.

Lighting Run Times: The controller tracks the amount of time (in hours) each lighting output has been ON. The run time of each output can be viewed and reset using the local display.

Audit Trails: The controller logs user interaction with the HMI. The Audit Trail feature allows the controller administrator to view all changes and button presses executed on the HMI. The Audit Trail is conveniently saved as a .csv file that can be downloaded from the control system and viewed on a .csv compatible device.

Construction (All dimensions are nominal) Exterior Dimensions: 104"W x 35"D x 78"H (2640mmW x 890mmD x 1980mmH).

Interior Dimensions: 73¼"W x 31 ¾"D x 65¼"H (1870mmW x 805mmD x 1655mmH) Floor: Perforated aluminum channel floor for uniform upward air flow ‐ Unifloor®.

Growth Area: 16.1ft² (1.5m²).

Growth Capacity: 78ft³ (2220 liters).

Growth Height: Approximately 48” (1220mm).

Cabinet Construction: Bonded paneling using CFC‐free insulation.

Exterior Finish: Blue‐green enamel baked on patterned aluminum.

Interior Finish: Reflective white enamel baked on smooth aluminum.

Drain Pan: 22 gauge stainless steel drain pan under the entire growth area, sloped the full length toward the machine compartment. A drain fitting is installed and connected to an external drain tube.

Condensing Unit Access: Easy access to compressor and refrigeration components through an exterior panel on the end of the chamber.

Doors: Two reach‐in doors with keyed magnetic locks, each clear opening 30"W x 54"H (760mmW x 1375mmH).

Observation Window: Dual pane with light tight cover 11" x 15" (280mm x 380mm). Note: With Ultra Low Temperature (ULT) option, observation window is unavailable.

Shelf: Corrosion resistant wire adjustable on ½" (12mm) centers.

Control Panel: Left hand (right hand model optional).

Instrument Ports: Two ports, 1" (25mm) with light tight caps.

Factory assembled, tested and fully crated.

Lighting Intensity: Approximately 1,000 micromoles/m²/s (higher light intensities are optional) Programming and Control: Automatic adjustment of light intensity is provided within the programmed range with broadband sources, responding to an in‐chamber quantum light sensor.

Light: Balanced spectrum for plant growth using Elixia LED broadband fixtures.

Light Fixture: Light canopy is fixed in place, designed to work with high efficiency electronic drivers with dimming capability.

Light Heat: Removed by refrigeration system.

Power Supply Unit: High efficiency electronic.

Light Meter: LI‐190R Quantum light meter for display and recording of light output, displayed through user interface. Sensor includes amplifier and 50’ cable length.

Temperature Control (Maximum design ambient temperature is +35°C) Range: +4°C to +35°C lights OFF, +10°C to +35°C lights ON. (Full fresh air.)

Control4: ±0.5°C, at control point.

Temperature Safety Limits:

Primary: A programmable min and max temperature or a limit tracking alarm that automatically follows the programmed set point.

Secondary: An independent factory‐set high and low temperature limit is also provided for increased assurance.

Audible alarms are included based on user‐defined limits.

Aspirator: Located in the growth area for accurate controlling and recording of chamber conditions.

Refrigeration Remote Outdoor Air‐Cooled Condenser: Comes complete with all‐weather housing, low ambient operation controls and low noise level operation. Remote location (up to 50' [15m] combined horizontal and vertical distance) of condenser only. Compressor, receiver, and other refrigeration components remain in cabinet machine compartment.

1. Inter‐connecting refrigeration and electrical lines are included up to a maximum of 50’ (15m).

Refrigeration pipe insulation shall be unjacketed black Armaflex.

2. For remote location distances over 50' (15m) (combined horizontal and vertical distance) please consult factory.

Valve: Electronic 3‐way modulating valve that regulates the temperature function of the chamber. The only moving part of this valve is a floating component within the pressure system, which is totally sealed.

Heat Exchanger Coil(s): Copper‐tube construction.

Refrigerant: Condensing unit is charged with CFC‐free refrigerant.

Monitoring: Refrigeration system operation monitored by control system, including visual and audible alarm. Pressure transducers allow for real‐time diagnostics for preventative maintenance and repair.

Air Flow Vertical: Uniform upward through Unifloor® less than 50ft/min (15.2m/min).

Fresh Air: Filtered inlet and adjustable exhaust to 20ft³/min (0.57m³/min).

Capable of adding optional Humidity Control Range: No control on basic unit.

Capable of adding optional Carbon Dioxide Additive Control Range: No control on basic unit.

Utility Requirements: (rating increases with some options.)

Electrical Service: (Alternative services available, consult factory) Control Panel: 120/208‐3Æ‐60Hz‐4 wire plus ground Condenser: 208‐230‐1Ø‐60Hz‐3wire plus ground Drain: Floor drain must be provided outside footprint of cabinet.

Provided with a U.S. electrical inspection by an inspection authority accredited by OSHA.

Warranty and After Sales Service Warranty and after sales service are handled by our dedicated team of service professionals. They provide a 24‐hour manned service phone to facilitate any repair emergency. In addition, the team is available, during regular business hours, to assist with any service or warranty repairs, replacement parts or questions. Company also offers a fully trained international network of independent third party service providers who can affect repairs wherever your chamber is located. Two‐year warranty (extended warranty optional).

Preventative Maintenance & Service Program (Optional) Offers comprehensive preventative maintenance and service programs for its equipment. Program frequency and structure are customized based on your requirements and desired level of service.

Extended Warranty: (Optional) Extended Warranty is an extension of the original Standard Warranty, including all of the original terms and conditions. An Extended Warranty may be purchased at any time during the Standard Warranty period or at a later date for an After Sale Extended Warranty. Note: Extended warranties must be purchased together with Preventative Maintenance. Consult factory for further details.

Installation Company responsible for/provide the following:

1. Submit a Site Inspection Report for client completion prior to mobilization on site (sent 1‐ month prior).

2. Transport the equipment from the off‐loading or storage area to the assembly area. Uncrate equipment and dispose of crating material in customer supplied disposal bins.

3. Assemble and make work the supplied equipment using qualified personnel. Start‐up and check equipment performance and operation to factory specifications.

a) Verification and function testing of building power, all input devices, all controlled devices, all non controlled devices, refrigeration and safety components and if applicable all special control sequences and devices; and

b) A 24‐hour test to demonstrate performance of all controlled parameters to their specified limits.

‐Provide four hours of instruction on use and maintenance of the equipment. Additional instruction and/or training available by agreement.

‐Operator Manual(s): Access to an electronic manual (PDF) available ‐Customer to be responsible for/provide the following:

Complete and return Site Inspection Report at minimum one (1) week prior to installation mobilization on site. Delay in receipt may negatively affect installation time‐line and/or charges noted.

‐Receive and off‐load equipment. If there will be a delay between the receipt and building utility connections of the equipment, the customer is responsible for providing adequate and secure storage.

Customer must provide forklift type equipment to remove equipment from common carrier highway transport truck/s.

‐The customer must prepare a clean and unobstructed route, including sufficient clearance through doorways, openings, elevators, etc. between the off‐loading or storage area and the location of chamber assembly.

‐The customer must ensure that installers have adequate on‐site storage and staging areas. If no such space is available, company may arrange for storage containers at the customer’s expense. Pricing is based on customer providing on‐site access to authorized personnel for a 10‐hr work‐day, Monday to

Friday, with 8hr to 10 hr work‐day on Saturday. Alterative work hours / days available upon request, consult factory prior to order placement. Note: Company shall not be responsible for compensation of client, hired consults/contractors for any extended or overtime hours to accommodate onsite access.

Disposal of refuse transferred to customer‐supplied bin(s), which are to be located within 50’ (15m) of exit door of building.

‐Supply and connection of building services to the equipment in accordance with local codes, restrictions and legislation. These services must be fully functional and sized in accordance with the equipment specifications in a timely manner. The customer shall supply and connect all building services to the equipment, including any cutting, penetrations, or patching required.

Building services may include some or all of the following:

‐ Electrical service (more than one may be required) ‐ Purified water for humidity system ‐ Fresh air and/or exhaust air connections ‐ Carbon dioxide supply ‐ Floor drain connections ‐ Conduit or cable trays or other special requirements for Ethernet support (applicable to CM only) Equipment must be installed on level floors. The customer shall ensure that all floors are level to within ½ inch.

Exceptions: Offer does not include (unless noted) road permits, street closures and/or traffic control, by others. No assumption has been made for crane rental(s). Roofing work, penetrations, curbing or pitch pockets, by others.

The quotation for installation services assumes that all on‐site responsibilities of the customer have been fulfilled upon company commencing the installation such that company can complete the installation in a single mobilization. This includes the provision of all building services, operating as specified. Should site delays result in company having to suspend its installation work, the customer shall reimburse company for all downtime, including the cost of return visits and associated expenses.

Expenses shall be charged on a cost plus 15% basis, and time shall be charged at company’s standard installation rates.

General: Company understands there is no requirement to provide unionized labor in the performance of its work. As such, company may supply non‐union or unionized trade labor at its sole discretion.

Local Regulatory Codes: No allowances have been made in the equipment design or manufacture to consider local regulatory codes including seismic requirements. Building code requirements for your area should be investigated prior to equipment installation.

Company provides installation services with the understanding that said services are to install and make work, within a given period of time, the company supplied equipment. In no way does company guarantee, or imply to guarantee, to supply a minimum number of man‐hours and/or installation personnel at any time during the process of this installation.