6640--MOLECULAR Analyzer for SARS-COV-2

Awarded Award Notice Posted

Solicitation number
36C24223Q1208
Agency
Veterans Integrated Service Network 2 Veterans Health Administration, Department of Veterans Affairs
Awarded
to Hologic Sales And Service, LLC
Set-aside
No set-aside

Opportunity facts

Contract number
36C24223P1863 Federal contract award
NAICS code
334516 Analytical Laboratory Instrument Manufacturing
PSC
6640 Laboratory Equipment And Supplies

Notice details come from SAM.gov. Updated .

About this opportunity

This Award Notice from the Department of Veterans Affairs documents a $255,000 contract awarded to Hologic Sales and Service, LLC on September 21, 2023. The contract is for the purchase of a molecular real-time PCR analyzer and reagents that will enable the Syracuse VA Medical Center laboratory to expand its testing capacity for SARS-CoV-2, influenza A and B, respiratory syncytial virus, and sexually transmitted infections. Delivery of the analyzer, which must be able to process a minimum of 200 samples within 8 hours, is required within 90 days along with a one-year warranty and training services.

Notice text

4

1

Statement of Work

Molecular Analyzer for SARS-CoV-2 (COVID) and Respiratory Testing

GENERAL INFORMATION

Title of Project: Purchase of Molecular Analyzer and Reagents for SARS-CoV-2 (COVID) and Respiratory Testing

Type of Contract Contemplated: One time purchase for multiple assay molecular analyzer

SCOPE

The Syracuse Molecular laboratory requires capacity to expand SARS-COV-2/Influenza/RSV molecular testing to Veteran s in the community in order to provide appropriate care in an inpatient and outpatient setting. Our current analyzer respiratory testing output is not sufficient. Procuring additional equipment and tests for molecular respiratory testing will allow the Syracuse VA Medical Center to manage patient symptoms sooner, decrease inpatient admissions and length of stay, and limit out of pocket expenses to Veteran s whom require respiratory testing but are not admitted to the facility. Larger testing capacities will increase patient satisfaction and prevent some patient visits to the Emergency Room for those needing testing. The analyzer should enhance the laboratory s ability expand PCR testing for other infectious diseases, such as STI s and HPV.

REQUIREMENT/SALIENT CHARACTERISTICS/SPECIFICATIONS

Common Nomenclature (commercial description): : Molecular Real-Time PCR Analyzer

Principles of operation: The analyzer of choice will allow the laboratory to run real-time PCR, TMA, and RT-TMA assays on a single, automated platform. Test assays allow for nucleic acid amplification for qualitative detection of RNA for specific targets. Assays may combine the technologies of target capture, Transcription Mediated Amplification (TMA), and Dual Kinetic Assay (DKA). The analyzer must have the ability to run FDA approved testing for Respiratory and EUA for COVID. The analyzer must also have the ability to run FDA approved testing for STIs, Viral Loads, and BV, as well as have the functionality to run Laboratory Developed Tests (LDTs) all in one single integrated platform.

Intended use: To identify infectious diseases in clinical specimens by molecular methods.

Equipment with which the item is used :

Validation materials, including control materials, must be included with contract purchase. A back-up UPS battery must be included with purchase.

Original equipment manufacturer part number:

Other pertinent information that describes the item, material or service required:

The analyzer must have the capability to test for the following using PCR methods:

SARS-CoV-2 virus

Influenza A virus

Influenza B virus

Respiratory Syncytial Virus (RSV)

Mycoplasma genitalium

Trichomonas vaginalis

Chlamydia trachomatis

Neisseria gonorrhoeae

Pierceable foil caps, no uncapping prior to sample processing.

The analyzer must be high throughput with at least 200 samples able to be processed within 8 hours.

Laboratory preparedness system to rapidly respond to emerging pathogen(s) in an emergency

The analyzer must be fully automated to allow samples to run with minimal to no technologist intervention.

A continuous fluids/waste module to the lower bay of the system, allowing access to replace a set of universal fluids when depleted and dispose solid and liquid waste containers during processing, increasing flexibility and throughput to the system.

Continuous and random-access sample loading that allows both samples and reagents to be loaded at any time during the day

The testing should NOT include extractions or pre-treatment prior to analyzing specimen.

The analyzer should NOT require plumbing, as the laboratory does not have the ability to provide plumbing at this time.

All tests listed above must be FDA Approved or approved for Emergency Use Authorization (EUA).

OTHER UNIQUE REQUIREMENTS

Compatibility with Existing Equipment:

Applicable DICOM compliance, Directives or IT Policies:

Ensure that contractors, third party partners, and servicers implement the VA security and privacy requirements, as defined in the contract. These requirements can also be added to the contract Statement of Work (SOW). The requirements apply to applicable contracts in which VA sensitive information is stored, generated, transmitted, or exchanged by VA, a contractor, subcontractor or a third-party, or on behalf of any of these entities regardless of format or whether it resides on a VA system or contractor or subcontractor s electronic information system(s) operating for or on the VA s behalf.

Warranty: One year warranty all parts and labor included. Products shall be free from defects in workmanship and materials for a period of one (1) year from the date of purchase.

Training: An application Specialist must be on-site for installation of assay-specific applications and initial verification of assay-specific runs. In addition to verification support, it is required that the application specialist provide primary operator training. It is also requested that additional operator training is available for changes in personnel.

DELIVERY

Deliver to Syracuse VA loading dock/warehouse

Inspection by BioMed/HTM

Inspection and acceptance will occur in C322

Delivery within 90 days of award.

PLACE OF PERFORMANCE

Syracuse VA Laboratory 3rd Floor

BILLING/INVOICING Tungsten Network Portal

CUSTOMER INFORMATION - Clinical Laboratory, Syracuse VA Medical Center Awarded contract 36C24223P1863 to HOLOGIC SALES AND SERVICE, LLC for 255000.00000000 on 2023-09-21.

Attachments

Files attached to this notice, newest first
File Type Posted
36C24223P1863_1.docx DOCX document

Notice history

Notices posted for this opportunity, newest first
Notice Type Posted
6640--MOLECULAR Analyzer for SARS-COV-2 This notice · Award Award Notice
6640--528-23-4-7085-0256 alinity system - abbott molecular Latest solicitation Solicitation

On GovTribe

Work this opportunity on GovTribe

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity