3-axis Magnetic Field Cancelation System

Awarded Award Notice Posted

Solicitation number
NB686080-16-00594
Agency
National Institute of Standards and Technology Department of Commerce
Awarded
to Integrated Dynamics Engineering Inc.
Set-aside
No set-aside

Opportunity facts

Contract number
DOCSB134116SU0127 Federal contract award
NAICS code
333517 Machine Tool Manufacturing
PSC
Not on record
Place of performance
325 Broadway Boulder, Colorado 80305, United States

Notice details come from SAM.gov. Updated .

Notice text

Added: Dec 18, 2015 11:00 am NB686080-16-00594
3-AXIS MAGNETIC FIELD CANCELATION SYSTEM

PURPOSE


The Applied Physics Division (div. 686), in conjunction with the National Institute of Standards and Technology (NIST) require a 3-axis magnetic field cancelation system for a preexisting cryogenic microwave probe station recently installed at NIST Boulder.


TECHNICAL SPECIFICATIONS:


1. The system must include a 3-axis sensor whose dimensions should be less than or approximately equal to 120x35x35mm. The sensor band width must be from DC to at least 100 kHz.
2. The system must be capable of nulling fields over a wide bandwidth (DC - 10kHz) with at least -40 dB attenuation at DC and -25 dB attenuation at 1 kHz.
3. The magnetic field sensor noise must be below 1nT/sqrt(Hz) for frequencies above 1 Hz.
4. The system must effectively cancel DC fields up to 70uT (given 36" square field cancelation coils).
5. System controller: the system must come with a system controller that can provide up to 6 channels of output current to control field null coils. The controller must have at least 16 bit output resolution to control coils. The controller must include a USB or RS 232 control interface. The controller must be rack mountable in a standard 19" equipment rack.
6. Coil interface and coil power: The system must include a simple interface to connect to a set of six NIST fabricating nulling coils. The maximum output power per coil must be at least 20 W/coil.
7. Software: the system must come with Windows based software to display sensed fields, set tuning parameters, and display operation parameters.


TECHNICAL CONSIDERATIONS


The system must be capable of nulling the earth's magnetic field, which typically varies from 25 to 65 microTesla to less than 100 nTesla.
INSPECTION AND ACCEPTANCE
The system must reduce both the DC and AC magnetic fields to the level specified above when installed on the existing NIST cryogenic probe station for acceptance.


DELIVERABLE SCHEDULE


The Contractor shall deliver no later than sixteen (16) weeks after receipt of order.

Attachments

Files attached to this notice, newest first
File Type Posted
Statement_of_Work_for_NB686080-16-00594.pdf PDF
Extension_Combined_Synop_Solicit_for_NB686080-16-00594.pdf PDF
Statement_of_Work_for_NB686080-16-00594.pdf PDF
Combined_Synop_Solicit_for_NB686080-16-00594.docx.pdf PDF

On GovTribe

Work this opportunity on GovTribe

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity