Specifications for Curie Point Analytical Pyrolyzer
Requirement: To acquire a state of the art Analytical pyrolyzer unit to interface to an Agilent 7890 Gas Chromatograph.
The system required is a JCI-55 model analytical pyrolyzer. The system must operate on a curie point heating system and be compatible with an Agilent 7890 Gas Chromatograph.
System specifications must include must include the following:
Probe
Heating System Curie point heating
Pyrolysis Pyrofoil (160℃~1040℃)
Oven Heating 150℃
Transfer line (Needle upper part)
300℃ (at the time of curie point heating)
Sample Tube Quartz glass, OD 6mm, Length 36mm
Needle Exclusive needle (OD 0.47mm, Length
42mm)
Connection Cable 2m
Dimensions OD 35mm, Length 180mm
Weight 110g
Power Supplied from controller
Controller
RF Power 48W, 600kHz
Pyrolysis time 5sec, 10sec, 15sec
Output Signal GC/start signal
Mode selection Qualitative analysis Mode, Quantitative analysis Mode, Solvent purge Mode、Baking Mode
Injection Gas (He) 5liter Helium carlrige or Supplying 0.5MPa
Helium
Dimension of Mainframe (mm)
150(W)×246(H)×367(D)
Weight 7Kg