Specifications for Curie Point Analytical Pyrolyzer

Requirement: To acquire a state of the art Analytical pyrolyzer unit to interface to an Agilent 7890 Gas Chromatograph.

The system required is a JCI-55 model analytical pyrolyzer. The system must operate on a curie point heating system and be compatible with an Agilent 7890 Gas Chromatograph.

System specifications must include must include the following:

Probe

Heating System Curie point heating

Pyrolysis Pyrofoil (160℃~1040℃)

Oven Heating 150℃

Transfer line (Needle upper part)

300℃ (at the time of curie point heating)

Sample Tube Quartz glass, OD 6mm, Length 36mm

Needle Exclusive needle (OD 0.47mm, Length

42mm)

Connection Cable 2m

Dimensions OD 35mm, Length 180mm

Weight 110g

Power Supplied from controller

Controller

RF Power 48W, 600kHz

Pyrolysis time 5sec, 10sec, 15sec

Output Signal GC/start signal

Mode selection Qualitative analysis Mode, Quantitative analysis Mode, Solvent purge Mode、Baking Mode

Injection Gas (He) 5liter Helium carlrige or Supplying 0.5MPa

Helium

Dimension of Mainframe (mm)

150(W)×246(H)×367(D)

Weight 7Kg

Power AC85~240V, 500VA(Max)