LPTA SUBMITTAL FORM-BIOREACTOR
FOR “LOWEST PRICE, TECHNICALLY ACCEPTABLE”
| Determination | Definition |
| ACCEPTABLE | A quote is acceptable if it meets or exceeds the minimum requirements listed in the Salient Characteristics column |
| UNACCEPTABLE | A quote will be deemed unacceptable if it does not meet or exceed the minimum requirements as stated in the Salient Characteristics. A lack of minimum content or failure to address all evaluated areas could render a quote unacceptable. PLEASE NOTE: Blank submittals in the Proposed column will deem the proposal unacceptable. |
SOLICITED: PARALLEL BIOREACTOR SYSTEM
| SALIENT CHARACTERISTICS | "MEETS REQUIREMENT" OR "DOES NOT MEET REQUIREMENT" | |
| 1 | PARALLEL BIOREACTOR SYSTEM |
2 Eight bioreactor vessels run simultaneously under independent conditions
3 Autoclavable bioreactor vessels (glass body with metal headplate)
4 Bioreactor minimum working volume of ≤ 25mL
5 Bioreactor maximum working volume of ≥ 150mL
6 Headplates configuration with ports/connectors for the following:
- Two liquid feed lines
- Gas inlet (sparger) and outlet (with Peltier cooled condenser)
- Temperature probe
- pH probe
- ORP probe
- Optical density probe
- Sampling port with dip tube
- Injection port with septum
- Impeller or suspended magnetic stirrer
7 Temperature control of each vessel from ambient to ≥ 70°C with precision of ± 0.1°C and accompanying probes (1/vessel)
8 Stirring speed of 50-500rpm
9 Controlled gas delivery by integrated mass flow controller
10 Peltier cooled condensers (1/vessel)
11 Autoclavable ORP probes (1/vessel)
12 Autoclavable pH probes (1/vessel)
13 Variable-speed peristaltic pumps with tubing (2/vessel)
14 Autoclavable optical density probes (1/vessel)
15 Control software and compatible PC to monitor and control stirrer speed, temperature, pH, and ORP
NOTES: (Please add additional Notes, as appropriate.