Attachment 2.
The requirements listed in the table below shall be accommodated by the responder offering a complete LGR Liquid Water isoptope Analyzer. Please indicate Y (yes) or N (No) if your system meets each requirement in the requirements list below, and use the comments column to further justify how your product meets the requirement.
REQUIREMENTS/ SPECIFICATIONS
Meets Requirement?
(Y/N) Comments
| 1 |
| Shall be Stand-alone, table-top spectrometer including auto-injector |
| 1 |
| Method of detection shall be: cavity-enhanced laser absorption |
| 2 |
| Performance: isotopic ratios δ2H, δ18O in water simultaneously by auto-injection of liquid specimens. |
Precision in High Throughput mode:
δ2H: 0.5‰ δ18O: 0.1‰
| 3 |
| High throughput: 800 injections per day |
| 4 |
| Must be compatible with nested disposable tips that are in stacks of 8 racks of 96 disposable tips to greatly increase efficiency for availability of disposable tips |
| 5 |
| Serviceable by user on site |
| 6 |
| Access to high-resolution absorption spectra during acquisition for real time diagnostics |
| 7 |
| Low sample volume per injection: |
1 microliter
| 8 |
| Thermal stability: Internally controlled thermal stabilization to avoid drifts while used in the field. samples, one at a time and 8 in parallel. Tip positioning precision must allow access into high-density microplates, up to and including 1536 well plates |
| 9 |
| Software: Stand-alone software for system control. |
| 10 |
| Maximum Power Consumption for field use: Not to exceed 200 watts |
| 11 |
| Temperature Range for Field Operation: 0-45 degrees C |
(End)